← Product Code [OXZ](/productcode/OXZ) · K213427

# PRIMED Pediatric Facemask (K213427)

_Primed Medical Products, Inc. · OXZ · Nov 22, 2022 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K213427

## Device Facts

- **Applicant:** Primed Medical Products, Inc.
- **Product Code:** [OXZ](/productcode/OXZ.md)
- **Decision Date:** Nov 22, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Pediatric

## Indications for Use

The PRIMED level 1 Pediatric Facemask is a single use, disposable device that is intended to be used by patient/child 4 to 12 years old in healthcare setting under an adult supervision to provide a barrier for the respiratory tract from microorganisms, body fluids and particulate material.

## Device Story

Single-use, disposable pediatric facemask; 4-layer non-woven construction; pleated design; sized for children ages 4-12. Provides respiratory barrier against microorganisms, body fluids, and particulates. Used in healthcare settings under adult supervision. Protects patient by reducing transmission of contaminants. Non-sterile.

## Clinical Evidence

Bench testing only. Device tested against ASTM F2100:2020 (Level 1), ASTM F2101 (BFE >99%), ASTM F2299 (PFE >99%), ASTM F1862 (>80 mmHg), EN 14683 (differential pressure <3.7 mmH2O/cm²), and 16 CFR 1610 (Class 1 flammability). Biocompatibility confirmed via ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation/sensitization). Safety for pediatric use verified via CPSC standards for sharp edges, lead content, and phthalates.

## Technological Characteristics

4-layer non-woven fabric construction; ultrasonically bonded; pleated form factor. Sized for pediatric faces (ages 4-12). Non-sterile. Latex-free. Meets ASTM F2100 Level 1 requirements. No electronic components or software.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- Dukal level 1 Pediatric face mask ([K210321](/device/K210321.md))
- Kimberly-Clark Child's Facemask ([K113340](/device/K113340.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 22, 2022

PRIMED Medical Products Inc. Mitra Fard Regulatory Affairs Specialist 200, 2003-91 St. SW Edmonton, AB T6X 0W8 Canada

# Re: K213427

Trade/Device Name: PRIMED Pediatric Facemask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: OXZ Dated: October 26, 2022 Received: October 27, 2022

### Dear Ms. Fard:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Katherine D. Kavlock -S

for

Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K213427

Device Name PRIMED Pediatric Facemask

#### Indications for Use (Describe)

The PRIMED level 1 Pediatric Facemask is a single use, disposable device that is intended to be used by patient/child 4 to 12 years old in healthcare setting under an adult supervision to provide a barrier for the respiratory tract from microorganisms, body fluids and particulate material.

| Type of Use (Select one or both, as applicable)                                                               |                                                                                                                         |
|---------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| <div style="display:inline-block"><input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div style="display:inline-block"><input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

| Summary Date:                                          | November 21, 2022                                                                                                                                                                                                                       |
|--------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer Information:                              | PRIMED Medical Products Inc.                                                                                                                                                                                                            |
| Address:                                               | 200, 2003-91 St. SW, Edmonton, AB,<br>Canada, T6X 0W8                                                                                                                                                                                   |
| Phone Number<br>Fax Number<br>Email<br>Contact person: | +1.877.877.4633<br>+1.780.497.7670<br>mitra.fard@primed.ca<br>Mitra Fard                                                                                                                                                                |
| Common name of the device                              | Pediatric Facemask                                                                                                                                                                                                                      |
| Trade or Proprietary Name:                             | PRIMED Pediatric Facemask                                                                                                                                                                                                               |
| Classification name for the device                     | Surgical Apparel                                                                                                                                                                                                                        |
| Classification regulation:<br>Class:                   | 878.4040<br>2                                                                                                                                                                                                                           |
| Product Code:                                          | OXZ (Pediatric/Child Facemask)                                                                                                                                                                                                          |
| Predicate device(s)                                    | Dukal level 1 Pediatric face mask 510(k)#: K210321<br>Kimberly-Clark Child's Facemask 510(k)#: K113340                                                                                                                                  |
| Device Description                                     | The PRIMED level 1 Pediatric Facemask is a single use, disposable<br>device intended to be used by patient/child 4 to 12 years old in<br>healthcare setting under an adult supervision to prevent the<br>transmission of microorganism. |

{4}------------------------------------------------

#### Product Description:

PRIMED Pediatric Mask (PG4-1680) is composed of 4 layers of non-woven material bounded together and pleated to form a mask. This mask is appropriately sized to the smaller faces of children between ages of 4 to 12 years old. PRIMED pediatric mask meets the requirements of ASTM F2100 for level 1 procedure mask while providing a lower differential pressure this mask is single use, disposable and is provided non-sterile. All of materials used in this device are typical materials commonly used in the construction of legally marketed surgical/ procedure masks with a safe history of use. The mask is not made with natural rubber latex.

#### Available Models

PG4-1680 Pediatric Mask, ASTM Level 1, Earloop, Dinosaur pattern,

#### Indication for Use:

The PRIMED level 1 Pediatric Facemask is a single use, disposable device that is intended to be used by patient/child 4 to 12 years old in healthcare setting under an adult supervision to provide a barrier for the respiratory tract from microorganisms, body fluids and particulate material.

| Elements of<br>comparison                                | Proposed PRIMED Pediatric mask                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Primary Predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Secondary Predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Comparison |
|----------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Manufacturer                                             | PRIMED Medical Products Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Kimberly-Clark                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Dukal                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |            |
| 510(k) Number                                            | K213427                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | K113340                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | K210321                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |            |
| Product Code:                                            | OXZ                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | OXZ                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | OXZ                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Similar    |
| Regulation<br>number                                     | 21 CFR 878.4040                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | 21 CFR 878.4040                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 21 CFR 878.4040                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Similar    |
| Indication for<br>use                                    | The PRIMED level 1 Pediatric<br>Facemask is a single use,<br>disposable device that is intended<br>to be used by patient/child 4 to 12<br>years old in healthcare setting<br>under an adult supervision to<br>provide a barrier for the<br>respiratory tract from<br>microorganisms, body fluids and<br>particulate material.                                                                                                                                                                                                                                                      | The Kimberly-Clark Pediatric/ Child<br>Facemask is intended to be worn by<br>the patient/child (recommended<br>ages 4-12) to provide protection for<br>the respiratory tract. It is a single use,<br>disposable device that is provided<br>non-sterile. This Face Mask is<br>recommended for use in a healthcare<br>selling with appropriate adult<br>supervision                                                                                                                                                                                                                            | The Dukal Level 1 Pediatric Face Masks<br>are intended to be worn by the<br>patient/child to cover the nose and<br>mouth to provide a barrier for the<br>respiratory tract from microorganisms,<br>body fluids, and particulate material.<br>Recommended ages are 4-12. The face<br>mask is specifically for use with patients<br>whose age or illness may prevent them<br>from taking necessary precautions in<br>situations where transfer of<br>microorganisms, body fluids, and<br>particulates can occur. The face masks<br>are recommended for use in a healthcare<br>setting with appropriate adult<br>supervision. The Dukal Level 1 Pediatric                                                                       | Similar    |
| Elements of<br>comparison                                | Proposed PRIMED Pediatric mask                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Primary Predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Secondary Predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Comparison |
|                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Face Masks are single use, disposable<br>devices that are provided non-sterile.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |            |
| Material                                                 | Non- woven                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Non-woven                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Non-woven                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Similar    |
| Bacterial<br>Filtration<br>(ASTM F2101)                  | >99%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | 99.6%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | >99%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Similar    |
| Particulate<br>Filtration<br>(ASTM F2299)                | >99%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | 98.5%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | >99%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Similar    |
| Differential<br>Pressure (MIL<br>M 3654C)                | <3.7 mmH2O/cm²                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | 2.6 mmH2O/cm²                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | <3.1 mmH2O/cm²                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Similar    |
| Fluid<br>Resistance<br>(ASTM F1862)                      | > 80mmHg                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Not applicable                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | > 80mmHg                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Similar    |
| Flammability<br>(16 CFR 1610)                            | Class I                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Class I                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Class I                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Similar    |
| Cytotoxicity                                             | Non-cytotoxic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Non-cytotoxic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Non-cytotoxic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Similar    |
| Irritation                                               | Not an irritant                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Not an irritant                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Not an irritant                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Similar    |
| Sensitization                                            | Not a sensitizer                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Not a sensitizer                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Not a sensitizer                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Similar    |
| Migration of<br>certain<br>Elements)<br>EN71-3           | Pass                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Pass                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Tested under EN 71.3                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Similar    |
| Total lead<br>content                                    | Tested under CPSC-CH-E1002-08.3                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Not Available                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Tested under CPSC-CHE1002-08.3 &<br>CPSC-CH-E1001-08.3                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Similar    |
| Phthalate<br>Analysis                                    | Tested under CPSC-CH-C1001-09.3                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Not Available                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Tested under CPSC-CHC1001-09.4                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Similar    |
| Sharp edge<br>hazard                                     | Meet the requirements of 16 CFR<br>1500.48 and 16 CFR 1500.49 when<br>after simulating use and abuse of<br>toys as per 16CFR 1500.53                                                                                                                                                                                                                                                                                                                                                                                                                                               | Not Available                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | ested under ASTM F963-17 (sec. 4.6, 4.7,<br>and 4.9), 16 CFR Part 1500, 16 CFR Part<br>1501                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Similar    |
| Mask<br>construction<br>and<br>technological<br>features | PRIMED Pediatric Child mask is<br>generally composed of a filter<br>layer that is ultrasonically bonded<br>between layers of nonwoven<br>fabric. This mask is appropriately<br>sized to the smaller faces of<br>children between ages of 4 to 12<br>years old. The PRIMED single use,<br>non-sterile Pediatric/Child<br>Facemask meet requirements of<br>level 1 masks as per ASTM F2100.<br>All of materials used in this device<br>are typical materials commonly<br>used in the construction of legally<br>marketed surgical masks with a<br>safe history of use. The masks are | The Kimberly-Clark Pediatric/Child<br>Facemask is a three layer mask,<br>constructed of nonwoven polyester<br>blends and polypropylene materials.<br>Bindings are nonwoven polyester<br>and earloops are knitted<br>polyester/lycra. A malleable<br>nosepiece is placed within the<br>bindings for comfort and<br>individualized fit around the<br>wearer's nose. The Pediatric/Child<br>Facemask is appropriately sized to<br>the smaller faces of children across a<br>diverse population. The<br>Pediatric/Child Facemask is a single<br>use, disposable device, provided<br>non-sterile. | The Dukal Level 1 Pediatric Face Mask is<br>manufactured using ultrasonic hot<br>sealing, composed of three layers of<br>materials and pleated to form the mask.<br>The inner layer is composed of<br>Polypropylene (Spunbond), the middle<br>layer is Polypropylene (Meltblown) filter<br>material, and the outer layer is<br>polypropylene (Spunbond). Decorative<br>patterns are printed with colored inks.<br>Masks are held in place on wearer with<br>polyester/spandex elastic earloop and<br>contain a malleable aluminum wire nose<br>piece. The Pediatric/Child's Face Mask is<br>appropriately sized to the smaller faces<br>of children across a diverse population.<br>All of the materials used in this device | Similar    |
| Elements<br>of<br>comparison                             | Proposed PRIMED Pediatric mask                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Primary Predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Secondary Predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Comparison |
|                                                          | not made with natural rubber<br>latex.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | are typical materials commonly used in<br>the construction of Surgical Masks and<br>are being used in current legally<br>marketed devices. This product is not<br>made with natural rubber latex.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |            |
| Anthropometry                                            | Based on an anthropometric study<br>sample that is representative of<br>the US population in both gender<br>and racial distribution, the Dukal<br>Level 1 Pediatric Face Mask will<br>provide adequate coverage to<br>children 4 to 12 years old with face<br>width of 8.5 to 12.8 cm and lower<br>face length of 8.5 to 11.4 cm                                                                                                                                                                                                                                                   | Not available                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Based on an anthropometric study<br>sample that is representative of the US<br>population in both gender and racial<br>distribution, the Dukal Level 1 Pediatric<br>Face Mask will provide adequate<br>coverage to children between the ages<br>of 4-12 years old of weight between 25-<br>176 pounds, and of height between<br>3′00″ — 5′08″".                                                                                                                                                                                                                                                                                                                                                                              | Similar    |

#### Summary of the Technological Characteristics of PRIMED Pediatric Facemask and the predicate

{5}------------------------------------------------

{6}------------------------------------------------

# 510(k) Summary

# Summary of Testing:

PRIMED pediatric masks are tested to below listed standards:

| Standard             |                                                                                                                                                                                                    | Result                                                   |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------|
| ASTM F2100:2020      | Standard Specification for Performance of Materials<br>Used in Medical Face Masks                                                                                                                  | Meet and exceed requirement<br>of Level 1 procedure mask |
| ASTM F2101: 2019     | Standard Test Method for Evaluating the Bacterial<br>Filtration Efficiency (BFE) of Medical Face Mask<br>Materials, Using a Biological Aerosol of Staphylococcus<br>aureus                         | >99 %                                                    |
| ASTM F2299:2003      | Test Method for Determining the Initial Efficiency of<br>Materials Used in Medical Face Masks to Penetration by<br>Particulates Using Latex Spheres                                                | >99%                                                     |
| ASTM F1862: 2017     | Standard Test Method for Resistance of Medical Face<br>Masks to Penetration by Synthetic Blood                                                                                                     | >80 mmHg                                                 |
| EN 14683:2019        | Medical Face Masks—Requirements and Test Methods:<br>Determine breathing resistance or differential pressure<br>as directed in EN 14683:2019, Annex C.                                             | < 3.7 mmH2O/cm²                                          |
| 16 CFR part 1610     | Standard for the Flammability of Clothing Textiles                                                                                                                                                 | Class 1                                                  |
| ISO 10993-5:2009     | Biological evaluation of medical devices - Part 5: Tests<br>for in vitro cytotoxicity                                                                                                              | Pass                                                     |
| ISO 10993-10:2010    | Biological evaluation of medical devices - Part 10: Tests<br>for irritation and skin sensitization                                                                                                 | Pass                                                     |
| EN 71-3: 2013        | Migration of certain elements                                                                                                                                                                      | pass                                                     |
| CPSC 16 CFR 1500.48  | Technical requirements for determining a sharp point in<br>toys and other articles for use by children under 8 years<br>old after simulating use and abuse of device as per CPSC<br>16 CFR 1500.53 | pass                                                     |
| CPSC 16 CFR 15003.49 | Technical requirements for determining a sharp metal or<br>glass edge in toys and other articles for use by children                                                                               | Pass                                                     |

{7}------------------------------------------------

# PRIMED Pediatric Facemask

# 510(k) Summary

| Standard           |                                                                                           | Result |
|--------------------|-------------------------------------------------------------------------------------------|--------|
|                    | under 8 years old, after simulating use and abuse of<br>device as per CPSC 16 CFR 1500.53 |        |
| CPSC-CH-E1002-08.3 | Standard operation procedure for determining total<br>Lead in non-metal children product  | Pass   |
| CPSC-CH-C1001-09.3 | Analysis of Phthalate                                                                     | Pass   |

#### Conclusion:

Based on the test results and comparison of PRIMED Pediatric Facemasks with the predicate devices, it can be concluded that PRIMED Pediatric mask is substantially equivalent and as effective as the predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K213427](https://fda-staging.innolitics.com/device/K213427)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).