← Product Code [QKB](/productcode/QKB) · K213628

# VBrain (K213628)

_Vysioneer, Inc. · QKB · Dec 16, 2021 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K213628

## Device Facts

- **Applicant:** Vysioneer, Inc.
- **Product Code:** [QKB](/productcode/QKB.md)
- **Decision Date:** Dec 16, 2021
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** AI/ML, Software as a Medical Device

## Indications for Use

VBrain is a software device intended to assist trained medical professionals, during their clinical workflows of radiation therapy treatment planning, by providing initial object contours of known (diagnosed) brain tumors and organs at risk in the brain (i.e., the region of interest, ROI) on axial T1 contrast-enhanced brain MRI images. VBrain is intended to be used on adult patients only. VBrain uses an artificial intelligence algorithm (i.e., deep learning neural networks) to contour (segment) brain tumor and organs at risk (brain stem, eyes, optic nerves, optic chiasm) in the brain on MRI images for trained medical professionals' attention, which is meant for informational purposes only and not intended for replacing their current standard practice of manual contouring process. VBrain does not alter the original MRI image, nor does it intend to detect tumors for diagnosis. VBrain is intended only for generating Gross Tumor Volume (GTV) contours of brain metastases, meningiomas, and acoustic neuromas, and contours of organs at risk in the brain; it is not intended to be used with images of other brain tumors or other body parts. The user must know the tumor type when they use VBrain. VBrain also contains the automatic image registration feature to register volumetric medical image data. (e.g., MR, CT). It allows rigid image registration to adjust the spatial position and orientation of two images. Medical professionals must finalize (confirm or modify) the contours generated by VBrain, as necessary, using an external platform available at the facility that supports DICOM-RT viewing/editing functions, such as image visualization software and treatment planning system.

## Device Story

VBrain is a software application for radiation therapy treatment planning; assists clinicians in contouring brain tumors and organs at risk (OAR). Input: axial T1 contrast-enhanced brain MRI images. Operation: utilizes deep learning neural networks to perform automated segmentation of GTV (metastases, meningiomas, acoustic neuromas) and OARs; includes rigid image registration module for volumetric data (MR, CT). Output: DICOM-RT structure sets. Workflow: integrates with PACS; pulls scans, executes algorithm modules, sends results back to PACS. Clinical use: clinician reviews, confirms, or modifies generated contours on external DICOM-RT compatible visualization/treatment planning systems. Benefit: streamlines manual contouring process; improves efficiency in radiation therapy planning. Used in clinical settings by trained medical professionals.

## Clinical Evidence

Bench testing only. Software verification and validation testing performed in accordance with FDA guidance for Major Level of Concern software. Testing confirmed device performs within design specifications and meets all performance requirements.

## Technological Characteristics

AI-based (deep learning) software for image segmentation and rigid registration. Components: workflow management module, tumor contouring module, OAR contouring module, registration module. Connectivity: PACS network integration. Output: DICOM-RT structure sets. Software class: Major Level of Concern.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- VBrain ([K203235](/device/K203235.md))
- VBrain-OAR ([K212116](/device/K212116.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the words "U.S. FOOD & DRUG" on one line and "ADMINISTRATION" on the line below. The FDA logo is in blue.

Vysioneer Inc. % Vicki Lin Regulatory Specialist 33 Rogers. St., # 308 Cambridge. MA 02142

Re: K213628

Trade/Device Name: VBrain Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: QKB Dated: November 12, 2021 Received: November 17, 2021

Dear Vicki Lin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR

{1}------------------------------------------------

803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) k213628

Device Name V Brain

#### Indications for Use (Describe)

V Brain is a software device intended to assist trained medical professionals, during their clinical workflows of radiation therapy treatment planning, by providing initial object contours of known (diagnosed) brain tumors and organs at risk in the brain (i.e., the region of interest, ROI) on axial T1 contrast-enhanced brain is intended to be used on adult patients only.

V Brain uses an artificial intelligence algorithm (i.e., deep learning neural networks) to contour (segment) brain tumor and organs at risk (brain stem, eyes, optic chiasm) in the brain on MRI images for trained medical professionals' attention, which is meant for informational purposes only and not intended for replacing their current standard practice of manual contouring process. VBrain does not alter the original MRI image, nor does it intend to detect tumors for diagnosis. VBrain is intended only for generating Gross Tumor Volume (GTV) contours of brain metastases,

meningiomas, and acoustic neuromas, and contours of organs at risk in the brain; it is not intended to be used with images of other brain tumors or other body parts. The user must know the tumor type when they use VBrain.

V Brain also contains the automatic image registration feature to register volumetric medical image data. (e.g., MR, CT). It allows rigid image registration to adjust the spatial position and orientation of two images.

Medical professionals must finalize (confirm or modify) the contours generated by V Brain, as necessary, using an external platform available at the facility that supports DICOM-RT viewing/editing functions, such as image visualization software and treatment planning system.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D) | <span> <span style="font-size: 20px;">☑</span> </span> |
|----------------------------------------------|--------------------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C)  | <span> <span style="font-size: 20px;">☐</span> </span> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for Vysioneer. The logo consists of a stylized letter V above the word "VYSIONEER". The V is made up of two diagonal lines that do not quite meet at the bottom, creating a gap. The color of the V and the text is a muted purple.

# 510(k) Summary

This 510(k) Summary is submitted in accordance with 21 CFR 807.92

## Submitter

Vysioneer Inc.

33 Rogers St. #308, Cambridge, MA 02142

| Contact Person:        | Vicki Lin               |
|------------------------|-------------------------|
| Phone:                 | 609-865-8659            |
| Email:                 | vicki.lin@vysioneer.com |
| Date Summary Prepared: | November 12, 2021       |

## Device Name

| Trade Name:          | VBrain                                                              |
|----------------------|---------------------------------------------------------------------|
| Common Name:         | Radiological Image Processing Software for<br>Radiation Therapy     |
| Classification Name: | Medical image management and processing<br>system (21 CFR 892.2050) |
| Regulatory Class:    | II                                                                  |
| Product Code:        | QKB                                                                 |

## Predicate Devices

Predicate Device: VBrain

510(k) Holder/Submitter: Vysioneer Inc.

510(k) Number: K203235 (Cleared on 03/19/2021)

Predicate Device: VBrain-OAR

510(k) Holder/Submitter: Vysioneer Inc.

510(k) Number: K212116 (Cleared on 10/12/2021)

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for Vysioneer. The logo consists of the word "VYSIONEER" in a sans-serif font, with a horizontal line above it. Above the line is a stylized "V" shape, also in a sans-serif font. The color of the "V" and the word "VYSIONEER" is a shade of purple.

## Intended Use / Indications for Use

VBrain is a software device intended to assist trained medical professionals, during their clinical workflows of radiation therapy treatment planning, by providing initial object contours of known (diagnosed) brain tumors and organs at risk in the brain (i.e., the region of interest, ROI) on axial T1 contrast-enhanced brain MRI images. VBrain is intended to be used on adult patients only.

VBrain uses an artificial intelligence algorithm (i.e., deep learning neural networks) to contour (segment) brain tumor and organs at risk (brain stem, eyes, optic nerves, optic chiasm) in the brain on MRI images for trained medical professionals' attention, which is meant for informational purposes only and not intended for replacing their current standard practice of manual contouring process. VBrain does not alter the original MRI image, nor does it intend to detect tumors for diagnosis. VBrain is intended only for generating Gross Tumor Volume (GTV) contours of brain metastases, meningiomas, and acoustic neuromas, and contours of organs at risk in the brain; it is not intended to be used with images of other brain tumors or other body parts. The user must know the tumor type when they use VBrain.

VBrain also contains the automatic image registration feature to register volumetric medical image data. (e.g., MR, CT). It allows rigid image registration to adjust the spatial position and orientation of two images.

Medical professionals must finalize (confirm or modify) the contours generated by VBrain, as necessary, using an external platform available at the facility that supports DICOM-RT viewing/editing functions, such as image visualization software and treatment planning system.

# Device Description

VBrain is a software application system intended for use in the contouring (segmentation) of brain MRI images and in the registration of multi-modality images. The device consists of a workflow management module and 3 algorithm modules, which are the tumor contouring algorithm module, OAR contouring algorithm module, and registration algorithm modules can work independently, and yet can be integrated with each other.

The tumor contouring (segmentation) algorithm module consists of image preprocessing, deep learning neural networks, and postprocessing components, and is intended to contour brain tumor on the axial T1 contrast-enhanced MR images. It utilizes deep learning neural networks to generate contours for the detected/diagnosed brain tumors and export the results as DICOM-RT objects (using the RT Structure Set ROI Contour attribute, RTSTRUCT).

The OAR contouring (segmentation) algorithm module consists of image preprocessing, deep learning neural networks, and postprocessing components, and is intended to contour organs at risk in the brain on the axial T1 contrast-enhanced MR images. It utilizes deep learning neural networks to generate contours for the organs at risk in the brain and export the results as DICOM-RT objects (using the RT Structure Set ROI Contour attribute, RTSTRUCT).

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image contains the logo for Vysioneer. The logo consists of the word "VYSIONEER" in a sans-serif font, with a stylized "V" shape above it. The color of the text and the "V" shape is a dark purple.

The registration algorithm module registers volumetric medical image data (e.g., MR, CT). It allows rigid image registration to adjust the spatial position and orientation of two images.

The workflow management module is configured to work on a PACS network. Upon user's request, it will pull patient scans from a PACS, and it will trigger a predefined workflow, in which different algorithm modules are executed to generate the DICOM output. The DICOM output of a workflow can be sent back to the PACS.

### Comparison of Technological Characteristics with the Predicate Devices

The intended use and features of the modified device VBrain are a combination of those of the two predicate devices. The modified device uses VBrain's (K203235) tumor contouring algorithm, VBrain-OAR's (K212116) organs at risk contouring algorithm, registration algorithm and application. Each algorithm works independently and does not interfere with the others, yet they could be integrated in a predefined pipeline to work consecutively.

Same as the predicate devices, the modified device, VBrain is an AI-based (deep learning) software devices intended to be used in the workflow of radiation therapy treatment planning by providing tools to automatically segment/contour tumors and organs at risk on images as well as perform image registration (image fusion). VBrain uses the identical Deep Neural Networks and registration method as the predicates.

Minor modifications are made to increase convenience of use and improve functionality. Risk analysis and V&V testing were performed, and the results confirm that the modifications do not change the device's safety and effectiveness.

The modified device is substantially equivalent to the predicate devices because it has identical intended use, technological characteristics, and principles of operation as the predicates.

## Performance Data

Regression testing was performed to verify and validate the changes of V Brain in accordance with FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" for software devices identified as Major Level of Concern related to radiation therapy treatment planning.

The protocol, methods and acceptance criteria of software verification and validation testing used to evaluate the changes were not modified from those used in the predicate submission. The acceptance criteria and a summary of the results were provided for each test. VBrain passed all V&V testing, performance requirements and specifications are met.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for Vysioneer. The logo consists of a stylized letter V in a dark blue color, with the word "VYSIONEER" written in a sans-serif font below it. The logo is simple and modern.

## Substantially Equivalent (SE) Conclusion

VBrain is considered substantially equivalent to the predicate devices.

Verification and validation testing and hazard analysis demonstrate that VBrain performs within its design specifications and is as safe and effective as the predicate. The minor modifications of the device do not introduce any new potential risks. Based on the information presented in these 510(k) premarket notifications, it could be concluded that the subject device VBrain is as safe and effective as the predicate devices, with the same intended use, technological characteristics, and principles of operation.

---

**Source:** [https://fda-staging.innolitics.com/device/K213628](https://fda-staging.innolitics.com/device/K213628)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
