K213536 · Deka Research and Development · QFG · Jul 25, 2023 · Clinical Chemistry
Device Facts
Record ID
K213536
Device Name
DEKA ACE Pump System
Applicant
Deka Research and Development
Product Code
QFG · Clinical Chemistry
Decision Date
Jul 25, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5730
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The DEKA ACE Pump System is intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin, ages 13 and above. The pump is able to reliably and securely communicate with compatible, digitally connected devices, including automated insulin dosing software, to receive, execute, and confirm commands from these devices. The pump is intended for single patient, home use and requires a prescription. The bolus calculator is indicated for use for aiding the user in determining the bolus insulin dosage for management of diabetes mellitus based on consumed carbohydrates, operator-entered blood glucose, insulin to carbohydrate ratio, target glucose values, and current insulin on board.
Device Story
DEKA ACE Pump System is an Alternate Controller Enabled (ACE) infusion pump for subcutaneous insulin delivery; manages diabetes mellitus in patients 13+. System components: durable pump (rechargeable Li-ion battery), single-use cassette (microfluidic valves, reservoir, Acoustic Volume Sensing), and wireless remote controller (touch display). Inputs: user-entered data (carbs, blood glucose, insulin parameters) and commands from compatible automated insulin dosing software/iCGM. Operation: pump executes basal/bolus delivery based on programmed rates or external controller commands; AVS technology monitors fluid volume. Used in home environment by patient. Output: insulin delivery; data logs (delivery, alarms, connectivity). Benefits: precise, automated insulin management; interoperability with iCGM/automated dosing systems. Healthcare providers use logs for clinical decision-making. Safety features: occlusion detection, failsafe communication protocols, and alarm systems.
Clinical Evidence
No clinical data was obtained in support of this premarket submission. Substantial equivalence is supported by bench testing, including basal/bolus accuracy, occlusion testing, fault insertion, reliability, drug compatibility, and environmental testing per IEC 60601 series standards.
Technological Characteristics
Microprocessor-controlled micro-dosing pump with Acoustic Volume Sensing (AVS) feedback. Powered by rechargeable lithium-ion battery. Ingress protection IP28. Materials compliant with ISO-10993. Sterilization via gamma radiation (ISO 11137-1). Connectivity: wireless communication with iCGM and iAGC. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11, IEC 60601-2-24, IEC 62304, ISO 14971.
Indications for Use
Indicated for subcutaneous insulin delivery for diabetes mellitus management in patients ages 13 and older. Requires prescription. Bolus calculator indicated for dosage determination based on carbohydrates, blood glucose, insulin sensitivity, insulin-to-carbohydrate ratio, target glucose, and insulin on board.
Regulatory Classification
Identification
An alternate controller enabled infusion pump (ACE pump) is a device intended for the infusion of drugs into a patient. The ACE pump may include basal and bolus drug delivery at set or variable rates. ACE pumps are designed to reliably and securely communicate with external devices, such as automated drug dosing systems, to allow drug delivery commands to be received, executed, and confirmed. ACE pumps are intended to be used both alone and in conjunction with digitally connected medical devices for the purpose of drug delivery.
Special Controls
Alternate controller enabled infusion pumps must comply with the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include the following:
(i) Evidence demonstrating that device infusion delivery accuracy conforms to defined user needs and intended uses and is validated to support safe use under actual use conditions.
(A) Design input requirements must include delivery accuracy specifications under reasonably foreseeable use conditions, including ambient temperature changes, pressure changes (
*e.g.,* head-height, backpressure, atmospheric), and, as appropriate, different drug fluidic properties.(B) Test results must demonstrate that the device meets the design input requirements for delivery accuracy under use conditions for the programmable range of delivery rates and volumes. Testing shall be conducted with a statistically valid number of devices to account for variation between devices.
(ii) Validation testing results demonstrating the ability of the pump to detect relevant hazards associated with drug delivery and the route of administration (
*e.g.,* occlusions, air in line, etc.) within a clinically relevant timeframe across the range of programmable drug delivery rates and volumes. Hazard detection must be appropriate for the intended use of the device and testing must validate appropriate performance under the conditions of use for the device.(iii) Validation testing results demonstrating compatibility with drugs that may be used with the pump based on its labeling. Testing must include assessment of drug stability under reasonably foreseeable use conditions that may affect drug stability (
*e.g.,* temperature, light exposure, or other factors as needed).(iv) The device parts that directly or indirectly contact the patient must be demonstrated to be biocompatible. This shall include chemical and particulate characterization on the final, finished, fluid contacting device components demonstrating that risk of harm from device-related residues is reasonably low.
(v) Evidence verifying and validating that the device is reliable over the ACE pump use life, as specified in the design file, in terms of all device functions and in terms of pump performance.
(vi) The device must be designed and tested for electrical safety, electromagnetic compatibility, and radio frequency wireless safety and availability consistent with patient safety requirements in the intended use environment.
(vii) For any device that is capable of delivering more than one drug, the risk of cross-channeling drugs must be adequately mitigated.
(viii) For any devices intended for multiple patient use, testing must demonstrate validation of reprocessing procedures and include verification that the device meets all functional and performance requirements after reprocessing.
(2) Design verification and validation activities must include appropriate design inputs and design outputs that are essential for the proper functioning of the device that have been documented and include the following:
(i) Risk control measures shall be implemented to address device system hazards and the design decisions related to how the risk control measures impact essential performance shall be documented.
(ii) A traceability analysis demonstrating that all hazards are adequately controlled and that all controls have been validated in the final device design.
(3) The device shall include validated interface specifications for digitally connected devices. These interface specifications shall, at a minimum, provide for the following:
(i) Secure authentication (pairing) to external devices.
(ii) Secure, accurate, and reliable means of data transmission between the pump and connected devices.
(iii) Sharing of necessary state information between the pump and any digitally connected alternate controllers (
*e.g.,* battery level, reservoir level, pump status, error conditions).(iv) Ensuring that the pump continues to operate safely when data is received in a manner outside the bounds of the parameters specified.
(v) A detailed process and procedure for sharing the pump interface specification with digitally connected devices and for validating the correct implementation of that protocol.
(4) The device must include appropriate measures to ensure that safe therapy is maintained when communications with digitally connected alternate controller devices is interrupted, lost, or re-established after an interruption (
*e.g.,* reverting to a pre-programmed, safe drug delivery rate). Validation testing results must demonstrate that critical events that occur during a loss of communications (*e.g.,* commands, device malfunctions, occlusions, etc.) are handled appropriately during and after the interruption.(5) The device design must ensure that a record of critical events is stored and accessible for an adequate period to allow for auditing of communications between digitally connected devices and to facilitate the sharing of pertinent information with the responsible parties for those connected devices. Critical events to be stored by the system must, at a minimum, include:
(i) A record of all drug delivery
(ii) Commands issued to the pump and pump confirmations
(iii) Device malfunctions
(iv) Alarms and alerts and associated acknowledgements
(v) Connectivity events (
*e.g.,* establishment or loss of communications)(6) Design verification and validation must include results obtained through a human factors study that demonstrates that an intended user can safely use the device for its intended use.
(7) Device labeling must include the following:
(i) A prominent statement identifying the drugs that are compatible with the device, including the identity and concentration of those drugs as appropriate.
(ii) A description of the minimum and maximum basal rates, minimum and maximum bolus volumes, and the increment size for basal and bolus delivery, or other similarly applicable information about drug delivery parameters.
(iii) A description of the pump accuracy at minimum, intermediate, and maximum bolus delivery volumes and the method(s) used to establish bolus delivery accuracy. For each bolus volume, pump accuracy shall be described in terms of the number of bolus doses measured to be within a given range as compared to the commanded volume. An acceptable accuracy description (depending on the drug delivered and bolus volume) may be provided as follows for each bolus volume tested, as applicable: Number of bolus doses with volume that is <25 percent, 25 percent to <75 percent, 75 percent to <95 percent, 95 percent to <105 percent, 105 percent to <125 percent, 125 percent to <175 percent, 175 to 250 percent, and >250 percent of the commanded amount.
(iv) A description of the pump accuracy at minimum, intermediate, and maximum basal delivery rates and the method(s) used to establish basal delivery accuracy. For each basal rate, pump accuracy shall be described in terms of the amount of drug delivered after the basal delivery was first commanded, without a warmup period, up to various time points. The information provided must include typical pump performance, as well as worst-case pump performance observed during testing in terms of both over-delivery and under-delivery. An acceptable accuracy description (depending on the drug delivered) may be provided as follows, as applicable: The total volume delivered 1 hour, 6 hours, and 12 hours after starting delivery for a typical pump tested, as well as for the pump that delivered the least and the pump that delivered the most at each time point.
(v) A description of delivery hazard alarm performance, as applicable. For occlusion alarms, performance shall be reported at minimum, intermediate, and maximum delivery rates and volumes. This description must include the specification for the longest time period that may elapse before an occlusion alarm is triggered under each delivery condition, as well as the typical results observed during performance testing of the pumps.
(vi) For wireless connection enabled devices, a description of the wireless quality of service required for proper use of the device.
(vii) For any infusion pumps intended for multiple patient reuse, instructions for safely reprocessing the device between uses.
Predicate Devices
t:slim X2 insulin pump with interoperable technology (DEN180058)
Submission Summary (Full Text)
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
## I Background Information:
A 510(k) Number
K213536
B Applicant
DEKA Research and Development
C Proprietary and Established Names
DEKA ACE Pump System
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| QFG | Class II | 21 CFR 880.5730 - Alternate Controller Enabled Pump | CH - Clinical Chemistry |
E Purpose for Submission:
New device
## II Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
The DEKA ACE Pump System is intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin, ages 13 and above. The pump is able to reliably and securely communicate with compatible, digitally connected devices, including automated insulin dosing software, to receive, execute, and confirm commands from these devices. The pump is intended for single patient, home use and requires a prescription.
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
The bolus calculator is indicated for use for aiding the user in determining the bolus insulin dosage for management of diabetes mellitus based on consumed carbohydrates, operator-entered blood glucose, insulin sensitivity, insulin to carbohydrate ratio, target glucose values, and current insulin on board.
## C Special Conditions for Use Statement(s):
This device is for prescription use only.
The DEKA ACE Pump System is not intended for anyone unable or unwilling to:
- Test blood glucose (BG) levels as recommended by their healthcare team.
- Maintain sufficient diabetes self-care skills.
- See their healthcare team regularly.
- Demonstrate adequate carbohydrate-counting skills (preferred, not required).
The user must not:
- Have difficulty seeing.
- Have difficulty hearing.
- Have deficient cognitive capabilities.
- Have physical impairments which would make operating the pump or controller difficult.
The device must be removed before Magnetic Resonance Imaging (MRI). Exposure to MRI may cause the pump and controller to move or lead to electric shocks and may result in severe injury.
The device is designed to use rapid-acting U-100 insulin. The DEKA ACE Pump System is compatible with the following U-100 insulins: Humalog.
## III Device Description
The DEKA ACE Pump is an Alternate Controller Enabled (ACE) infusion pump intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin, ages 13 and above. The pump is able to reliably and securely communicate with compatible, digitally connected devices, including automated insulin dosing software, to receive, execute, and confirm commands from these devices. The pump is intended for single patient, home use and requires a prescription. The system is able to be integrated with a Dexcom G6 interoperable Continuous Glycemic Controller (iCGM).
The DEKA ACE Pump System consists of the following components:
- Pump: A durable pump that incorporates fluid delivery algorithms and interfaces to an DEKA ACE Pump cassette, Remote Interface, iCGM, and iAGC. The pump is powered by a rechargeable lithium-ion battery.
- Cassette: A single use pumping cassette that combines microfluidic valves, a pump chamber, insulin reservoir, and Acoustic Volume Sensing (AVS) measurement chamber. The cassette interfaces to an DEKA ACE Pump and off-the-shelf infusion set.
- Remote Interface (Controller): A wireless controller that serves as the user interface to the DEKA ACE Pump system. This includes a large color touch display for ease of use.
K213536 - Page 2 of 11
{2}
The Controller is used in conjunction with the pump, Dexcom G6 CGM, and blood glucose monitor devices to program and monitor insulin delivery and blood glucose levels in your body.
## IV Substantial Equivalence Information:
A Predicate Device Name(s):
t:slim X2 insulin pump with interoperable technology
B Predicate 510(k) Number(s):
DEN180058
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K213536 | DEN180058 |
| --- | --- | --- |
| Device Trade Name | DEKA ACE Pump System | t:slim X2 insulin pump with interoperable technology |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | Intended for the subcutaneous delivery of insulin at variable rates for the management of diabetes mellitus in people requiring insulin. Intended to be interoperable with connected devices including CGMs and automated insulin dosing algorithms. | Same |
| Operating Environment | Home Use | Same |
| Insulin Delivery Modes | Both basal and bolus | Same |
| General Device Characteristic Differences | | |
K213536 - Page 3 of 11
{3}
| Specific Drug/Biologic Use | U-100 Insulin.
System has been tested with Humalog | U-100 Insulin.
System has been tested with NovoLog and Humalog. |
| --- | --- | --- |
| Intended Population | Persons with Diabetes Mellitus ages 13 and up | Persons with Diabetes Mellitus ages 6 and up |
| Insulin Basal Rate Delivery Range | 0 – 30 U/hour | 0 – 15 U/hour |
| Insulin Bolus Delivery Range | Programmable from 0.05 - 25.00 Units in 0.01 Unit increments. | 0.01 U at volumes greater than 0.05 U, Max bolus volume 25 U |
V Standards/Guidance Documents Referenced:
- FDA Guidance “Infusion Pumps Total Product Life Cycle” dated December 2, 2014
- FDA Guidance “Applying Human Factors and Usability Engineering to Medical Devices” dated February 3, 2016
- FDA Guidance “Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices” dated May 11, 2005
- FDA Guidance “Content of Premarket Submissions for Management of Cybersecurity in Medical Devices” dated October 2, 2014
- FDA Guidance “General Principles of Software Validation” dated January 11, 2002
- ISO 10993-1:2009 Biological Evaluation of Medical Devices: Evaluation and Testing
- ISO 10993-3:2014 Biological Evaluation of Medical Devices – Part 3: Tests for Genotoxicity, Carcinogenicity and Reproductive Toxicity
- ISO 10993-4:2002(A-2006) Biological Evaluation of Medical Devices – Part 4: Selection of Tests for Interactions with Blood
- ISO 10993-5:2009 Biological Evaluation of Medical Devices – Part 5: Tests for in Vitro Cytotoxicity
- ISO 10993-10:2010 Biological Evaluation of Medical Devices – Part 10: Tests for Irritation and Skin Sensitization
- ISO 10993-11:2006 Biological Evaluation of Medical Devices – Part 11: Tests for Systemic Toxicity
- ISO 14971 Second Edition 2007-03-01 Medical Devices – Application of Risk Management to Medical Devices
- IEC 60601-1:2005+AMD1:2012 Medical Electrical Equipment – Part 1: General Requirements for Basic Safety and Essential Performance
- IEC 60601-1-2 Ed. 4.0 b: 2014 Medical Electrical Equipment – Part 1-2: General Requirements for Basic Safety and Essential Performance – Collateral Standard: Electromagnetic Disturbances- Requirements and Tests
K213536 - Page 4 of 11
{4}
- IEC 60601-1-6 Ed. 3.1 b. 2013 Medical electrical equipment – Part 1-6: General requirements for basic safety and essential performance – Collateral standard: Usability
- IEC 60601-1-8 Ed. 2.1 b: 2012 Medical Electrical Equipment – Part 1-8: General Requirements for Basic Safety and Essential Performance – Collateral Standard: General Requirements, Tests and Guidance for Alarm Systems in Medical Electrical Equipment and Medical Electrical Systems
- IEC 60601-1-11 Issued: 2015/01/20 Ed.2 Medical Electrical Equipment – Part 1-11: General Requirements for Basic Safety and Essential Performance – Collateral Standard- Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment
- IEC 62304:2006 Medical Device Software – Software Life Cycle Processes
- IEC 62366-1 Ed. 1.0 Medical Devices – Part 1: Application Of Usability Engineering To Medical Devices
- ISO 11607-1:2006 Packaging for Terminally Sterilized Medical Devices – Part 1: Requirements for Materials, Sterile Barrier Systems and Packaging Systems
- ISO 11137-1: 2006 Sterilization of health care productions – Radiation – Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices.
- ISO 11137-2: 2013 Sterilization of health care products – Radiation – Part 2: Establishing the sterilization dose
## VI Performance Characteristics:
### A. Analytical Performance
#### 1. Basal delivery accuracy
To assess basal delivery accuracy, 32 pumps were tested by delivering at minimum, intermediate, and maximum basal rates (0.1, 1.0, and 30 U/hr). Sixteen of the pumps were unaged and 16 had been aged to simulate three years of regular use. For both aged and unaged pumps, eight pumps were tested with an unaged cassette, and eight with a cassette which underwent one year of accelerated aging. Delivery accuracy was assessed by pumping insulin into a container on a scale and measuring the weight of the liquid at 1h, 6h, and 12h intervals for minimum and intermediate rates (0.1 and 1.0 U/hr) and 1h and 6h intervals for maximum rate (30 U/hr).
The following tables report the typical basal performance (median) observed, along with the lowest and highest results observed for minimum, intermediate, and maximum basal rate settings for all pumps tested.
K213536 - Page 5 of 11
{5}
Table 1: Amount of fluid delivered after 1, 6, and 12 hours with 0.1 U/hr (minimum) basal rate setting
| 0.1 U/hr Basal Duration | 1 hour | 6 hours | 12 hours |
| --- | --- | --- | --- |
| Total expected delivery volume | 0.1 U | 0.6 U | 1.2 U |
| Median amount delivered [min, max] | 0.12 U
[0.09, 0.17] | 0.62 U
[0.57, 0.66] | 1.22 U
[1.16, 1.31] |
Table 2: Amount of fluid delivered after 1, 6, and 12 hours with 1 U/hr (intermediate) basal rate setting
| 1 U/hr Basal Duration | 1 hour | 6 hours | 12 hours |
| --- | --- | --- | --- |
| Total expected delivery volume | 1.0 U | 6.0 U | 12.0 U |
| Median amount delivered [min, max] | 1.02 U
[0.98, 1.09] | 6.05 U
[5.84, 6.22] | 12.07 U
[11.73, 12.33] |
Table 3: Amount of fluid delivered after 1, 6, and 12 hours with 30 U/hr (high) basal rate setting
| 30 U/hr Basal Duration | 1 hour | 6 hours |
| --- | --- | --- |
| Total expected delivery volume | 30 U | 180 U |
| Median amount delivered [min, max] | 30.16 U
[29.80, 30.61] | 181.05 U
[178.94, 184.46] |
2. Bolus delivery accuracy
To assess bolus delivery accuracy, 32 pumps were tested by delivering at minimum, intermediate, and maximum bolus volumes (0.05, 5.0, and 25 U). Each pump delivered 25 minimum and intermediate volume boluses and 7 maximum volume boluses interspersed during periods of active basal delivery. For each pump and cassette combination, a head height, orientation, infusion set, and insulin type was chosen. An approximately even distribution of parameters were used across the 32 combinations. Sixteen of the pumps were unaged and sixteen had been aged to simulate three years of regular use. For both aged and unaged pumps, eight pumps were tested with an unaged cassette, and eight with a cassette which underwent one year of accelerated aging.
Tables 4-6 below show the number (and %) of boluses within the specified range of each target bolus volume.
K213536 - Page 6 of 11
{6}
Table 4: Amount of fluid delivered after a 0.05 U bolus request (n=800 boluses)
| Units delivered after a 0.05 U bolus request (% of commanded units) | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | <0.0125 | 0.0125 - 0.0375 | 0.0375- 0.045 | 0.045- 0.0475 | 0.0475- 0.0525 | 0.0525- 0.0550 | 0.055- 0.0625 | 0.0625- 0.0875 | 0.0875- 0.125 | >0.125 |
| | (<25%) | (25-75%) | (75-90%) | (90-95%) | (95-105%) | (105-110%) | (110-125%) | (125-175%) | (175-250%) | (>250%) |
| Number and percent of boluses | 0/800 (0.0%) | 53/800 (6.6%) | 202/800 (25.3%) | 107/800 (13.4%) | 278/800 (34.8%) | 80/800 (10.0%) | 69/800 (8.6%) | 11/800 (1.4%) | 0/800 (0%) | 0/800 (0%) |
Table 5: Amount of fluid delivered after a 5.0 U bolus request (n=800 boluses)
| Units delivered after a 5.0 U bolus request (% of commanded units) | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | <1.25 | 1.25-3.75 | 3.75-4.50 | 4.50-4.75 | 4.75-5.25 | 5.25-5.50 | 5.50-6.25 | 6.25-8.75 | 8.75-2.50 | >12.50 |
| | (<25%) | (25-75%) | (75-90%) | (90-95%) | (95-105%) | (105-110%) | (110-125%) | (125-175%) | (175-250%) | (>250%) |
| Number and percent of boluses | 0/800 (0.0%) | 0/800 (0.0%) | 0/800 (0.0%) | 0/800 (0.0%) | 800/800 (100%) | 0/800 (0.0%) | 0/800 (0.0%) | 0/800 (0.0%) | 0/800 (0.0%) | 0/800 (0.0%) |
Table 6: Amount of fluid delivered after a 25 U bolus request (n=224 boluses)
| Units delivered after a 25.0 U bolus request (% of commanded units) | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | <6.25 | 6.25-18.75 | 18.75-22.5 | 22.5-23.75 | 23.75-26.25 | 26.25-27.5 | 27.5-31.25 | 31.25-43.75 | 43.75-62.5 | >62.5 |
| | (<25%) | (25-75%) | (75-90%) | (90-95%) | (95-105%) | (105-110%) | (110-125%) | (125-175%) | (175-250%) | (>250%) |
| Number and percent of boluses | 0/224 (0.0%) | 0/224 (0.0%) | 0/224 (0.0%) | 0/224 (0.0%) | 222/224 (99.1%) | 2/224 (0.9%) | 0/224 (0.0%) | 0/224 (0.0%) | 0/224 (0.0%) | 0/224 (0.0%) |
K213536 - Page 7 of 11
{7}
K213536 - Page 8 of 11
# 3. Occlusion detection
To assess occlusion detection performance and unintended bolus upon occlusion release performance, 32 pumps were tested. Sixteen of the pumps were unaged and 16 had been aged. For both aged and unaged pumps, eight pumps were tested with an unaged cassette, and eight with a cassette which underwent one year of accelerated aging.
The following table depicts the typical (average) time to occlusion detection for three different situations when using U-100 insulin.
Table 7: Timing of occlusion detection alarms
| | Typical time to occlusion detection | Maximum time to occlusion detection |
| --- | --- | --- |
| 2.0 U Bolus | 1 minute 52 seconds | 2 minutes 17 seconds |
| 1.0 U/hr Basal | 19 minutes 33 seconds | 27 minutes 22 seconds |
| 0.1 U/hr Basal | 1 hour 55 minutes | 3 hours 9 minutes |
After pumps alarmed, the occlusions were cleared and the total amount of fluid delivered was measured. Typical volumes were 0.25 U, and maximum volumes were 0.4 U.
# B. Other Supportive Instrument Performance Characteristics Data
## 1. Hazard Analysis
A comprehensive hazard analysis was reviewed, in which design inputs and outputs, risks, and risk mitigations for hardware and software associated with proper functioning of the insulin pump were reviewed. The sponsor performed a hazard analysis to account for the unique intended use, design elements, and risks of their ACE pump. This analysis identified hazards which could reasonably be anticipated to impact the proper use of the device, traced all identified risks to adequate design controls, and demonstrated that design features were appropriately implemented and validated.
## 2. Human Factors
Human factors validation tests were conducted with the DEKA ACE Pump System. All study participants received training that was consistent with the training that patients would receive with the commercial product. Usability evaluations assessed comprehension and usability of the device for critical device tasks. Results of the study demonstrated that the DEKA ACE Pump System is validated for its intended use.
{8}
K213536 - Page 9 of 11
3. Biocompatibility
Biocompatibility testing was performed per ISO 10993-1:2018, FDA Guidance Document: Use of International Standard ISO 10993-1 “Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process,” and FDA special controls for alternate controller enabled infusion pumps for the disposable cassette, pump and the disposable cassette packaging and dust cover. All endpoints were tested adequately, and results were acceptable.
4. Sterility
The sterility information was reviewed and found to be acceptable.
5. Insulin Compatibility and Stability
The DEKA ACE Pump system is compatible with U-100 insulin Humalog (insulin lispro) for up to 3 days.
6. Additional Bench Testing
In addition to the performance testing described above, mechanical testing, simulated use testing, and other device verification testing was conducted to demonstrate that the system meets its intended use and is safe, reliable, and all safety and reliability critical requirements have been adequately verified. Summaries for reliability, safety, and verification testing follow:
| Testing to Support System Reliability |
| --- |
| Durability of pumping mechanism |
| Acoustic Volume Sensor |
| Pump Button |
| Pump Alarm |
| Pump and cassette mating |
| Pump Battery |
| Charger |
| Power Button |
| Remote Interface |
| Testing to Support System Safety |
| --- |
| Environmental Safety Testing to 60601-1-11 |
| Safety and Essential Performance Testing to 60601-1 |
| Occlusion Detection Testing |
| Design specific tests: Volume Cross Checks, Inter and Intra Delivery Leaks |
| Alarms Testing |
| Data Handling Testing |
| Pump Activation and Deactivation Testing |
{9}
| Pump/Controller Connectivity Testing |
| --- |
| User Guide Testing |
| Insulin Delivery Verification Testing |
| Incidental Delivery |
| Worst Case Accuracy |
| Fault Insertion Testing |
| Testing to Support System Verification |
| --- |
| Environmental Safety Testing to 60601-1-11 |
| Safety and Essential Performance Testing to 60601-1 |
| Occlusion Detection Testing |
| Alarms Testing |
| Data Handling Testing |
| Pump Activation and Deactivation Testing |
| Pump/Controller Connectivity Testing |
| User Guide Testing |
| Insulin Delivery Verification Testing |
| Environmental conditions Testing |
# 7. Electromagnetic Compatibility and Wireless Coexistence
Electromagnetic compatibility, electromagnetic immunity and wireless coexistence testing was performed for the pump. All tests demonstrated that the device would perform as expected in the home healthcare environment.
# 8. Basic Safety and Essential Performance (Electrical Safety)
The sponsor provided verification evidence for compliance with the IEC 60601-1 and applicable collateral standards. Verification results support the finding of substantial equivalent for this device.
# 9. Data Logging
The sponsor provided a summary of pump and controller logging capability which enable the device to record critical events including insulin delivery, pump commands and confirmations, connectivity states, malfunctions, and alarms. These were reviewed and found to be adequate.
# 10. Interoperability
A plan and approach for interoperability were provided according to the FDA Guidance "Design Considerations and Pre-market Submission Recommendations for Interoperable Medical Devices - Guidance for Industry and Food and Drug Administration Staff" and determined to be adequate to support and clearly specify expectations, requirements, and interface specifications to potential interoperable devices. In addition, their plan covered their approach to working with connected device companies regarding contractual
K213536 - Page 10 of 11
{10}
approaches, interfaces for data communication and exchange, and post-market reporting procedures and responsibilities (e.g., who is responsible for investigating and reporting complaints, malfunctions, and adverse events).
The sponsor additionally provided validated software protocols intended to ensure secure, accurate, and reliable communication with digital interfacing devices, as well as failsafe design features to mitigate the risks associated with interruption of communication with digitally connected devices. These protocols were reviewed and found to be adequate.
## 11. Software and Cybersecurity
Detailed Information on software and cybersecurity of the device was reviewed and found acceptable.
## VII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
## VIII Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K213536 - Page 11 of 11
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.