ADD (Digital Flat Panel X-Ray Detector) is indicated for digital imaging solution designed for providing general radiographic diagnosis of human anatomy targeting both adult and children. It is intended to replace film based radiographic diagnostic systems. Not to be used for mammography.
Device Story
ADD is a wired/wireless digital flat panel X-ray detector; replaces film-based systems. Input: X-ray radiation; utilizes amorphous silicon TFT sensor array with CsI scintillator to convert X-rays to electrical signals. Features auto-trigger signal sensing (AED) for generator-independent operation. Output: Digital radiographic images for clinical review. Used in clinical settings by healthcare professionals. Output facilitates general radiographic diagnosis. Benefits: faster, streamlined digital imaging workflow.
Clinical Evidence
Bench testing included electrical safety (AAMI ANSI ES60601-1), EMC (IEC 60601-1-2), and performance testing (IEC 62220-1 for DQE). Clinical images were provided as supplemental evidence to demonstrate the device functions as intended, though not required for substantial equivalence given identical core sensor technology to the predicate.
Indicated for digital radiographic diagnosis of human anatomy in adults and children. Not for mammography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
November 15, 2021
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
H&abyz Co., Ltd. % Hyunook (David) Han Regulatory Affairs Manager 1F, 2-Dong, 41-16 Cheoinseong-ro Namsa-myeon, Cheoin-gu Yongin-si. Gyeonggi-do 17118 REPUBLIC OF KOREA
Re: K213497
Trade/Device Name: ADD Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: MQB Dated: August 21, 2021 Received: November 1, 2021
Dear Hyunook (David) Han:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
Page 2
801; medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
, for
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213497
Device Name ADD
Indications for Use (Describe)
ADD (Digital Flat Panel X-Ray Detector) is indicated for digital imaging solution designed for providing general radiographic diagnosis of human anatomy targeting both adult and children. It is intended to replace film based radiographic diagnostic systems. Not to be used for mammography.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| <div> <span> <svg height="12" width="12"> <path d="M0 0 L12 12 M12 0 L0 12" stroke="black" stroke-width="2"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
5. 510(K) Summary
{4}------------------------------------------------
# 510(k) Summary
[As Required by 21 CFR 807.92] K213497
## 1. Date Prepared [21 CFR 807.92(a)(a)]
October 28, 2021
## 2. Submitter's Information [21 CFR 807.92(a)(1)]
- Name of Manufacturer: H&abyz Co., Ltd.
- Address: 1F, 2-Dong, 41-16 Cheoinseong-Ro, Namsa-Myeon, Cheoin-Gu, Yongin-Si, Gyeonggi-Do, Republic of Korea [17118] ● Contact Name: Namkyu Hur / Quality Manager
- Telephone No.: +82 070-4658-9300
- Email Address: hho@abyzr.com
- Registration No.: 3016674851
## 3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]
| Trade/Device/Model Name | ADD |
|-----------------------------|-----------------------------------|
| Reference No. | HAD1417MCW, HAD1717MCW |
| Common Name | Digital Flat Panel X-ray Detector |
| Device Classification Name | Stationary X-ray System |
| Regulation Number | 21 CFR 892.1680 |
| Classification Product Code | MQB |
| Device Class | II |
| 510(k) Review Panel | Radiology |
{5}------------------------------------------------
## 4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)]
The identified predicate device within this submission are shown as follow;
## Predicate Device
| ● | 510(k) Number: | K203188 |
|---|--------------------------------|---------------------------------|
| ● | Applicant: | H&abyz Co., Ltd. |
| ● | Trade/Device Name: | ADD |
| ● | Common Name: | Digital Diagnostic X-ray System |
| ● | Classification Name: | System. X-ray, Stationary |
| ● | Regulation Number: | 21 CFR 892.1680 |
| ● | Classification Product<br>Code | MQB |
| ● | Device Class: | II |
| ● | 510(k) Review Panel: | Radiology |
These predicate devices have not been subject to a design-related recall
{6}------------------------------------------------
## 5. Description of the Device [21 CFR 807.92(a)(4)]
The ADD are wired or wireless digital flat panel detectors that have been designed for faster, more streamlined approach to digital radiography systems. The ADD detector utilize a combination of propriety TFT glass and scintillators(CsI), and those and electronics are housed in one package. The detectors support an auto-trigger signal sensing technology that allows the detectors to be used without generator integration.
The flat panel sensors of the ADD are fabricated using thin film technology based on amorphous silicon technology. Electronically, the sensors are much like conventional photodiode arrays. Each pixel in the array consists of a light-sensing photodiode and a switching Thin Film Transistor (TFT) in the same electronic circuit. Amorphous silicon photodiodes are sensitive to visible light, with a response curve roughly comparable to human vision. The sensitivity of amorphous silicon photodiodes peaks in green wavelengths, well matched to scintillators such as CsI. The response has the excellent linearity of a charge-integrating-biased photodiode.
SDK-MCW is the software of Detector that performs image acquisition, image correction, and preprocessing. According to the FDA quidance document entitled "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices," it has a moderate level of concern. Also, it does not base on previously-cleared software and was originally coded.
{7}------------------------------------------------
## 6. Indications for use [21 CFR 807.92(a)(5)]
ADD (Digital Flat Panel X-Ray Detector) is indicated for digital imaging solution designed for providing general radiographic diagnosis of human anatomy targeting both adult and children. It is intended to replace film based radiographic diagnostic. Not to be used for mammography.
## 7. Technological Characteristics (Equivalence to Predicate Device) [21 CFR 807.92(a)(6)]
There are no significant differences in the technological characteristics of these devices compared to the predicate device which adversely affect safety or effectiveness. Provided below is a table summarizing and comparing the technological characteristics of the ADD and the predicate device:
| | Predicate Device | Proposed Device | Note |
|----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| K Number | K203188 | | - |
| Manufacturer | H&abyz Co., Ltd. | H&abyz Co., Ltd. | - |
| Trade Name | ADD | ADD | - |
| Common Name | Digital Flat Panel X-ray Detector | Digital Flat Panel X-ray Detector | Same |
| Product Code | MQB | MQB | Same |
| Regulation<br>Number | 21 CFR 892.1680 | 21 CFR 892.1680 | Same |
| 510(k) Review<br>Panel | Radiology | Radiology | Same |
| Indications for<br>Use | ADD (Digital Flat Panel X-Ray<br>Detector) is indicated for digital<br>imaging solution designed for<br>providing general radiographic<br>diagnosis of human anatomy<br>targeting both adult and children.<br>It is intended to replace film<br>based radiographic diagnostic.<br>Not to be used for mammography. | ADD (Digital Flat Panel X-Ray<br>Detector) is indicated for digital<br>imaging solution designed for<br>providing general radiographic<br>diagnosis of human anatomy<br>targeting both adult and children.<br>It is intended to replace film<br>based radiographic diagnostic.<br>Not to be used for mammography | Same |
| Detector name | HAD1417MCW | HAD1417MCW<br>HAD1717MCW | Difference |
| Scintillator | CsI | CsI | Same |
| Effective Pixel<br>Area | [HAD1417MCW]<br>345.24 x 425.6 mm | [HAD1417MCW]<br>345.24 x 425.6 mm<br>[HAD1717MCW]<br>425.04 x 425.6 mm | Difference |
| Total Pixel<br>Number | [HAD1417MCW]<br>2,560 x 3,072 pixels | [HAD1417MCW]<br>2,560 x 3,072 pixels<br>[HAD1717MCW]<br>3072 x 3,072 pixels | Difference |
| Pixel Pitch | 140um | 140um | Same |
| High Contrast<br>Limiting<br>Resolution<br>(LP/mm) | Max. 3.5 | Max. 3.5 | Same |
[Table 1. Comparison of Proposed Device to Predicate Devices]
{8}------------------------------------------------
| | Predicate Device | Proposed Device | Note |
|---------------------|---------------------------------------------------------|------------------------------------------------------|------|
| Communication | Wired/Wireless | Wired/Wireless | Same |
| DQE | 50% (0.1lp/mm, min.) | 50% (0.1lp/mm, min.) | Same |
| MTF | 97% (0.1lp/mm, min.) | 97% (0.1lp/mm, min.) | Same |
| Anatomical<br>Sites | General | General | Same |
| Exposure Mode | Normal Mode(Manual),<br>AED Mode<br>(Auto<br>Detection) | Normal Mode(Manual),<br>AED Mode (Auto<br>Detection) | Same |
| Wireless | IEEE 802.11a/b/g/n | IEEE 802.11a/b/g/n | Same |
The proposed device is substantially equivalent to the previously cleared detector provided as part of the complete imaging system.
It is substantially equivalent to these devices in design, function, materials, operational principles, and intended use. The proposed device has been tested about electrical safety, EMC, and performance, and the software has been validated. Also, the clinical data has been provided to support the substantial equivalence to the predicate device.
The only difference is the size of the detectors, and it is a difference only in the area where the xray image can be taken. And a result of verification, it was confirmed that there is no difference in other performance. Therefore, the difference in the sizes of the detectors does not affect the safety and effectiveness. These differences do not raise different questions of safety and effectiveness than the predicate.
## 8. Non-Clinical Test summary
The ADD complies with voluntary standards for electrical safety, electromagnetic compatibility. The following data were provided in support of the substantial equivalence determination:
- 1) Electrical Safety, Electromagnetic Compatibility and Performance:
The ADD complies with the electrical safety and electromagnetic compatibility requirements established by the standards.
| Standards<br>No. | Standards<br>Organization | Standard Title | Version | Publication<br>Year |
|------------------|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|---------------------|
| ES60601-1 | AAMI ANSI | Medical Electrical Equipment - Part 1: General<br>Requirements for Basic Safety and Essential<br>Performance (IEC 60601-1:2005, MOD) | ES60601-1:<br>2005(R)201<br>2 and<br>A1:2012 | 2014 |
| 60601-1-2 | IEC | Medical Electrical Equipment - Part 1-2: General<br>Requirements for Safety – Collateral Standard:<br>Electromagnetic Compatibility - Requirements<br>and Tests | 60601-1-2<br>Edition 4.0<br>2014-02 | 2016 |
| - | FDA | Radio Frequency Wireless Technology in Medical<br>Devices | August 14 | 2013 |
{9}------------------------------------------------
## 2) Software Validation
The ADD contains MODERATE level of concern software. The software was designed and developed according to a software development process and was verified and validated. Software information is provided in accordance with FDA quidance:
- The content of premarket submissions for software contained in medical devices, on May 11, 2005
## 3) Biocompatibility
- ISO 10993-1 and series. Biological evaluation of medical devices .
## 4) Performance Test
Imaging performance test has been conducted according to:
- IEC 62220-1, Medical Electrical Equipment – Characteristics of Digital X-ray Imaging Devices – Part 1-1: Determination of the Detective Quantum Efficiency – Detectors Used in Radiographic Imaging.
We select the predicate device in order to demonstrate adequate DQE performance of the ADD detector. According to the above comparison table, subject device shows similar or same DQE.
5) Cybersecurity
- . Content of Premarket Submissions for Management of Cybersecurity in Medical Devices, on October2, 2014
6) Label
- CFR Part 801
- Pediatric Information for X-ray Imaging Device Premarket Notifications, on November 28, 2017
## 9. Clinical Test Summary
Clinical data has been provided according to FDA guidance document "Guidance for the Submission of 510(k)s for Solid Sate X-ray Imaging Devices". Clinical images were provided; these images were not necessary to establish substantial equivalence based on the differences from the predicate (note TFT technology with CsI scintillator that is identical to the predicate image plate) but they provide further evidence in addition to the laboratory performance data to show that the subject digital detector works as intended.
## 10. Substantial Equivalence [21 CFR 807.92(b)(1) and 807.92]
There are no significant differences between the ADD and the predicate device, K203188 that would adversely affect the use of the product. It is substantially equivalent to these devices in indications for use and technology characteristics.
{10}------------------------------------------------
## 11. Conclusion [21 CFR 807.92(b)(3)]
In according with the Federal Food & Drug and cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification, concludes that the subject device is substantially equivalent in safety and effectiveness to the predicate device as described herein.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.