K213534 · Resurgeonz, LLC · OLI · Mar 11, 2022 · General, Plastic Surgery
Device Facts
Record ID
K213534
Device Name
PHantom Trilogy
Applicant
Resurgeonz, LLC
Product Code
OLI · General, Plastic Surgery
Decision Date
Mar 11, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Resurgeonz pHantom Trilogy™ is indicated for use as a non-surgical, non-invasive aesthetic treatment for the temporary reduction of the circumference of waist, hips, thighs through the process of disrupting adipocyte cells, within the fat layer, for the release of fat and lipids from the targeted cells.
Device Story
Non-surgical, non-invasive fat-reducing low-level laser system; uses red and infrared light (660nm, 808nm, 980nm) to disrupt adipocyte cells in dermal/adipose tissue; releases fat/lipids. Device features LCD touchscreen; continuous operation mode; user-controlled treatment time, energy, pulse, and delay. Includes emergency safety cut-off switch. System uses multi-diode paddles (8, 10, or 12) with forced-air cooling; two smaller paddles available for hard-to-reach areas. Used in healthcare provider offices, Medi-Spas, and clinics by trained operators. Healthcare providers use device to provide aesthetic body contouring; patients benefit from temporary circumference reduction. Device powered by 120V, 60Hz; no wireless/external communication.
Clinical Evidence
Bench testing only. No clinical data provided. Performance testing confirmed device meets functional claims and safety requirements.
Technological Characteristics
Diode laser system; wavelengths 660nm (red), 808nm, 980nm (infrared). Power output up to 72mW per diode. ABS plastic console; TFT color touchscreen. Forced-air paddle cooling. Biocompatibility per ISO 10993 (Parts 1, 5, 10). Non-sterile. IPX0 rating. 120V/60Hz power source.
Indications for Use
Indicated for non-surgical, non-invasive aesthetic treatment for temporary reduction of waist, hip, and thigh circumference in patients 18 years and older.
Regulatory Classification
Identification
A Low Level Laser System for Aesthetic Use is a device using low level laser energy for the disruption of adipocyte cells within the fat layer for the release of fat and lipids from these cells for noninvasive aesthetic use.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Industry and Food and Drug Administration Staff; Class II Special Controls Guidance Document: Low Level Laser System for Aesthetic Use.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 11, 2022
Resurgeonz LLC Julie Lalonde CEO 1550 Monte Carlo Court Merritt Island, Florida 32952
Re: K213534
Trade/Device Name: PHantom Trilogy(TM) Regulation Number: 21 CFR 878.5400 Regulation Name: Low Level Laser System For Aesthetic Use 878.5650 Regulatory Class: Class II Product Code: OLI Dated: January 20, 2022 Received: January 24, 2022
Dear Julie Lalonde:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K213534
Device Name Resurgeonz pHantom Trilogy(TM)
#### Indications for Use (Describe)
Resurgeonz pHantom Trilogy(TM) is indicated for use as a non-surgical, non-invasive aesthetic treatment for the temporary reduction of the circumference of waist, hips, thighs through the process of disrupting adipocyte cells, within the fat layer, for the release of fat and lipids from the targeted cells.
Type of Use (Select one or both, as applicable)
| <div> <span> <b> </b> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|-----------------------------------------------------------------------------------------------------|
| <div> <span> <b> </b> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
## 510(k) Summary of notification application K213534
| Submitter Information | |
|---------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitted By: | Resurgeonz |
| Address | 1550 Monte Carlo Court<br>Merritt Island, Florida 32952 |
| Phone Number | 321-222-5959 |
| Fax Number | N/A |
| Establish Registration No. | 622317 |
| Are You a Small Business? | Yes - SBD22819 |
| Name of Contact Person | Julie LaLonde |
| Contact Person Address | 1550 Monte Carlo Court<br>Merritt Island, Florida 32952 |
| Contact Person Phone No. | 407-233-6292 |
| Contact Person email | Admin@resurgeonz.com |
| Date Prepared | November 29, 2020 |
| Name of Medical Device | |
| Trade or proprietary name | pHantom Trilogy™ |
| Common or usual name | Fat reducing Low Level Laser |
| Classification name | Fat reducing Low Level Laser system for aesthetic use |
| Classification panel | General & Plastic Surgery |
| Regulation | 21 CFR 880.5440 |
| Product Code(s) | OLI |
| Device Classification | II |
| Legally marketed device(s) to<br>which equivalence is claimed | The Resurgeonz, pHantom Trilogy™ has the following<br>similarities to the predicate device, IR Technology's invisa-<br>RED ELITE (K192275) and Laser LipoLTD, Strawberry and<br>Strawberry and Cream (K130341). |
| Reason for 510(k) | New device offering to the non-invasive aesthetic<br>enhancement market. |
| Device Description | The Resurgeonz pHantom Trilogy™ is a non-surgical, non-<br>invasive, fat reducing low level laser system using a<br>combination of Red and Infrared light spectrums to<br>propagate laser frequencies of 660nm, 808nm and 980nm<br>into a patient's dermal and adipose tissues for the reduction<br>of the measured circumference of the patients waist, hips,<br>and thighs. The device features an LCD touchscreen. Mode of<br>operation is continuous. Allows individual patient session<br>controls of: treatment time, energy, pulse and delay.<br>Emergency safety cut off switch. Treatment applicators (multi<br>diode paddles) connected to the device console can number<br>8,10, or 12. Paddles include built in air fan cooling.<br>Combination of Coherent Frequencies of red and infrared: 26<br>diodes per big paddle, (13 red diodes at 660nm) and thirteen<br>(13 infrared diodes) of the 13 infrared diodes, (6 diodes) are<br>808nm and (7 diodes) are 980nm. Two smaller paddles are<br>available to swap out in place of two big paddles. The two<br>smaller paddles only have 2 diodes each of red diodes at<br>660nm and are used for smaller hard to reach body areas.<br>Power output or each diode is variable up to 72mW to per<br>diode for all red and infrared diodes. The device does not<br>have wireless and/or external wired communication. The<br>device is powered by 120V, 60Hz. There is a side holder for<br>paddles when not in use. |
| Technological Characteristics | For comparison of technological characteristics with the<br>predicates please see the table below. |
| Performance Data | Resurgeonz pHantom TrilogyTM overall performance testing<br>passed all specified test requirements of non-clinical (bench)<br>tests conducted for determination of Substantial Equivalence<br>(SE) results, and show of Substantial Equivalence (SE) with the<br>predicate devices. |
| Indication for Use | Resurgeonz pHantom Trilogy™ is indicated for use as a non-<br>surgical, non-invasive aesthetic treatment for the temporary<br>reduction of the circumference of waist, hips, thighs through<br>the process of disrupting adipocyte cells, within the fat layer,<br>for the release of fat and lipids from the targeted cells. |
| Manufacturing Site | 1550 Monte Carlo Court<br>Merritt Island, FL 32952 |
| Environment for use | Healthcare provider office, Medi-Spas, and Clinics. |
| Single patient use | No |
| Sterilization process | N/A - The Resurgeonz PHantom Trilogy™ is supplied a non-sterile device. Cleaning instructions are provided in the User Manual. |
| Biocompatibility | Meets acceptance criteria per ISO 10993: Parts 1, 5, and 10 requirements |
| Conclusion on Substantial Equivalence | The Resurgeonz PHantom Trilogy™ described and comparatively analyzed in this submission is determined to be Substantially Equivalent (SE) to the predicate device, for intended use, design, component and materials, operating principles, environment for use, performance, safety, effectiveness, product labeling and packaging, and is supported by the information provided in the 510(k) submission. |
| Conclusions drawn from non-clinical and clinical data | The Resurgeonz PHantom Trilogy™ meets the functional claims, and Intended Use as described in the product Labeling. |
| Statement of Safety and Efficacy | The safety and effectiveness are equivalent to the predicate device K192275 and predicate device K130341. The claim for Substantial Equivalence (SE) is supported by the information provided in the 510(k) submission. |
{4}------------------------------------------------
{5}------------------------------------------------
{6}------------------------------------------------
| Characteristic | Subject<br>Device<br>Resurgeonz,<br>LLC<br>PHantom Trilogy™ | Predicate Device (K192275)<br>IR Technology<br>invisa-RED ELITE ™ | Predicate Device<br>Strawberry/Strawberry &<br>Cream (K130341) |
|-----------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Class | II | II | II |
| Regulation Number | 21 CFR 878.5400 | 21 CFR 878.5400 | 21 CFR 878.5400 |
| Product Code(s) | OLI | OLI | OLI |
| Intended Use | Resurgeonz pHantom Trilogy™ is<br>indicated for use as a non-surgical,<br>non-invasive aesthetic treatment for<br>the temporary reduction of the<br>circumference of waist, hips, thighs<br>through the process of disrupting<br>adipocyte cells within the fat layer for<br>the release of fat and lipids from the<br>targeted cells. | The invisa-RED ™ is indicated for use<br>as a non-invasive aesthetic treatment<br>for the temporary reduction of the<br>circumference of waist, hips, thighs<br>through the process of<br>photobiomodulation affecting<br>adipocyte cells within the adipose<br>layer for the release of lipids from<br>these cells. | The Strawberry/Strawberry & Cream<br>is indicated for use as a non-invasive<br>aesthetic treatment for the<br>temporary reduction in waist<br>circumference by the disruption of<br>adipocyte cells within the fat layer for<br>the release of fat and lipids from<br>these cells. |
| Technology | Laser lights in the visible red and<br>infrared spectrums | Same | Laser lights in the visible red<br>spectrum |
| Light Spectrum-<br>Wavelength | Red light (Near Infrared) at 660nm and<br>Infrared at 808nm and 980nm | Red light (Near Infrared) at 680nm and<br>Infrared at 980nm | Red light (Near Infrared) 660 +/-<br>15nm |
| Light Waveform | Pulsed and Continuous | Same | Continuous |
| Laser Type | Diode Laser | Same | Same |
| Source Power | 120VAC / 60Hz | 120-240VAC / 50-60Hz | 100-240V, 50-60Hz |
| Light Power Output | 72mW each for Laser Diodes 660nm,<br>808nm and 980 nm | 200mW per diode each for Laser<br>Diodes both 680nm and 980nm | 40 mW +/- 15% for each laser diode<br>660 nm |
| Number of Pads /<br>Paddles | 6, 8, 10 or 12 Big Paddles (optional)<br>(size 130 x 66mm) | 6, 8, 10, or 12 Big Paddles (optional)<br>(size 130 x 66mm) | 4,6,8 or 10 Big paddles (optional)<br>(size similar) |
| Number of<br>paddles/probes<br>(size small) | 2 (optional) with two 660nm diodes<br>used to replace 2 big paddles | NA | 2 with one 660nm diode each,<br>included with big paddles |
| Number of diodes<br>per big paddle | 26 total, 13 red diodes and 13 infrared<br>diodes | Same | 6 red diodes per paddle |
| Number of diodes<br>per small paddle | 2 total, 2 red diodes only 660NM | NA | Similar<br>1 red diode per small paddle 660NM |
| Max number of Near<br>Infrared diodes per<br>system (red) | 156 | Same | 62 |
| Max number of<br>Infrared diodes per<br>system | 156 | Same | NA |
| Energy Delivery | Machine mounted paddles | Same | NA |
| Display | TFT Color Touchscreen LCD | Same | same |
| Console/Housing | ABS Plastic | Same | similar |
| Safety Features | Emergency stop button/key switch…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.