← Product Code [FMI](/productcode/FMI) · K213407

# UltiCare Disposable Pen Needles (K213407)

_Ultimed Incorporated · FMI · Dec 16, 2021 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K213407

## Device Facts

- **Applicant:** Ultimed Incorporated
- **Product Code:** [FMI](/productcode/FMI.md)
- **Decision Date:** Dec 16, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5570
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic, 3rd-Party Reviewed

## Indications for Use

The UltiCare™ Disposable Pen Needles are used with pen injector devices for the subcutaneous injection of drugs.

## Device Story

UltiCare Disposable Pen Needles are sterile, single-use, hypodermic single lumen needles; designed for use with pen injector devices for subcutaneous drug delivery. Device consists of double-ended cannula, needle hub, needle shield, and primary container. Available in gauges 29G to 32G and lengths 4mm to 12.7mm. Used by patients or clinicians in home or clinical settings. Attaches to pen injectors via screw threads; compatible with injectors complying with ISO 11608-1. Facilitates subcutaneous drug administration; benefits patient through reliable, sterile delivery mechanism.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including particulate evaluation (USP <788>), functional performance per ISO 11608-2:2012 (flow rate, bond strength, needle point sharpness, assembly ease), and biocompatibility per ISO 10993-1 (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, hemocompatibility).

## Technological Characteristics

Materials: Stainless Steel 304 (cannula), High-Density polyethylene (cover). Dimensions: 29G-32G, 4mm-12.7mm. Sterilization: Ethylene Oxide (EtO), SAL 10^-6. Standards: ISO 11608-2, ISO 9626, ISO 7864, ISO 10993-1. Connectivity: None (mechanical).

## Regulatory Identification

A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.

## Predicate Devices

- BD pen needles ([K162516](/device/K162516.md))

## Reference Devices

- UltiCare™ Disposable Pen Needles ([K100812](/device/K100812.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 16, 2021

UltiMed Incorporated % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114

Re: K213407

Trade/Device Name: UltiCare Disposable Pen Needles Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic single lumen needle Regulatory Class: Class II Product Code: FMI Dated: October 18, 2021 Received: October 19, 2021

Dear Prithul Bom:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

CAPT Alan M. Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

### K213407

Device Name UltiCare™ Disposable Pen Needles

Indications for Use (Describe)

The UltiCare™ Disposable Pen Needles are used with pen injector devices for the subcutaneous injection of drugs.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### TRADITIONAL 510(K) SUMMARY 5.0

| Submitted by:                      | UltiMed Inc.<br>350 Highway 7, Suite 100<br>Excelsior, MN 55331                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:                    | Cori Ragan<br>Principal Advisor<br>Labcorp<br>5353 Wayzata Boulevard, Suite 505<br>Minneapolis, Minnesota  55416<br>Telephone: 612-268-8746<br>Cori.Ragan@labcorp.com                                                                                                                                                                                                                                                                                                                                                                                     |
| Date of Summary:                   | 15 December 2021                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Device Trade Name:                 | UltiCare™ Disposable Pen Needles                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Common or Usual Name:              | Pen Needles                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Regulation Number:                 | 21 CFR §880.5570                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Classification Name:               | Hypodermic single lumen needles                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Product Code:                      | FMI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Device Classification:             | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Review Panel:                      | General Hospital                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Predicate/ Reference<br>Device(s): | Predicate Device - K162516 - BD pen needles<br>Reference Device - K100812 - UltiCare™ Disposable Pen<br>Needles                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Model Numbers:                     | 71006 (29G, 12.7mm)<br>61005 (30G, 5mm)<br>61008 (30G, 8mm)<br>71013 (31G, 5mm)<br>71004 (31G, 6mm)<br>71005 (31G, 8mm)<br>71003 (32G, 4mm)<br>71014 (32G, 6mm)                                                                                                                                                                                                                                                                                                                                                                                           |
| Device Description:                | The UltiCare™ Disposable Pen Needles are sterile, single-<br>use, hypodermic single lumen needles designed for use with<br>pen injectors for subcutaneous injection of a desired dose of<br>drugs approved for delivery using a pen needle. The<br>UltiCare™ Disposable Pen Needles consist of a double-<br>ended cannula, a needle hub, a needle shield and the needle<br>primary container. The UltiCare™ Disposable Pen Needles<br>are non-toxic and are available in a variety of needle sizes<br>(29 gauge to 32 gauge) and lengths (5/32" to 1/2"). |

{4}------------------------------------------------

#### The UltiCare™ Disposable Pen Needles are used with pen Intended Use: injector devices for the subcutaneous injection of drugs.

#### Technological A comparison of characteristics of UltiCare Disposable Pen Characteristics: Needles and the predicate device is shown in the table below.

| Feature                         | UltiCare™ Disposable Pen<br>Needle (Pending)                                                                                 | BD Pen Needle<br>(K162516)                                                                                    | Conclusion |
|---------------------------------|------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|------------|
| Product Code                    | FMI                                                                                                                          | FMI                                                                                                           | Identical  |
| Device<br>Description           | Pen needle                                                                                                                   | Pen needle                                                                                                    | Identical  |
| Design Needle<br>and Hub        | Image: UltiCare Disposable Pen Needle                                                                                        | Image: BD Pen Needle                                                                                          | Similar    |
| Intended use(s)                 | The UltiCare™ Disposable<br>Pen Needles are used with<br>pen injector devices for the<br>subcutaneous injection of<br>drugs. | BD Pen Needle is intended for<br>use with pen injector devices<br>for the subcutaneous injection<br>of drugs. | Identical  |
| Type of use                     | Over-the Counter Use                                                                                                         | Over-the Counter Use                                                                                          | Identical  |
| Specific drug<br>use            | Drugs                                                                                                                        | Drugs                                                                                                         | Identical  |
| Tip type                        | Center tip                                                                                                                   | Center tip                                                                                                    | Identical  |
| Needle lumen                    | Single Lumen                                                                                                                 | Single Lumen                                                                                                  | Identical  |
| Needle gauge<br>(mm)            | 32 G (0.23 mm)<br>to<br>29 G (0.33 mm)                                                                                       | 32 G (0.23 mm)<br>to<br>29 G (0.33 mm)                                                                        | Identical  |
| Needle length in<br>inches (mm) | 5/32" (4 mm)<br>to<br>1/2" (12.7 mm)                                                                                         | 5/32" (4 mm)<br>to<br>1/2" (12.7 mm)                                                                          | Identical  |
| Needle<br>connector type        | Compatible with pen injectors<br>that comply with ISO 11608-1                                                                | Compatible with pen injectors<br>that comply with ISO 11608-1                                                 | Identical  |
| Biocompatibility                | Conforms to ISO<br>10993-1                                                                                                   | Conforms to ISO<br>10993-1                                                                                    | Identical  |
| Sterilization                   | Ethylene Oxide (EtO)                                                                                                         | Gamma Irradiation                                                                                             | Different  |

### Table 5-1: Characteristic Comparison

{5}------------------------------------------------

| Feature                                    | UltiCare™ Disposable Pen<br>Needle (Pending) | BD Pen Needle<br>(K162516) | Conclusion |
|--------------------------------------------|----------------------------------------------|----------------------------|------------|
| Method of<br>attachment to<br>pen injector | Screw threads                                | Screw threads              | Identical  |
| Sterility                                  | $SAL = 10^{-6}$                              | $SAL = 10^{-6}$            | Identical  |
| Needle                                     | Stainless Steel 304                          | Stainless Steel            | Identical  |
| Needle cover<br>(primary<br>container)     | High-Density polyethylene                    | Unknown                    | Unknown    |

### Purpose of the reference device:

The design of the needle and hub as well as the needle cover (primary container) materials and dimensions of the subject device were leveraged from the reference device. All test methods used on the reference device for biocompatibility, performance, shelf life, sterilization, distribution, and package integrity were leveraged for the subject device.

## Non-Clinical Performance Data:

UltiCare Disposable Pen Needles successfully passed all the required non-clinical testing which included the following:

- Testing to evaluate particulates per USP <788> .
- Testing for compliance with the requirements of 11608-2:2012 Needle-based . injection systems for medical use -- Requirements and test methods -- Part 2: Needles
- The table below presents the requirements of the 11608-2:2012 Needle-based injection systems for medical use -- Requirements and test methods -- Part 2: Needles standard and the result of the testing conducted.

{6}------------------------------------------------

| Table 5-2: Requirements of 11608-2:2012                                       |                                                                                                                                                                                                                                         |                    |
|-------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------|
| Test Parameter                                                                | Clause no. & requirement of ISO 11608-2:2012                                                                                                                                                                                            | Results            |
| Materials                                                                     | 4.1 The needle shall be made of tubing materials specified in ISO 9626.                                                                                                                                                                 | Meets requirements |
| Dimensions                                                                    | 4.2 The needles shall fit the test apparatus specified in item 7.3 of ISO 11608-2.                                                                                                                                                      | Meets requirements |
| Determination of flow rate through the needle                                 | 4.3 The needle was tested in accordance with Annex A to ISO 11608-2 to determine flow rate through the needle.                                                                                                                          | Meets requirements |
| Bond between hub and needle tube                                              | 4.4 The union of the hub and needle tube shall not break when tested in accordance with Clause 9 of ISO 11608-2.                                                                                                                        | Meets requirements |
| Needle points                                                                 | 4.5 When examined under a magnification of x2,5, needle points shall appear sharp and free from feather edges, burrs and hooks.                                                                                                         | Meets requirements |
| Freedom from defects                                                          | 4.6 The needle tube shall fulfill the requirements of ISO 7864, 11.3.                                                                                                                                                                   | Meets requirements |
| Lubrication                                                                   | 4.7 The needle tube should be lubricated at both the patient end and the cartridge end. The lubricant shall not, under normal or corrected-to-normal vision, be visible as droplets of fluid on the outside surface of the needle tube. | Meets requirements |
| Dislocation of measuring point at patient end                                 | 4.8 Dislocation of the cannula point at the patient end shall be in accordance with ISO 11608-2 Table 2 when tested as per Clause 8 (of ISO 11608-2).                                                                                   | Meets requirements |
| Determination of functional compatibility with needle-based injection systems | 4.9 Compatibility with any NIS shall be claimed only after testing in accordance with Clause 11.                                                                                                                                        | Meets requirements |
| Ease of assembly and disassembly                                              | 4.10 Attachment of the needle shall be possible without removing the needle from its opened unit packaging. Compliance is checked according to the requirements of Clause 11.                                                           | Meets requirements |
| Sterility                                                                     | 4.11 The needle in its unit packaging shall has been subjected to a validated sterilization process.                                                                                                                                    | Meets requirements |
| Pre-conditioning of needles                                                   | 6 All requirements of the standard related to preconditioning of needles were met                                                                                                                                                       | Meets requirements |

{7}------------------------------------------------

| Biocompatibility<br>testing:  | The Biological Tests selected to be performed on the sterile final<br>product that has direct contact with the end-user according to the<br>2020 FDA guidance Use of International Standard ISO 10993- 1,<br>"Biological evaluation of medical devices - Part 1: Evaluation and<br>testing within a risk management process" considering contact<br>type and duration were as follows:<br>1. Cytotoxicity<br>2. Sensitization<br>3. Irritation or Intracutaneous Reactivity<br>4. Acute Systemic Toxicity<br>5. Material-Mediated Pyrogenicity<br>6. Hemocompatibility<br>7. Subacute/subchronic toxicity<br>Biocompatibility tests selected as per the requirements of FDA<br>guidance Use of International Standard ISO 10993-1, "Biological<br>evaluation of medical devices - Part 1: Evaluation and testing<br>within a risk management process" considering contact type and<br>duration for the UltiCare Disposable Pen Needles did not show<br>any adverse biological / biocompatibility reactions. |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Clinical Performance<br>Data: | Clinical data is not required.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Conclusion:                   | Testing has demonstrated that UltiCare Disposable Pen<br>Needles are substantially equivalent to the predicate device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |

---

**Source:** [https://fda-staging.innolitics.com/device/K213407](https://fda-staging.innolitics.com/device/K213407)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
