K213674 · Pulse Biosciences, Inc. · GEI · Sep 22, 2022 · General, Plastic Surgery
Device Facts
Record ID
K213674
Device Name
CellFX System
Applicant
Pulse Biosciences, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Sep 22, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CellFX® System is indicated for dermatological procedures requiring ablation and resurfacing of the skin. Specific Indication: The CellFX System with the CellFX Treatment Tip 2.5mm is indicated for the treatment of sebaceous hyperplasia (SH) in patients with Fitzpatrick skin types I-III.
Device Story
CellFX System is a platform technology for dermatological ablation and resurfacing. It consists of a console, handpiece, sterile treatment tips with stainless steel microneedles, and software. The device delivers nanosecond-duration electrical pulses (Nano-Pulse Stimulation) to targeted tissue, disrupting cellular function to induce cell death while sparing non-cellular structures. The physician selects treatment parameters via a touchscreen interface on the console. The system is used in clinical settings. By ablating targeted lesions, the device provides a therapeutic benefit for conditions like sebaceous hyperplasia. The output is the physical ablation of skin tissue, which the clinician observes to assess treatment progress and aesthetic improvement.
Clinical Evidence
Prospective, multi-center, randomized, split-face clinical study (n=59 subjects, 121 CellFX-treated lesions vs. 107 electrodessication-treated lesions). Primary safety endpoints (pigmentary changes/scarring at 60 days; skin textural changes at 30 days) showed CellFX non-inferiority to electrodessication. Primary effectiveness endpoint (GAIS improvement at 60 days) showed 76.9% responder rate for CellFX vs. 75.5% for electrodessication. Adverse events were mild, transient swelling (2.5% of lesions), resolving without intervention.
Technological Characteristics
Electrosurgical ablation device (21 CFR 878.4400). Delivers nanosecond electrical pulses (100-700 ns width, 300V-15kV amplitude) via stainless steel microneedle arrays. Frequency: 1-10 Hz. Power: 30W output. Sterile tips (EO sterilization). System dimensions: 132x46x53 cm; weight 54 kg. Connectivity: Touchscreen interface. No changes in materials or fundamental technology from predicate.
Indications for Use
Indicated for dermatological ablation and resurfacing; specifically sebaceous hyperplasia (SH) treatment in patients with Fitzpatrick skin types I-III.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
September 22, 2022
Pulse Biosciences, Inc. Punam Gollamudi Director US Regulatory Affairs 3957 Point Eden Way Hayward. California 94545
Re: K213674
Trade/Device Name: CellFX System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: August 25, 2022 Received: August 26, 2022
Dear Punam Gollamudi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213674
Device Name CellFX® System
#### Indications for Use (Describe)
The CellFX® System is indicated for dermatological procedures requiring ablation and resurfacing of the skin.
Specific Indication: The CellFX System with the CellFX Treatment Tip 2.5mm is indicated for the treatment of sebaceous hyperplasia (SH) in patients with Fitzpatrick skin types I-III.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| <span style="font-size:12px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:12px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments reqarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(K) SUMMARY
K213674
# 510(k) Summary
## I. SUBMITTER
| Applicant: | Pulse Biosciences, Inc.<br>3957 Point Eden Way<br>Hayward, CA 94545<br>Phone: (510) 906-4649 |
|-----------------|----------------------------------------------------------------------------------------------------------|
| Contact Person: | Punam Gollamudi<br>Director of US Regulatory Affairs<br>Pulse Biosciences, Inc.<br>Phone: (415) 305-4404 |
| Date Prepared: | September 19, 2022 |
### II. DEVICE INFORMATION
| Trade Name: | CellFX® System |
|-----------------------|-------------------------------------------------------------------|
| Regulation Number | 21 CFR § 878.4400 |
| Regulation Name: | Electrosurgical Cutting and Coagulation Device and<br>Accessories |
| Regulation Class: | Class II |
| Product Code: | GEI |
| Classification Panel: | General and Plastic Surgery |
#### III. PREDICATE DEVICE
K211444 - CellFX® System, Pulse Biosciences, Inc. No reference devices were used in this submission.
# IV. DEVICE DESCRIPTION
The Pulse Biosciences® CellFX® System is a proprietary platform technology. The CellFX System consists of the CellFX Console, CellFX Handpiece, CellFX Treatment Tips, and CellFX Software. The CellFX System delivers nanosecond duration electrical pulses that disrupt the function of cells leading to cell death, while sparing non-cellular tissue. The CellFX System delivers a series of timed, nanosecond electrical pulses (referred to as a "Cycle") to ablate and resurface tissue areas in dermatologic conditions.
{4}------------------------------------------------
The CellFX Console delivers short electric pulses at amplitudes up to 15 kV and pulse widths up to 700 ns. The electrical energy pulses are applied directly to targeted tissue using sterile Treatment Tips with stainless steel microneedles. The treatment parameters are selected by the user through a user interface on the Touchscreen Display of the CellFX Console.
# V. INDICATIONS FOR USE STATEMENT
The CellFX® System is indicated for dermatological procedures requiring ablation and resurfacing of the skin.
Specific Indication: The CellFX System with the CellFX Treatment Tip 2.5mm is indicated for the treatment of sebaceous hyperplasia (SH) in patients with Fitzpatrick skin types I-III.
# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The subject device has the same technological characteristics as the predicate device. The device design, materials, fundamental scientific technology, materials and processes for packaging and sterilization have not been changed from the predicate device.
| | Subject Device | Predicate Device (K211444) |
|----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | CellFX® System | CellFX® System |
| Prescription Only | Yes | Yes |
| Regulation<br>Number | 878.4400 | 878.4400 |
| Product Code | GEI | GEI |
| Indications for Use | | |
| Indications for Use<br>Statement | The CellFX System is indicated<br>for dermatological procedures<br>requiring ablation and resurfacing<br>of the skin.<br>Specific Indication: The CellFX<br>System with the CellFX Treatment<br>Tip 2.5mm is indicated for the<br>treatment of sebaceous<br>hyperplasia (SH) in patients with<br>Fitzpatrick skin types I-III. | The CellFX System is intended for<br>dermatological procedures<br>requiring ablation and resurfacing<br>of the skin. |
| | Subject Device | Predicate Device (K211444) |
| Technological Characteristics | | |
| Mechanism of<br>Action | Nano-Pulse Stimulation: ultrafast<br>nanosecond electrical pulses to the<br>targeted tissue via treatment tips<br>with an array of microneedles | Nano-Pulse Stimulation: ultrafast<br>nanosecond electrical pulses to the<br>targeted tissue via treatment tips<br>with an array of microneedles |
| Pulse Frequency | 1 Hz -10 Hz | 1 Hz -10 Hz |
| Pulse<br>Amplitude/Voltage | 300V to 15kV | 300V to 15kV |
| Pulse Width | 100 ns – 700 ns | 100 ns – 700 ns |
| Power Input | 100-240 VAC, 50/60 Hz, 2A | 100-240 VAC, 50/60 Hz, 2A |
| Power Output | 30 Watts | 30 Watts |
| Energy Density<br>Range | 2.5 x 2.5 mm – 30-60 mJ/mm³ | 1.5 x 1.5 mm – 110-190 mJ/mm³<br>2.5 x 2.5 mm – 60-110 mJ/mm³<br>5.0 x 5.0 mm – 45-85 mJ/mm³ |
| Use of Local<br>Anesthesia | Yes | Yes |
| Sterile Tip | EO Sterilization | EO Sterilization |
| Handpiece | Single Handpiece for 2-Row Tips | Single Handpiece for 2-Row Tips |
| System Physical Characteristics | | |
| Height | 132 cm | 132 cm |
| Length | 46 cm | 46 cm |
| Width | 53 cm | 53 cm |
| Weight | 54 kg | 54 kg |
| Power Input | 100-240 VAC, 50/60 Hz, 2A | 100-240 VAC, 50/60 Hz, 2A |
| Tip Characteristics | | |
| Treatment Tip | 2-Row | 2-Row |
| Microneedle<br>Configuration | | |
| 1.5 x 1.5 mm<br>2.5 x 2.5 mm<br>5.0 x 5.0 mm | N/A<br>2 rows of 3 or 4 needles per row<br>N/A | 2 rows of 3 needles per row<br>2 rows of 3 or 4 needles per row<br>2 rows of 4 or 6 needles per row |
| Tip Insertion Depth | 1 mm and 2 mm | 1 mm and 2 mm |
{5}------------------------------------------------
{6}------------------------------------------------
# VII. PERFORMANCE DATA
The following performance data was referenced or provided in support of the substantial equivalence determination.
#### Biocompatibility Testing
Biocompatibility testing was not required as there were no changes to the device compared to the predicate.
### Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing was not required as there were no changes to the device compared to the predicate.
#### Bench Testing
Performance testing was not required as there were no changes to the device compared to the predicate.
#### Software Verification and Validation Testing
Software verification and validation testing was not required as there were no changes to the device compared to the predicate.
#### Animal Study
Animal testing was not required to support the change to the Indications for Use.
### Clinical Study
A prospective, multi-center, randomized, comparison clinical study was conducted to support safety and effectiveness of the CellFX System for the treatment of sebaceous hyperplasia (SH). A total of 59 subjects treated with the 2.5mm Tip were mostly female (79.7%) and white (91.5%). All treatments were performed at a depth of 2mm on subjects with Fitzpatrick skin types I-III. A total of 121 SH lesions were randomized to the CellFX System and 107 to Electrodessication in a split-face study design.
### Safety Endpoints
The study was designed to evaluate co-primary safety and effectiveness endpoints. The first primary safety endpoint of lesions showing hyperpigmentation and scarring and assessed by the blinded Investigator at 60 days after the last treatment demonstrated CellFX to be non-inferior to Electrodessication. This safety endpoint found with 19.2% of lesions treated with CellFX and 12.1% of lesions treated with electrodessication having pigmentary changes or scarring. The second co-primary safety endpoint of skin textural changes (crusting, or other skin textural changes) as assessed by the blinded Investigator at 30 days after the last treatment showed rates of 6.8% of lesions treated with CellFX and 10.3% of lesions treated with Electrodessication.
{7}------------------------------------------------
# Effectiveness Endpoint
The primary effectiveness endpoint of improvement at 60-days post-last treatment as assessed "live" by the blinded site investigator using the 5-point GAIS (1 = much worse, 2 = worse, 3 = no change, 4 = improved, 5 = much improved, with scores of 4 or 5 classified as responders) showed rates of aesthetic appearance to be improved or much improved for 76.9% of lesions treated with CellFX and 75.5% of lesions treated with electrodessication.
# Adverse Events
There were 3 (2.5%, n=120) treatment-related non-serious adverse events (AEs) in 3 subjects with mild swelling at the lesion site treated with the CellFX 2.5mm treatment tip. None of the events required medical intervention and all resolved by the 7-day follow-up visit. One subject was instructed to use ice for swelling.
### Summary
Evaluation of primary safety and effectiveness endpoints showed CellFX to be noninferior to Electrodessication. Adverse events associated with the procedure were relatively mild, transient, and not unexpected for this type of procedure, demonstrating that the risks associated with CellFX and procedure are minimal. There were no device-related adverse events and no unanticipated adverse events. The clinical study demonstrated that the CellFX System using the CellFX Treatment Tip 2.5mm was safe and effective for the treatment of sebaceous hyperplasia (SH).
### CONCLUSION
The CellFX System has the same intended use, technological characteristics and principles of operation and similar indications for use as its predicate device. Clinical studies of the CellFX System have demonstrated the safety and effectiveness profile of the device in the intended population. Thus, the CellFX System is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.