The HemoSphere Advanced Monitoring Platform is intended to be used by qualified personnel or trained clinicians in a critical care environment in a hospital setting. The Advanced Viewfinder Remote mobile application can be used for supplemental near real-time remote display of monitored hemodynamic parameter data as well as Faults, Alerts and Notifications generated by the HemoSphere Advanced Monitoring Platform. The HemoSphere Advanced Monitoring Platform is intended for use with compatible Edwards Swan-Ganz and Oximetry Catheters, FloTrac sensors, Acumen IQ sensors, TruWave DPT sensors, ForeSight sensors, and ClearSight/Acumen IQ finger cuffs.
Device Story
HemoSphere Advanced Monitoring Platform is a modular hemodynamic monitoring system used in hospital critical care settings (OR, ICU, ER) by clinicians. It processes inputs from various sensors (Swan-Ganz catheters, oximetry cables, pressure cables, finger cuffs, ForeSight sensors) to calculate and display hemodynamic parameters (CO, CI, SV, SVR, SvO2, ScvO2, StO2, blood pressure). The system includes software features: Acumen Hypotension Prediction Index (HPI) for predicting hypotensive events (MAP < 65 mmHg) and Acumen Assisted Fluid Management (AFM) for fluid responsiveness suggestions. Data is visualized on the monitor and can be streamed to the Viewfinder Remote mobile application for remote viewing. The device provides quantitative physiological data to assist clinicians in hemodynamic optimization and goal-directed therapy. It does not make therapeutic decisions; clinicians review data to guide fluid administration or other interventions. Benefits include continuous assessment of cardiac function, fluid status, and tissue oxygenation, potentially improving patient management in ischemic or hemodynamic instability states.
Clinical Evidence
No clinical testing was performed. Evidence consists of bench testing, system-level verification, software verification, and a usability study. Bench simulations verified measured and derived parameters. Electrical safety and EMC testing confirmed compliance with IEC 60601-1, 60601-1-2, 60601-2-34, 60601-2-57, 60601-2-49, and 80601-2-49. Wireless coexistence testing followed ANSI C63.27/D1.0. Usability studies confirmed that intended users could perform critical tasks without usability issues leading to harm.
Technological Characteristics
Modular hemodynamic monitor with external sub-system modules. Connectivity via wired/wireless interfaces to Hospital Information Systems and mobile devices (Viewfinder Remote). Software includes HPI and AFM algorithms. Complies with IEC 60601-1, 60601-1-2, 60601-1-6, 60601-1-8, 62304, 62366, 60601-2-34, 60601-2-57, 60601-2-49, and 80601-2-49. Cybersecurity includes Windows 10 IoT, OS/App hardening, and security logs.
Indications for Use
Indicated for adult and pediatric critical care patients requiring hemodynamic monitoring (cardiac output, venous oxygen saturation, blood pressure, vascular resistance, fluid responsiveness, and tissue hemoglobin oxygenation) in hospital environments (OR, ICU, ER). Specific indications vary by module (Swan-Ganz, Oximetry, Pressure, Technology/ForeSight, ClearSight). Acumen HPI is indicated for surgical/non-surgical patients receiving advanced hemodynamic monitoring. Acumen AFM is indicated for surgical patients ≥18 years. ForeSight sensors are indicated for regional hemoglobin oxygen saturation monitoring based on weight/age (Adults/adolescents >40kg; Pediatrics >3kg; Cerebral <8kg/Non-cerebral <5kg).
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 22, 2022
Edwards Lifesciences, LLC Manthan Damani Manager, Regulatory Affairs One Edwards Way Irvine, California 92614
Re: K213682
Trade/Device Name: HemoSphere Advanced Monitor, HemoSphere Swan-Ganz Module, HemoSphere Oximetry Cable, HemoSphere Pressure Cable, HemoSphere ClearSight Module, HemoSphere Technology Module, HemoSphere ForeSight Oximeter Cable, Acumen Hypotension Prediction Index Feature, Acumen Assisted Fluid Management Feature, Viewfinder Remote Regulation Number: 21 CFR 870.1425
Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II Product Code: DQK, DQE, QAQ, MUD, DXN, DSB, QMS, FLL Dated: November 19, 2021 Received: November 22, 2021
Dear Manthan Damani:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
#### 510(k) Number (if known)
#### K213682
#### Device Name
HemoSphere Advanced Monitor, HemoSphere Swan-Ganz Module, HemoSphere Pressure Cable, HemoSphere ClearSight Module, HemoSphere Technology Module, HemoSphere ForeSight Oximen Hypotension Prediction Index Feature, Acumen Assisted Fluid Management Feature, Viewfinder Remote
#### Indications for Use (Describe)
HemoSphere Advanced Monitor with HemoSphere Swan-Ganz Module:
The HemoSphere Advanced Monitor when used with the HemoSphere Swan-Ganz Module and Edwards Swan-Ganz Catheters is indicated for use in adult and pediatric critical care patients requiring monitoring of cardiac output [continuous (CO) and intermittent (iCO)] and derived hemodynamic parameters. Pulmonary artery blood temperature monitoring is used to compute continuous and intermittent CO with thermodilution technologies. It may also be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the Edwards Swan-Ganz catheter indications for use statement for information on target patient population specific to the catheter being used. Refer to the Edwards Swan-Ganz catheter indications for use statement for information on target patient population specific to the catheter being used.
Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.
HemoSphere Advanced Monitor with HemoSphere Oximetry Cable:
The HemoSphere Advanced Monitor when used with the HemoSphere Oximetry cable and Edwards is indicated for use in adult and pediatric critical care patients requiring of venous oxygen saturation (SvO2 and ScvO2) and derived hemodynamic parameters in a hospital environment. Refer to the Edwards oximetry catheter indications for use statement for information on target patient population specific to the catheter being used.
Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.
HemoSphere Advanced Monitor with HemoSphere Pressure Cable:
The HemoSphere Advanced Monitor when used with the HemoSphere Pressure Cable is indicated for use in critical care patients in which the balance between cardiac function, fluid status, vascular resistance and pressure needs continuous assessment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the Edwards FloTrac, Acumen IQ and TruWave DPT sensor indications for use statement for information on target patient population specific to the sensor being used.
The Edwards Acumen Hypotension Prediction Index feature provides the clinician with physiological insight into a patient's likelihood of future hypotensive events (defined as mean arterial pressure < 65 mmHg for at least one minute in duration) and the associated hemodynamics. The Acumen HPI feature is intended for use in surgical or non-surgical patients receiving advanced hemodynamic monitoring. The Acumen HPI feature is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Acumen Hypotension Prediction Index (HPI) parameter.
Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.
{3}------------------------------------------------
HemoSphere Advanced Monitor with Acumen Assisted Fluid Management Feature and Acumen IQ Sensor:
The Acumen Assisted Fluid Management (AFM) software feature provides the clinician with physiological insight into a patient's estimated response to fluid therapy and the associated hemodynamics. The Acumen AFM software is intended for use in surgical patients ≥18 years of age, that require advanced hemodynamic monitoring. The Acumen AFM software feature offers suggestions regarding the paysiological condition and estimated response to fluid therapy. Acumen AFM fluid administration suggestions are offered to the clinician; the decision to administer a fluid bolus is made by the clinician, based upon review of the patient's hemodynamics. No therapeutic decisions should be made based solely on the Assisted Fluid Management suggestions.
HemoSphere Advanced Monitor with HemoSphere Technology Module and ForeSight Oximeter Cable:
The non-invasive ForeSight Oximeter Cable is intended for use as an adjunct monitor of absolute regional hemoglobin oxygen saturation of blood under the sensors in individuals at risk for reduced-flow or no-flow ischemic states. The ForeSight Oximeter Cable is also intended to monitor relative changes of total hemoglobin of blood under the sensors. The ForeSight Oximeter Cable is intended to allow for the display of StO2 on the HemoSphere advanced monitor.
- When used with large sensor, the ForeSight Oximeter Cable is indicated for use on adults and transitional adolescents >40 kg.
· When used with Medium Sensors, the ForeSight Oximeter Cable is indicated for use on pediatric subjects 23 kg.
· When used with Small Sensors, the ForeSight Oximeter Cable is indicated for cerebral use on pediatric subjects < 8 kg and non-cerebral use on pediatric subjects <5kg.
Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.
HemoSphere Advanced Monitor with HemoSphere ClearSight Module:
The HemoSphere advanced monitor when used with the HemoSphere ClearSight module, pressure controller and a compatible Edwards finger cuff are indicated for patients over 18 years of age in which the balance between cardiac function, fluid status and vascular resistance needs continuous assessment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. In addition, the noninvasive system is indicated for use in patients with co-morbidities for which hemodynamic optimization is desired and invasive measurements are difficult. The HemoSphere advanced monitor and compatible Edwards finger cuffs noninvasively measures blood pressure and associated hemodynamic parameters.
The Edwards Lifesciences Acumen Hypotension Index feature provides the clinician with physiological insight into a patient's likelihood of future hypotensive events (defined as mean arterial pressure < 65 mmHg for at least one minute in duration) and the associated hemodynamics. The Acumen HPI feature is intended for use in surgical patients receiving advanced hemodynamic monitoring. The Acumen HPI feature is considered to be additional quanitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Hypotension Prediction Index (HPI) parameter.
Refer to the ClearSight finger cuff and Acumen IQ finger cuff indications for use statements for information on target patient population specific to the finger cuff being used.
Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.
{4}------------------------------------------------
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# K213682
| Sponsor: | Edwards Lifesciences LLC<br>One Edwards Way<br>Irvine, CA 92614 | | |
|-------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|-----------------|
| Establishment<br>Registration<br>Number: | 2015691 | | |
| Contact<br>Person: | Manthan Damani<br>Senior Manager, Regulatory Affairs<br>One Edwards Way<br>Irvine, CA 92614<br>manthan_damani@edwards.com<br>Telephone: (949) 250-7853 | | |
| Date: | June 22, 2022 | | |
| Platform<br>Name | HemoSphere Advanced Monitoring Platform | | |
| Trade Name: | HemoSphere Advanced Monitor<br>HemoSphere Swan-Ganz Module<br>HemoSphere Oximetry Cable<br>HemoSphere Pressure Cable<br>HemoSphere Technology Module ( <i>subject</i> )<br>HemoSphere ForeSight Oximeter Cable ( <i>subject</i> )<br>HemoSphere ClearSight Module<br>Acumen Hypotension Prediction Index software feature ( <i>subject</i> )<br>Acumen Assisted Fluid Management software feature ( <i>subject</i> )<br>Viewfinder Remote ( <i>subject</i> ) | | |
| Common<br>Name: | Cardiac Output/Oximetry/Ejection Fraction Computer | | |
| Classification<br>Name for<br>HemoSphere<br>Monitor and<br>Accessories: | Programmable Diagnostic Computer<br>Fiberoptic Oximeter Catheter<br>Adjunctive Predictive Cardiovascular Indicator<br>Oximeter, Tissue Saturation (Non-Invasive)<br>System, Measurement, Blood-Pressure, Non-Invasive<br>Plethysmograph, Impedance<br>Adjunctive Open Loop Fluid Therapy Recommender<br>Thermometer, Electronic, Clinical | 21 CFR 870.1425 | 21 CFR 870.1230 |
| | | 21 CFR 870.2210 | 21 CFR 870.2700 |
| | | 21 CFR 870.1130 | 21 CFR 870.2770 |
| | | 21 CFR 870.5600 | 21 CFR 880.2910 |
# 510(k) Summary – HemoSphere Advanced Monitoring Platform
{6}------------------------------------------------
| Product Code<br>for<br>HemoSphere<br>Monitor and<br>Accessories: | DQK, Class II<br>DQE, Class II<br>QAQ, Class II<br>MUD, Class II<br>DXN, Class II<br>DSB, Class II<br>QMS, Class II<br>FLL, Class II |
|------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Primary<br>Predicate<br>Device for<br>HemoSphere<br>Monitor and<br>Accessories: | HemoSphere Advanced Monitoring Platform, manufactured by Edwards Lifesciences,<br>K203687 cleared May 28, 2021, is being utilized for substantial equivalence to the device<br>modularity, basic device functionality, graphical user interface (GUI) used, core predictive<br>algorithm for the Acumen Hypotension Prediction Index (HPI) feature, and cybersecurity<br>features. The indications for use are also similar to the subject device. |
| Secondary<br>Predicate<br>Devices for<br>HemoSphere<br>Monitor and<br>Accessories: | FORE-SIGHT ELITE Tissue Oximeter Module (K190270, cleared October 21, 2019),<br>manufactured by Edwards Lifesciences, is being utilized for substantial equivalence to the<br>tissue hemoglobin feature and indications for use. |
| | Acumen Assisted Fluid Management (AFM) feature on EV1000A Clinical Platform<br>(DEN190029, granted November 13, 2020), manufactured by Edwards Lifesciences, is<br>being utilized for substantial equivalence to the Acumen Assisted Fluid Management<br>(AFM) software feature and indications for use. |
| | HemoSphere Advanced Monitoring Platform (K211465, cleared July 08, 2021),<br>manufactured by Edwards Lifesciences, is being utilized for substantial equivalence to the<br>connectivity features of the HemoSphere Advanced monitor to allow it to connect to the<br>Viewfinder Remote mobile application as well as for the Viewfinder Remote application<br>itself. |
| Device<br>Description: | The HemoSphere Advanced Monitoring platform was designed to simplify the customer<br>experience by providing one platform with modular solutions for their hemodynamic<br>monitoring needs. The user can choose from the available optional sub-system modules or<br>use multiple sub-system modules at the same time. This modular approach provides the<br>customer with the choice of purchasing and/or using specific monitoring applications based<br>on their needs. Users are not required to have all of the modules installed at the same time<br>for the platform to function. |
| | HemoSphere Advanced Monitoring Platform, subject of this submission, consists of the<br>HemoSphere Advanced Monitor that provides a means to interact with and visualize<br>hemodynamic and volumetric data on the monitor screen and its five (5) optional external<br>modules: the HemoSphere Swan-Ganz Module (K163381 Cleared, April 14, 2017), the<br>HemoSphere Oximetry Cable (K163381 Cleared, April 14, 2017), HemoSphere Pressure<br>Cable (K180881 Cleared, November 16, 2018), HemoSphere Technology Module<br>(previously referred to as "Tissue Oximetry Module"; K190305 cleared, August 29, 2019), |
{7}------------------------------------------------
HemoSphere ForeSight Module (K180003, May 10, 2018), and the HemoSphere ClearSight Module (K203687 cleared, May 28, 2021). The HemoSphere Advanced Monitor also has wired and wireless capabilities which was originally used only for connecting to a Hospital Information System (HIS) for data charting purposes. This capability is now used to allow it to stream continuously monitored data to the Viewfinder Remote, a mobile device-based application, for remote viewing the information (K211465 cleared July 8, 2021). The remotely transmitted data from the patient monitoring sessions include all hemodynamic parameter data and the associated physiological alarm notifications, historical trend data, and parameter waveform data.
A high-level overview of the new features and modifications that have been implemented as part of this subject submission are listed below:
#### -Addition of new features, algorithm, and/or technology to the HemoSphere Platform:
- o Tissue Hemoglobin
- The updated ForeSight Oximeter Cable (previously referred to as "FORE-SIGHT ELITE Oximeter Module/FSM") with the Tissue Hemoglobin algorithm (K190270 cleared October 21, 2019) has been incorporated into the HemoSphere Monitoring Platform. The tissue hemoglobin parameter includes relative changes in oxygenated and deoxygenated hemoglobin concentrations for assessment of oxygenation changes as well as their sum total hemoglobin concentration, under the sensor when used with a compatible host monitor. As such, the indications for use and intended use of the existing HemoSphere Advanced Monitor have been expanded to include the ability to measure relative change in total hemoglobin. Additionally, since the HemoSphere Advanced Monitor will act as the host monitor, the graphical user interface (GUI) of the Monitor is being updated to allow display of the hemoglobin parameter.
- Assisted Fluid Management software feature: O
- The Edwards' Acumen™ Assisted Fluid Management (AFM) ■ software feature (DEN190029 granted on November 13, 2020) has been incorporated into the HemoSphere Advanced Monitoring Platform to recognize patterns of fluid responsiveness (i.e., hemodynamic data and past responses to fluid) and recommend when fluid administration is expected to improve the patient's hemodynamic state. This AFM software feature (AFM algorithm + AFM GUI) is now being incorporated as part of the HemoSphere Advanced Monitor. No modifications have been made to the previously granted AFM algorithm.
{8}------------------------------------------------
- Updates to existing features, algorithm, and/or technology to the HemoSphere Platform:
- HPI Smart Alerts and Smart Trends o
. The Acumen Hypotension Prediction Index (HPI) algorithm defines hypotensive events as mean arterial pressure (MAP) < 65 mmHg for at least one minute in duration. The software triggers an alert screen whenever the HPI parameter value is greater than 85 for two consecutive readings. From the alert screen, the user has the option to go to a secondary screen, where additional data related to alert can be viewed.
Modifications have been made to the HPI alert popup and HPI secondary screen. The appearance and behavior of the HPI alert popup (referred to as a Smart Alert popup) has been modified to display the physiological mechanisms of preload, contractility, and afterload that may have changed in accordance with user-defined settings. The HPI secondary screen options also now include a new Smart Trends screen that displays values and trends for user-defined parameters associated with preload, contractility, and afterload when they have reached the user-defined targets. The Smart Settings menu allows clinician to select parameters to be monitored for each physiological mechanism (preload, contractility, afterload), and customize when the mechanism is triggered based on the percent change threshold over a pre-set time interval. There are no changes to the core HPI algorithm, the behavior of the HPI parameter display, nor the indications for use and intended use of the HPI parameter due to this Smart Trends/ Smart Alerts modification.
- Connectivity feature O
- As part of its connectivity development program, Edwards had . modified the HemoSphere Advanced Monitor (K211465 cleared July 8,2021) to allow it to stream (wired or wireless) continuously monitored non-invasive hemodynamic data to the Viewfinder Remote, a mobile device-based application, for remote viewing the information.
Modifications have been made to the connectivity feature on HemoSphere Advanced Monitoring Platform to also allow the remote display of minimally invasive and invasive hemodynamic data on the physicians/clinicians mobile/handheld devices such as cell phone using Viewfinder Remote application. Additional connectivity feature modifications include workflow-based updates to the HemoSphere monitor to support the functionality of remote viewing, including the ability to end a session, additional metadata support, multiple setup wizards for initial configuration, as well as remote software and cybersecurity patch updates.
{9}------------------------------------------------
- Cybersecurity updates O
- Cybersecurity updates on HemoSphere Monitor include Windows 10 IoT and cybersecurity tools, OS and App level hardening, and dedicated security logs.
- Miscellaneous Updates: о
- ForeSight Sensor surface temperature detection ■
- Non-Pulsatile Mode
- CASMED IFMOut/EMR protocol integration
- Status Screen Updates
- I Default CVP Values
- I Nomenclature changes due to rebranding
- Viewfinder Remote Application updates O
- 트 The intended use of the Viewfinder Remote mobile application, which is an accessory to the HemoSphere Monitor (K211465, cleared July 08, 2021), has been expanded to allow for remote display of all existing hemodynamic data from the HemoSphere monitor including minimally invasive and invasive parameters on the physicians/clinicians mobile/handheld devices such as cell phone when using Viewfinder Remote application (previously only allowed display of the non-invasive parameters). Please Note: The AFM feature is not offered or displayed on Viewfinder Remote.
#### HemoSphere Advanced Monitor with HemoSphere Swan-Ganz Module Indications for
Use:
The HemoSphere Advanced Monitor when used with the HemoSphere Swan-Ganz Module and Edwards Swan-Ganz Catheters is indicated for use in adult and pediatric critical care patients requiring monitoring of cardiac output [continuous (CO) and intermittent (iCO)] and derived hemodynamic parameters. Pulmonary artery blood temperature monitoring is used to compute continuous and intermittent CO with thermodilution technologies. It may also be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the Edwards Swan-Ganz catheter indications for use statement for information on target patient population specific to the catheter being used.
Refer to the Intended Use statement below for a complete list of measured and derived parameters available for each patient population.
## HemoSphere Advanced Monitor with HemoSphere Oximetry Cable
The HemoSphere Advanced Monitor when used with the HemoSphere Oximetry Cable and Edwards oximetry catheters is indicated for use in adult and pediatric critical care patients requiring monitoring of venous oxygen saturation (SvO2 and ScvO2) and derived hemodynamic parameters in a hospital environment. Refer to the Edwards oximetry catheter indications for use statement for information on target patient population specific to the catheter being used.
{10}------------------------------------------------
Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.
## HemoSphere Advanced Monitor with HemoSphere Pressure Cable
The HemoSphere Advanced Monitor when used with the HemoSphere Pressure Cable is indicated for use in critical care patients in which the balance between cardiac function, fluid status, vascular resistance and pressure needs continuous assessment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the Edwards FloTrac, Acumen IQ, and TruWave DPT sensor indications for use statement for information on target patient population specific to the sensor being used.
The Edwards Lifesciences Acumen Hypotension Prediction Index feature provides the clinician with physiological insight into a patient's likelihood of future hypotensive events (defined as mean arterial pressure < 65 mmHg for at least one minute in duration) and the associated hemodynamics. The Acumen HPI feature is intended for use in surgical or nonsurgical patients receiving advanced hemodynamic monitoring. The Acumen HPI feature is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Hypotension Prediction Index (HPI) parameter.
Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.
## HemoSphere Advanced Monitor with Acumen Assisted Fluid Management Feature and Acumen IQ Sensor:
The Acumen Assisted Fluid Management (AFM) software feature provides the clinician with physiological insight into a patient's estimated response to fluid therapy and the associated hemodynamics. The Acumen AFM software feature is intended for use in surgical patients >18 years of age, that require advanced hemodynamic monitoring. The Acumen AFM software feature offers suggestions regarding the patient's physiological condition and estimated response to fluid therapy. Acumen AFM fluid administration suggestions are offered to the clinician; the decision to administer a fluid bolus is made by the clinician, based upon review of the patient's hemodynamics. No therapeutic decisions should be made based solely on the Assisted Fluid Management suggestions.
## HemoSphere Advanced Monitor with HemoSphere Technology Module and ForeSight Oximeter Cable
The noninvasive ForeSight Oximeter Cable is intended for use as an adjunct monitor of absolute regional hemoglobin oxygen saturation of blood under the sensors in individuals at risk for reduced-flow or no-flow ischemic states. The ForeSight Oximeter Cable is also intended to monitor relative changes of total hemoglobin of blood under the sensors. The ForeSight Oximeter Cable is intended to allow for the display of StO2 and relative change in total hemoglobin on the HemoSphere advanced monitor.
{11}------------------------------------------------
- . When used with large sensors, the ForeSight Oximeter Cable is indicated for use on adults and transitional adolescents >40 kg.
- . When used with medium sensors, the ForeSight Oximeter Cable is indicated for use on pediatric subjects >3 kg.
- When used with small sensors, the ForeSight Oximeter Cable is indicated for ● cerebral use on pediatric subjects <8 kg and non-cerebral use on pediatric subjects <5kg.
Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.
## HemoSphere Advanced Monitor with HemoSphere ClearSight Module
The HemoSphere Advanced Monitor when used with the HemoSphere ClearSight module, pressure controller and a compatible Edwards finger cuff are indicated for patients over 18 years of age in which the balance between cardiac function, fluid status and vascular resistance needs continuous assessment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. In addition, the noninvasive system is indicated for use in patients with comorbidities for which hemodynamic optimization is desired and invasive measurements are difficult. The HemoSphere advanced monitor and compatible Edwards' finger cuffs noninvasively measures blood pressure and associated hemodynamic parameters.
The Edwards Lifesciences Acumen Hypotension Prediction Index feature provides the clinician with physiological insight into a patient's likelihood of future hypotensive events (defined as mean arterial pressure < 65 mmHg for at least one minute in duration) and the associated hemodynamics. The Acumen HPI feature is intended for use in surgical patients receiving advanced hemodynamic monitoring. The Acumen HPI feature is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Hypotension Prediction Index (HPI) parameter.
Refer to the ClearSight finger cuff and Acumen IQ finger cuff indications for use statements for information on target patient population specific to the finger cuff being used.
Refer to the ClearSight finger cuff indications for use statements for information on target patient population specific to the finger cuff being used.
#### Intended Use- HemoSphere Advanced Monitoring Platform: Intended Use
of the The HemoSphere Advanced Monitoring Platform is intended to be used by qualified HemoSphere personnel or trained clinicians in a critical care environment in a hospital setting. The Advanced Viewfinder Remote mobile application can be used for supplemental near real-time remote Monitor: display of monitored hemodynamic parameter data as well as Faults, Alerts and Notifications generated by the HemoSphere Advanced Monitoring Platform.
{12}------------------------------------------------
The HemoSphere Advanced Monitoring Platform is intended for use with compatible Edwards Swan-Ganz and Oximetry Catheters, FloTrac sensors, Acumen IQ sensors, TruWave DPT sensors, ForeSight sensors, and ClearSight/Acumen IQ finger cuffs.
A comprehensive list of parameters available while monitoring with the HemoSphere advanced monitor and a connected HemoSphere Swan-Ganz Module are listed in the table below. Only iCO, iCI, iSVR, and iSVRI are available to the pediatric patient population.
| Parameter | Description | Sub-System<br>Module<br>Used | Patient<br>Population | Hospital<br>Environment |
|-----------|---------------------------------------------------------|-----------------------------------|------------------------|--------------------------------------------------------------------|
| CO | continuous cardiac output | | | |
| sCO | STAT cardiac output | | | |
| CI | continuous cardiac index | | | |
| sCI | STAT cardiac index | | | |
| EDV | right ventricular end<br>diastolic<br>volume | | | |
| sEDV | STAT right ventricular<br>end<br>diastolic volume | | | |
| EDVI | right ventricular end<br>diastolic<br>volume index | | | |
| sEDVI | STAT right ventricular<br>end<br>diastolic volume index | | | |
| HRavg | averaged heart rate | | Adult only | Operating<br>Room,<br>Intensive Care<br>Unit,<br>Emergency<br>Room |
| LVSWI | left ventricular stroke work index | HemoSphere<br>Swan-Ganz<br>Module | | |
| PVR | pulmonary vascular resistance | | | |
| PVRI | pulmonary vascular resistance<br>index | | | |
| RVEF | right ventricular ejection fraction | | | |
| sRVEF | STAT right ventricular ejection<br>fraction | | | |
| RVSWI | right ventricular stroke work index | | | |
| SV | stroke volume | | | |
| SVI | stroke volume index | | | |
| SVR | systemic vascular resistance | | | |
| SVRI | systemic vascular resistance index | | | |
| BT | pulmonary artery blood temperature | | | |
| iCO | intermittent cardiac output | | | |
| iCI | intermittent cardiac index | | | |
| iSVR | intermittent systemic<br>vascular<br>resistance | | Adult and<br>Pediatric | |
| iSVRI | intermittent systemic<br>vascular<br>resistance index | | | |
A comprehensive list of parameters available for adult and pediatric patient populations while monitoring with the HemoSphere Advanced Monitor and a connected HemoSphere oximetry cable are as listed below:
{13}------------------------------------------------
| Parameter | Description | Sub-System Module Used | Patient Population | Hospital Environment |
|-----------|----------------------------------|---------------------------------|------------------------|-----------------------------------------------------------|
| SvO2 | Mixed Venous Oxygen Saturation | HemoSphere<br>Oximetry<br>Cable | Adult and<br>Pediatric | Operating Room,<br>Intensive Care Unit,<br>Emergency Room |
| ScvO2 | Central Venous Oxygen Saturation | | | |
A comprehensive list of additional parameters that are available for adult and pediatric patient populations on the HemoSphere Advanced Monitor and a connected HemoSphere Swan-Ganz Module and a connected HemoSphere Oximetry Cable are as listed below:
| Parameter | Description | Sub-System<br>Module<br>Used | Patient<br>Population | Hospital<br>Environment |
|-----------|------------------------------------------------------------------------|--------------------------------------------------------------------------|------------------------|--------------------------------------------------------------------|
| DO2 | Oxygen Delivery | HemoSphere<br>Swan-Ganz<br>Module and<br>HemoSphere<br>Oximetry<br>Cable | Adult and<br>Pediatric | |
| DO2I | Oxygen Delivery Indexed | | | Operating<br>Room,<br>Intensive Care<br>Unit,<br>Emergency<br>Room |
| VO2 | Oxygen Consumption | | | |
| VO2e | Estimated Oxygen Consumption<br>when ScvO2 is being monitored | | | |
| VO2I | Oxygen Consumption Index | | | |
| VO2Ie | Estimated Oxygen Consumption<br>Index when ScvO2 is being<br>monitored | | | |
A comprehensive list of parameters available for adult patient populations while monitoring with the HemoSphere Advanced Monitor and a connected HemoSphere pressure cable are as listed below:
| Parameter | Description | Sub-System<br>Module<br>Used | Patient<br>Population | Hospital<br>Environment |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------|---------------------------------|-----------------------|--------------------------------------------------------------------|
| CO/<br>CI | Continuous Cardiac Output¹/<br>Continuous Cardiac Index¹ | | | Operating<br>Room,<br>Intensive Care<br>Unit,<br>Emergency<br>Room |
| CVP | Central Venous Pressure | | | |
| DIA | Systemic arterial diastolic blood<br>pressure | | | |
| DIAPAP | pulmonary artery diastolic blood<br>pressure | | | |
| dP/dt | Systolic slope² | | | |
| Eadyn | Dynamic Arterial Elastance² | | | |
| MAP | Mean Arterial Pressure | | | |
| MPAP | Mean Pulmonary Arteria¹ Pressure | HemoSphere<br>Pressure<br>Cable | Adult only | |
| PPV | pulse pressure variation¹ | | | |
| PRART | Pulse rate | | | |
| SV/<br>SVI | Stroke Volume¹/<br>Stroke Volume Index¹ | | | |
| SVR/<br>SVRI | Systemic Vascular Resistance¹/<br>Systemic Vascular Resistance¹<br>Index | | | |
| SVV | Stroke Volume Variation¹ | | | |
| SYSART | Systemic Arterial Systolic Blood<br>Pressure | | | |
| SYSPAP | Pulmonary Artery Systolic Blood<br>Pressure | | | |
| HPI | Acumen Hypotension Prediction<br>Index² | | | |
| ¹FloTrac parameters are available when using a FloTrac/Acumen IQ sensor and if the FloTrac feature is enabled.<br>²HPI parameters are available when using an Acumen IQ sensor and if the HPI feature is activated. | | | | |
{14}------------------------------------------------
A list of Acumen Assisted Fluid Management (AFM) outputs available for surgical patients ≥18 years of age while monitoring with the HemoSphere Advanced Monitor and a connected HemoSphere pressure cable are as listed below:
| AFM outputs | Sub-System<br>Module Used | Patient<br>Population | Hospital<br>Environment |
|-----------------------------------------------------------------------------------------------|---------------------------------|--------------------------|-------------------------|
| Fluid Bolus Suggested | | | |
| Test Bolus Suggested | | ≥18 years of age<br>only | Operating<br>Room |
| Fluid Not Suggested | | | |
| Suggestions Suspended | HemoSphere<br>Pressure<br>Cable | | |
| Bolus in Progress... | | | |
| Bolus Complete | | | |
| Bolus Complete; Analyzing Hemodynamic Response | | | |
| Tracked Case Vol. | | | |
| AFM outputs are available when using an Acumen IQ sensor and if the AFM feature is activated. | | | |
A comprehensive list of parameters available for adult patient populations while monitoring with the HemoSphere Advanced Monitor and a connected HemoSphere pressure cable and a connected HemoSphere oximetry cable are as listed below:
| Parameter | Description | Sub-System<br>Module<br>Used | Patient<br>Population | Hospital<br>Environment |
|-----------|------------------------------------------------------------------------|--------------------------------------------------------------------------|-----------------------|--------------------------------------------------------------------|
| DO2 | Oxygen Delivery | | | |
| DO2I | Oxygen Delivery Indexed | HemoSphere<br>Swan-Ganz<br>Module and<br>HemoSphere<br>Oximetry<br>Cable | Adult only | Operating<br>Room,<br>Intensive Care<br>Unit,<br>Emergency<br>Room |
| VO2 | Oxygen Consumption | | | |
| VO2e | Estimated Oxygen Consumption<br>when ScvO2 is being monitored | | | |
| VO2I | Oxygen Consumption Index | | | |
| VO2Ie | Estimated Oxygen Consumption<br>Index when ScvO2 is being<br>monitored | | | |
A comprehensive list of parameters available while monitoring with the HemoSphere advanced monitor and both a connected HemoSphere Swan-Ganz Module and pressure cable are listed below:
| Parameter | Description | Sub-System<br>Module<br>Used | Patient<br>Population | Hospital<br>Environment |
|-----------|-------------------------------|-----------------------------------------------|-----------------------|--------------------------------------------------------------------|
| CO20s | 20-second cardiac output | HemoSphere | Adult only | Operating<br>Room,<br>Intensive Care<br>Unit,<br>Emergency<br>Room |
| CI20s | 20-second cardiac index | Swan-Ganz | | |
| SV20s | 20-second stroke volume | Module and<br>HemoSphere<br>Pressure<br>Cable | | |
| SVI20s | 20-second stroke volume index | | | |
A comprehensive list of parameters available while monitoring with the HemoSphere advanced monitor and a connected HemoSphere Technology Module, and ForeSight Oximeter Cable are listed below.
{15}------------------------------------------------
| Parameter | Description | Sub-System<br>Module<br>Used | Patient<br>Population | Hospital<br>Environment |
|-----------|---------------------------------------------------------------------------------|------------------------------------|------------------------|-----------------------------------------------|
| StO2 | Absolute regional hemoglobin<br>oxygen saturation of blood under<br>the sensors | ForeSight<br>Oximeter<br>Cable and | Adult and<br>Pediatric | Operating<br>Room,<br>Intensive Care<br>Unit, |
| ΔctHb | Relative change in Total<br>Hemoglobin | HemoSphere<br>Technology<br>Module | | Emergency<br>Room |
A comprehensive list of parameters available while monitoring with the HemoSphere advanced monitor and a connected HemoSphere ClearSight module are listed below.
| Parameter | Description | Sub-System<br>Module<br>Used | Patient<br>Population | Hospital<br>Environment |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|------------------------------------|-----------------------|-------------------------|
| CO/CI | Continuous Cardiac Output/<br>Continuous Cardiac Index | | | |
| DIA | Noninvasive arterial diastolic blood<br>pressure | | | |
| MAP | Noninvasive Mean Arterial Pressure | | | |
| PPV | pulse pressure variation | | | Operating<br>Room, |
| PR | Noninvasive Pulse rate | | | Intensive<br>Care Unit, |
| SV/ | Stroke Volume/ Stroke | HemoSphere<br>ClearSight<br>Module | Adult only | Emergency<br>Room |
| SVI | Volume Index | | | |
| SVR/ | Systemic Vascular Resistance Systemic | | | |
| SVRI | Vascular Resistance Index | | | |
| SVV | Stroke Volume Variation | | | |
| SYS | Systolic Blood Pressure | | | |
| dP/dt | Maximal slope of the arterial<br>pressure upstroke1 | | | |
| Eadyn | Dynamic Arterial Elastance1 | | | |
| HPI | Acumen Hypotension Prediction<br>Index1 | | | Operating<br>Room only |
| 1HPI parameters are available when using an Acumen IQ cuff and if the HPI feature is activated. | | | | |
| Note: CO/CI and SV/SVI are measured using a reconstructed brachial arterial waveform. All other monitored parameters use a<br>reconstructed radial arterial waveform. SVR/SVRI are derived from CO/CI and | | | | |
| MAP along with an entered or monitored CVP value. | | | | |
A comprehensive list of parameters available for adult patient populations while monitoring with the HemoSphere advanced monitor and both a connected HemoSphere ClearSight module and oximetry cable are listed below:
| Parameter | Description | Sub-System<br>Module<br>Used | Patient<br>Population | Hospital<br>Environment |
|-----------|------------------------------------------------------------------------|---------------------------------------------------------------------------|-----------------------|----------------------------------------------|
| DO2 | Oxygen Delivery | HemoSphere<br>ClearSight<br>Module and<br>HemoSphere<br>Oximetry<br>Cable | | Operating<br>Room,<br>Intensive Care<br>Unit |
| DO2I | Oxygen Delivery Indexed | | | |
| VO2 | Oxygen Consumption | | Adult only | |
| VO2e | Estimated Oxygen Consumption<br>when ScvO2 is being monitored | | | |
| VO2I | Oxygen Consumption Index | | | |
| VO2Ie | Estimated Oxygen Consumption<br>Index when ScvO2 is being<br>monitored | | | |
{16}------------------------------------------------
## Intended Use of Viewfinder Remote:
Viewfinder Remote is a mobile application, which provides supplemental remote near realtime display of hemodynamic data measured by a connected HemoSphere advanced monitoring platform. Viewfinder Remote allows clinicians to view continuous monitoring data and alarms/alerts remotely for multiple patients. All displayed data is generated by connected HemoSphere advanced monitoring platforms, and not by Viewfinder Remote. Viewfinder Remote is intended for use by clinicians as a supportive visual aid, and not as a replacement for in-person patient monitoring with connected HemoSphere advanced monitoring platforms.
Comparison of The existing HemoSphere Advanced Monitoring Platform, K203687 (cleared May 28, Technological 2021) consists of:
Characteristics with Predicate Devices:
- HemoSphere Advanced Monitor
- HemoSphere Swan-Ganz Module
- HemoSphere Oximetry Cable ●
- HemoSphere Pressure Cable
- HemoSphere Technology Module (previously known as "Tissue Oximetry ● Module")
- . HemoSphere ForeSight Oximeter Cable (previously known as "FORE-SIGHT ELITE Tissue Oximeter Module")
- HemoSphere ClearSight Module ●
- Acumen Hypotension Prediction Index (HPI) for Minimally Invasive and Non-. Invasive technology
- . Viewfinder Remote Mobile Application (for non-invasive parameters)
Per the FDA Guidance, Deciding When to Submit a 510(k) for a Change to an Existing Device (October 25, 2017), the modifications made to the HemoSphere Advanced Monitoring Platform as part of this submission are considered as technology, engineering, and performance changes. A summary of the new and modified technological characteristics of the subject device along with a reference to the predicate device(s) for those characteristics are provided below:
| New/Modified<br>Feature or<br>Technology | Predicate Device Chosen with Justification |
|---------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| HemoSphere<br>Advanced<br>Monitoring Platform | HemoSphere Advanced Monitoring Platform (K203687, cleared<br>May 28, 2021)<br>This is the base platform to which the new features, technology additions<br>or modifications (inclusive of miscellaneous GUI/ functionality<br>modifications) are being made and has similar intended use and<br>indications and hence is the Primary Predicate. |
| Tissue Hemoglobin<br>Software Feature<br>and Parameters | FORE-SIGHT ELITE Tissue Oximeter Module (K190270 cleared<br>October 21, 2019)<br>This device (smart cable) was chosen as a secondary predicate for the<br>new indication on the HemoSphere platform related to the hemoglobin<br>parameters (oxygenated hemoglobin, deoxygenated hemoglobin, and<br>their summation total hemoglobin, of blood under the sensors). The<br>algorithm for this parameter is directly being ported over from what was<br>cleared for the FORE-SIGHT ELITE Tissue Oximeter Module (also now<br>referred to as HemoSphere ForeSight Oximeter Cable (FSOC) as part<br>of rebranding that is occurring). There are GUI modifications being made<br>to integrate it with the HemoSphere Advanced Monitoring Platform. |
| Assisted Fluid<br>Management and<br>Acumen AFM Cable | Acumen Assisted Fluid Management (AFM) Software Feature<br>(DEN190029 granted November 13, 2020)<br>This software feature was chosen as the secondary predicate since the<br>Assisted Fluid Management algorithm and feature, as granted in this<br>DeNovo will be used on the HemoSphere Advanced Monitor. |
| HPI Smart Trends<br>and Alerts | HemoSphere Advanced Monitoring Platform (K203687 cleared May<br>28, 2021)<br>This device was chosen as the predicate since it is the base HemoSphere<br>platform for the HPI feature to which modifications are being made.<br>There are no changes to the core HPI algorithm, the behavior of the HPI<br>parameter display, nor the indications for use and intended use of the HPI<br>parameter due to this Smart Trends/ Smart Alerts modification. |
| Connectivity<br>Solutions | HemoSphere Advanced Monitoring Platform (K211465 cleared July 8,<br>2021)<br>This device was chosen as the predicate for the connectivity features that<br>allow the HemoSphere Advanced Monitor to connect to the Viewfinder<br>Remote mobile application. Modifications have been made to the<br>connectivity feature on HemoSphere Advanced Monitor to enable<br>remote software and cybersecurity patch updates and also allow the<br>remote display of minimally invasive and invasive hemodynamic<…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.