The Kogent Disposable Copper Bipolar Forceps are a single use product sold sterile and are intended for use in electrosurgery for coagulation of tissue. The Kogent Disposable Copper Irrigating Bipolar Forceps are a single use product sold sterile and are intended for use in electrosurgery for coagulation and irrigation of tissue. The Kogent Disposable Copper Illuminating Bipolar Forceps are a single use product sold sterile and are intended for use in electrosurgery for coagulation and to supply light for surgical procedures.
Device Story
Disposable bipolar forceps for general surgical procedures; used to grasp, manipulate, coagulate, irrigate, and illuminate tissue. Device connects to electrosurgical generator via bipolar cable; coagulation activated by footswitch. Irrigating models deliver sterile fluid to tips; illuminating models use fibers to deliver light to surgical site. Used in clinical/surgical settings by physicians. Provides coagulation, irrigation, and illumination to assist surgical visualization and hemostasis; benefits patient through precise tissue management and reduced cross-contamination risk due to single-use design.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including electrical safety (IEC 60601-1, 60601-2-2), biocompatibility (ISO 10993-5, -10, -11), sterilization validation (ISO 11135), and performance testing (thermal conductivity, non-stick ability, flow rate, illumination output, and thermal effects on tissue).
Technological Characteristics
Bayonet-style bipolar forceps; silver-plated copper base with PVDF insulation. Energy source: electrosurgical generator. Features: irrigation channels, illumination fibers. Sterilization: Ethylene Oxide (ETO) to 10^-6 SAL. Standards: IEC 60601-1, IEC 60601-2-2, IEC 60601-1-6, ISO 10993, ISO 11135, ASTM F1980.
Indications for Use
Indicated for use in electrosurgery for coagulation of tissue, irrigation of tissue, and/or illumination of the surgical site during surgical procedures. Single-use, sterile, prescription-only devices.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 7, 2022
Katalyst Surgical, LLC Liz Morgan Ouality and Regulatory Specialist 754 Goddard Ave Chesterfield, Missouri 63005
Re: K213610
Trade/Device Name: Kogent Disposable Copper Forceps, Kogent Disposable Irrigating Copper Forceps, Kogent Disposable Illuminating Copper Forceps Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: September 6, 2022 Received: September 6, 2022
Dear Liz Morgan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known) K213610
Device Name Kogent Disposable Copper Bipolar Forceps Kogent Disposable Irrigating Copper Bipolar Forceps Kogent Disposable Illuminating Copper Bipolar Forceps
Indications for Use (Describe)
The Kogent Disposable Copper Bipolar Forceps are a single use product sold sterile and are intended for use in electrosurgery for coagulation of tissue. The Kogent Disposable Copper Irrigating Bipolar Forceps are a single use product sold sterile and are intended for use in electrosurgery for coagulation and irrigation of tissue. The Kogent Disposable Copper Illuminating Bipolar Forceps are a single use product sold sterile and are intended for use in electrosurgery for coagulation and to supply light for surgical procedures.
Type of Use (Select one or both, as applicable)
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------|
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Manufacturer: | Katalyst Surgical, LLC<br>754 Goddard Avenue<br>Chesterfield, MO 63005<br>636-536-5950 (phone)<br>636-787-0603 (fax) |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Liz Morgan<br>Katalyst Surgical, LLC<br>636-536-5950 (phone)<br>636-787-0603(fax)<br>e.morgan@katalystsurgical.com |
| Date Prepared: | November 4, 2021 |
| 510(k) Number: | K213610 |
| Device Trade Name: | Kogent Disposable Copper Bipolar Forceps<br>Kogent Disposable Irrigating Copper Bipolar Forceps<br>Kogent Disposable Illuminating Copper Bipolar Forceps |
| Common Name: | Bipolar Forceps |
| Classification: | 21 CFR 878.4400 |
| Classification Name: | Electrosurgical cutting and coagulation device and accessories |
| Class: | II |
| Product Code: | GEI |
Indications for Use:
The Kogent Disposable Copper Bipolar Forceps are a single use product sold sterile and are intended for use in electrosurgery for coagulation of tissue. The Kogent Disposable Copper Irrigating Bipolar Forceps are a single use product sold sterile and are intended for use in electrosurgery for coagulation and irrigation of tissue. The Kogent Disposable Copper Illuminating Bipolar Forceps are a single use product sold sterile and are intended for use in electrosurgery for coagulation and to supply light for surgical procedures.
{4}------------------------------------------------
## Device Description:
These devices are disposable bipolar forceps, designed for single use in general surgical procedures. They require connection with a suitable bipolar cable to the bipolar output of an electrosurgical generator. These forceps are designed to grasp, manipulate, coagulate, irrigate and illuminate, when applicable, selected tissues. The irrigation tube is designed to carry sterile fluid to the tips of the instrument. The illumination style forceps contain illumination fibers on each tine to deliver light to the surgical site. Coagulation is achieved using electrosurgical energy generated by the electrosurgical generator and activated by a footswitch. The forceps are provided sterile by ethylene oxide and in sterile pouches.
### Predicate Device:
The Kogent Disposable Copper Bipolar Forceps was shown to be substantially equivalent to the previously cleared devices: Synergetics Disposable Spetzler Malis Dual Irrigating Bipolar Forceps K110924, Synergetics Spetzler Malis Dual Illuminating Bipolar Forceps K160103, and Kogent Bipolar Forceps K123172.
| | Subject Device | Predicate Device |
|-------------------------------|------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|
| Device Name | Kogent Disposable Copper<br>Bipolar Forceps | Kogent Bipolar Forceps |
| Company | Katalyst Surgical, LLC | Katalyst Surgical, LLC |
| K Number | K213610 | K123172 |
| Classification | Class II | Class II |
| Product Code | GEI | GEI |
| Indications for Use | Single use product sold sterile and are intended for use in electrosurgery for coagulation of tissue | Single use product sold sterile and are intended for use in electrosurgery for coagulation of tissue |
| Forceps Design | Bayonet Style | Bayonet Style |
| Patient Contact Material | Silver Plated copper base with PVDF insulation | Silver Plated aluminum base with PVDF insulation |
| Electrical Safety Testing | IEC 60601-1<br>IEC 60601-1-6<br>IEC 60601-2-2 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-2 |
| Size Offerings Overall Length | 6.25, 7, 8, and 9 inches | 7, 8, and 9 inches |
| Size Offerings Distal Tip | 0.5mm, 1.0mm, 1.5mm | 0.5mm, 1.0mm, 1.5mm |
| Packaging Configuration | Double Tyvek/Poly pouches | Double Tyvek/Poly pouches |
| Biocompatibility | Biocompatible per test reports | Biocompatible per test reports |
| Method of Sterilization | ETO | ETO |
| Sterility Assurance Level | 10-6 | 10-6 |
#### Comparison of Technical Characteristics:
{5}------------------------------------------------
| | Subject Device | Predicate Device |
|-------------------------------|---------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|
| Device Name | Kogent Disposable Copper<br>Irrigating Bipolar Forceps | Synergetics Disposable<br>Spetzler Malis Dual<br>Irrigating Bipolar Forceps |
| Company | Katalyst Surgical, LLC | Synergetics, Inc |
| K Number | K213610 | K110924 |
| Classification | Class II | Class II |
| Product Code | GEI | GEI |
| Indications for Use | Single use product sold<br>sterile and are intended for<br>use in electrosurgery for<br>coagulation and irrigation of<br>tissue | Sterile single use, for use in<br>electrosurgery for<br>coagulation and irrigation of<br>tissue |
| Forceps Design | Bayonet Style | Bayonet Style |
| Patient Contact Material | Silver Plated copper base<br>with PVDF insulation | Silver Plated aluminum base<br>with PVDF insulation |
| Electrical Safety Testing | IEC 60601-1<br>IEC 60601-1-6<br>IEC 60601-2-2 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-2 |
| Size Offerings Overall Length | 7,8 and 9 inch handles | 7,8 and 9 inch handles |
| Size Offerings Distal Tip | 0.5mm, 1.0mm, 1.5mm | 0.5mm, 1.0mm, 1.5mm |
| Packaging Configuration | Double Tyvek/Poly pouches | Rigid PETG Tray with<br>Tyvek 1073B Lid |
| Biocompatibility | Biocompatible per test<br>reports | Biocompatible per test<br>reports |
| Method of Sterilization | ETO | ETO |
| Sterility Assurance Level | 10-6 | 10-6 |
| | Subject Device | Predicate Device |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|
| Device Name | Kogent Disposable Copper<br>Illuminating Bipolar<br>Forceps | Synergetics Spetzler Malis<br>Dual Illuminating Bipolar<br>Forceps |
| Company | Katalyst Surgical, LLC | Synergetics, Inc |
| K Number | K213610 | K160103 |
| Classification | Class II | Class II |
| Product Code | GEI | GEI |
| Indications for Use | Single use product sold<br>sterile and are intended for<br>use in electrosurgery for<br>coagulation and to provide<br>light for surgical procedures | Single use devices sold<br>sterile and are intended for<br>use in electrosurgery for<br>coagulation of tissue and to |
{6}------------------------------------------------
| | | supply light for surgical<br>procedures |
|----------------------------------|---------------------------------------------------|-----------------------------------------------------|
| Forceps Design | Bayonet Style | Bayonet Style |
| Patient Contact Material | Silver Plated copper base<br>with PVDF insulation | Silver Plated aluminum base<br>with PVDF insulation |
| Electrical Safety Testing | IEC 60601-1<br>IEC 60601-1-6<br>IEC 60601-2-2 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-2 |
| Size Offerings Overall<br>Length | 7,8 and 9 inch handles | 8 and 9 inch handles |
| Size Offerings Distal Tip | 0.5mm, 1.0mm, 1.5mm | 0.5mm, 1.0mm, 1.5mm |
| Packaging Configuration | Double Tyvek/Poly pouches | Rigid PETG Tray with<br>Tyvek 1073B Lid |
| Biocompatibility | Biocompatible per test<br>reports | Biocompatible per test<br>reports |
| Method of Sterilization | ETO | ETO |
| Sterility Assurance Level | 10-6 | 10-6 |
### Performance Data:
Clinical Testing was not required to prove substantial equivalence. Bench testing performed between the candidate device and the predicate devices that the Kogent Copper Bipolar Forceps are substantially equivalent to predicate devices.
The product required the following other tests which were completed successfully:
#### Medical Electrical Equipment:
IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012, IEC 60601-1:2005/AMD2:2020 IEC 60601-2-2: 2017 for use in conjunction with IEC 60601-1:2005, COR1:2006, COR 2:2007, AMD1:2012 or IEC 60601-1:2012 IEC 60601-1-6:2010, AMD1:2013 for use in conjunction with IEC 62366:2007, AMD1:2014 and IEC 60601-1:2005, COR1:2006, COR2:2007, AMD1:2012
#### Biocompatibility:
ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5: Tests for In Vitro Cytotoxicity
ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests for Irritation and Skin Sensitization
ISO 10993-11:2017 Biological Evaluation of Medical Devices - Part 11: Tests for Systemic Toxicity
#### Sterilization:
ISO 11135:2014 Sterilization of health-care products-Ethylene oxide-Requirements for the development, validation and routine control of a sterilization process for medical devices
#### Shelf Life:
{7}------------------------------------------------
# K213610
ASTM F1980 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices
Bench Testing Thermal Conductivity Non-Stick Ability Average Flow Rate Average Illumination Output Average Temperature Rise due to Illumination Output Thermal Effects on Tissue
## Conclusion
The Kogent Disposable Copper Bipolar Forceps were shown to be substantially equivalent to previously cleared devices with respect to intended use, indications for use, technological characteristics, performance characteristics, and biocompatibility.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.