K213491 · Ctl Medical Corporation · OLO · May 12, 2022 · Neurology
Device Facts
Record ID
K213491
Device Name
Universal Pilot Guidance Instrument System
Applicant
Ctl Medical Corporation
Product Code
OLO · Neurology
Decision Date
May 12, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Indications for Use
The CTL Amedica Navigation Instrument System is intended to be used during the preparation and placement of CTL Amedica screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. The CTL Amedica Navigation Instrument System is specifically designed for use with the Medtronic StealthStation System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure such as vertebra can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks for the anatomy.
Device Story
The CTL Amedica Navigation Instrument System consists of manual surgical instruments—drivers, taps, awls, and probes—designed for use with the Medtronic StealthStation Navigation System. These instruments are used by surgeons in spinal surgery to assist in the precise placement of pedicle screw implants. The system functions by registering the instruments with the StealthStation, which provides real-time, multiplane 3D and 2D visualization of anatomical structures relative to CT/MR models or fluoroscopy. By providing visual confirmation of hardware placement, the system aids surgeons in complex and minimally invasive procedures, potentially improving accuracy and patient outcomes. The instruments are provided non-sterile and require steam sterilization by the user.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by non-clinical bench testing, including compatibility testing with the StealthStation system, dimensional comparisons, and rigidity testing of the NavLock engagement feature.
Technological Characteristics
Manual surgical instruments (drivers, taps, awls, probes) manufactured from Stainless Steel (ASTM F899) and Aluminum (ASTM B221). Designed for compatibility with Medtronic StealthStation navigation hardware/software. Non-sterile, intended for user steam sterilization.
Indications for Use
Indicated for use during spinal surgery to assist surgeons in precisely locating anatomical structures for the preparation and placement of CTL Amedica screws in open or minimally invasive procedures, specifically when used with the Medtronic StealthStation System for conditions where stereotactic surgery is appropriate.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
Medtronic Navigated Manual Reusable Instruments for Use with StealthStation (K153442)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath in a smaller font.
May 12, 2022
CTL Medical Corporation % Dhaval Saraiya Regulatory/Quality Consultant Omni Strategic Solutions, LLC 700 Pennsylvania Ave SE 2nd Floor Washington, District of Columbia 20003
Re: K213491
Trade/Device Name: CTL Amedica Navigation Instruments Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: April 8, 2022 Received: April 12, 2022
Dear Dhaval Saraiya:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213491
#### Device Name
CTL Amedica Navigation Instrument System
#### Indications for Use (Describe)
The CTL Amedica Navigation Instrument System is intended to be used during the preparation and placement of CTL Amedica screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. The CTL Amedica Navigation Instrument System is specifically designed for use with the Medtronic StealthStation System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure such as vertebra can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks for the anatomy.
Type of Use (*Select one or both, as applicable*)
| <span> <b> ☑ </b> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|--------------------------------------------------------------------------------------------|
| <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
In accordance with 21 CFR \$807.92 and the Safe Medical Devices Act of 1990, the following information is provided for the CTL Amedica Navigation Instrument System 510(k) premarket notification. The submission was prepared in accordance with the FDA guidance document, 'Format for Traditional and Abbreviated 510(k)s', issued on September 13, 2019.
| Sponsor: | CTL Medical Corporation<br>Sean Suh<br>4550 Excel Pkwy<br>Ste 300<br>Addison, TX 75001 | | |
|------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|---------------|
| Contact Person: | Dhaval S.<br>Omni Strategic Solutions, LLC.<br>Regulatory/Quality Consultant<br>Email: omniregsolutions@gmail.com<br>Tel: 213-400-2576 | | |
| Date: | 11 May 2022 | | |
| Subject Device: | Trade Name: CTL Amedica Navigation Instrument System<br>Common Name: Orthopedic Stereotaxic Instrument<br>Classification Name: OLO – Stereotaxic Instrument (21 CFR 882.4560) | | |
| Predicate Device(s): | | | |
| Primary | K172115 | Navigated Instrument System | Orthofix Inc. |
| | K153442 | Medtronic Navigated<br>Manual Reusable<br>Instruments for Use with<br>StealthStation | Medtronic |
| Purpose and<br>Device Description: | The purpose of this submission is to request clearance for the new<br>CTL Amedica Navigation Instrument System.<br>The CTL Amedica Navigation Instrument System are manual<br>surgical instruments for use with the Medtronic® StealthStation™<br>Navigation System to assist surgeons in precisely locating | | |
{4}------------------------------------------------
anatomical structures in either open, minimally invasive, or percutaneous procedures for preparation and placement of pedicle screw system implants. This surgical imaging technology provides surgeons visualization for complex and MIS procedures and confirms the accuracy of advanced surgical procedures. Use of this navigation system provides the surgeon access to real-time, multiplane 3D images (and 2D images) providing confirmation of hardware placement.
The Navigation Instruments include the following:
- Drivers ●
- . Taps
- Awls
- Probes .
Use of the Navigated Instrument System is limited to Taps ranging in sizes from 4.5mm to 8.5mm in diameter. The Navigation Instruments are to be used with the following CTL Amedica Systems but not limited to:
- RAPHAELTM Pedicle Screw System Family
- PICASSO II™ MIS Spinal System Family
- TAURUS™ Pedicle Screw System Family
The CTL Amedica Navigation Instruments are designed for use only with the hardware and software of the Medtronic StealthStation Navigation System.
The CTL Amedica Navigation Instruments are manufactured from Stainless Steel and Aluminum. All of these materials conform to ASTM F899.
#### Intended Use and Indications for Use:
The CTL Amedica Navigation Instrument System is intended to be used during the preparation and placement of CTL Amedica screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. The CTL Amedica Navigation Instrument System is specifically designed for use with the Medtronic StealthStation System, which are indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure such as vertebra can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks for the anatomy.
# Summary of Technological
Characteristics:
The rationale for substantial equivalence is based on consideration of the following characteristics:
{5}------------------------------------------------
- Intended Use: The intended use is similar to the intended use ● cleared in K172115 and K153442.
- . Indications for Use: The indications for use are similar to the indications for use cleared in K172115 and K153442.
- . Materials: The CTL Amedica Navigation Instrument System are manufactured from Stainless Steel (per ASTM F899) and Aluminum (per ASTM B221) which are commonly used materials in orthopedic instruments and similar to materials used in K172115 and K153442.
- Design Features: The design features for the CTL Amedica ● Navigation Instrument System is similar to those in currently marketed devices cleared in K172115 and K153442. The design differences have not identified any issues that would impact the safety and effectiveness of the device.
- Sterilization: The CTL Amedica Navigation Instrument System are offered to the user in the non-sterile configuration. The non-sterile instruments will be required to be steam sterilized by the user prior to use is similar to the devices cleared in K172115 and K153442.
### Summary of Performance Data (Nonclinical and/or Clinical):
#### Non-Clinical Tests: ●
- Compatibility Testing Testing included physical testing with all subject instruments to demonstrate that the subject instruments can be regsitered with the StealthStation system and are compatible with the StealthStation system software in terms of accuracy, function and performance in a simulated surgical navigation use environment
- o Dimensional Comparison Engineering analysis and tolerance analysis included dimensional measurements and geometrical comparisons of both the predicate devices and subject devices
- o Rigidity Comparison Testing to demonstrate that the NavLock engagement feature on the subject instruments is equivalent to the predicate engagement feature
The non-clinical testing demonstrated that the subject instruments registered successfully with the StealthStation system and are substantially equivalent to the predicate devices with regards to compatibility, accuracy, function and performance. Therefore, it can be concluded that the compatibility, accuracy, function, and performance of the subject devices are substantially equivalent to the predicate devices.
- Clinical Tests: ●
o N/A
{6}------------------------------------------------
## Substantial Equivalence Conclusion:
The CTL Amedica Navigation Instrument System has shown to be substantially equivalent to the predicate device with regards to indications, design, technology, funcitionality and performance. Results of the non-clinical tests indicate that the device will perform within the intended uses and no new issues of safety and effectiveness have been raised.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.