iSR'obot Mona Lisa 2.0 is a user-controlled, stereotaxic accessory intended to guide physicians in the planning and positioning of insertion tools, such as a third-party needle or a probe, during image-guided diagnostic and interventional procedures in conjunction with the guidance of transrectal ultrasound involving the prostate gland in a clinical setting. Examples of such procedures include, but are not limited to, image fusion for diagnostic clinical examinations and procedures, soft tissue biopsies, and soft tissue ablations. The iSR'obot Mona Lisa 2.0 provides 2D and 3D visualization of Ultrasound images and the ability to fuse and register these images with those from other imaging modalities such as Magnetic Resonance, etc. It also provides the ability to display a simulated image of an insertion tool on a computer monitor screen, the target organ, and the current and projected future path of the insertion tool taking into account patient movement. Other software features include multi-planar reconstruction, segmentation, image measurements, 2D/3D image registration, and reporting.
Device Story
iSR'obot Mona Lisa 2.0 is a stereotaxic accessory for prostate procedures; guides physicians in planning/positioning insertion tools (needles/probes). Inputs: live 2D ultrasound feeds from third-party systems; imported MRI images. Operation: workstation constructs 3D ultrasound stacks; fuses/registers MRI with ultrasound; displays simulated tool trajectory on monitor. Robotic arm (8-axis) holds/moves ultrasound probe and orients needle guidance mechanism. Physician manually inserts needle. Features: multi-planar reconstruction, segmentation, image measurements, patient movement adjustment (re-aligns prostate model to live ultrasound). Output: visual guidance for needle placement; procedure reports. Used in clinical settings by urologists/physicians. Benefits: facilitates accurate needle placement for biopsies/ablations; enables fusion of multi-modality imaging; allows for patient movement compensation.
Clinical Evidence
No clinical data submitted. Evidence based on bench testing, system verification, software validation, usability testing, and biocompatibility testing.
Technological Characteristics
Stereotaxic accessory with 8-axis motorized robotic arm and bed rail stabilizer. Materials: biocompatible needle guide holder (ISO 10993). Connectivity: PACS (DICOM). Software: multi-planar reconstruction, segmentation, 2D/3D registration, elastic fusion. Standards: IEC 62304 (software), IEC 62366-1 (usability), ANSI UL 2900-1/2-1 (cybersecurity), IEC 60601-1/1-2/1-6 (electrical safety/EMC).
Indications for Use
Indicated for physicians performing image-guided diagnostic and interventional procedures (e.g., biopsies, ablations) of the prostate gland using transrectal ultrasound guidance in a clinical setting.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG ADMINISTRATION" in blue as well. The FDA is a federal agency responsible for regulating and supervising the safety of food, drugs, and other products.
Biobot Surgical Pte Ltd % Lim Yan Shin Regulatory Affairs 79 Ayer Rajah Crescent, #04-05 Singapore. Singapore 139955 SINGAPORE
December 22, 2021
#### Re: K213411
Trade/Device Name: iSR'obot Mona Lisa 2.0 Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: Class II Product Code: IYO, ITX, LLZ, OIJ Dated: October 15, 2021 Received: October 19, 2021
Dear Lim Yan Shin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Jessica Lamb, Ph.D. Assistant Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
#### K213411
Device Name iSR'obot Mona Lisa 2.0
#### Indications for Use (Describe)
iSR'obot Mona Lisa 2.0 is a user-controlled, stereotaxic accessory intended to guide physicians in the planning and positioning of insertion tools, such as a third-party needle or a probe, during image-guided diagnostic and interventional procedures in conjunction with the guidance of transrectal ultrasound involving the prostate gland in a clinical setting. Examples of such procedures include, but are not limited to, image fusion for diagnostic clinical examinations and procedures, soft tissue biopsies, and soft tissue ablations.
The iSR'obot Mona Lisa 2.0 provides 2D and 3D visualization of Ultrasound images and the ability to fuse and register these images with those from other imaging modalities such as Magnetic Resonance, etc. It also provides the ability to display a simulated image of an insertion tool on a computer monitor screen, the target organ, and the current and projected future path of the insertion tool taking into account patient movement. Other software multiplanar reconstruction, segmentation, image measurements, 2D/3D image registration, and reporting.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------|----------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Biobot. The word "biobot" is written in lowercase letters. The "bio" portion of the word is in orange, and the "bot" portion of the word is in blue. The "bio" portion of the word is also slightly larger than the "bot" portion of the word.
# 510(k) Summarv
# K213411
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
This 510(k) summary of Safety and Effectiveness information is submitted in accordance with the requirement of 21 CFR Part 807.87(h).
| Date: | October 15, 2021 |
|--------------------|--------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Biobot Surgical Pte Ltd<br>79 Ayer Rajah Crescent,<br>#04-05, Singapore 139955 |
| Primary Contact: | Lim Yan Shin, Regulatory Affairs<br>Biobot Surgical Pte Ltd<br>Cell: +65 97263992<br>E-mail: yanshin.lim@ziggroup.com.sg |
| Secondary Contact: | Georgiann Keyport, Regulatory Consultant, U.S.<br>Email: gkeyport@canopyregulatory.com<br>Cell: 952-994-8267 |
#### Product Identification
| Device Trade Name: | iSR'obot Mona Lisa 2.0 |
|-----------------------|---------------------------------------------------------------------------------------------|
| Common / Usual Name: | System, image processing, radiological |
| Classification Names: | 892.1560 Ultrasonic pulsed echo imaging system<br>892.1570 Diagnostic ultrasonic transducer |
| Product Code: | IYO, ITX, LLZ, OIJ |
| Manufacturer / Design | Biobot Surgical Pte Ltd |
| Location: | 79 Ayer Rajah Crescent,<br>#04-05, Singapore 139955 |
### Device Description
The iSR'obot Mona Lisa 2.0 is user-controlled, stereotaxic accessory intended to guide physicians in the planning and positioning of insertion tools, such as a needle or a probe, during image-guided diagnostic and interventional procedures involving the prostate gland in a clinical setting. The device displays the 2D live image feeds from commercially available ultrasound systems and constructs 3D ultrasound image stacks. The system allows the importation of an MRI image to create a model of patient prostate by providing fusion between Ultrasound and Magnetic Resonance Imaging (MRI). The system is compatible with commercially available ultrasound systems, transrectal ultrasound bi-
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Biobot. The word "biobot" is written in blue, with the "b" and "i" connected. Above the word is an orange oval shape, with a small dot on the upper left side. The logo is simple and modern.
plane probes, and commercially available needle devices. Other software features include multi-planar reconstruction, segmentation, image measurements and 2D/3D image registration.
### Workstation
The workstation is connected to the ultrasound system via a standard cable and hardware connector and displays the 2D live image feed in a format that is compatible with the iSR 'obot Mona Lisa 2.0. Using the application software, the urologist or physician defines the apex and base limits of the prostate gland and the robotic arm moves the ultrasound probe within those limits to capture multiple 2D slices of the prostate gland to construct the 3D image stack. The urologist or physician may refine the constructed 3D image stack by indicating and confirming the planned lesion core or tumor location, anatomical markers within or around the prostate gland, and the prostate gland contour. In addition, the application software can utilize previously acquired images of the patient's prostate, which may include other image modalities like magnetic resonance images and register to this 3D image stack. The completed procedure information can be stored on a location selected by the urologist or physician, such as in the workstation or in a PACS server. The stored procedure information can be used for future review with the patient or examination. Previously stored biopsy procedural plan can be recalled and its 3D model may be aligned and registered to current 3D model of the prostate to facilitate planning of the interventional procedure.
#### Robotic Navigation Module
The robotic navigation module comprises of a robotic arm and bed rail stabilizer. The stabilizer is a mechanical device and is able to lock and release to position the robotic arm. One end of the stabilizer is first attached to the bed rail while the other end is used to mount the robotic arm so that the robotic arm is able to be positioned close to the patient's perineum while the patient is in a lithotomy position. The robotic arm is a motorized mechanical structure and has two key functions, which is to hold and move the ultrasound probe of a commercially available ultrasound system to display 2D live ultrasound image feeds and to orientate its needle guidance mechanism to facilitate insertion of a commercially available needle (based on a planned and simulated needle trajectory) by the urologist or physician. The movement of the ultrasound probe and the needle guidance mechanism is motorized. As a result, the workstation is able to construct and display a 3D image stack and rendered surface model of the prostate.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Biobot Analytics. The logo features the word "biobot" in a sans-serif font. The "bio" portion of the word is in a gradient orange color, while the "bot" portion is in blue. The logo has a modern and clean design.
During the procedure, the real time 2D ultrasound image is visible on the iSR 'obot Mona Lisa 2.0 display. After the robotic arm has orientated its trajectory according to the planned location, the urologist or physician manually inserts the needle into the prostate via the needle guidance mechanism. The 2D live ultrasound image may be marked up to record the actual locations where the needle has been inserted. The system is able to pivot and facilitate re-insertion of the needle if the marked up actual location is distant from the planned location. The urologist or physician is able to observe prostate shift on live ultrasound image, re-register the prostate model and needle location plan to the live ultrasound prostate image.
#### Needle Guide Holder
The needle guide holder is designed to be used with the iSR obot Mona Lisa 2.0. It is a component installed on the robotic arm functioning as a channel for insertion tools to go through. The robotic arm has a motorized mechanism to close and open the needle guide holder in order to hold and release the insertion tools. The needle guide holder may be manufactured to fit insertion tools with varying diameters.
#### Disposables
The iSR 'obot Kit is intended to be used with the iSR'obot Mona Lisa 2.0 as a component of a system for performing transperineal prostate procedure. The kit consists of probe sheath, drape and biopsy papers. The probe sheath functions as a sheath to a third-party ultrasound probe for ultrasound scanning of the prostate organ.
| Predicate Device Name | Predicate 510(k) Submission Reference |
|------------------------|---------------------------------------|
| iSR'obot Mona Lisa 1.0 | K203659 (Primary) |
| iSR'obot MRI-US Fusion | K161109 (Reference) |
| iSR'obot Biopsy Kit | K163502 (Reference) |
| Artemis | K162474 (Reference) |
## Predicate Device Information and Comparison
# Intended Use / Indications for Use
iSR'obot Mona Lisa 2.0 is a user-controlled, stereotaxic accessory intended to guide physicians in the planning and positioning of insertion tools, such as a third-party needle or a probe, during image-guided diagnostic and interventional procedures in conjunction with the guidance of transrectal ultrasound involving the prostate gland in a clinical setting.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Biobot. The word "biobot" is written in a sans-serif font, with "bio" in a bright orange color and "bot" in a dark blue color. The "bio" portion of the logo is set against a larger orange oval shape, giving the impression of a stylized sun or planet.
Examples of such procedures include, but are not limited to, image fusion for diagnostic clinical examinations and procedures, soft tissue biopsies, and soft tissue ablations.
The iSR 'obot Mona Lisa 2.0 provides 2D and 3D visualization of Ultrasound images and the ability to fuse and register these images with those from other imaging modalities such as Magnetic Resonance, etc. It also provides the ability to display a simulated image of an insertion tool on a computer monitor screen, the target organ, and the current and projected future path of the insertion tool taking into account patient movement. Other software features include multi-planar reconstruction, segmentation, image measurements, 2D/3D image registration, and reporting.
# Technology Characteristics Compared to Predicate Devices
iSR'obot Mona Lisa 2.0 employs the same fundamental scientific technology (design, function and specifications) as that of its legally marketed predicate device, iSR'obot Mona Lisa 1.0 (K203659).
Similarities in technology characteristics include:
- Platform-hosted motorized devices and are able to provide 2D and 3D views of . the prostate gland;
- Use the same technology to acquire a transrectal ultrasound image to plan and . guide a needle for diagnostic procedure;
- Control the trajectory and depth for needle placement via needle guiding • mechanism;
- . Fusion of Magnetic Resonance and Ultrasound images for mapping planning information of the prostate gland
- . Re-positioning of subsequent needle insertion for the same target lesion after the user has indicated the actual needle landing position from the prior insertion.
Modifications to iSR'obot Mona Lisa 2.0 include:
- The iSR'obot Mona Lisa 2.0 is intended for image-guided diagnostic and . interventional procedures while the legally marketed predicate device is intended only for image-guided diagnostic procedure;
- . Visualization of simulated images in an interventional procedure of 1) the lesions and their margins and 2) treatment zone;
- . Needle guide holder, which acts as a channel for insertion tool positioning, is detachable and can be cleaned and sterilized for subsequent use, and enables multi-needle insertion trajectory planning for interventional procedures;
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for Biobot Analytics. The logo features the word "biobot" in a stylized font. The "bio" portion of the word is in orange, while the "bot" portion is in blue. The "o" in "bio" is shaped like an oval.
- . Patient movement adjustment function for re-alignment of the prostate model with the live-ultrasound image in the event that prostate shifts due to patient movement during a procedure;
- . Connectivity to picture archiving and communication system (PACS) to download DICOM images and download/upload of procedure case; and
- . Provide an offline post-procedure review
Other changes include minor user interface variations such as graphics user interface (GUI) design. These differences do not significantly affect the function or use of the device, nor do they raise new or additional safety risks.
These changes are being implemented as a product improvement effort and not due to a corrective action or field action.
iSR'obot Mona Lisa 2.0 has the following technological characteristics, similar to the Artemis (K162474):
- . Same intended use for insertion tool planning and navigation for ultrasound imageguided prostate diagnostic and interventional procedures.
- . Connecting to standard ultrasound system to display and acquire 2D live ultrasound images of the prostate gland
- Perform other viewing and imaging-processing function such as image segmentation, image multi planar reconstruction, image overlays, and measurements
- . Image registration and elastic fusion of ultrasound image with previously created MRI prostate model and/or prior biopsy plan on application software
- . Generate 2D/3D model of a prostate model and trajectory of insertion tools
- . Plan insertion tool positions on the fused prostate model by physicians and mark the actual position of insertion tools
- . Connectivity to PACS to download DICOM images and download/upload of procedure case
- Artemis has a needle guide mechanism attachment similar to the iSR'obot Mona . Lisa 2.0. which allows insertion of third-party insertion tools
The differences between iSR'obot Mona Lisa 2.0 and Artemis (K162474) are that:
- . Anatomical Access - iSR'obot Mona Lisa 2.0 is intended for insertion tools positioned transperineally while Artemis is intended for insertion tools positioned transperineally and transrectally procedure.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for Biobot. The logo consists of the word "biobot" in blue, with the "bio" portion of the word in a larger font size than the "bot" portion. To the left of the word "biobot" is an orange oval shape that is slightly tilted to the left.
- . Positioning of insertion tools - iSR'obot Mona Lisa 2.0 has an 8 axis motorized robotic arm that navigates the needle guide holder based on the clinician's insertion tool position plan. The Artemis has a semi-robotic arm (mechanical arm with encoders) that clinicians may control to track and visualize the insertion tool as it is being positioned. Despite the slight difference, the clinician is in control throughout and manually inserts the insertion tool for both devices.
- Adjustment of prostate model in the event of patient movement iSR'obot ● Mona Lisa 2.0 provides an adjustment panel for users to manually adjust the prostate model to align with the ultrasound live image of the shifted prostate. Artemis tracks the position and orientation of the ultrasound probe, and users mark landmark features on the ultrasound live image to enable its software to automatically adjust the prostate model to align with the ultrasound live image of the shifted prostate.
- Information for active surveillance - Artemis is intended to be used for patients in active surveillance to keep track of previous procedures information and outcomes. iSR'obot Mona Lisa 2.0 allows previous procedures information to be accessed for patients under active surveillance, but the histopathological outcome is not currently captured in the software.
The technological characteristic above are features that assists physicians to plan and insert an insertion tool. The physician has full control of the tasks to be performed. According to the risk analysis, non-clinical performance tests, device verification and validation tests, the above differences do not result in additional or increase to existing known risks for iSR'obot Mona Lisa 2.0.
# Substantial Equivalence
The technological characteristics such as intended use, method of operation, general function and application of the iSR'obot Mona Lisa 2.0 are equivalent to the primary predicate iSR'obot Mona Lisa 1.0 (K203659),
iSR'obot Mona Lisa 1.0 (K203659), iSR'obot MRI-US Fusion (K161109) and iSR'obot Biopsy Kit (K163502) shall be collectively known as iSR 'obot Mona Lisa 1.0 System.
Risk analysis was conducted to evaluate the modifications and features update in iSR'obot Mona Lisa 2.0. All verification and validation activities were performed and results demonstrated substantial equivalence. Table 1 indicates the comparison between the subject device and predicate devices.
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the word "biobot" in a stylized font. The word is written in lowercase letters. The "bio" portion of the word is colored in orange, while the "bot" portion is in blue. The orange color appears to fade into the blue color.
| Table 1: Comparison Between Predicate Devices and Subject Device iSR'obot Mona Lisa<br>2.0 | | | |
|--------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technological<br>Characteristic | Reference Device: | Predicate Device: | Submitted Device: |
| | Eigen's Artemis<br>(K162474) | iSR'obot Mona Lisa<br>1.0 System (Primary -<br>K203659, Reference -<br>K161109, K163502) | iSR'obot Mona Lisa<br>2.0 |
| Intended Use | Artemis along with the<br>Needle Guide<br>Attachment is used for<br>image-guided<br>interventional and<br>diagnostic procedures<br>of the prostate gland. It<br>provides 2D and 3D<br>visualization of<br>Ultrasound (US)<br>images and the ability<br>to fuse and register<br>these images with those<br>from other imaging<br>modalities such as<br>Ultrasound, Magnetic<br>Resonance, Computed<br>Tomography, etc. It<br>also provides the ability<br>to display a simulated<br>image of a tracked<br>insertion tool such as a<br>biopsy needle,<br>guidewire or probe on a<br>computer monitor<br>screen that shows<br>images of the target<br>organ and the current<br>and the projected future<br>path of the<br>interventional<br>instrument taking into<br>account patient<br>movement. The<br>software also provides<br>a virtual grid on the live<br>ultrasound for | iSR'obot Mona Lisa 1.0<br>is intended for use by<br>trained urologist or<br>physician to perform<br>the computer-assisted<br>transperineal prostate<br>biopsy under transrectal<br>ultrasound guidance.<br>The device serves as a<br>biopsy needle guide<br>only. It shall be used in<br>conjunction with a third<br>party ultrasound<br>machine and endorectal<br>probe that supports B-<br>Mode, and a third party<br>prostate biopsy gun and<br>needle. The insertion of<br>biopsy needle will be<br>done by urologist. The<br>patient is administered<br>general anesthesia and<br>placed in a lithotomy<br>position. | iSR'obot Mona Lisa<br>2.0 is a user-controlled,<br>stereotaxic accessory<br>intended to guide<br>physicians in the<br>planning and<br>positioning of insertion<br>tools, such as a third-<br>party needle or a probe,<br>during image-guided<br>diagnostic and<br>interventional<br>procedures in<br>conjunction with the<br>guidance of transrectal<br>ultrasound involving<br>the prostate gland in a<br>clinical setting.<br>Examples of such<br>procedures include, but<br>are not limited to,<br>image fusion for<br>diagnostic clinical<br>examinations and<br>procedures, soft tissue<br>biopsies, and soft tissue<br>ablations.<br>The iSR'obot Mona<br>Lisa 2.0 provides 2D<br>and 3D visualization of<br>Ultrasound images and<br>the ability to fuse and<br>register these images<br>with those from other<br>imaging modalities<br>such as Magnetic |
| | performing systematic<br>sampling of the target<br>organ. Other software<br>features include patient<br>data management,<br>multi-planar<br>reconstruction,<br>segmentation, image<br>measurements, 2D/3D<br>image registration,<br>reporting, and<br>pathology management.<br>Artemis is intended for<br>treatment planning and<br>guidance for clinical,<br>interventional and/or<br>diagnostic procedures.<br>The device is intended<br>to be used in<br>interventional and<br>diagnostic procedures<br>in a clinical setting.<br>Example procedures<br>include, but are not<br>limited to image fusion<br>for diagnostic clinical<br>examinations and<br>procedures, soft tissue<br>biopsies, soft tissue<br>ablations and<br>placement of fiducial<br>markers. Artemis is<br>also intended to be used<br>for patients in active<br>surveillance to keep<br>track of previous<br>procedures information<br>and outcomes. | | Resonance, etc. It also<br>provides the ability to<br>display a simulated<br>image of an insertion<br>tool on a computer<br>monitor screen, the<br>target organ, and the<br>current and projected<br>future path of the<br>insertion tool taking<br>into account patient<br>movement. Other<br>software features<br>include multi-planar<br>reconstruction,<br>segmentation, image<br>measurements, 2D/3D<br>image registration, and<br>reporting. |
| Product Code | LLZ | IYO, ITX, LLZ, OIJ | IYO, ITX, LLZ, OIJ |
| Class | II | II | II |
| Target<br>Anatomy | Prostate | Prostate | Prostate |
| | | | |
| Anatomy<br>Access | Transperineal<br>Transrectal | Transperineal<br>Not available | Transperineal<br>Not available |
| Patient<br>population | Patients for a biopsy<br>procedure | Patients for a biopsy<br>procedure | Patients for a biopsy<br>procedure |
| | Patients for an<br>interventional<br>procedure | | Patients for an<br>interventional<br>procedure |
| | Patients in active<br>surveillance | | |
| Clinical Utility | Soft tissue biopsies, | Soft tissue biopsies | Soft tissue biopsies |
| | Soft tissue ablations | | Soft tissue ablations |
| | Placement of fiducial<br>markers | | |
| | Pathology management | | |
| | Patient data<br>management | | |
| Software | | | |
| Window OS | Yes | Yes | Yes |
| Medical<br>Imaging<br>Software | Yes | Yes | Yes |
| Compliance<br>with FDA<br>Cybersecurity | Yes | Yes | Yes (Updated)…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.