Replacement of the radial head for degenerative or post-traumatic disabilities presenting pain, crepitation, and decreased motion at the radio-humeral and/or proximal radio-ulnar joint with: joint destruction and/or subluxation visible on x-ray; and/or resistance to conservative treatment. Primary replacement after fracture of the radial head. Symptomatic sequelae after radial head resection. Revision following failed radial head arthroplasty.
Device Story
One-piece, stemmed radial head replacement implant; manufactured from wrought CoCrMo alloy. Used in orthopedic surgery for radial head arthroplasty. Implanted by surgeons to replace damaged radial head; restores joint function; alleviates pain/crepitation. System includes accessory instruments for site preparation and insertion. Subject submission updates labeling to MR Conditional status; no changes to device design, materials, or intended use.
Clinical Evidence
No clinical data; substantial equivalence supported by non-clinical bench testing regarding MR compatibility.
Technological Characteristics
One-piece stemmed prosthesis; wrought CoCrMo alloy (ASTM F1537). Head diameters: 19, 22, 25mm; stem diameters: 5.5–8.5mm; stem lengths: 21–24mm. MR Conditional. No software or electronic components.
Indications for Use
Indicated for patients requiring radial head replacement due to degenerative or post-traumatic disabilities (pain, crepitation, decreased motion), joint destruction/subluxation, fracture, symptomatic sequelae post-resection, or failed prior arthroplasty. Contraindications: infection, sepsis, osteomyelitis; relative: uncooperative patients, neurologic disorders, osteoporosis, metabolic bone disorders, osteomalacia, distant infection, rapid joint destruction, or marked bone loss.
Regulatory Classification
Identification
An elbow joint radial (hemi-elbow) polymer prosthesis is a device intended to be implanted made of medical grade silicone elastomer used to replace the proximal end of the radius.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 18, 2022
DePuy Synthes Quinn Mccarthy Senior Regulatory Affairs Specialist 1301 Goshen Parkway West Chester. Pennsylvania 19380
Re: K213563
Trade/Device Name: DePuy Synthes Radial Head Replacement System Regulation Number: 21 CFR 888.3170 Regulation Name: Elbow Joint Radial (Hemi-Elbow) Polymer Prosthesis Regulatory Class: Class II Product Code: KWI Dated: January 28, 2022 Received: January 31, 2022
Dear Quinn Mccarthy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Lixin Liu, Ph.D. Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K213563
Device Name
DePuy Synthes Radial Head Replacement System
Indications for Use (Describe)
•Replacement of the radial head for degenerative or post-traumatic disabilities presenting pain, crepitation, and decreased motion at the radio-humeral and/or proximal radio-ulnar joint with:
o joint destruction and/or subluxation visible on x-ray; and/or
- o resistance to conservative treatment.
- · Primary replacement after fracture of the radial head.
- · Symptomatic sequelae after radial head resection.
- · Revision following failed radial head arthroplasty.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a logo for DePuy Synthes, a company that specializes in orthopedics. The logo features a red graphic element on the left, followed by the company name in gray. Below the company name, it says "THE ORTHOPAEDICS COMPANY OF Johnson & Johnson" in smaller letters, with "Johnson & Johnson" in red.
# 510(k) Summary
| Sponsor | DePuy Synthes<br>Quinn McCarthy<br>1301 Goshen Parkway<br>West Chester, PA 19380<br>Phone: (732) 947 2251 |
|------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared | March 04, 2022 |
| Proprietary Name | DePuy Synthes Radial Head Replacement System – MR Conditional |
| Classification Name | Elbow joint radial (hemi elbow) polymer prosthesis |
| Classification | Class II<br>Regulation Number: 21 CFR §888.3170<br>Product Code: KWI |
| Predicate Device | Revolution Radial Head - K183618 |
| Device Description | This document is regarding the DePuy Synthes Radial Head<br>Replacement System-MR Conditional. The Radial Head Replacement<br>System Implant is a one-piece, stemmed radial head replacement. The<br>radial head is available in diameters of 19, 22 and 25mm. The radial<br>stems range from 5.5 to 8.5mm in diameter and from 21 –24mm in<br>length. The Radial Head is manufactured from wrought CoCrMo alloy.<br>The system is provided with a set of accessory instruments designed for<br>preparation of the implant site and insertion of the implants into bone.<br>The devices in scope of the subject submission are being reviewed for<br>MR Conditional labeling in addition to the previously cleared<br>indications for use. |
| Indications for Use | Replacement of the radial head for degenerative or post-<br>traumatic disabilities presenting pain, crepitation, and<br>decreased motion at the radio-humeral and/or proximal radio-<br>ulnar joint with: joint destruction and/or subluxation visible on x-ray; and/or resistance to conservative treatment. Primary replacement after fracture of the radial head. Symptomatic sequelae after radial head resection. |
| Image: THE ORTHOPAEDICS COMPANY OF Johnson & Johnson | |
| Contraindications | Absolute contraindications include: Infection, sepsis, osteomyelitis.<br><br>Relative contraindications include: Uncooperative patient or patient<br>with neurologic disorders who are incapable of following directions,<br>osteoporosis, metabolic disorders which may impair bone formation,<br>osteomalacia, distant foci of infections which may spread to the<br>implant site, rapid joint destruction, marked bone loss or bone<br>resorption apparent on roentgenogram |
| Comparison to<br>Predicate | The purpose of this submission is to add MR Conditional information to<br>the device labeling for the DePuy Synthes Radial Head Replacement<br>System - MR Conditional, K183618. The intended use and<br>technological characteristics of the devices remains unchanged. The<br>device name has been updated to reflect the current company name,<br>DePuy Synthes, previously cleared as the Revolution Radial Head and<br>now renamed to the DePuy Synthes Radial Head Replacement System -<br>MR Conditional.<br><br>The materials of the devices remain unchanged from K183618 (CoCrMo<br>alloy conforming to ASTM F1537).<br><br>The technical characteristics of the device remain unchanged from the<br>predicate K183618:<br>One-piece Head diameters of 19, 22 and 25mm Head Heights of 9-15, 10-16, 11-17mm Stem diameters 5.5 – 8.5mm Stem lengths 21-24mm Radial head thickness 9 – 17mm Articular depth 1.6 – 2.6mm |
| Non-Clinical<br>Performance Data | This submission includes non-clinical testing to support the conditional<br>safety of the DePuy Synthes Radial Head Implants in the Radial Head<br>Replacement System in the MR environment. The testing includes<br>magnetically induced displacement force (ASTM F2052-15) and torque<br>(ASTM F2213-17), radio frequency (RF) heating (ASTM F2182-11a), and<br>image artifacts (ASTM F2119-07 (2013)). The non-clinical performance<br>data demonstrate that the subject devices, when exposed to the MR<br>environment under specific MR conditions of use, raise no different<br>questions of safety or effectiveness. |
| Clinical<br>Performance Data | Clinical testing was not necessary for the determination of substantial<br>equivalence. |
| Substantial<br>Equivalence | The subject device has the same intended use compared to the<br>predicate device. The technological characteristics of the device remain<br>unchanged. |
| | It is concluded that the information provided herein supports<br>substantial equivalence of the subject devices to the predicate device,<br>K183618. |
{4}------------------------------------------------
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for DePuy Synthes, a company that specializes in orthopedics. The logo features a red symbol on the left, followed by the company name in gray. Below the company name, it says "THE ORTHOPAEDICS COMPANY OF Johnson & Johnson" in a smaller font.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.