K213401 · DiaDent Group International · KLE · Jan 31, 2022 · Dental
Device Facts
Record ID
K213401
Device Name
DIAPLUS Universal
Applicant
DiaDent Group International
Product Code
KLE · Dental
Decision Date
Jan 31, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3200
Device Class
Class 2
Indications for Use
1) All direct restorations 2) All indirect restorations 3) Desensitizing/sealing of tooth 4) Intra-oral repairs
Device Story
DIAPLUS Universal is a 7th generation dentin adhesive system; functions as conditioner, primer, and bonding resin. Applied to tooth surface by clinician; solvent volatilized via drying; photopolymerized. Contains 10-MDP monomer with hydrophobic long carbon chain and hydrophilic phosphate group; enhances surface activity and permeability into dentin and enamel; increases bonding to hydroxyapatite. Used in dental clinics for direct/indirect restorations, desensitizing, and intra-oral repairs. Output is a polymerized adhesive layer facilitating restoration retention. Benefits patient by providing stable, durable dental bonding.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing and biocompatibility studies conforming to ISO 4049, ISO 7405, ISO 29022, and ISO 10993 series (genotoxicity, cytotoxicity, irritation, systemic toxicity).
Technological Characteristics
Resin-based dentin adhesive. Composition includes 10-MDP, HEMA, BisGMA/TEGDMA, and photoinitiators. Photopolymerization principle. Conforms to ISO 4049 (restorative materials) and ISO 29022 (shear bond strength). Biocompatibility per ISO 10993. Supplied in 1ml/5ml bottles.
Indications for Use
Indicated for all direct and indirect dental restorations, desensitizing or sealing of tooth structure, and intra-oral repairs.
Regulatory Classification
Identification
A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 31, 2022
Diadent Group International Kab Lee Quality Assurance Manager 16, Osongsaengmyeong 4-ro, Osong-eup, Heungdeok-gu Cheongju-si, Chungcheongbuk-do 28161 SOUTH KOREA
Re: K213401
Trade/Device Name: DIAPLUS Universal Regulation Number: 21 CFR 872.3200 Regulation Name: Resin Tooth Bonding Agent Regulatory Class: Class II Product Code: KLE Dated: August 29, 2021 Received: October 18, 2021
Dear Kab Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213401
Device Name DIAPLUS Universal
Indications for Use (Describe)
- 1. All direct restorations
- 2. All indirect restorations
- 3. Desensitizing/sealing of tooth
- 4. Intra-oral repairs
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K213401
# 510(k) Summary
### 1 Application Information
| Date Prepared: | Jan 27, 2022 |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Company Name and Address: | DiaDent Group International<br>16, Osongsaengmyeong 4-ro, Osong-eup, Heungdeok-gu, Cheongju-<br>si, Chungcheongbuk-do, 28161, Republic of Korea |
| Contact Person: | Kab Sun, Lee<br>Quality Assurance Manager<br>Phone: +82-43-266-2315<br>FAX: +82-43-235-2315<br>Email: diadent32@diadent.co.kr |
#### 2 Device Information
| 510(k) Number | K213401 |
|---------------------|-----------------------------|
| Classification Name | Agent, Tooth Bonding, Resin |
| Common Name | Resin tooth bonding agent. |
| Review Panel: | Dental |
| Regulation Number: | 21 CFR 872.3200 |
| Product Code: | KLE |
| Device Class: | II |
| Trade Name | DIAPLUS Universal |
### 3 Predicate Devices
The legally marketed devices to which substantial equivalence is being claimed are:
| | Primary Predicate Device |
|--------------------|--------------------------|
| 510(k) Number: | K161051 |
| Applicant: | BISCO, Inc. |
| Device Name: | All-Bond Universal w/BAC |
| Regulation Number: | 21 CFR 872.3200 |
| Product Code: | KLE |
| Device Class: | II |
### 4 Device Configuration
DIAPLUS Universal: 1 bottle (5ml, 1ml)
### 5 Device Description
It is a 7th generation dentin adhesive system that handles all the functions of the conditioner primer bonding resin. The polymerization method is photopolymerization type, and after being applied to the tooth surface, the solvent is volatilized and polymerized in the drying process to perform the function as a dentine adhesive.
# 6 Intended Use/Indications for Use
- 1) All direct restorations
- 2) All indirect restorations
- 3) Desensitizing/sealing of tooth
- 4) Intra-oral repairs
{4}------------------------------------------------
# 7 Comparison Table and Discussion
This device compares to the legally marketed devices as follows:
| | Subject Device | Primary Predicate Device | |
|----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Common name | Resin tooth bonding agent. | Resin tooth bonding agent. | |
| Trade Name | DIAPLUS Universal | ALL-BOND Universal w/BAC | Discussion |
| 510(k)<br>Number | K213401 | K161051 | - |
| Indication for<br>Use | 1. All direct restorations<br>2. All indirect restorations<br>3. Desensitizing/sealing of tooth<br>4. Intra-oral repairs | 1. all direct restorations<br>2. all indirect restorations<br>3. intra-oral repairs (i.e. repair of any<br>fixed dental prosthesis containing<br>zirconia, alumina, metals, glass<br>ceramics, tooth<br>structure, and composites)<br>4. desensitizing/sealing of tooth<br>structure<br>5. protective varnish for glass ionomer<br>fillings<br>6. priming of enamel for orthodontic<br>use | Equivalent |
| Composition | - 10-methacryloyloxydecyl<br>dihydrogenphosphate<br>- 2-Hydroxyethyl methacrylate<br>- Bis[2-(methacryloyloxy)ethyl]<br>phosphate<br>- 3-(Trimethoxysilyl)propyl<br>methacrylate<br>- Trimethylolpropane<br>trimethacrylate<br>- BisGMA/TEGDMA Monomer<br>Blend<br>- Poly(ethylene glycol)<br>dimethacrylate<br>- Diphenyliodonium<br>hexafluorophosphate<br>- Phenyl Bis(2,4,6-<br>Trimethylbenzoyl)phosphineoxide<br>- (+/-)Camphorquinone<br>- Ethyl 4-dimethyl aminobenzoate<br>- tert-butyl hydroquinone<br>- Ethyl alcohol<br>- Silicon dioxide | - 10-Methacryloyloxydecyl dihydrogen<br>phosphate<br>- 2-hydroxyethyl methacrylate<br>- Bisphenol A diglycidylmethacrylate<br>- water<br>- ethanol | |
| Principle of<br>operation | After being applied to the tooth<br>surface, the solvent is volatilized<br>and photopolymerized in the<br>drying process to perform the<br>function as a dentin adhesive. The<br>main material, MDP contains a<br>hydrophobic long carbon chain<br>and a hydrophilic phosphate<br>group, so it has excellent surface<br>activity and can easily permeates<br>dentin & enamel. It increases<br>bonding to hydroxyapatite<br>remaining in dentin. | All-Bond Universal is an<br>ethanol/water-based dental adhesive<br>which bonds to dentin and cut and un-<br>cut enamel. | Equivalent |
| Performance<br>Standard<br>Conformance | Conformed<br>ISO 4049 and ISO 29022 | Conformed<br>ISO 4049 and ISO 29022 | Equivalent |
| | Subject Device | Primary Predicate Device | |
| Common name | Resin tooth bonding agent. | Resin tooth bonding agent. | |
| Trade Name | DIAPLUS Universal | ALL-BOND Universal w/BAC | Discussion |
| Physical<br>properties | - Film thickness<br>- Sensitivity to light<br>- Shear bond strength<br>- Occlusion of dental tubules | - Film thickness<br>- Shear bond strength | Equivalent |
| Biocompatibil<br>ity | Biocompatible | Biocompatible | Equivalent |
| Package<br>Contents | One bottle | One bottle | Equivalent |
| RX only/<br>OTC | RX Only | RX Only | equivalent |
{5}------------------------------------------------
### Differences
## -Composition
| Subject Device<br>(DIAPLUS Universal) | Predicate Device<br>(ALL-BOND Universal w/BAC) | Discussion |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| - 10-methacryloyloxydecyl<br>dihydrogen phosphate<br>- 2-Hydroxyethyl methacrylate<br>- Bis[2-(methacryloyloxy)ethyl]<br>phosphate<br>- 3-(Trimethoxysilyl)propyl<br>methacrylate<br>- Trimethylolpropane<br>trimethacrylate<br>- BisGMA/TEGDMA Monomer<br>Blend<br>- Poly(ethylene glycol)<br>dimethacrylate<br>- Diphenyliodonium<br>hexafluorophosphate<br>- Phenyl Bis(2,4,6-<br>Trimethylbenzoyl)phosphineoxide<br>- (+/-) Camphor quinone<br>- Ethyl 4-dimethyl aminobenzoate<br>- tert-butyl hydroquinone<br>- Ethyl alcohol<br>- Silicon dioxide | - 10-Methacryloyloxydecyl<br>dihydrogen phosphate<br>- 2-hydroxyethyl methacrylate<br>- Bisphenol A<br>diglycidylmethacrylate<br>- water<br>- ethanol | The main ingredients (10-<br>methacryloyloxydecyl dihydrogen<br>phosphate, 2-Hydroxyethyl methacrylate)<br>are similar, with some different ingredients.<br>However, the results of biocompatibility<br>and performance tests confirm that<br>DIAPLUS Universal similar products are<br>equivalent. Performance tests confirm that<br>DIAPLUS Universal similar products are<br>equivalent. |
### 8. Non-Clinical performance data
This device has demonstrated conformance with non-clinical performance requirements through evaluation and testing in accordance with the following harmonized standards:
| -ISO 4049 | Dentistry -- Polymer-based restorative materials |
|--------------|------------------------------------------------------------------------------------------------------------------------|
| -ISO 7405 | Dentistry - Evaluation of Biocompatibility of Medical Devices Used In Dentistry<br>[Including: Amendment 1 (2013)] |
| -ISO 29022 | Dentistry - Adhesive - Notched-edge sheer bond strength test |
| -ISO 10993-1 | Biological evaluation of medical devices - Part 1: Evaluation and testing within a<br>risk management process |
| -ISO 10993-3 | Biological evaluation of medical devices - Part 3: Tests for genotoxicity<br>carcinogenicity and reproductive toxicity |
{6}------------------------------------------------
| -ISO 10993-5 | Biological evaluation of medical devices - Part 5: Tests for in vitro Cytotoxicity |
|---------------|----------------------------------------------------------------------------------------------------|
| -ISO 10993-10 | Biological evaluation of medical devices - Part 10: Tests for irritation and skin<br>sensitization |
| -ISO 10993-11 | Biological evaluation of medical devices - Part 11: Tests for Systemic toxicity |
### The following Performance tests were conducted.
- Film Thickness
- Shear Bonding Strength
- Sensitivity to Ambient Light
- Occlusion of dentinal tubules
#### Biocompatibility test:
| Test | Standard | Test result |
|------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|--------------------------|
| Cytotoxicity Study | • 10993-5 - Biological Evaluation of Medical Devices -<br>Part 5: Tests for In Vitro Cytotoxicity | Not<br>Compatible<br>BSE |
| Bacterial Reverse Mutation Study | • 10993-3 - Biological Evaluation of Medical Devices –<br>Part 3: Tests for genotoxicity, carcinogenicity and<br>reproductive toxicity | Compatible |
| In vitro Mammalian Chromosomal Aberration Test | • 10993-3 - Biological Evaluation of Medical Devices -<br>Part 3: Tests for genotoxicity, carcinogenicity and<br>reproductive toxicity | Compatible |
| Mammalian Erythrocyte Micronucleus | • 10993-3 - Biological Evaluation of Medical Devices -<br>Part 3: Tests for genotoxicity, carcinogenicity and<br>reproductive toxicity | Compatible |
| Skin Sensitization Test | • 10993-10 - Biological Evaluation of Medical Devices -<br>Part 10: Tests for Irritation and Skin Sensitization | Compatible |
| Oral Mucosa Irritation Test | • 10993-10 - Biological Evaluation of Medical Devices -<br>Part 10: Tests for Irritation and Skin Sensitization | Compatible |
| Acute Systemic Toxicity | • 10993-11 - Biological Evaluation of Medical Devices -<br>Part 11: Tests for Systemic Toxicity | Compatible |
(Among the components of this product, 10-MDP is an acidic monomer, and depending on the acidity, it is a substance that causes cytotoxicity according to the principle of action of this product, so cytotoxicity test setting the test item is meaningless.
### 9 Clinical Performance Data
No clinical data was collected or provided to support substantial equivalence between the subject and predicate devices.
### 10. Conclusion
Based on the above information and all data provided in this submission, the comparison of intended uses, technological characteristics, and non-clinical performance testing demonstrates that the subject device is substantially equivalent to the legally marketed devices identified in this submission.
{7}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.