← Product Code [KRD](/productcode/KRD) · K213385

# GEM (K213385)

_Obsidio, Inc. · KRD · Jul 1, 2022 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K213385

## Device Facts

- **Applicant:** Obsidio, Inc.
- **Product Code:** [KRD](/productcode/KRD.md)
- **Decision Date:** Jul 1, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.3300
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

Obsidio GEM™ is indicated for use in the embolization of: · Hypervascular tumors, · Blood vessels to occlude blood flow for controlling bleeding/hemorrhaging in the peripheral vasculature.

## Device Story

GEM is a pre-hydrated, bioresorbable embolization agent composed of gelatin, layered silicate (Laponite), and tantalum powder. Supplied in a 1 ml syringe with a luer lock tip, it is injected by a physician (e.g., interventional radiologist) through an intravascular catheter into a target vessel. The material acts as a mechanical barrier to occlude blood flow. Used in clinical settings for treating hypervascular tumors and controlling peripheral hemorrhaging. The tantalum powder provides radiopacity for visualization during delivery. The device is intended for single use and provides a permanent or semi-permanent occlusion to benefit patients by reducing blood supply to tumors or stopping active bleeding.

## Clinical Evidence

No human clinical data. Evidence consists of bench testing and two animal studies in swine. Study 1 (n=19) evaluated hypervascular tumor embolization across 4 timepoints (4-180 days) comparing GEM to EmboCube; assessed occlusion, arterial flow, and tissue response. Study 2 (n=10) evaluated simulated hemorrhage control over 7 days; assessed hemostatic success and time to hemostasis. Bench testing included rheometry, injection force, catheter deliverability, biocompatibility (ISO 10993-1), and sterility/packaging validation.

## Technological Characteristics

Composition: pre-hydrated bioresorbable gelatin, layered silicate (Laponite), and tantalum powder. Form factor: 1 ml syringe with luer lock tip. Principle: mechanical occlusion via injectable material. Sterilization: provided sterile. Biocompatibility: ISO 10993-1 compliant. Standards: ISO 13408-1 (aseptic fill), ANSI/AAMI/ISO 11737-2 (sterility), ANSI/AAMI ST72 (pyrogenicity).

## Regulatory Identification

A vascular embolization device is an intravascular implant intended to control hemorrhaging due to aneurysms, certain types of tumors (e.g., nephroma, hepatoma, uterine fibroids), and arteriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in neurovascular applications are also not included in this classification, see § 882.5950 of this chapter.

## Special Controls

*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 870.1(e).

## Predicate Devices

- EmboCube™ Embolization Gelatin ([K183120](/device/K183120.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 1, 2022

Obsidio. Inc. % Janice Hogan Partner Hogan Lovells US LLP 1735 Market Street, Suite 2320 Philadelphia, Pennsylvania 19103

Re: K213385

Trade/Device Name: GEM Regulation Number: 21 CFR 870.3300 Regulation Name: Vascular Embolization Device Regulatory Class: Class II Product Code: KRD Dated: June 8, 2022 Received: June 8, 2022

# Dear Janice Hogan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 

{1}------------------------------------------------

801): medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Misti Malone Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K213385

Device Name GEMTM

Indications for Use (Describe)

Obsidio GEM™ is indicated for use in the embolization of:

· Hypervascular tumors,

· Blood vessels to occlude blood flow for controlling bleeding/hemorrhaging in the peripheral vasculature.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|                                                 |  |

|X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary Obsidio GEM™

# Obsidio, Inc. Address, Telephone Number, Contact Person and Date Prepared

Obsidio, Inc.

1520 Main Street, Unit 4-C Columbia, SC 29201

(267)-235-5951 Phone: Contact Person: Ehsan Jabbarzadeh Date Prepared: June 3, 2022

## Name of Device and Name/Address of Sponsor

GEMTM

## Common or Usual Name

Vascular Embolization Device

## Classification Name

Device, Vascular, For Promoting Embolization

## Predicate Devices

Biosphere Medical, S.A. EmboCube™ Embolization Gelatin (K183120)

### Intended Use / Indications for Use

Obsidio GEM™ is indicated for use in the embolization of:

- Hypervascular tumors ●
- Blood vessels to occlude blood flow for controlling bleeding/hemorrhaging in the peripheral . vasculature

### Technological Characteristics

GEM™ consists of a pre-hydrated bioresorbable gelatin, layered silicate (Laponite) mixture, and tantalum powder packaged in a 1 ml syringe with a standard luer lock tip. The device is provided ready to use. The device is injected into a target vessel through intravascular catheters to provide a mechanical barrier to blood flow. The device is intended for single use and is provided sterile.

### Performance Data

No performance standards have been established under Section 514 of the Food, Drug and Cosmetic Act for these devices. However, vascular embolization devices are subject to the special controls specified in "Guidance for Industry and FDA Staff - Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices", issued on December 29, 2004. In addition, the subject device follows the FDA Draft Guidance on Medical Devices Containing

{4}------------------------------------------------

Materials Derived from Animal Sources (Except for In Vitro Diagnostic Devices), issued on January 23, 2014.

GEM™ was evaluated for various performance attributes to demonstrate the substantial equivalence of the device to the predicate. Non-clinical performance testing included:

- Macroscopic Inspection (Visual Inspection) .
- . Component Characterization
- Mechanical Property (Rheometry) ●
- pH test
- . Injection Force
- . Catheter Deliverability
- Component Sterilization Validations ●
- . Aseptic Fill Validations (ISO 13408-1:2008)
- . Packaging Validations (Shelf-life, Transit, Sealing)
- Sterility (ANSI/AAMI/ISO 11737-2:2019)
- Pyrogenicity (ANSI/AAMI ST72:2019)
- Biocompatibility (ISO 10993-1:2018) .
	- o Cytotoxicity
	- Sensitization o
	- Irritation o
	- Material Mediated Pvrogenicity O
	- Hemolysis O
	- Complement Activation o
	- Chemical Characterization/Toxicological Risk Assessment o

Two animal studies were conducted; one for hypervascular tumors and a second for the indication to use in blood vessels to occlude blood flow to control bleeding/hemorrhaging in the peripheral vasculature.

The first animal study was conducted on ten test (GEM™) and nine control (EmboCube™ Embolization Gelatin, K183120) adult swine (evenly distributed male and female) in 38 renal arteries and 19 hepatic arteries over 4 timepoints (4 days, 30 days, 180 days) to address the indication for embolization of hypervascular tumors. The animal study evaluated overall in life health, test article performance, occlusion, arterial flow scale, and regional and downstream response to GEM™ along with an analysis of explanted tissues to evaluate for downstream layered silicate presence/toxicity.

The second animal study for simulated hemorrhage was conducted on five test (GEM™) and five control (EmboCube™ Embolization Gelatin, K183120) adult swine in eight renal arteries and eight splenic arteries over a period of 7 days for the evaluation of persistence of hemostasis. The animal study included prespecified endpoints for overall in life health, test article performance, hemostatic success, macroscopic regional tissue response, time to hemostasis, and embolization effectiveness of GEM™. The product met its specifications. All test results confirmed that the technological differences between the subject and the predicate device did not adversely impact its performance.

{5}------------------------------------------------

# Substantial Equivalence

GEM™ has the same intended uses, indications for use, similar technological characteristics, and principle of operation as its predicate device. The technological differences involving composition and preparation between GEM™, and its predicate device, raise no new issues of safety or effectiveness. Performance data demonstrate that GEM™ is substantially equivalent to the predicate EmboCube™ Embolization Gelatin.

## Conclusions

Based on the indications for use, design, safety, and performance testing, GEM™ is substantially equivalent to the predicate device, EmboCube™ Embolization Gelatin.

---

**Source:** [https://fda-staging.innolitics.com/device/K213385](https://fda-staging.innolitics.com/device/K213385)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
