K213577 · Apex Medical Corp. · JOW · Jun 28, 2022 · Cardiovascular
Device Facts
Record ID
K213577
Device Name
VenAir, Sequential Compression System
Applicant
Apex Medical Corp.
Product Code
JOW · Cardiovascular
Decision Date
Jun 28, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5800
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The intended use of the VenAir Sequential Compression System (hereby referenced as "VenAir system") is to help prevent Deep Vein Thrombosis (DVT) and pulmonary embolism. The garments are single patient use - do not reuse. The VenAir system is intended for use only in professional healthcare facility environment by trained medical staff. It is not for use in the home healthcare environment. The VenAir system should be used as part of a prescribed plan of care.
Device Story
The VenAir Sequential Compression System is a pneumatic device that applies sequential and gradient pressure to a patient's limbs (calf, thigh, or foot) to increase venous blood flow and circulation. The system consists of a controller unit, tubing, and single-patient-use disposable garments. It is operated by trained medical staff in professional healthcare facilities. The device includes digital sensors to monitor garment connections, pressure output, and power supply. Operators use the control panel to manage settings and refer to the user manual for error codes. By enhancing blood circulation, the device aids in the prevention of DVT and pulmonary embolism in at-risk patients.
Clinical Evidence
No clinical test data was used to support the decision of substantial equivalence. The device was evaluated through bench testing, including biocompatibility (ISO 10993 series), software validation, electromagnetic compatibility, electrical safety, and performance testing (alarm function, battery, pressure accuracy, and venous refill detection).
Technological Characteristics
Sequential pneumatic compression system; Class I, Type BF electrical classification; IP23 ingress protection. Materials include polyester, PVC, polyamide, nylon, and polyurethane. Features digital sensors for pressure and connection monitoring. Powered by 100-240V AC or internal lithium battery. Complies with IEC 62304 for software and ISO 10993 for biocompatibility.
Indications for Use
Indicated for at-risk patients to help prevent Deep Vein Thrombosis (DVT) and pulmonary embolism in professional healthcare facilities.
Regulatory Classification
Identification
A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG ADMINISTRATION" in blue as well. The FDA is a federal agency responsible for regulating and supervising the safety of food, drugs, and other products.
Apex Medical Corp. Chieh Yang Quality Engineering Manager No. 9, Min Sheng St., Tu-Cheng, New Taipei City, 23679 Taiwan
Re: K213577
Trade/Device Name: VenAir, Sequential Compression System Regulation Number: 21 CFR 870.5800 Regulation Name: Compressible Limb Sleeve Regulatory Class: Class II Product Code: JOW Dated: May 31, 2022 Received: June 2, 2022
Dear Chieh Yang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Nicole Gillette Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K213577
#### Device Name
VenAir Sequential Compression System
#### Indications for Use (Describe)
The intended use of the VenAir Sequential Compression System (hereby referenced as "VenAir system") is to help prevent Deep Vein Thrombosis (DVT) and pulmonary embolism. The garments are single patient use - do not reuse. The VenAir system is intended for use only in professional healthcare facility environment by trained medical staff. It is not for use in the home healthcare environment. The VenAir system should be used as part of a prescribed plan of care.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{3}------------------------------------------------
VenAir Sequential Compression System Appendix 5 - 510(k) Summary
# 510(k) Summary
| 5.1 Type of submission: | Traditional |
|-------------------------|-------------------------------------------------------------------|
| 5.2 Date of summary: | May 31, 2022 |
| 5.3 Submitter: | Apex Medical Corp. |
| Address: | No. 9, Min Sheng St., Tu-Cheng, New Taipei<br>City, 23679, Taiwan |
| Phone: | +886-2-2268-5568 |
| Fax: | +886-2-2268-9662 |
| Contact: | Chieh Yang (meow.yang@apexmedicalcorp.com) |
| Job title: | Quality Engineering Manager |
# 5.4 Identification of the device:
| Proprietary/Trade name: | VenAir Sequential Compression System |
|---------------------------------|--------------------------------------|
| Classification product<br>Code: | JOW |
| Regulation number: | 870.5800 |
| Regulation description: | Compressible limb sleeve |
| Review panel: | Cardiovascular |
| Device class: | II |
# 5.5 Identification of the Predicate Device:
| Predicate device name: | Kendall SCD 700 Sequential Compression<br>Controller |
|------------------------------|------------------------------------------------------|
| Manufacturer: | Covidien LLC. |
| Classification product code: | JOW |
| Regulation number: | 870.5800 |
| Device class: | II |
| 510(k) number: | K120944 |
{4}------------------------------------------------
## 5.6 Indication for use
The intended use of the VenAir Sequential Compression System (hereby referenced as "VenAir system") is to help prevent Deep Vein Thrombosis (DVT) and pulmonary embolism. The garments are single patient use - do not reuse. The VenAir system is intended for use only in professional healthcare facility environment by trained medical staff. It is not for use in the home healthcare environment. The VenAir system should be used as part of a prescribed plan of care.
# 5.7 Device description
The VenAir Sequential Compression System is a sequential pneumatic compression system by applying sequential and gradient pressure to increase venous blood flow and circulation in at-risk patients to help prevent deep vein thrombosis and pulmonary embolism.
The product consists of the machine, tubing sets, and disposable (single patient use) garments (thigh, calf, and foot optional purchase) and focuses on compressing the limbs to enhance better blood circulation.
The system has digital sensors to check the garment connections, pressure output and power supply. On the other hand, operator can follow the user manual to check the error code and action as advised.
# 5.8 Non-clinical testing
A series of tests were performed to assess the safety and effectiveness of VenAir Sequential Compression System. All the test results demonstrate that subject device meets the requirements of its pre-defined acceptance criteria and intended use.
- . Shelf life test
- Biocompatibility test .
- Cytotoxicity test -
- -Skin irritation test
- Skin sensitization test -
Test results performed in biocompatibility test reports demonstrated that subject device complies with ISO 10993-1: 2018, ISO 10993-2: 2006, ISO 10993-5: 2009, ISO 10993-10: 2010, ISO 10993-12:2021, USP <87>: 2020, OECD404: 2015, OECD406: 1992.
{5}------------------------------------------------
- . Software Validation
- . Electromagnetic compatibility and electrical safety
- . Performance test
- Alarm function test report -
- Battery charging and discharging test report -
- Cycle time test report -
- -Pressure resistance test report
- Pressure accuracy test report -
- Air tightness and pull force test report -
- Venous refill detection function test report -
- · Usability test
All the test results demonstrate VenAir Sequential Compression System meets the requirements of its pre-defined acceptance criteria and intended use, and is substantially equivalent to the predicate device.
## 5.9 Clinical testing
No clinical test data was used to support the decision of substantial equivalence.
## 5.10 Substantial equivalence determination
VenAir Sequential Compression System submitted in this 510(k) file is substantially equivalent to the predicate device. Differences between the devices cited in this section do not raise any new issue of substantial equivalence.
| Item | Subject device | Predicate device | Substantial<br>equivalence<br>determination |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| Proprietary name | VenAir Sequential<br>Compression System | Kendall SCD 700 Sequential<br>Compression Controller | Similar.<br>Indication for use of both devices has slight different.<br>However, both devices are designed to help prevent deep vein thrombosis and |
| 510(k) No. | K213577 | K120944 | |
| Indication for use | The intended use of the VenAir<br>Sequential Compression System<br>(hereby referenced as “VenAir<br>system”) is to help prevent Deep<br>Vein Thrombosis (DVT) and<br>pulmonary embolism. The | The Kendall SCD TM 700 Sequential<br>Compression System (hereby<br>referenced as "Kendall SCD TM 700<br>Series") is designed to apply<br>intermittent pneumatic compression<br>to increase venous blood flow in | |
| Item | Subject device | Predicate device | Substantial<br>equivalence<br>determination |
| Proprietary name<br>510(k) No. | VenAir Sequential<br>Compression System<br>K213577 | Kendall SCD 700 Sequential<br>Compression Controller<br>K120944 | |
| | garments are single patient use - do<br>not reuse. The VenAir system is<br>intended for use only in<br>professional healthcare facility<br>environment by trained medical<br>staff. It is not for use in the home<br>healthcare environment. The<br>VenAir system should be used as<br>part of a prescribed plan of care. | at-risk patients in order to help<br>prevent deep vein thrombosis and<br>pulmonary embolism. The Kendall<br>SCD ™ 700 Series is a prescription<br>device for use in a clinical setting or<br>in the home. | pulmonary embolism. It<br>doesn't raise any new<br>issues of substantial<br>equivalence. |
| Type of use<br>Mechanism of<br>action | Prescription Use<br>The device is a sequential<br>pneumatic compression system by<br>applying sequential and gradient<br>pressure to increase venous blood<br>flow and circulation in at-risk<br>patients that aids in the prevention<br>of deep vein thrombosis (DVT) a<br>potentially life threatening<br>condition which can lead to<br>pulmonary embolism. | Prescription Use<br>The device is a sequential pneumatic<br>compression system by applying<br>sequential and gradient pressure to<br>increase venous blood flow and<br>circulation in at-risk patients that<br>aids in the prevention of deep vein<br>thrombosis (DVT) a potentially life<br>threatening condition which can lead<br>to pulmonary embolism. | Equivalence<br>Equivalence |
| Intended use<br>environment | Healthcare facilities. | Clinical, home use | Similar.<br>It does not raise new issue<br>of substantial equivalence. |
| Application | Non-invasive, external | Non-invasive, external | Equivalence |
| Anatomic location | Calf, thigh, foot | Leg (calf and thigh), foot | Equivalence |
| Item | Subject device | Predicate device | Substantial<br>equivalence<br>determination |
| Proprietary name | VenAir Sequential<br>Compression System | Kendall SCD 700 Sequential<br>Compression Controller | |
| 510(k) No. | K213577 | K120944 | |
| Dimension of<br>pump (mm) | 195×178×186 | 196x173x185 (free standing)<br>196x173x114 (place on foot board) | Different.<br>It does not raise new issue<br>of substantial equivalence. |
| Weight | 2.765 kg<br>2.977 kg (Inc. Battery) | 2.3 kg | Different.<br>It does not raise new issue<br>of substantial equivalence. |
| Power supply | 100-240 V A.C., 50/60 Hz | 100-240 V A.C., 50/60 Hz | Equivalence |
| Battery | Yes | Yes | Equivalence |
| Battery type | Lithium Battery | Lithium Battery | Equivalence |
| Electrical<br>classification | Class I, Type BF | Class I, Type BF | Equivalence |
| Ingress of water<br>protection | IP23 | IP23 | Equivalence |
| Control panel | Yes | Yes | Equivalence |
| Mode of operation | Continuous | Continuous | Equivalence |
| Set pressure | Calf, thigh: 45 mmHg<br>Foot: 130 mmHg | Leg (calf and thigh): 45 mmHg<br>Foot: 130 mmHg | Equivalence |
| Inflation time | Calf / thigh: 11 sec<br>Foot: 5 sec | Leg (calf and thigh): 11 sec<br>Foot: 5 sec | Equivalence |
| Deflation time | Calf / thigh:<br>Based on Venous Reflux Sensing<br>Technology<br>Foot: 60 sec | Leg (calf and thigh):<br>Based on Vascular refill detection<br>measurement technology<br>Foot: 60 sec | Both devices complied<br>with the IEC 62304.<br>Differences do not raise<br>new issue of substantial<br>equivalence. |
| Item | Subject device | Predicate device | Substantial<br>equivalence<br>determination |
| Proprietary name | VenAir Sequential<br>Compression System | Kendall SCD 700 Sequential<br>Compression Controller | |
| 510(k) No. | K213577 | K120944 | |
| Applied part | Calf garment,<br>Thigh garment,<br>Foot garment | Leg (calf and thigh) sleeves,<br>Foot cuff | Similar<br>It does not raise new issue<br>of substantial equivalence. |
| Main material of<br>applied part | Polyester, Polyvinyl Chloride,<br>Polyamide, Nylon, Polyurethane | Meet ISO10993-1 requirement | Similar<br>Both devices complied<br>with the ISO10993. It does<br>not raise new issue of<br>substantial equivalence. |
| Applied<br>part<br>chamber | Calf, thigh: 3<br>Foot: 1 | Leg (calf and thigh): 3<br>Foot: 1 | Equivalence |
| Operating<br>conditions | - | - | |
| Temperature | 5° C to 40° C | 10°C to 40°C | Different.<br>It does not raise new issue<br>of substantial equivalence. |
| Relative humidity | 30% to 75% non-condensing | 85% Maximum, non-condensing | |
| Atmospheric<br>pressure | 752 hPa to 1,060 hPa | 700 hPa to 1060 hPa | |
{6}------------------------------------------------
#### VenAir Sequential Compression System Appendix 5 - 510(k) Summary
{7}------------------------------------------------
# VenAir Sequential Compression System
{8}------------------------------------------------
#### VenAir Sequential Compression System Appendix 5 - 510(k) Summary
# 5.11 Similarity and difference
The VenAir Sequential Compression System has been compared with Kendall SCD 700 Sequential Compression Controller (K120944). The subject device has the similar indication for use, intended use environment, intended use anatomical location, set pressure and applied part as the predicate devices.
Although the dimension, weight, inflation time and deflation time are different between the subject device and predicate devices, a series testing were demonstrate that the differences do not raise any new issue of substantial equivalence.
{9}------------------------------------------------
In conclusion, the subject device has undergone a series of testing, and the results complied with the test requests; therefore, the difference between the subject device and the predicate device did not raise any problem of safe and effectiveness. The subject device is substantially equivalent to the predicate devices as it claims.
# 5.12 Conclusion
After analyzing non-clinical laboratory studies and safety testing data, it can be concluded that VenAir Sequential Compression System is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.