← Product Code [KTI](/productcode/KTI) · K213434

# Elation Pulmonary Balloon Dilation Catheter (K213434)

_Merit Medical Systems, Inc. · KTI · Mar 8, 2022 · Ear, Nose, Throat · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K213434

## Device Facts

- **Applicant:** Merit Medical Systems, Inc.
- **Product Code:** [KTI](/productcode/KTI.md)
- **Decision Date:** Mar 8, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 874.4680
- **Device Class:** Class 2
- **Review Panel:** Ear, Nose, Throat
- **Attributes:** Therapeutic

## Indications for Use

The Elation™ Pulmonary Balloon Dilation Catheter is intended to be used to endoscopically dilate strictures of the trachea and bronchi.

## Device Story

Multi-lumen catheter with distal dilation balloon; designed for passage through bronchoscope working channel. Compatible with 0.035 in guidewire. Features two redundant inflation lumens for balloon control. Includes endoscopic and radiopaque marker bands; glow-in-the-dark shaft tag indicates balloon diameter and pressure. Used by physicians in clinical settings to dilate airway strictures. Device expands strictures via balloon inflation, potentially improving patient airway patency.

## Clinical Evidence

Bench testing only; no human clinical data. Non-clinical testing included balloon length, diameter, burst pressure, pressure cycling (10 cycles), leak testing, material attachment, and tensile testing (tip, proximal, wire). Design validation performed in porcine lungs to demonstrate safety of dilating bronchi 6mm or smaller. All tests met predetermined acceptance criteria.

## Technological Characteristics

Multi-lumen catheter; 7.5F diameter; 100 cm working length. Features radiopaque and endoscopic marker bands. Includes glow-in-the-dark shaft tag for diameter/pressure identification. Mechanical dilation principle via balloon inflation. Compatible with 0.035 in guidewire.

## Regulatory Identification

A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.

## Predicate Devices

- Elation™ Pulmonary Balloon Dilation Catheter ([K161392](/device/K161392.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 8, 2022

Merit Medical Systems, Inc. David Thomas Principal Regulatory Affairs Specialist 1600 West Merit Parkway South Jordan, Utah 84095

### Re: K213434

Trade/Device Name: Elation Pulmonary Balloon Dilation Catheter Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible Or Rigid) And Accessories Regulatory Class: Class II Product Code: KTI Dated: January 14, 2022 Received: January 18, 2022

#### Dear David Thomas:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

Submission Number (if known)

K213434

Device Name

Elation™ Pulmonary Balloon Dilation Catheter

Indications for Use (Describe)

The Elation™ Pulmonary Balloon Dilation Catheter is intended to be used to endoscopically dilate strictures of the trachea and bronchi.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

| 510(k) #: K213434 | 510(k) Summary | Prepared on: 2022-01-13 |
|-------------------|----------------|-------------------------|
|-------------------|----------------|-------------------------|

| Contact Details | 21 CFR 807.92(a)(1) |
|-----------------|---------------------|
|-----------------|---------------------|

| Applicant Name              | Merit Medical Systems, Inc.                                 |
|-----------------------------|-------------------------------------------------------------|
| Applicant Address           | 1600 West Merit Parkway South Jordan UT 84095 United States |
| Applicant Contact Telephone | 8013164956                                                  |
| Applicant Contact           | Mr. David Thomas                                            |
| Applicant Contact Email     | david.thomas@merit.com                                      |

| Device Name | 21 CFR 807.92(a)(2) |
|-------------|---------------------|
|-------------|---------------------|

| Device Trade Name   | Elation™ Pulmonary Balloon Dilation Catheter     |
|---------------------|--------------------------------------------------|
| Common Name         | Bronchoscope (flexible or rigid) and accessories |
| Classification Name | Bronchoscope Accessory                           |
| Regulation Number   | 874.4680                                         |
| Product Code        | KTI                                              |

| Legally Marketed Predicate Devices | 21 CFR 807.92(a)(3) |
|------------------------------------|---------------------|
|------------------------------------|---------------------|

| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
|-------------|----------------------------------------------------------|--------------|
| K161392     | Elation™ Pulmonary Balloon Dilation Catheter             | KTI          |

## Device Description Summary

The Pulmonary Balloon Catheters are multi-lumen catheters with a dilation balloon on the distal tip. The catheters have two extension legs with luer connectors on the proximal end. The catheter is designed to pass through the working channel of a bronchoscope and accept a 0.035 in (0.89 mm) X 260 cm guidewire lumen. Two redundant inflation lumens are used to inflate and deflate the dilation balloon.

The dilation balloon will be available in a 7.5F catheter with a length of 1.5 cm and in one additional balloon diameter of 4mm range that will be added to the current range from to 20 mm. Each balloon size will inflate to three different diameters for the specified inflation pressures. The balloon will be identifiable with both endoscopic and radiopague marker bands. A glow-in-the-dark tag that can be read in low light conditions is attached to the catheter shaft. The tag indicates diameter and corresponding pressure of the balloon.

## Intended Use/Indications for Use

The Elation™ Pulmonary Balloon Dilation Catheter is intended to endoscopically dilate strictures of the trachea and bronchi.

## Indications for Use Comparison

The indications for use has not changed from the predicate device.

# 21 CFR 807.92(a)(4)

ures of the trachea and bronchi.

21 CFR 807.92(a)(5)

## 21 CFR 807.92(a)(5)

{4}------------------------------------------------

## Technological Comparison

The Elation® Pulmonary Balloon Dilation Catheter (4mm) size is the same as the currently marketed Elation Pulmonary Balloon Dilation Catheter except for the smaller diameter range for the balloon of (4-5-6) mm diameter ranges offered and the shorter length of 15mm. The currently marketed Elation® Pulmonary Balloon Dilation Catheter is offered in lengths of 20, 30 and 55mm. There are no changes to the materials, principal of operation or indications for uses the same 7.5F catheter size with a working length of 100 cm as the current Elation® Pulmonary Balloon Dilation Catheter.

#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)

The following non-clinical tests were performed to support the substantial equivalence for this product:

- · Balloon length
- · Balloon diameter
- · Burst pressure
	- o Pressure cycling and leak testing (10 cycles at largest diameter/pressure)
	- o Leak testing
	- o Material attachment
- · Tip Tensile
- · Proximal Tensile
- · Wire Tensile

In addition, a design validation study was performed with porcine lungs to demonstrate the safety of diation of smaller (6mm diameter or below) bronchi.

All tests passed the predetermined acceptance criteria.

---

**Source:** [https://fda-staging.innolitics.com/device/K213434](https://fda-staging.innolitics.com/device/K213434)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
