← Product Code [FXX](/productcode/FXX) · K213406

# Savvy Mask - Level 3 Surgical Face Mask (K213406)

_Texas Medplast, LLC · FXX · Nov 18, 2021 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K213406

## Device Facts

- **Applicant:** Texas Medplast, LLC
- **Product Code:** [FXX](/productcode/FXX.md)
- **Decision Date:** Nov 18, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

The Disposable Surgical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device provided non-sterile.

## Device Story

Savvy Mask is a disposable, 3-layer, flat-pleated surgical face mask. Construction includes outer/inner layers of thermal-bonded polyethylene/polyester nonwoven fabric and a middle layer of melt-blown polypropylene filter material. Features include a bendable nosepiece and options for head ties or ear loops (spandex/nylon). Device is non-sterile and latex-free. Used in clinical settings by healthcare personnel or patients to provide a physical barrier against microorganisms, body fluids, and particulates. Operates as a passive barrier; no electronic or mechanical components. Performance meets ASTM F2100 Level 3 standards for fluid resistance, filtration efficiency, and breathability.

## Clinical Evidence

No clinical data. Bench testing only. Performance verified against ASTM F1862 (fluid resistance), ASTM F2299 (particulate filtration), ASTM F2101 (bacterial filtration), EN 14683:2019 (differential pressure), and 16 CFR 1610 (flammability). Biocompatibility confirmed per ISO 10993-1, ISO 10993-5, and ISO 10993-10.

## Technological Characteristics

3-layer flat-pleated mask. Materials: Polyethylene/polyester nonwoven (inner/outer), melt-blown polypropylene (middle). Nosepiece: Galvanized wire coated with polyethylene. Ear loops: Spandex/nylon. Dimensions: 9.5 cm x 17.5 cm. Non-sterile. Passive device; no energy source, software, or connectivity.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- DemeMASK Surgical Mask ([K201479](/device/K201479.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right of this symbol, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.

November 18, 2021

Texas MedPlast LLC % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114

Re: K213406

Trade/Device Name: Savvy Mask - Level 3 Surgical Face Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: October 18, 2021 Received: October 19, 2021

Dear Prithul Bom:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Clarence W. Murray, III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K213406

Device Name Savvy Mask - Level 3 Surgical Mask

### Indications for Use (Describe)

The Disposable Surgical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device provided non-sterile.

| Type of Use (Select one or both, as applicable) |                                             |
|-------------------------------------------------|---------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D)    | Over-The-Counter Use (21 CFR 801 Subpart C) |

_ | Prescription Use (Part 21 CFR 801 Subpart D)

|X Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# K213406 510(k) SUMMARY

This summary of 510(k) is submitted in accordance with the requirements of 21 CFR 8807.92:

#### l. SUBMITTER

Texas MedPlast LLC 6630 Roxburgh Drive #171 Houston, TX 77041 USA Tel: +1.832.288.2106 Fax: N/A

| Contact Person: | Diego Olmos        |
|-----------------|--------------------|
| Date Prepared:  | September 15, 2021 |

### ll. DEVICE

| Name of Device:              | Savvy Mask - Level 3 Surgical Mask |
|------------------------------|------------------------------------|
| Classification Name:         | Surgical Apparel                   |
| Regulation:                  | 21 CFR §878.4040                   |
| Regulatory Class:            | Class II                           |
| Product Classification Code: | FXX                                |

### . PREDICATE DEVICE

| Predicate Manufacturer: | DemeTECH Corporation   |
|-------------------------|------------------------|
| Predicate Trade Name:   | DemeMASK Surgical Mask |
| Predicate 510(k):       | K201479                |

No reference devices were used in this submission.

### IV. DEVICE DESCRIPTION

The Savvy Surgical Face Mask is a disposable, single-use surgical mask. The surgical face masks are non-sterile, 3 layers, flat-pleated style. The outer and inner layers of the face masks consist of a thermal bonded bi-component - polyethylene/polyester nonwoven fabric. The middle layer consists of a melt blown polypropylene filter material. Each mask contains head ties, ear loops or ear bands to secure the mask to the user's face and mouth, as well as a fully enclosed, soft, bendable nosepiece to fit over the nose. This device is not made from natural rubber latex.

### V. INDICATIONS FOR USE

The Disposable Surgical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device provided non-sterile.

{4}------------------------------------------------

### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVCE

The following characteristics were compared between the subject device and the predicate device in Table 1 below.

| Feature                      | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                              | Predicate Device<br>(K201479)                                                                                                                                                                                                                                                                                                                                                                                                        | Result                                |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------|
| Indications for Use          | The Disposable Surgical Face<br>Masks are intended to be<br>worn to protect both the<br>patient and healthcare<br>personnel from transfer of<br>microorganisms, body fluids<br>and particulate material.<br>These face masks are<br>intended for use in infection<br>control practices to reduce the<br>potential exposure to blood<br>and body fluids. This is a<br>single use, disposable device<br>provided non-sterile. | The Disposable Surgical<br>Face Masks are intended<br>to be worn to protect both<br>the patient and<br>healthcare personnel<br>from transfer of<br>microorganisms, body<br>fluids and particulate<br>material. These face<br>masks are intended for<br>use in infection control<br>practices to reduce the<br>potential exposure to<br>blood and body fluids.<br>This is a single use,<br>disposable device<br>provided non-sterile. | Same                                  |
| Materials                    |                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                       |
| Inner Facing Layer           | Polyethylene/Polyester<br>Nonwoven                                                                                                                                                                                                                                                                                                                                                                                          | Spun-bond polypropylene                                                                                                                                                                                                                                                                                                                                                                                                              | Similar (see<br>conclusion)           |
| Middle Layer                 | Melt blown polypropylene                                                                                                                                                                                                                                                                                                                                                                                                    | Melt blown polypropylene                                                                                                                                                                                                                                                                                                                                                                                                             | Same                                  |
| Outer Facing Layer           | Polyethylene/Polyester<br>Nonwoven                                                                                                                                                                                                                                                                                                                                                                                          | Spun-bond polypropylene                                                                                                                                                                                                                                                                                                                                                                                                              | Similar (see<br>conclusion)           |
| Nose piece                   | Galvanized wire coated with<br>polyethylene                                                                                                                                                                                                                                                                                                                                                                                 | Galvanized wire coated<br>with polyethylene                                                                                                                                                                                                                                                                                                                                                                                          | Same                                  |
| Ear loop                     | Spandex and Nylon - Not<br>made from natural rubber<br>latex                                                                                                                                                                                                                                                                                                                                                                | Spandex/nylon                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                                  |
| Head ties- Tie On            | Polyethylene/Polyester nonwoven                                                                                                                                                                                                                                                                                                                                                                                             | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                  | Predicate<br>offers only ear<br>loops |
| Ear Band                     | Polyethylene/Polyester<br>Nonwoven                                                                                                                                                                                                                                                                                                                                                                                          | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                  | Predicate<br>offers only ear<br>loop  |
| Mask Style                   | Flat Pleated                                                                                                                                                                                                                                                                                                                                                                                                                | Flat Pleated                                                                                                                                                                                                                                                                                                                                                                                                                         | Same                                  |
| Color                        | White                                                                                                                                                                                                                                                                                                                                                                                                                       | White                                                                                                                                                                                                                                                                                                                                                                                                                                | Same                                  |
| Dimension (Width),<br>Adult  | 9.5 cm ± 1.0 cm                                                                                                                                                                                                                                                                                                                                                                                                             | 9.5 cm ± 1.0 cm                                                                                                                                                                                                                                                                                                                                                                                                                      | Same                                  |
| Dimension (Length),<br>Adult | 17.5 cm ± 1.0 cm                                                                                                                                                                                                                                                                                                                                                                                                            | 17.5 cm ± 1.0 cm                                                                                                                                                                                                                                                                                                                                                                                                                     | Same                                  |
| OTC Use                      | Yes                                                                                                                                                                                                                                                                                                                                                                                                                         | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                                  |
| Sterility                    | Non-Sterile                                                                                                                                                                                                                                                                                                                                                                                                                 | Non-Sterile                                                                                                                                                                                                                                                                                                                                                                                                                          | Same                                  |
| Use                          | Single Use                                                                                                                                                                                                                                                                                                                                                                                                                  | Single Use                                                                                                                                                                                                                                                                                                                                                                                                                           | Same                                  |
| ASTM F2100 Level             | Level 3                                                                                                                                                                                                                                                                                                                                                                                                                     | Level 3                                                                                                                                                                                                                                                                                                                                                                                                                              | Same                                  |

Table 1 – Comparison of Technological Characteristics

{5}------------------------------------------------

## VII. PERFORMANCE DATA

The subject device was tested and conformed to the following standards and the requirements stated in the Guidance for Industry and FDA Staff: Surgical Masks – Premarket Notification [510(k)] Submission issued on March 5, 2004. A summary of the benchtop performance testing results is provided below in Table 2.

| Item                                                       | Proposed Device -<br>Savvy Mask                                                                                                                       | Acceptance Criteria                                                  | Predicate Device<br>(K201479)                     | Result |
|------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|---------------------------------------------------|--------|
| Level 3 Fluid<br>Resistance<br>Performance<br>ASTM F1862   | Three non-sequential lots<br>of 32 (total of 96, AQL<br>4.0) passed at 160mmHg<br>Lot 1: 31/32 pass<br>Lot 2: 30/32 pass<br>Lot 3: 31/32 pass         | AQL 4%, single<br>sampling plan, 29<br>out of 32 Pass at<br>160mmHg  | Pass at 160 mmHg<br>(Level 3 Fluid<br>Resistance) | Pass   |
| Particulate<br>Filtration<br>Efficiency<br>ASTM F2299      | Three non-sequential lots<br>of 32 (total of 96, AQL<br>4.0) passed at ≥98%<br>Lot 1:<br>32/32 pass<br>Lot 2:<br>32/32 pass<br>Lot 3:<br>32/32 pass   | ≥ 98%                                                                | Pass at ≥99%                                      | Pass   |
| Bacterial<br>Filtration<br>Efficiency<br>ASTM F2101        | Three non-sequential lots<br>of 32 (total of 96, AQL<br>4.0) passed at ≥98% -<br>Lot 1:<br>32/32 pass<br>Lot 2:<br>32/32 pass<br>Lot 3:<br>32/32 pass | ≥ 98%                                                                | Pass at ≥99%                                      | Pass   |
| Differential<br>Pressure<br>ASTM<br>F2100/EN<br>14683:2019 | Three non-sequential lots<br>of 32 (total of 96, AQL<br>4.0) passed at <6.0<br>$H2O/cm2$<br>Lot 1:<br>32/32 pass                                      | AQL 4%, single<br>sampling plan, 29/32<br>pass<br>< 5.0 mm $H20/cm²$ | MIL-M-36954C<br>Average 3.6<br>mm $H2O/cm2$       | Pass   |

Table 2 – Benchtop Performance Testing

{6}------------------------------------------------

|                                        | 30/32 pass<br>Lot 3:<br>31/32 pass                                                                                                                              |                                      |         |      |
|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|---------|------|
| Class 1<br>Flammability<br>16 CFR 1610 | Three non-sequential lots<br>of 32 (total of 96, AQL<br>4.0) passed Class 1<br>16 CFR 1610<br>Lot 1: Class 1, IBE<br>Lot 2: Class 1, IBE<br>Lot 3: Class 1, IBE | Class 1<br>< 3.5 second<br>burn time | Class 1 | Pass |

(Vol 018)

## Sterilization & Shelf-life Testing

Not Applicable (This is a non-sterile device and shelf-life is not applicable to this device because of low likelihood of time-dependent product degradation.)

### Biocompatibility Testing

Biocompatibility testing was performed in accordance with ISO 10993-1:2018. Specifically, the following testing endpoints were evaluated.

| Table 3 - Biocompatibility Testing |  |  |
|------------------------------------|--|--|
|------------------------------------|--|--|

| Biocompatibility Testing Endpoints | Acceptance Criteria | Result |
|------------------------------------|---------------------|--------|
| Cytotoxicity - ISO 10993-5         | Non-Cytotoxic       | Pass   |
| Skin Sensitization - ISO 10993-10  | Non- Sensitizing    | Pass   |
| Skin Irritation - ISO 10993-10     | Non-Irritating      | Pass   |

(Vol 015)

## Table 4 - Summary of Non-Clinical Performance Testing

The following standards have been used to evaluate the Savvy Mask Surgical Mask:

{7}------------------------------------------------

| ASTM F2100       | Standard Specification for Performance of Materials Used in<br>Medical Face Masks                                                                                     |
|------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ASTM F1862       | Standard Test Method for Resistance of Medical Face Masks to<br>Penetration by Synthetic Blood (Horizontal Projection of Fixed<br>Volume at a Known Velocity)         |
| ASTM F2299       | Standard Test Method for Determining the Initial Efficiency of<br>Materials<br>Used in Medical Face Masks to Penetration by Particulates Using<br>Latex Spheres       |
| ASTM F2101       | Standard Method for Evaluating the Bacterial Filtration Efficiency<br>(BFE)<br>of Medical Face Mask Materials, Using a Biological<br>Aerosol of Staphylococcus aureus |
| EN 14683:2019    | Standard Test Method for Differential Pressure                                                                                                                        |
| 16 CFR Part 1610 | Standard for Flammability                                                                                                                                             |
| ISO 10993-1      | Biological evaluation of medical devices - Part 1: Evaluation and<br>testing within a risk management process                                                         |
| ISO 10993-5      | Biological evaluation of medical devices - Part 5: Tests for in<br>vitro cytotoxicity of medical devices                                                              |
| ISO 10993-10     | Biological evaluation of medical devices - Part 10: Tests for<br>irritation and skin sensitization                                                                    |

## Software Verification and Validation Testing

Not Applicable (Passive Device)

Electrical safety and electromagnetic compatibility (EMC)

Not Applicable (Passive Device)

### Animal Study

Animal performance testing was not required to demonstrate safety and effectiveness of the device.

### Human Clinical Performance Testing

Clinical testing was not required to demonstrate the safety and effectiveness of the device.

## VIII. CONCLUSIONS

The conclusions drawn from the performance data demonstrate that the subject device is as safe, as effective, and performs as well as or better than the legally marketed device K201479, DemeMASK Surqical Face Mask manufactured by DemeTECH Corporation.

---

**Source:** [https://fda-staging.innolitics.com/device/K213406](https://fda-staging.innolitics.com/device/K213406)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).