K213436 · Clarius Mobile Health Corp. · IYN · Nov 15, 2021 · Radiology
Device Facts
Record ID
K213436
Device Name
Clarius Ultrasound Scanner
Applicant
Clarius Mobile Health Corp.
Product Code
IYN · Radiology
Decision Date
Nov 15, 2021
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
AI/ML, Pediatric
Indications for Use
The Clarius Ultrasound Scanner is a software-based ultrasound imaging system and accessories, intended for diagnostic imaging. It is indicated for diagnostic ultrasound imaging and fluid flow analysis in the following applications: ophthalmic, fetal, abdominal, intra-operative (non-neurological), pediatric, small organ, cephalic (adult), trans-rectal, trans-vaginal, musculo-skeletal (conventional, superficial), urology, gynecology, cardiac (adult, pediatric), peripheral vessel, carotid, and procedural guidance of needles into the body. The system is a transportable ultrasound system intended for use in environments where is provided by trained healthcare professionals.
Device Story
Portable, software-controlled ultrasound system; acquires/displays high-resolution, real-time ultrasound data. Piezoelectric transducers transmit non-ionizing sound waves; convert echoes to electronic signals; construct anatomical images. Images transmitted wirelessly via Wi-Fi/Bluetooth to off-the-shelf iOS or Android devices. Used by qualified healthcare practitioners (doctors, nurses, sonographers) in clinical and EMS environments. Output viewed on mobile device screen; assists in diagnostic imaging and procedural guidance. Benefits include portability and real-time visualization for clinical decision-making.
Clinical Evidence
Bench testing only. Performance and safety verified through testing against recognized standards (IEC 60601 series, ISO 10993-1, etc.). No clinical data presented.
Technological Characteristics
Portable ultrasound system; wireless transducers (convex, linear, phased array, intracavity). Piezoelectric sensing. Internal integrated lithium-ion battery. Connectivity via Wi-Fi/Bluetooth to iOS/Android devices. Complies with IEC 60601-1, 60601-1-2, 60601-1-6, 60601-1-12, 60601-2-37, ISO 10993-1, AIUM/NEMA UD 2/3, IEC 62304, IEC 62366-1, ISO 14971, IEC 60529, IEC 62133, IEC 61157.
Indications for Use
Indicated for diagnostic ultrasound imaging and fluid flow analysis in ophthalmic, fetal, abdominal, intra-operative (non-neurological), pediatric, small organ, cephalic (adult), trans-rectal, trans-vaginal, musculoskeletal (conventional, superficial), urology, gynecology, cardiac (adult, pediatric), peripheral vessel, carotid, and needle guidance. For use by trained healthcare professionals in clinical and EMS environments.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".
Clarius Mobile Health Corp. % Ms. Agatha Szeliga Director, Regulatory Affairs 130-2985 Virtual Way Vancouver, British Columbia V5M 4X7 CANADA
# Re: K213436
Trade/Device Name: Clarius Ultrasound Scanner Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN Dated: October 21, 2021 Received: October 22, 2021
November 15, 2021
Dear Ms. Szeliga:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR
{1}------------------------------------------------
803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213436
Device Name Clarius Ultrasound Scanner
#### Indications for Use (Describe)
The Clarius Ultrasound Scanner is a software-based ultrasound imaging system and accessories, intended for diagnostic imaging. It is indicated for diagnostic ultrasound imaging and fluid flow analysis in the following applications: ophthalmic, fetal, abdominal, intra-operative (non-neurological), pediatric, small organ, cephalic (adult), trans-rectal, trans-vaginal, musculo-skeletal (conventional, superficial), urology, cardiac (adult, pediatric), peripheral vessel, carotid, and procedural guidance of needles into the body.
The system is a transportable ultrasound system intended for use in environments where is provided by trained healthcare professionals.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary - K213436
This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR § 807.92.
| Subject Device Trade Name: | Clarius Ultrasound Scanner |
|-------------------------------|-----------------------------------------------------------|
| Subject Device Model Numbers: | C3 HD3, C7 HD3, EC7 HD3, L7 HD3, L15 HD3, PA HD3, L20 HD3 |
| Common Name: | Diagnostic Ultrasound System and Accessories |
#### Regulation Number, Name and Product Codes:
| Regulation Number | Regulation Name | Product Code |
|-------------------|------------------------------------------|--------------|
| 21 CFR § 892.1550 | Ultrasonic Pulsed Doppler Imaging System | IYN |
| 21 CFR § 892.1560 | Ultrasound Pulsed Echo Imaging System | IYO |
| 21 CFR § 892.1570 | Diagnostic Ultrasonic Transducer | ITX |
| FDA 510(k) Review Panel: | Radiology |
|-------------------------------|-------------------------------------------------------------------------------------|
| Classification: | Class II |
| Manufacturer: | Clarius Mobile Health Corp.<br>130-2985 Virtual Way<br>Vancouver, BC V5M 4X7 Canada |
| Contact Name: | Agatha Szeliga<br>Director, Regulatory Affairs<br>agatha.szeliga@clarius.com |
| Date 510(k) Summary Prepared: | October 20, 2021 |
# Predicate Device Information:
| Device Trade Name: | Clarius Ultrasound Scanner |
|---------------------------------|---------------------------------------------------------|
| 510(k) Reference: | K192107 |
| Submitter Name: | Clarius Mobile Health Corp. |
| Regulation Name: | Ultrasonic Pulsed Doppler Imaging System |
| Classification Product Code(s): | IYN |
| Subsequent Product Codes | IYO, ITX |
| Regulation Number: | 21 CFR § 892.1550; 21 CFR § 892.1560; 21 CFR § 892.1570 |
| Classification: | Class II |
Note: The predicate device has not been subject to a design-related recall.
{4}------------------------------------------------
#### Device Description
The Clarius Ultrasound Scanner is a portable, general-purpose, software-controlled, diagnostic ultrasound system used to acquire and display high-resolution, real-time ultrasound data through an offthe-shelf (OTS) iOS or Android device. The Clarius Ultrasound Scanner comprises a series of wireless transducers employing Bluetooth and Wi-Fi-based technology to communicate with traditional tablet/smartphone devices via direct Wi-Fi. This allows the user to export ultrasound images and display them across a range of portable personal devices.
The system is a transportable ultrasound system intended for use in environments where is provided by trained healthcare professionals, including the emergency medical services (EMS) environment.
The Clarius Ultrasound Scanner is intended for use by qualified healthcare practitioners (e.g., doctors, nurses, sonographers) who are trained in the use of ultrasound imaging technology.
| Transducers/ Scanners (various models) | C3 HD3, C7 HD3, EC7 HD3, L7 HD3, L15 HD3, PA HD3, L20 HD3 |
|----------------------------------------|---------------------------------------------------------------------------------------------------|
| Software | Clarius Ultrasound App (Clarius App) for iOS;<br>Clarius Ultrasound App (Clarius App) for Android |
| Accessories | Clarius Charger<br>Clarius Fan |
The Clarius Ultrasound Scanner comprises the following:
#### Indications for Use for the Clarius Ultrasound Scanner
The Clarius Ultrasound Scanner is a software-based ultrasound imaging system and accessories, intended for diagnostic imaging. It is indicated for diagnostic ultrasound imaging and fluid flow analysis in the following applications: ophthalmic, fetal, abdominal, intra-operative (non-neurological), pediatric, small organ, cephalic (adult), trans-rectal, trans-vaginal, musculo-skeletal (conventional, superficial), urology, gynecology, cardiac (adult, pediatric), peripheral vessel, carotid, and procedural guidance of needles into the body.
The system is a transportable ultrasound system intended for use in environments where is provided by trained healthcare professionals.
| Criteria | SUBJECT DEVICE | PREDICATE DEVICE |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Clarius Ultrasound Scanner | Clarius Ultrasound Scanner |
| 510(k) Holder/<br>Manufacturer | Clarius Mobile Health Corp. | Clarius Mobile Health Corp. |
| Submission<br>Reference | Current Submission | K192107 |
| 510(k) Track | Track 3 | Track 3 |
| Product Codes | IYN1, IYO2, ITX3 | IYN1, IYO2, ITX3 |
| Criteria | SUBJECT DEVICE | PREDICATE DEVICE |
| | Clarius Ultrasound Scanner | Clarius Ultrasound Scanner |
| Regulation | 21 CFR 892.15501 | 21 CFR 892.15501 |
| Number | 21 CFR 892.15602 | 21 CFR 892.15602 |
| | 21 CFR 892.15703 | 21 CFR 892.15703 |
| Regulation Name | Ultrasonic Pulsed Doppler Imaging System1; | Ultrasonic Pulsed Doppler Imaging System1; |
| | Ultrasonic Pulsed Echo Imaging System2; | Ultrasonic Pulsed Echo Imaging System2; |
| | Diagnostic Ultrasonic Transducer3 | Diagnostic Ultrasonic Transducer3 |
| Transducer | C3 HD3, C7 HD3, EC7 HD3, L7 HD3, L15 HD3, | C3 HD, C7 HD, EC7 HD, L7 HD, L15 HD, PA HD, |
| Models | PA HD3, L20 HD3 | L20 HD |
| Transducer Types | Convex Array | Convex Array |
| | Linear Array | Linear Array |
| | Phased Array | Phased Array |
| | Intracavity | Intracavity |
| Intended Use | Diagnostic ultrasound imaging and fluid flow | Diagnostic ultrasound imaging and fluid flow |
| Indications for | analysis of the human body<br>The Clarius Ultrasound Scanner is a software- | analysis of the human body<br>The Clarius Ultrasound Scanner is a software- |
| Use and Clinical | based ultrasound imaging system and | based ultrasound imaging system and |
| Usage | accessories, intended for diagnostic imaging. | accessories, intended for diagnostic imaging. |
| | It is indicated for diagnostic ultrasound | It is indicated for diagnostic ultrasound |
| | imaging and fluid flow analysis in the | imaging and fluid flow analysis in the |
| | following applications: ophthalmic, fetal, | following applications: ophthalmic, fetal, |
| | abdominal, intra-operative (non- | abdominal, intra-operative (non- |
| | neurological), pediatric, small organ, cephalic | neurological), pediatric, small organ, cephalic |
| | (adult), trans-rectal, trans-vaginal, musculo- | (adult), trans-rectal, trans-vaginal, musculo- |
| | skeletal (conventional, superficial), urology, | skeletal (conventional, superficial), urology, |
| | gynecology, cardiac (adult, pediatric), | gynecology, cardiac (adult, pediatric), |
| | peripheral vessel, carotid, and procedural | peripheral vessel, carotid, and procedural |
| | guidance of needles into the body. | guidance of needles into the body. |
| | The system is a transportable ultrasound | The system is a transportable ultrasound |
| | system intended for use in environments | system intended for use in environments |
| | where healthcare is provided by trained | where healthcare is provided by trained |
| | healthcare professionals. | healthcare professionals. |
| | Ophthalmic Fetal Abdominal Intraoperative (Ab/Vasc) Pediatric Small Organ Adult cephalic Trans-rectal Trans-vaginal Musculoskeletal (conventional) Musculoskeletal (superficial) Urology Gynecology Cardiac adult | Ophthalmic Fetal Abdominal Intraoperative (Ab/Vasc) Pediatric Small Organ Adult cephalic Trans-rectal Trans-vaginal Musculoskeletal (conventional) Musculoskeletal (superficial) Urology Gynecology Cardiac adult |
| Criteria | SUBJECT DEVICE | PREDICATE DEVICE |
| | Clarius Ultrasound Scanner | Clarius Ultrasound Scanner |
| | Cardiac pediatric Peripheral vessel Carotid Needle guidance | Cardiac pediatric Peripheral vessel Carotid Needle guidance |
| Principle of<br>Operation | Piezoelectric material in the system's<br>transducer transmits high frequency, non-<br>ionizing sound waves to the designated<br>region of the body and converts the<br>subsequent echoes detected to electronic<br>signals in order to construct an image of the<br>internal structures of an anatomical field. The<br>image is sent wirelessly from the transducer<br>to an external iOS or Android viewing device<br>on which the image can be displayed. | Piezoelectric material in the system's<br>transducer transmits high frequency, non-<br>ionizing sound waves to the designated<br>region of the body and converts the<br>subsequent echoes detected to electronic<br>signals in order to construct an image of the<br>internal structures of an anatomical field. The<br>image is sent wirelessly from the transducer<br>to an external iOS or Android viewing device<br>on which the image can be displayed. |
| Power Source | Internal integrated built-in (non-removable)<br>lithium-ion battery | Removable lithium-ion battery |
| Display | iOS or Android mobile device | iOS or Android mobile device |
| Wireless<br>Capability | Communicates wirelessly via Wi-Fi and<br>Bluetooth | Communicates wirelessly via Wi-Fi and<br>Bluetooth |
| Portability | Portable ultrasound system | Portable ultrasound system |
| System<br>Components | Transducers/scanners<br>Software (Clarius App)<br>Accessories (Charger and Fan) | Transducers/scanners<br>Software (Clarius App)<br>Accessories (Charger and Fan) |
| Modes of<br>Operation | B-mode<br>M-mode<br>Color Doppler<br>Power Doppler<br>Pulse-Wave Doppler (PWD)<br>Combined (B+M; B+CD; B+PD; B+PWD) | B-mode<br>M-mode<br>Color Doppler<br>Power Doppler<br>Pulse-Wave Doppler (PWD)<br>Combined (B+M; B+CD; B+PD; B+PWD) |
| Safety Standards | The Clarius Ultrasound Scanner complies with<br>the following safety standards:<br>60601-1<br>60601-1-2<br>60601-1-6<br>60601-1-12<br>60601-2-37 | The Clarius Ultrasound Scanner complies with<br>the following safety standards:<br>60601-1<br>60601-1-2<br>60601-1-6<br>60601-1-12<br>60601-2-37 |
| Standard | Title of Standard | |
| IEC 60601-1:2012 | Medical electrical equipment – Part 1: General requirements for basic safety | |
| | and essential performance | |
| IEC 60601-1-2:2014 | Medical electrical equipment - Part 1-2: General requirements for basic safety | |
| | and essential performance - Collateral Standard: Electromagnetic | |
| | Compatibility - Requirements and tests | |
| IEC 60601-1-6 Edition 3.1 2013- | Medical electrical equipment - Part 1-6: General Requirements for Basic Safety | |
| 10 | and Essential Performance - Collateral Standard: Usability | |
| IEC 60601-1-12 Edition 1.0 2014- | Medical electrical equipment - Part 1-12: General Requirements for Basic | |
| 06 | Safety and Essential Performance - Collateral Standard: Requirements for | |
| | Medical Electrical Equipment and Medical Electrical Systems Intended for Use | |
| | in the Emergency Medical Services Environment | |
| IEC 60601-2-37:2015 | Medical electrical equipment - Part 2-37: Particular requirements for the basic | |
| | safety and essential performance of ultrasonic medical diagnostic and | |
| | monitoring equipment | |
| ISO 10993-1:2018 | Biological evaluation of medical devices – Part 1: Evaluation and testing within | |
| | a risk management process | |
| AIUM/NEMA UD 2-2004 | NEMA Standards Publication UD 2-2004 (R2009) Acoustic Output | |
| | Measurement Standard for Diagnostic Ultrasound Equipment Revision 3. | |
| | (Radiology) | |
| AIUM/NEMA UD 3-2004 | NEMA Standards Publication UD 3-2004 (R2009) Standard for Real-Time | |
| | Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.