← Product Code [IYN](/productcode/IYN) · K213523

# X-CUBE i8, X-CUBE i9 (K213523)

_Alpinion Medical Systems Co., Ltd. · IYN · Jan 28, 2022 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K213523

## Device Facts

- **Applicant:** Alpinion Medical Systems Co., Ltd.
- **Product Code:** [IYN](/productcode/IYN.md)
- **Decision Date:** Jan 28, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1550
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

The X-CUBE i8, X-CUBE i9 diagnostic ultrasound systems are intended for use by, or by the order the supervision of, a licensed physician who is qualified for the evaluation of soft tissue and blood flow in the clinical applications of Fetal; Abdominal(renal & GYN/pelvic); Pediatric; Small Organ(breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Trans-raginal; Musculo-skeletal(Conventional); Musculo-skeletal(Superficial); Cardiac(adult& pediatric); Trans-esoph. (Cardiac); Peripheral Vessel(PV); and Urology(including prostate). And, in the imaging modes of 2D(B) mode; Harmonic mode(HAR); M mode; Anatomical M mode; Color Flow Doppler(CF) Mode; Power Doppler(PD) Mode; Directional PD mode; Pulsed Wave Doppler(PWD) Mode; Continuous Wave Doppler(CWD) Mode; High PRF Doppler mode; Tissue Doppler Imaging(TDI) Mode. The X-CUBE i8. X-CUBE i9 are intended to be used in a hospital or medical clinic.

## Device Story

X-CUBE i8 and i9 are diagnostic ultrasound systems for medical imaging. Systems accept acoustic signals via transducers, processing them to generate 2D, M-mode, and Doppler (CF, PD, PWD, CWD, TDI) images. Operated by physicians or under their supervision in hospitals/clinics. Features include Xpeed, Full SRI, Spatial Compounding, Panoramic, Stress Echo, Cube Strain, Needle Vision Plus, Elastography, Cube view, CEUS, Cube Note, Auto EF, Auto NT, and Microvascular Imaging. Output displayed on integrated flat panel for clinical diagnosis of soft tissue and blood flow. Benefits include optimized workflow and image quality for diagnostic assessment.

## Clinical Evidence

No clinical studies were required to support substantial equivalence. Evidence is based on non-clinical bench testing, including biocompatibility, thermal, electrical, electromagnetic, and mechanical safety, confirming conformance to applicable standards (IEC, AAMI/ANSI/ISO, NEMA, AIUM).

## Technological Characteristics

Ultrasound imaging system; pulsed Doppler technology. Materials: Biocompatible patient-contact materials. Connectivity: Wireless LAN. Safety standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, NEMA UD 2, NEMA UD 3. Features: Xpeed, Full SRI, Spatial Compounding, Elastography, Auto EF, Auto NT, Microvascular Imaging. Power: Max 220W.

## Regulatory Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- E-CUBE i7 ([K182594](/device/K182594.md))

## Reference Devices

- X-CUBE 90 ([K211300](/device/K211300.md))
- E-CUBE 12 ([K181277](/device/K181277.md))
- E-CUBE 15 ([K150773](/device/K150773.md))
- E-CUBE 8 ([K181617](/device/K181617.md))
- E-CUBE 11 ([K161439](/device/K161439.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Alpinion Medical Systems Co., Ltd. % Boyeon Cho Quality Management Representative 5FL, I dong, 77, heungan-daero 81 beon-gil dongan-gu Anyang-si. Gyeonggi-do 14117 REPUBLIC OF KOREA

January 28, 2022

#### Re: K213523

Trade/Device Name: X-CUBE i8, X-CUBE i9 Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: November 1, 2021 Received: November 3, 2021

Dear Boyeon Cho:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K213523

Device Name X-CUBE i8, X-CUBE i9

#### Indications for Use (Describe)

The X-CUBE i8, X-CUBE i9 diagnostic ultrasound systems are intended for use by, or by the order the supervision of, a licensed physician who is qualified for the evaluation of soft tissue and blood flow in the clinical applications of Fetal; Abdominal(renal & GYN/pelvic); Pediatric; Small Organ(breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Trans-raginal; Musculo-skeletal(Conventional); Musculo-skeletal(Superficial); Cardiac(adult& pediatric); Trans-esoph. (Cardiac); Peripheral Vessel(PV); and Urology(including prostate).

And, in the imaging modes of 2D(B) mode; Harmonic mode(HAR); M mode; Anatomical M mode; Color Flow Doppler(CF) Mode; Power Doppler(PD) Mode; Directional PD mode; Pulsed Wave Doppler(PWD) Mode; Continuous Wave Doppler(CWD) Mode; High PRF Doppler mode; Tissue Doppler Imaging(TDI) Mode.

The X-CUBE i8. X-CUBE i9 are intended to be used in a hospital or medical clinic.

| Type of Use (Select one or both, as applicable)                                                                                                 |                                                                                                                                 |
|-------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)     </span> | <span style="font-family: Arial, sans-serif;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)     </span> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### K213523

# 510(k) Summarv

In accordance with 21CFR807.92, the following summary of information is provided;

- Nov 4st, 2021 Date
- Submitter: ALPINION MEDICAL SYSTEMS Co., Ltd. Address: 5FL, I dong, 77, heungan-daero 81 beon-gil dongan-gu Anyang-si, Gyeonggi-do, 14117, REPUBLIC OF KOREA
- Primary Contact Boyeon CHO Person Quality Management Representative(QMR) Address: 5FL, I dong, 77, heungan-daero 81 beon-gil dongan-qu Anyang-si, Gyeonggi-do, 14117, REPUBLIC OF KOREA Phone: +82 70 7465 2104 Fax: +82 2 851 5595 Email: qa ra@alpinion.com
- Secondary Kevin CHUN Contact Person Address: 21222 30th Dr SE Ste C-122, Bothell, WA 98021, United States Phone: 425 949 1059 Fax: 425 949 4910 Email: kevin.chun@alpinionusa.com
	- Device Trade X-CUBE i8, X-CUBE i9 Name:
	- Common/ Ultrasonic Pulsed Doppler Imaging System Usual Name:
	- Classification System, Imaging, Pulsed Doppler Ultrasonic Names
- Product Code: Ultrasonic Pulsed Doppler Imaging System, 21CFR 892.1550 90-IYN Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYO Diagnostic Ultrasound Transducer, 21CFR 892.1570, 90-ITX
	- Primarv K182594 E-CUBE i7 Ultrasonic Pulsed Doppler Imaging System Predicate Device
	- Reference K211300 X-CUBE 90 Ultrasonic Pulsed Doppler Imaging System Devices K181277 E-CUBE 12 Ultrasonic Pulsed Doppler Imaging System K150773 E-CUBE 15 Ultrasonic Pulsed Doppler Imaging System K181617 E-CUBE 8 Ultrasonic Pulsed Doppler Imaging System K161439 E-CUBE 11 Ultrasonic Pulsed Doppler Imaging System
- The new technology of the subject device is as follows compared to Predicate Subject device New technology and Reference devices.
	- 1. Transducer C1-7GT

{4}------------------------------------------------

2. Feature Please see comparison table on pages F-3 to F-6

Device Description:

X-CUBE i8 and X-CUBE i9 products are an ultrasound imaging system for medical diagnosis. This system platform provides optimal patient diagnosis workflow with the wide flat panel display, ergonomic control panel with easy user interface, optimal image quality.

#### 1. Signal Mode:

2D(B) mode, Harmonic mode (HAR), M mode, Color M mode, Anatomical M mode, Color Flow Doppler(CF) Mode, Power Doppler(PD) Mode, Directional PD mode, Pulsed Wave Doppler(PWD) Mode, Continuous Wave Doppler(CWD) Mode, High PRF Doppler mode, Tissue Doppler Imaging(TDI) Mode

#### 2. Combination Mode:

B/Color Doppler, B/PWD, B/Color Doppler/PWD

Acoustic output track: Track 3

Indications The X-CUBE i8, X-CUBE i9 diaqnostic ultrasound systems are intended for For Use: use by, or by the order of, and under the supervision of, a licensed physician who is qualified for the evaluation of soft tissue and blood flow in the clinical applications of Fetal; Abdominal(renal & GYN/pelvic); Pediatric; Small Organ(breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Trans-rectal; Trans-vaginal; Musculo-skeletal(Conventional); Musculo-skeletal(Superficial); Cardiac(adult& pediatric); Trans-esoph. (Cardiac); Peripheral Vessel(PV); and Urology(including prostate).

> And, in the imaging modes of 2D(B) mode: Harmonic mode(HAR); M mode; Color M mode: Anatomical M mode: Color Flow Doppler(CF) Mode: Power Doppler(PD) Mode; Directional PD mode; Pulsed Wave Doppler(PWD) Mode; Continuous Wave Doppler(CWD) Mode; High PRF Doppler mode; Tissue Doppler Imaging(TDI) Mode.

The X-CUBE i8, X-CUBE i9 are intended to be used in a hospital or medical clinic.

{5}------------------------------------------------

Determination of Substantial Equivalence: Comparison table with Predicate devices:

| Model                                      | Proposed<br>X-CUBE i8,<br>X-CUBE i9<br>ALPINION Medical<br>Systems Co., Ltd. | Predicate<br>E-CUBE i7<br>ALPINION Medical<br>Systems Co., Ltd. | Reference<br>X-CUBE 90<br>ALPINION Medical<br>Systems Co., Ltd. | Reference<br>E-CUBE 12<br>ALPINION Medical<br>Systems Co., Ltd. | Reference<br>E-CUBE 15<br>ALPINION Medical<br>Systems Co., Ltd. | Reference<br>E-CUBE 8<br>ALPINION Medical<br>Systems Co., Ltd. | Reference<br>E-CUBE 11<br>ALPINION Medical<br>Systems Co., Ltd. |
|--------------------------------------------|------------------------------------------------------------------------------|-----------------------------------------------------------------|-----------------------------------------------------------------|-----------------------------------------------------------------|-----------------------------------------------------------------|----------------------------------------------------------------|-----------------------------------------------------------------|
| Feature                                    | K213523                                                                      | K182594                                                         | K211300                                                         | K181277                                                         | K150773                                                         | K181617                                                        | K161439                                                         |
|                                            | Indications for Use                                                          |                                                                 |                                                                 |                                                                 |                                                                 |                                                                |                                                                 |
| - Fetal                                    | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              | √                                                               |
| - Abdominal<br>(Renal&GYN/Pelvic)          | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              | √                                                               |
| - Intra-operative<br>(Specify, Neuro)      |                                                                              |                                                                 |                                                                 |                                                                 |                                                                 |                                                                |                                                                 |
| - Pediatric                                | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              | √                                                               |
| - Small Organ<br>(breast, testes, thyroid) | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              | √                                                               |
| - Neonatal Cephalic                        | √                                                                            |                                                                 | √                                                               | √                                                               |                                                                 | √                                                              |                                                                 |
| - Adult Cephalic                           | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              | √                                                               |
| - Trans-rectal                             | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              | √                                                               |
| - Trans-vaginal                            | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              | √                                                               |
| - Musculo-skeletal<br>(Conventional)       | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              | √                                                               |
| - Musculto skeletal<br>(Superficial)       | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              | √                                                               |
| - Cardiac (Adult)                          | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              | √                                                               |
| - Cardiac (Pediatric)                      | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              | √                                                               |
| - Trans-esoph.<br>(Cardiac)                | √                                                                            |                                                                 |                                                                 |                                                                 |                                                                 |                                                                |                                                                 |
| - Peripheral Vessel                        | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              | √                                                               |
| - Urology<br>(including prostate)          | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              | √                                                               |
|                                            | Dimensions and Weight                                                        |                                                                 |                                                                 |                                                                 |                                                                 |                                                                |                                                                 |
| Weight<br>(Excluding options)              | 6kg<br>(excluding Option)                                                    | 7.2kg<br>(excluding Option)                                     | 85kg                                                            | 94 kg                                                           | 105 kg                                                          | 55 kg                                                          | 94 kg                                                           |
| Height                                     | 62.5/352 mm                                                                  | 84.9 mm                                                         | 1325/1560 mm                                                    | 1,420/1,520 mm                                                  | 1,413/1,848 mm                                                  | 830~1,430 mm                                                   | 1,455/1,695 mm                                                  |
| Width                                      | 385 mm                                                                       | 402.6 mm                                                        | 554 mm                                                          | 590 mm                                                          | 585 mm                                                          | 532 mm                                                         | 590 mm                                                          |
| Depth                                      | 370 mm                                                                       | 366.5 mm                                                        | 815 mm                                                          | 895 mm                                                          | 670 mm                                                          | 787 mm                                                         | 895 mm                                                          |
| Electrical Power                           |                                                                              |                                                                 |                                                                 |                                                                 |                                                                 |                                                                |                                                                 |
| Voltage                                    | 20V, 11A                                                                     | 19V, 10.5A                                                      | 100-120V~,<br>200-240V~                                         | 100-120V~,<br>200-240V~                                         | 100-120V~,<br>200-240V~                                         | 100-120V~,<br>200-240V~                                        | 100-120V~,<br>200-240V~                                         |
| Frequency                                  | 50-60 Hz                                                                     | 50-60 Hz                                                        | 50-60 Hz                                                        | 50/60 Hz                                                        | 50-60 Hz                                                        | 50-60 Hz                                                       | 50/60 Hz                                                        |
| Power                                      | Max. 220W                                                                    | Max. 200W                                                       | Max. 700VA                                                      | Max. 600VA                                                      | Max. 900VA                                                      | Max. 450VA                                                     | Max. 600VA                                                      |
| Imaging Modes                              |                                                                              |                                                                 |                                                                 |                                                                 |                                                                 |                                                                |                                                                 |
| - 2D(B) mode                               | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              | √                                                               |
| - Harmonic mode                            | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              | √                                                               |
| - M mode                                   | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              | √                                                               |
| - Color M mode                             | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              | √                                                               |
| - Anatomical M mode                        | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              | √                                                               |
| - Color Flow Doppler<br>(CF) mode          | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              | √                                                               |
| - Power Doppler (PD)<br>mode               | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              | √                                                               |
| - Directional PD mode                      | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              | √                                                               |
| - Pulsed wave Doppler<br>(PWD) mode        | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              | √                                                               |
| - Continuous wave<br>Doppler (CWD) mode    | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              | √                                                               |
| - High PRF Doppler<br>mode                 | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              | √                                                               |
| - Tissue Doppler<br>imaging (TDI) mode     | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              | √                                                               |
| Features                                   |                                                                              |                                                                 |                                                                 |                                                                 |                                                                 |                                                                |                                                                 |
| - Xpeed™                                   | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              | √                                                               |
| - Full SRI™                                | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              | √                                                               |
| -Spatial Compounding<br>Image (SCI)        | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              | √                                                               |
| - Panoramic                                | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              | √                                                               |
| - Stress Echo                              | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              | √                                                               |
| - Cube Strain™                             | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              | √                                                               |
| - Needle Vision™ Plus                      | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              | √                                                               |
| - Elastography                             | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              | √                                                               |
| - Cube view™                               | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              | √                                                               |
| - Contrast Enhanced<br>Ultrasound (CEUS)   | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              |                                                                 |
| - Cube Note                                | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               |                                                                |                                                                 |
| - Auto EF                                  | √                                                                            | √                                                               | √                                                               |                                                                 |                                                                 |                                                                |                                                                 |
| - Auto NT                                  | √                                                                            | √                                                               | √                                                               | √                                                               |                                                                 | √                                                              |                                                                 |
| - Microvascular<br>Imaging (MVI)           | √                                                                            | √                                                               | √                                                               |                                                                 |                                                                 |                                                                |                                                                 |
| Accessories or Kits                        |                                                                              |                                                                 |                                                                 |                                                                 |                                                                 |                                                                |                                                                 |
| Color printer                              | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              | √                                                               |
| B/W printer                                | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              | √                                                               |
| DVD-RW                                     | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              | √                                                               |
| Foot switch                                | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              | √                                                               |
| Wireless LAN                               | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              | √                                                               |
| SC1-6 Biopsy guide kit                     | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              | √                                                               |
| L3-12 Biopsy guide kit                     | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              | √                                                               |
| EN3-10 Reusable<br>Biopsy needle guide     | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              | √                                                               |
| EN3-10 Disposable<br>Biopsy needle guide   | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              | √                                                               |
| ECG module / cable                         | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              | √                                                               |
| Disinfectant & Ultrasound Gel              |                                                                              |                                                                 |                                                                 |                                                                 |                                                                 |                                                                |                                                                 |
| Ultrasonic gel                             | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              | √                                                               |
| Cidex OPA<br>(Disinfectant agaents)        | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              | √                                                               |
|                                            |                                                                              |                                                                 |                                                                 |                                                                 |                                                                 |                                                                |                                                                 |
| Cidex Plus<br>(Disinfectant agaents)       | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              |                                                                 |
| Gigasept FF<br>(Disinfectant agaents)      | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              |                                                                 |
| Virkon<br>(Disinfectant agaents)           | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              |                                                                 |
| Wavicide-01<br>(Disinfectant agaents)      | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              |                                                                 |
| AIDAL PLUS<br>(Disinfectant agaents)       | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              |                                                                 |
| Cetylcide-G<br>(Disinfectant agaents)      | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              |                                                                 |
| Sporicidin<br>(Disinfectant agaents)       | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              |                                                                 |
| Thermal, mechanical and electrical safety  |                                                                              |                                                                 |                                                                 |                                                                 |                                                                 |                                                                |                                                                 |
| - NEMA UD2, UD3                            | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              |                                                                 |
| - AIUM Medical<br>Ultrasound Safety        | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              |                                                                 |
| - IEC 60601-1                              | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              |                                                                 |
| - IEC 60601-1-2                            | √                                                                            | √                                                               | √                                                               | √                                                               | √                                                               | √                                                              |                                                                 |

ALPINION MEDICAL SYSTEMS Co., Ltd.

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

Summary of Non-Clinical Tests:

X-CUBE i8 and X-CUBE i9 have been evaluated for biocompatibility, acoustic output as well as thermal, electrical, electromagnetic, and mechanical safety, and has been found to conform to applicable medical device safety standards. X-CUBE i8, X-CUBE i9 and its application comply with voluntary standards as detailed in this premarket submission.

- � IEC60601-1:2005(Third Edition)+CORR.1:2006+CORR.2:2007+A1:2012, Medical Electrical Equipment - Part 1: General Requirements for Safety
- � IEC60601-1-2:2014. Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
- IEC60601-2-37:2007/AMD1:2015, Medical Electrical Equipment Part 2-� 37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment
- � AAMI/ANSI/ISO10993-1:2009(R)2013, Biological Evaluation of Medical Devices - Part 1:Evaluation and Testing within a risk management process
- � AAMI/ANSI/ISO14971:2007/(R)2010, Medical devices-Application of risk management to medical devices
- ◆ AIUM MUS, Third edition, Medical Ultrasound Safety
- � NEMA UD 2-2004(R2009), Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment
- ◆ NEMA UD 3-2004(R2009), Standard for Real Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic ultrasound Equipment

The following quality management system measures were applied to the development of X-CUBE i8 and X-CUBE i9:

- Medical Device Risk Management �
- � Requirements Reviews
- � Desian Reviews
- � Component Verification
- ◆ Integration Review (System Verification)
- � Performance Testing (System Verification)
- � Safety Testing (Compliance Test)
- � Design Validation

Transducer materials and other patient contact materials are biocompatible.

#### Summary of Clinical Tests:

The subject of this premarket submission, X-CUBE i8 and X-CUBE i9, did not require clinical studies to support substantial equivalence.

Discussion:

X-CUBE i8 and X-CUBE i9 were compared with the predicate device. The subject devices are in conformance with applicable safety standards.

Therefore, the differences between X-CUBE i8 and X-CUBE i9, and the predicate device would not affect the safety, effectiveness and essential performance.

{10}------------------------------------------------

- Conclusion: The design, development and quality process of the manufacturer confirms with 21 CFR 820 and ISO 13485. The devices are designed to conform to applicable medical device safety standards and compliance. Therefore, ALPINION MEDICAL SYSTEMS Co., Ltd. considers X-CUBE i8 and X-CUBE i9 to be as safe, and effective as the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K213523](https://fda-staging.innolitics.com/device/K213523)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).