Surgical Mask

K213601 · Wuhan Zonsen Medical Products Co., Ltd. · FXX · Dec 16, 2022 · General, Plastic Surgery

Device Facts

Record IDK213601
Device NameSurgical Mask
ApplicantWuhan Zonsen Medical Products Co., Ltd.
Product CodeFXX · General, Plastic Surgery
Decision DateDec 16, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4040
Device ClassClass 2

Indications for Use

The Surgical Masks (model: ZSFM 23, ZSFM 24) are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

Device Story

Surgical masks (models ZSFM 23, ZSFM 24) designed for infection control; protect patients and healthcare personnel from microorganisms, body fluids, and particulates. ZSFM 23: four-layer, flat-pleated, tie-on mask. ZSFM 24: three-layer, flat-pleated, ear-loop mask. Both feature malleable polyethylene nose clips for facial fit. Materials include polypropylene non-woven fabric (inner/outer layers), polypropylene melt-blown non-woven fabric (middle layers), and spandex (ear loops). Single-use, disposable, non-sterile. Used in clinical settings to reduce exposure to blood and body fluids.

Clinical Evidence

No clinical data. Bench testing only. Performance verified against ASTM F2100-20, ASTM F1862-17 (synthetic blood penetration), EN 14683:2019 (differential pressure), ASTM F2101-19 (BFE), ASTM F2299-17 (particulate filtration), and 16 CFR Part 1610-2008 (flammability). Biocompatibility confirmed via ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation/sensitization).

Technological Characteristics

Materials: Polypropylene non-woven fabric, polypropylene melt-blown non-woven fabric, spandex, malleable polyethylene wire. Design: Flat-pleated, 3 or 4 layers. Dimensions: 17.5 cm x 9.5 cm. Standards: ASTM F2100-20, ASTM F1862-17, ASTM F2101-19, ASTM F2299-17, EN 14683:2019, ISO 10993-5, ISO 10993-10, 16 CFR 1610. Non-sterile, single-use.

Indications for Use

Indicated for use by patients and healthcare personnel to protect against transfer of microorganisms, body fluids, and particulate material in infection control practices.

Regulatory Classification

Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible. (ii) Analysis and nonclinical testing must: (A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and (B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device. (iii) NIOSH approved under its regulation. (2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. December 16, 2022 Wuhan Zonsen Medical Products Co., Ltd Cynthia Ye General Manager No 8 Jinchao Road, Zhucheng Street, Xinzhou District Wuhan, Hubei 431000 China Re: K213601 Trade/Device Name: Surgical Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: November 11, 2022 Received: November 21, 2022 Dear Ms. Ye: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Brent Showalter -S Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K213601 Device Name Surgical Mask Indications for Use (Describe) The Surgical Masks (model: ZSFM 23, ZSFM 24) are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(K) Summary ### 510k Number: K213601 Revised date: May 12, 2022 ### A. Applicant: Wuhan Zonsen Medical Products Co., Ltd Address: No 8 Jinchao Road, Zhucheng Street, Xinzhou District, Wuhan, Hubei, China Contact Person: Cynthia Ye Tel: +86- 27-82737771 Submission Correspondent: Primary contact: Cynthia Ye Tel: +86-27-82737771 Email: Cynthia@zonsenmed.com Secondary contact: Linna Ye Tel: +86-27-82737772 Email: registration01@zonsenmed.com #### B. Device: Trade Name: Surgical Mask Common Name: Surgical Mask Model(s): ZSFM23, ZSFM24 Regulatory Information Classification Name: Surgical Mask Classification: Class II Product code: FXX Regulation Number: 878.4040 Review Panel: Surgical Apparel ## C. Predicate device: K211827 Level 3 Fluid Resistant Procedure/Surgical Mask Zhejiang Lanhine Medical Products LTD. #### D. Indications for use : The Surgical Masks (model: ZSFM 23, ZSFM 24) are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile. #### E. Device Description: {4}------------------------------------------------ #### Wuhan Zonsen Medical Products Co., Ltd No 8 Jinchao Road, Zhucheng Street, Xinzhou District, Wuhan, Hubei, China The proposed device (model: ZSFM 23) is blue color, four-layer, and flat pleated type mask, utilizing tie-on way for wearing, and it has nose clips design for fitting the face mask around the nose. The proposed device (model: ZSFM 24) is blue color, three-layer, and flat pleated type mask, utilizing ear loops way for wearing, and it has nose clips design for fitting the face mask around the nose. The proposed device (model: ZSFM 23) is manufactured with four layers, the inner and outer layers are made of polypropylene non-woven fabric, and the two middle layers are made of polypropylene Melt blown non-woven fabric. The proposed device is held in place over the users' mouth and nose by two tie-on bands welded to the face mask. The tie-on bands are made of polypropylene non-woven fabric. The nose clip is fixed between the layers of face mask to allow the user to fit the face mask around their nose, which is made of Malleable polyethylene wire. The proposed device (model: ZSFM 24) is manufactured with three layers, the inner and outer layers are made of polypropylene non-woven fabric, and the middle layer is made of polypropylene Melt blown non-woven fabric. The proposed device is held in place over the users' mouth and nose by two elastic ear loops welded to the face mask. The elastic ear loops are made of Spandex. The nose clip is fixed between the layers of face mask to allow the user to fit the face mask around their nose, which is made of Malleable polyethylene wire. The surgical masks will be provided in blue. The surgical masks are sold non-sterile and are intended to be single use, disposable devices. | Comparison<br>Component | Proposed Device | Predicate Device | Comparis<br>on | | |------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------|---------| | Manufacturer | Wuhan Zonsen Medical Products<br>Co., Ltd | Zhejiang Lanhine Medical<br>Products LTD. | / | | | 510K number | K213601 | K211827 | / | | | Device name | Surgical Mask | Level 3 Fluid Resistant<br>Procedure/Surgical Mask | / | | | Classification | Class II Device, FXX (21<br>CFR878.4040) | Class II Device, FXX (21<br>CFR878.4040) | Same | | | Comparison<br>Component | Proposed Device | Predicate Device | Comparis<br>on | | | Indications for use | The Surgical Masks (model:<br>ZSFM 23, ZSFM 24) are intended<br>to be worn to protect both the<br>patient and healthcare personnel<br>from transfer of microorganisms,<br>body fluids and particulate<br>material. These masks are<br>intended for use in infection<br>control practices to reduce the<br>potential exposure to blood and<br>body fluids. This is a single use,<br>disposable device(s), provided<br>non-sterile. | The Level 3 Fluid<br>Resistant Procedure/Surgical<br>Mask (model: 15604F, 15704F)<br>are intended to be worn to<br>protect both the patient and<br>healthcare personnel from<br>transfer of microorganisms,<br>body fluids and particulate<br>material. These face masks<br>are intended for use in infection<br>control practices to reduce the<br>potential exposure to blood and<br>body fluids. This a single<br>use, disposable device(s),<br>provided non-sterile. | Same | | | Ear loop model and<br>tie-on model | Model: ZSFM 23, Tie-On, Flat<br>Pleated, 4 layers<br>Model: ZSFM 24, Ear Loops, Flat<br>Pleated, 3 layers | Ear Loops, Tie-On, Flat<br>Pleated, 3 layers | Similar | | | | | | Same | | | | Outer<br>layer | Polypropylene non-woven fabric | Non-woven Fabric<br>(Polypropylene) | Same | | | Middle<br>layer(s) | Polypropylene Melt blown<br>non-woven fabric | Melton brown Fabric<br>(Polypropylene) | Same | | Mater<br>ial | Inner<br>layer | Polypropylene non-woven fabric | Non-woven Fabric<br>(Polypropylene) | Same | | | Nose clip | Malleable polyethylene | Polypropylene coating<br>iron | Similar | | | Tie-on<br>bands | Polypropylene non-woven fabric | Non-woven Fabric<br>(Polypropylene) | Same | | | Ear loops | Spandex | Polyurethane | Same | | Color | Blue | Blue | Same | | | Dimension | 17.5 cm +/- 0.5 cm | 17.5 cm +/- 0.5 cm | Same | | | | 9.5 cm +/- 0.5 cm | 9.5 cm +/- 0.5 cm | Same | | | OTC use | Yes | Yes | Same | | | Sterility | Non-Sterile | Non-Sterile | Same | | | Use | Single use; disposable | Single use; disposable | Same | | | ASTM F2100 Level | Model: ZSFM 23, Level 3 | Level 3 | Same | | | Comparison<br>Component | Proposed Device | Predicate Device | Comparis<br>on | | | | Model: ZSFM 24, Level 1 | | Similar | | | Biocompatibility | Meet ISO10993-5<br>and ISO10993-10 | ISO10993 | Same | | # F. Comparison with predicate device {5}------------------------------------------------ #### Wuhan Zonsen Medical Products Co., Ltd No 8 Jinchao Road, Zhucheng Street, Xinzhou District, Wuhan, Hubei, China {6}------------------------------------------------ ### G. Summary of Non-Clinical Test Non-clinical tests were conducted to verify that the proposed device met all design specifications as was similar to the predicate device. The test results demonstrated that the proposed device complies with the following standards and the requirements stated in the Guidance for Industry and FDA Staff. Surgical Masks - Premarket Notification [510(k)] Submission issued on March 5, 2004: - A ISO 10993-5: 2009 Biological Evaluation of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity: - A ISO 10993-10: 2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization: - A ASTM F2100-20, Standard Specification for Performance of Materials Used In Medical Face Masks: - > ASTM F1862-17, Standard Test Method for Resistance of Medical Face Masks To Penetration by Synthetic Blood; - > EN 14683: 2019, Requirements and Test Methods of Differential pressure of Medical Face Mask: - > ASTM F2101-19, Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) Of Medical Face Mask Materials, Using A Biological Aerosol of Staphylococcus Aureus; - A ASTM F2299-17, Standard test method for determining the initial efficiency of materials used in medical face masks to penetration by particulates using latex spheres; - > 16 CFR Part 1610-2008, Standard for the Flammability of clothing textiles; | | | | Test results of Proposed device | | | |-----------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|--------------------------------| | Performance | Purpose | Requirement | Surgical Mask<br>(model: ZSFM 23) | Surgical Mask<br>(model: ZSFM 24) | | | Bacterial<br>Filtration<br>Efficiency<br>ASTM<br>F2101-19 | Assess the<br>performance of<br>a mask to<br>penetration by a<br>prepared solution<br>with known<br>concentration of an<br>indicator bacterial<br>organism | Level 1: ≥95%<br><br>Level 3: ≥98% | PASS<br>3 non-consecutive lots<br>tested, using a sample<br>size of 32/ lot.<br>Lot 1: 99.9%;<br>Lot 2: 99.9%;<br>Lot 3: 99.9%; | PASS<br>3 non-consecutive lots<br>tested, using a sample<br>size of 32/ lot.<br>Lot 1: 99.9%;<br>Lot 2: 99.9%;<br>Lot 3: 99.9%; | | | | Differential<br>Pressure | Assess the<br>performance of | Level 1:<<br>5.0mmH2O/cm² | PASS<br>3 non-consecutive lots | PASS<br>3 non-consecutive lots | Table 2 - Performance Testing {7}------------------------------------------------ #### Wuhan Zonsen Medical Products Co., Ltd No 8 Jinchao Road, Zhucheng Street, Xinzhou District, Wuhan, Hubei, China | EN 14683:2019,<br>Annex C | a mask for<br>resistance to<br>air movement<br>through the<br>materials of the face<br>of the mask | Level 3:<<br>6.0mmH2O/cm² | tested, using a sample<br>size of 32/ lot.<br>Lot 1: 4.86;<br>Lot 2: 4.83;<br>Lot 3: 4.87; | tested, using a sample<br>size of 32/ lot.<br>Lot 1: 4.5;<br>Lot 2: 4.5;<br>Lot 3: 4.5; | |---------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Sub-micron<br>particulate<br>filtration<br>efficiency<br>ASTM<br>F2299-17 | Assess the<br>performance of<br>a mask to<br>penetration by<br>sub-micron<br>polystyrene latex<br>particles of 0.1<br>micron | Level 1: ≥95%<br>Level 3: ≥98% | PASS<br>3 non-consecutive lots<br>tested, using a sample<br>size of 32/ lot.<br>Lot 1: 99.9%;<br>Lot 2: 99.9%;<br>Lot 3: 99.9%; | PASS<br>3 non-consecutive lots<br>tested, using a sample<br>size of 32/ lot.<br>Lot 1: 98.5%;<br>Lot 2: 98.6%;<br>Lot 3: 98.6%; | | Synthetic Blood<br>Penetration<br>ASTM<br>F1862-17 | Assess the<br>performance<br>of a mask to<br>resistance to a<br>synthetic blood<br>preparation targeted<br>toward the mask at<br>a set pressure | Level 1: 80 mm Hg<br>Level 3: 160 mm Hg | PASS<br>3 non-consecutive lots<br>tested, using a sample<br>size of 32/ lot.<br>Lot 1: 32 out of 32<br>pass at 160 mm Hg;<br>Lot 2: 32 out of 32<br>pass at 160 mm Hg;<br>Lot 3: 32 out of 32<br>pass at 160 mm Hg; | PASS<br>3 non-consecutive lots<br>tested, using a sample<br>size of 32/ lot.<br>Lot 1: 32 out of 32 pass<br>at 80 mm Hg;<br>Lot 2: 32 out of 32 pass<br>at 80 mm Hg;<br>Lot 3: 32out of 32 pass<br>at 80 mm Hg; | | Flammability<br>Test 16 CFR<br>Part 1610-2008 | Assess the<br>resistance of a mask<br>to ignition | Class 1 | PASS<br>3 non-consecutive lots<br>tested, using a sample<br>size of 32/ lot.<br>Class 1 | PASS<br>3 non-consecutive lots<br>tested, using a sample<br>size of 32/ lot.<br>Class 1 | | Conclusion | / | / | Meet the requirement<br>for Level 3 barrier | Meet the requirement<br>for Level 1 barrier | #### Table 3 -Biological Specifications: | Performance | Requirement | Results (ZSFM23 & ZSFM24) | |-------------------------------|----------------|-----------------------------------------------------------------------------| | Cytotoxicity<br>EN ISO10993-5 | Non-Cytotoxic | PASS<br>Under the conditions of the study, the device is<br>non-cytotoxic. | | Irritation<br>EN ISO 10993-10 | Non-Irritating | PASS<br>Under the conditions of the study, the device is<br>non-irritating. | {8}------------------------------------------------ | Sensitization<br>EN ISO 10993-10 | Non-Sensitizing | PASS<br>Under the conditions of the study, the device is<br>non-sensitizing. | |----------------------------------|-----------------|------------------------------------------------------------------------------| |----------------------------------|-----------------|------------------------------------------------------------------------------| #### H. Clinical Performance No clinical study is included in this submission. #### I. Conclusion The conclusion drawn from the nonclinical tests demonstrates that the subject device in this 510(K) submission, the Surgical Mask (model: ZSFM23, ZSFM24), is as safe, as effective, and performs as well as the legally marketed predicate device cleared under K211827.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...