4CIS BLACK MARLIN PLIF CAGE SYSTEM, 4CIS DOLPHIN TLIF CAGE SYSTEM

K213653 · Solco Biomedical Company India Private Limited · MAX · Dec 19, 2022 · Orthopedic

Device Facts

Record IDK213653
Device Name4CIS BLACK MARLIN PLIF CAGE SYSTEM, 4CIS DOLPHIN TLIF CAGE SYSTEM
ApplicantSolco Biomedical Company India Private Limited
Product CodeMAX · Orthopedic
Decision DateDec 19, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Spinal Fusion Cage System is an intervertebral body fusion devices intended for use to skeletally mature patients with Degenerative Disk Disease (DDD) of the lumbar spine with up to Grade 1 Spondylolisthesis at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. Spinal Fusion Cage System is indicated to be used with autologous bone graft to facilitate fusion and are intended to be used with supplemental fixation. The device is to be used in patients who have had six months of nonoperative treatment.

Device Story

Intervertebral body fusion cages (PLIF/TLIF) implanted via posterior or transforaminal approach; used singly or in pairs with supplemental fixation. Device restores disc height; maintains physiological lordotic angulation. Features hollow space for autologous bone graft to facilitate fusion; vertical square teeth on surfaces prevent subsidence and migration. Used by orthopedic/neurosurgeons in clinical settings. Provides mechanical stability for spinal fusion; benefits patients by addressing discogenic pain associated with DDD and spondylolisthesis.

Clinical Evidence

Bench testing only. No clinical data provided. Mechanical performance verified via ASTM F2077-14 and F2267-04 testing, including static axial compression, dynamic axial compression, and subsidence testing.

Technological Characteristics

Materials: PEEK Polymer (ASTM F2026) and Titanium Ti6Al4V Eli (ASTM F136). Tantalum markers (ASTM F560) for radiographic visualization. Design: Intervertebral cage with vertical square teeth for anchoring. Non-sterile, single-use. No software or electronic components.

Indications for Use

Indicated for skeletally mature patients with Degenerative Disk Disease (DDD) of the lumbar spine (L2-S1) with up to Grade 1 Spondylolisthesis at one or two contiguous levels. Requires six months of failed non-operative treatment. Used with autologous bone graft and supplemental fixation.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. December 19, 2022 SOLCO Biomedical Company India Private Limited Darshak Shah Director 5 & 6 Third Floor, B Jadav Chambers, Above Sales India Ahmedabad, Gujarat 380009 India Re: K213653 Trade/Device Name: 4CIS BLACK MARLIN PLIF CAGE SYSTEM, 4CIS DOLPHIN TLIF CAGE SYSTEM Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX Dated: December 7, 2022 Received: December 12, 2022 Dear Darshak Shah: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. # Katherine D. Kavlock -S for Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K213653 Device Name 4CIS BLACK MARLIN PLIF CAGE SYTEM 4CIS DOLPHIN TLIF CAGE SYSTEM #### Indications for Use (Describe) The Spinal Fusion Cage System is an intervertebral body fusion devices intended for use to skeletally mature patients with Degenerative Disk Disease (DDD) of the lumbar spine with up to Grade 1 Spondylolisthesis at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. Spinal Fusion Cage System is indicated to be used with autologous bone graft to facilitate fusion and are intended to be used with supplemental fixation. The device is to be used in patients who have had six months of nonoperative treatment. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------------------------|------------------------------------------------------------------| | <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the word "SOLCO" in large, bold, dark blue letters. The letters are sans-serif and appear to be a custom font. To the right of the word, there is a small red circle, which is likely part of the logo. The background is plain white. ## "510(K) SUMMARY" This summary of 510(k) substantial equivalence information isbeing submitted in accordacne with requirement of 21 CFR 807.92. | Submitter | SOLCO Biomedical Company India Private Limited<br>Survey No.1540, Beside Torrent Pharma, Village-Rajpur<br>Ahmedabad Mehsana Highway, Ta-Kadi, Dist-Mehsana<br>Gujarat- 382715, INDIA. | | | |--------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|---------------------------------------------------| | Contact Person | Darshak Shah<br>Director<br>SOLCO Biomedical Company India Private Limited<br>Phone : +91 98252 06091, Email: darshak@solco.co.in | | | | Submission Date | Thursday, December 15, 2022 | | | | Trade / Proprietary name | 4CIS BLACK MARLIN PLIF CAGE SYSTEM<br>4CIS DOLPHIN TLIF CAGE SYSTEM | | | | Classification Name | Intervertebral body Fusion Device | | | | Classification Code | MAX | | | | Regulatory Class | Class II | | | | Regulation Number | 21 CFR 888.3080 | | | | | 510K<br>Number | Trade or Proprietary or Model Name | Manufacturer | | Predicate Device | K221844<br>(Primary) | TDM Lumbar Interbody Fusion Cage<br>System | TDM Co. Ltd. Korea | | | K190563 | 4CIS® PEEK PLIF Cage,<br>4CIS® Pebble Beach PEEK PLIF Cage<br>4CIS® Torrey Pines PEEK TLIF Cage | Solco Biomedical Co., Ltd.<br>Republic of Korea | | | K202498<br>(Reference) | 4CIS SARA Spine System<br>4CIS VERTU Spine System | Solco Biomedical Company<br>India Private limited | | | K162358 | T-PAL Spacer System,<br>T-PAL Titanium Spacer System,<br>SYNFIXEvolution System | Synthes USA Products LLC | {4}------------------------------------------------ | | SOLCO BIOMEDICAL COMPANY INDIA PRIVATE LIMITED | K213653<br>PAGE NO 2 of 3 | |------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------| | | device intended for use as an intervertebral body fusion device in the lumbosacral region<br>(L2-S1) of the spine. It is implanted using a posterior or transforaminal approach and is<br>intended to be used singly or in pairs with supplemental fixation.<br>The Spinal Fusion Cage System is single component devices used to restore height of<br>disc and to facilitate lumbar intervertebral body fusion with maintaining physiological<br>lordotic angulation of lumbar spine. To allow maximum preservation and ensure ample<br>contact surfaces with bony endplate, a variety of shapes and sizes are available and each<br>device made withh PEEK Polymer has tantalum (ASTM F560) markers for ease of<br>visualization on radiographs. Vertical square teeth on the top and the bottom surface<br>prevent subsidence of the cage into the vertebral body while they increase the anchoring<br>and prevent slipping or expulsion. To make solid fusion of intervertebral body, hollow<br>space in the implant allows autologous bone graft material to be filled. The implant has<br>safety proven structure and material (Polyetheretherketone, ASTM F2026 & Titanium<br>Ti6Al4V Eli ASTM F136) to promote biological synostosis and assures mechanical safety<br>against load. | | | Indication for Use | The Spinal Fusion Cage System is an intervertebral body fusion devices intended for use<br>to skeletally mature patients with Degenerative Disk Disease (DDD) of the lumbar spine<br>with up to Grade 1 Spondylolisthesis at one or two contiguous levels from L2-S1. DDD is<br>defined as discogenic back pain with degeneration of the disc confirmed by history and<br>radiographic studies. Spinal Fusion Cage System is indicated to be used with autologous<br>bone graft to facilitate fusion and are intended to be used with supplemental fixation. The<br>device is to be used in patients who have had six months of non-operative treatment. | | | Comparison of<br>Technological<br>Characteristics<br>with the<br>Predicate Devices | The subject device and all the predicates have the same or similar indications for use<br>statements. The subject device is composed of the same material as the predicate<br>devices conforming to recognized industry standards for permanent implants and surgical<br>orthopedic instruments. All they have similar basic design features and functions as well<br>as those dimensions. The subject device and cited predicate devices are provided non-<br>sterile for single use only. The subject device demonstrated equivalent mechanical<br>performance to the cited predicate device under the same test conditions. | | | Performance Testing | Non-clinical testing was performed to demonstrate that the subject device is equivalent to<br>the predicate device. The following testing was performed in accordance with the ASTM | | | | F2077-14 and F2267-04:<br>- Static Axial Compression<br>- Dynamic Axial Compression<br>- Subsidence | | | Conclusion | The overall technology characteristics, material of construction, mechanical performance and design characteristics lead to the conclusion that subject device is substantially equivalent to legally marketed predicate devices for intended use,material composition, principles of operation, and design. | | {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for SOLCO. The logo is in blue, except for a red dot at the end. The font is sans-serif and bolded.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...