← Product Code [FXX](/productcode/FXX) · K213347

# ADI Medical Surgical Face Mask (K213347)

_Asia Dynamics Inc., D/B/A Adi Medical · FXX · Jan 25, 2022 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K213347

## Device Facts

- **Applicant:** Asia Dynamics Inc., D/B/A Adi Medical
- **Product Code:** [FXX](/productcode/FXX.md)
- **Decision Date:** Jan 25, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Indications for Use

ADI Medical Surgical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure of the wearer to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

## Device Story

ADI Medical Surgical Face Mask is a single-use, three-layer, flat-pleated mask; features spun-bond polypropylene inner/outer layers and melt-blown polypropylene filter middle layer; includes malleable polyethylene nose clamp and elastic ear loops (latex-free). Device worn by healthcare personnel and patients to provide physical barrier against microorganisms, body fluids, and particulates; reduces wearer exposure to blood/fluids. Used in clinical/infection control settings; non-sterile; disposable.

## Clinical Evidence

No clinical data. Bench testing only. Performance testing included fluid resistance (ASTM F1862), particulate filtration efficiency (ASTM F2299, 99.47%), bacterial filtration efficiency (ASTM F2101, 99.48%), differential pressure (EN 14683, 3.52 mmH2O/cm2), and flammability (16 CFR 1610, Class 1). Biocompatibility testing (ISO 10993) confirmed the device is non-cytotoxic, non-irritating, and non-sensitizing.

## Technological Characteristics

Three-layer construction: spun-bond polypropylene (inner/outer), melt-blown polypropylene (filter). Malleable polyethylene nose clamp. Polyester/spandex ear loops. ASTM F2100 Level 2 performance. Non-sterile. Biocompatibility per ISO 10993.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- Disposable Medical Face Mask ([K202513](/device/K202513.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

January 25, 2022

Asia Dynamics Inc., d/b/a ADI Medical Timothy Kania Official Correspondant Mdi Consultants Inc. 55 Northen Blvd, Suite 200 Great Neck. New York 11021

Re: K213347

Trade/Device Name: ADI Medical Surgical Face Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: December 27, 2021 Received: December 29, 2021

Dear Timothy Kania:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Clarence W. Murray III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K213347

Device Name ADI Medical Surgical Face Masks

#### Indications for Use (Describe)

ADI Medical Surgical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material.

These face masks are intended for use in infection control practices to reduce the posure of the wearer to blood and body fluids.

This is a single use, disposable device(s), provided non-sterile.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) SUMMARY

### 1. Submitter's Identification:

| Applicant:              | Asia Dynamics Inc., d/b/a ADI Medical       |
|-------------------------|---------------------------------------------|
| Address:                | 1565 South Shields Drive Waukegan, IL 60085 |
| Contact Person:         | Jonathan Zhu                                |
| Tel:                    | (847)-688-9968                              |
| Email:                  | Jon@adimedical.com                          |
| Date Summary Prepared:  | December 27, 2021                           |
| Official Correspondent: | Mr. Tim Kania                               |
|                         | Mdi Consultants, Inc.                       |
|                         | Address:                                    |
| Tel:                    | 732-796-4565                                |
| Email:                  | tim@mdiconsultants.com                      |

#### Name of the Device: ನ

| Proprietary Name:    | ADI Medical Surgical Face Mask |
|----------------------|--------------------------------|
| Classification Name: | Mask,Surgical                  |
| Common Name:         | Surgical Mask                  |
| Regulatory Class:    | II                             |
| Product Code:        | FXX                            |
| Regulation Name:     | Surgical Apparel               |
| Regulation Number:   | 878.4040                       |
| Review Panel:        | General Hospital               |

#### Information for the 510(k) Cleared Device (Predicate Device): ನೆ

Disposable Medical Face Mask, K202513

Surgical Face Mask

Shenzhen Jinko Industrial Co., Ltd..

No reference devices were used in this submission.

## 4. Device Description:

The ADI Medical Surgical Face Masks are single use, three-layer, flat -folded masks with ear loops and nose clamp. The ADI Medical Surgical Face Masks are manufactured with three layers, the inner and outer layers are made of spun-bond polypropylene, and the middle layer is made of melt blown polypropylene filter. The ear loops are held in place over the users' mouth and nose by two elastic ear loops welded to the facemask. The elastic ear loops are not made with natural rubber latex. The nose clamp in the layers of facemask is to allow the user to fit the facemask around their nose, which is made of malleable polyethylene wire. ADI Medical Surgical Face Masks will be

{4}------------------------------------------------

provided in blue. ADI Medical Surgical Face Masks are sold non-sterile and are intended to be single use, disposable devices.

# 5. Indications for Use:

ADI Medical Surgical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure of the wearer to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

#### Technological Characteristics Comnarison: la

| Device                         | Proposed Device                                                                                                                                                                                                                                                                                                                                                                                     | Predicate Device                                                                                                                                                                                                                                                                                                                                                                          | Result                                             |         |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------|---------|
| Manufacturer                   | Asia Dynamics Inc., d/b/a ADI Medical                                                                                                                                                                                                                                                                                                                                                               | Shenzhen Jinko Industrial Co., Ltd.                                                                                                                                                                                                                                                                                                                                                       | -                                                  |         |
| 510K number                    | K213347                                                                                                                                                                                                                                                                                                                                                                                             | K202513                                                                                                                                                                                                                                                                                                                                                                                   | -                                                  |         |
| Model Name                     | ADI Medical Surgical Face Mask                                                                                                                                                                                                                                                                                                                                                                      | Disposable Medical Face Mask                                                                                                                                                                                                                                                                                                                                                              | Similar                                            |         |
| Model No.                      | 76120                                                                                                                                                                                                                                                                                                                                                                                               | FM-04                                                                                                                                                                                                                                                                                                                                                                                     | -                                                  |         |
| Classification                 | Class II Device, FXX (21 CFR878.4040)                                                                                                                                                                                                                                                                                                                                                               | Class II Device, FXX (21 CFR878.4040)                                                                                                                                                                                                                                                                                                                                                     | Same                                               |         |
| Intend use                     | ADI Medical Surgical Masks are intended to be worn to protect both the patientand healthcare personnel from transfer of microorganisms, body fluids and particulate material.<br>These face masks are intended for use in infection control practices to reduce the potential exposure of the wearer to blood and body fluids.<br>This is a single use, disposable device(s), provided non-sterile. | The Medical Masks are intended to be worn to protect both the patientand healthcare personnel from transfer of microorganisms, body fluids and particulate material.<br>These face masks are intended for use in infection control practices to reduce the potential exposure of thewearer to blood and body fluids.<br>This is a single use, disposable device(s), provided non-sterile. | Same                                               |         |
| Model                          | Ear Loops, Flat Pleated, 3 layers                                                                                                                                                                                                                                                                                                                                                                   | Ear Loops, Flat Pleated, 3 layers                                                                                                                                                                                                                                                                                                                                                         | Same                                               |         |
| Material                       | Outer facing layer                                                                                                                                                                                                                                                                                                                                                                                  | Spun-bond polypropylene                                                                                                                                                                                                                                                                                                                                                                   | Spun-bond polypropylene                            | Same    |
|                                | Middle layer                                                                                                                                                                                                                                                                                                                                                                                        | Melt blown polypropylene filter                                                                                                                                                                                                                                                                                                                                                           | Melt blown polypropylene filter                    | Same    |
|                                | Inner Facing layer                                                                                                                                                                                                                                                                                                                                                                                  | Spun-bond polypropylene                                                                                                                                                                                                                                                                                                                                                                   | Spun-bond polypropylene                            | Same    |
|                                | Nose Clamp                                                                                                                                                                                                                                                                                                                                                                                          | Malleable polypropylene with steel wire                                                                                                                                                                                                                                                                                                                                                   | Malleable polypropylene with steel wire            | Same    |
|                                | Ear Loops                                                                                                                                                                                                                                                                                                                                                                                           | Polyester, Spandex                                                                                                                                                                                                                                                                                                                                                                        | Polyester, Spandex                                 | Same    |
| Color                          |                                                                                                                                                                                                                                                                                                                                                                                                     | Blue<br>CAS#147-14-8                                                                                                                                                                                                                                                                                                                                                                      | Variety (include blue)                             | Similar |
| Dimension<br>(Length)          |                                                                                                                                                                                                                                                                                                                                                                                                     | $175 \pm 2.5$ mm                                                                                                                                                                                                                                                                                                                                                                          | $175 \pm 8$ mm                                     | Similar |
| Dimension<br>(Width)           |                                                                                                                                                                                                                                                                                                                                                                                                     | $95 \pm 2.5$ mm                                                                                                                                                                                                                                                                                                                                                                           | 95+4.5 mm                                          | Similar |
| OTC use                        |                                                                                                                                                                                                                                                                                                                                                                                                     | Yes                                                                                                                                                                                                                                                                                                                                                                                       | Yes                                                | Same    |
| Sterility                      |                                                                                                                                                                                                                                                                                                                                                                                                     | Non-Sterile                                                                                                                                                                                                                                                                                                                                                                               | Non-Sterile                                        | Same    |
| Use                            |                                                                                                                                                                                                                                                                                                                                                                                                     | Single Use, Disposable                                                                                                                                                                                                                                                                                                                                                                    | Single Use, Disposable                             | Same    |
| ASTM F2100 Level               |                                                                                                                                                                                                                                                                                                                                                                                                     | Level 2                                                                                                                                                                                                                                                                                                                                                                                   | Level 2                                            | Same    |
| Biocompatibility<br>(ISO10993) |                                                                                                                                                                                                                                                                                                                                                                                                     | Non-Cytotoxic, Non-<br>Sensitizing, Non-Irritating                                                                                                                                                                                                                                                                                                                                        | Non-Cytotoxic, Non-<br>Sensitizing, Non-Irritating | Same    |

Table 1: Comparison to Predicate Device

{5}------------------------------------------------

{6}------------------------------------------------

## 7.Summary of Non-Clinical Tests:

The following testing was conducted to demonstrate that the subject device (3 nonconsecutive lots with each lot containing 32 samples) met the acceptance criteria and specification of the standard shown below.

| Standard                                               | Results of tests on<br>subject device                        | Pass Criteria                 | Results |
|--------------------------------------------------------|--------------------------------------------------------------|-------------------------------|---------|
| Fluid Resistance<br>PerformanceASTM<br>F1862           | 94 out of 96 pass at 120 mmHg, 3 non-consecutive lots tested | 29 out of 32 pass at 120 mmHg | PASS    |
| Particulate Filtration<br>Efficiency<br>ASTM F2299     | 96 out of 96 pass<br>Average: 99.47%<br>$\ge$ 98%            | $\ge$ 98%                     | PASS    |
| Bacterial Filtration<br>Efficiency<br>ASTM F2101       | 96 out of 96 pass<br>Average: 99.48%<br>$\ge$ 98%            | $\ge$ 98%                     | PASS    |
| Differential Pressure (Delta<br>P)<br>EN 14683 Annex C | 96 out of 96 pass<br>Average: 3.52<br>< 5.0mmH2O/cm2         | < 6.0mmH2O/cm2                | PASS    |
| Flammability 16 CFR 1610                               | Class 1                                                      | Class 1                       | PASS    |

|  | Table 2: Performance testing |  |
|--|------------------------------|--|
|--|------------------------------|--|

### Table 3: Biocompatibility Testing

| Item          | Proposed device                                                              | Acceptance Criteria | Result |
|---------------|------------------------------------------------------------------------------|---------------------|--------|
| Cytotoxicity  | Under the<br>conditions of the<br>study, the<br>device is non-<br>cytotoxic. | Non-Cytotoxic       | PASS   |
| Irritation    | Under the<br>conditions of the<br>study, the device<br>is non-irritating.    | Non-Irritating      | PASS   |
| Sensitization | Under the<br>conditions of the<br>study, the device<br>is non-sensitizing    | Non-Sensitizing     | PASS   |

### 8.Discussion of Clinical Tests Performed:

No clinical study is included in this submission.

## 9.Conclusions:

The conclusion drawn from the nonclinical tests demonstrates that the subject device in 510(k) submission K213347, the ADI Medical Surgical Masks, is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K202513.

---

**Source:** [https://fda-staging.innolitics.com/device/K213347](https://fda-staging.innolitics.com/device/K213347)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com