DENU Composite Resin is indications for use in: - Direct anterior and posterior restorations - Core buildup - Splinting - Indirect anterior and posterior restorations including inlays, onlays and veneers DENU Flow Resin is indications for use in: - Class III restorations - Class V restorations - Small Class I restorations (non stress-bearing restorations) - Pit and fissure sealing in molars and premolars - Repair of small defects in esthetic indirect inlays - Base/liner under direct restorations
Device Story
DENU Composite Resin and DENU Flow Resin are light-cured tooth shade resin materials used by dental professionals for restorative dentistry. The devices are applied to tooth structures and cured using a dental light source (400-500 nm, 400 mW/cm²) to harden the material. DENU Composite Resin is used for direct/indirect restorations, core buildup, and splinting; DENU Flow Resin is a lower-viscosity version used for Class I/III/V restorations, pit and fissure sealing, and as a base/liner. The materials function by filling and restoring tooth defects, providing structural support and esthetic correction. Clinical benefit is achieved through the restoration of tooth form and function.
Clinical Evidence
Bench testing only. Testing included ISO 4049 (depth of cure, flexural strength, water sorption, solubility, radio-opacity, color stability) and ISO 10993 (genotoxicity, cytotoxicity, skin sensitization, oral mucosa irritation, acute systemic toxicity). All results met preset acceptance criteria.
Technological Characteristics
Light-cured resin materials (BISGMA, UDMA, TEGDMA matrix). Fillers: Ba glass, silica. Photoinitiators: Camphorquinone, Ethyl 4-dimethylaminobenzoate. Pigments: Titanium(IV) oxide, iron oxides. Non-sterile. Curing: 400-500 nm wavelength, 400 mW/cm² intensity. Compliant with ISO 4049 (Type 1, Class 2, Group 1) and ISO 10993.
Indications for Use
Indicated for dental professionals performing direct/indirect anterior and posterior restorations, core buildup, splinting, Class I/III/V restorations, pit and fissure sealing, and base/liner applications.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 23, 2022
HDI, Inc. Taekyou Kim CEO A-1504, 14, Sagimakgol-ro, 45 Beon-gil, Jungwon-gu Seongnam-si, Gyeonggi-do 13209 SOUTH KOREA
Re: K213339
Trade/Device Name: DENU Composite Resin, DENU Flow Resin Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth Shade Resin Material Regulatory Class: Class II Product Code: EBF, EBC Dated: July 22, 2022 Received: July 26, 2022
Dear Taekyou Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### 4. INDICATION FOR USE STATEMENT
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known) K213339
Device Name DENU Composite Resin, DENU Flow Resin
Indications for Use (Describe) DENU Composite Resin is indications for use in: - Direct anterior and posterior restorations
- Core buildup
- Splinting
- Indirect anterior and posterior restorations including inlays, onlays and veneers
DENU Flow Resin is indications for use in:
- Class III restorations
- Class V restorations
- Small Class I restorations (non stress-bearing restorations)
- Pit and fissure sealing in molars and premolars
- Repair of small defects in esthetic indirect inlays
- Base/liner under direct restorations
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995,
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW .*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
Page 1 of 1
POC Publiching Servism (101)-649-4743 EF
{3}------------------------------------------------
### 5. 510(k) SUMMARY
#### K213339
#### 510(k) Summary
Date: July 22, 2022
## 1. SUBMITTER
HDI, Inc.
A-1504, 14, Sagimakgol-ro, 45 Beon-gil, Jungwon-gu, Seongnam-si, Gyeonggi-do, Republic of Korea TEL : +82-31-735-3510 FAX : +82-31-735-3511 Contact Name: Taekyou Kim Email: hdikorea@hanmail.net
### 2. DEVICE
· Trade Name: DENU Composite Resin, DENU Flow Resin
·Common Name: Tooth shade resin material
· Classification Name: Material, Tooth shade, Resin
· Regulation Number 872.3690
·Class: 2
·Classification Product Code:
DENU Composite Resin
Primary product code: EBF
DENU Flow Resin
Primary product code: EBF
Secondary product codes: EBC, EJK
# 3. CLEARED DEVICE (PREDICATE DEVICE)
K083610, FILTEK SUPREME ULTRA UNIVERSAL RESTORATIVE, 3M ESPE K100235, FILTEK SUPREME ULTRA FLOWABLE RESTORATIVE, 3M ESPE
### 4. DEVICE DESCRIPTION
{4}------------------------------------------------
DENU Composite Resin is light-cured composite resin which can be used in anterior and posterior teeth. This is classified into type 1, class 2, group 1 according to 4, ISO 4049:2019.
DENU Flow Resin is light-cured flowable composite resin which can be used in anterior and posterior teeth. This is classified into type 1, class 2 group 1 according to 4, ISO 4049:2019.
## 5. INDICATIONS FOR USE
DENU Composite Resin is indicated for use in:
- Direct anterior and posterior restorations
- Core buildup
- Splinting
- Indirect anterior and posterior restorations including inlays, onlays and veneers
DENU Flow Resin is indicated for use in:
- Class III restorations
- Class V restorations
- Small Class I restorations (non stress-bearing restorations)
- Pit and fissure sealing in molars and premolars
- Repair of small defects in esthetic indirect inlays
- Base/liner under direct restorations
## 6. PERFORMANCE TESTING (NON-CLINICAL)
The following test articles were tested based on the referenced standard. All the test results met the preset test criteria.
- . ISO 4049 - Sensitivity of ambient light, Depth of cure, Flexural Strength, Water Sorption, Solubility, Radio-opacity, Color/Color Stability
- ISO 10993-3 Genotoxicity
- . ISO 10993-5 - Cytotoxicity
- ISO 10993-10 Skin sensitization, Oral mucosa irritation •
- ISO 10993-11 - Acute systemic toxicity
## 7. SUBSTANITAL EQUIVALENCE
{5}------------------------------------------------
# DENU Composite Resin
| Descriptive<br>Information | New device | Predicate device | Discuss/Justify<br>the Differences | |
|----------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------|-----|
| 510(k) Number | New | K083610 | - | |
| Trade Name | DENU Composite Resin | FiltekTM Supreme Ultra Universal Restorative | - | |
| Manufacturer | HDI, Inc. | 3M ESPE | - | |
| Common Name | Tooth shade resin material | Tooth shade resin material | Equivalent | |
| Device Class | 2 | 2 | Equivalent | |
| Product Code | EBF | EBF | Equivalent | |
| Regulation<br>Number | 21 CFR 872.3690 | 21 CFR 872.3690 | Equivalent | |
| Device<br>Description | DENU Composite Resin is<br>light-cured composite resin<br>which can be used in anterior<br>and posterior teeth. This is<br>classified into type 1, class 2,<br>group 1 according to 4, ISO in<br>4049:2019. | 3MTM ESPETTM FiltekTM<br>Universal Restorative<br>material is a visible-light<br>activated, restorative<br>composite designed for use<br>in anterior and posterior<br>restorations | Equivalent | |
| Intended<br>Use(including<br>the indications<br>for use) | - Direct anterior and posterior<br>restorations<br>- Core buildup<br>- Splinting<br>- Indirect anterior and<br>posterior restorations<br>including inlays, onlays and<br>veneers | - Direct anterior and<br>posterior restorations<br>(including occlusal<br>surfaces)<br>- Core build-ups<br>- Splinting<br>- Indirect restorations<br>including inlays, onlays and | Equivalent | |
| | | veneers | | |
| Intended user | Dental professional | Dental professional | Equivalent | |
| Composition of Materials | BISGMA | BISGMA | Similar | |
| | UDMA | UDMA | | |
| | TEGDMA | TEGDMA | | |
| | Ba glass | BISEMA-6 | | |
| | Silane | PEGDMA | | |
| | Camphorquinone | Silane treated ceramic | | |
| | Ethyl 4-<br>dimethylaminobenzoate | Silane Treated Silica | | |
| | Titanium(IV) oxide | Silane Treated Zirconia | | |
| | Yellow ferric oxide | Phenyl bis(2,4,6-<br>trimethylbenzoyl)-<br>phosphine oxide | | |
| | Iron(III) oxide | | | |
| | Iron(II,III) oxide | | | |
| Applicable<br>standards | ISO 4049<br>ISO 10993 | ISO 4049<br>ISO 10993 | Equivalent | |
| Physical properties | | | | |
| Compressive<br>strength | 187.4(46) MPa | DEB Shade<br>- 370.56 (15.13) MPa<br>T Shade<br>- 394.01 (25.05) MPa | Both<br>acceptance<br>criteria | |
| Flexural<br>strength | 135.74(11.72) MPa | DEB Shade<br>- 165.14 (13.59) MPa<br>T Shade<br>- 157.98 (8.16) MPa | Both<br>acceptance<br>criteria | |
| Elastic modulus | 7950.5(464.6) MPa | DEB Shade<br>- 11348 (271) MPa<br>T Shade<br>- 9180 (431) MPa | Both<br>acceptance<br>criteria | |
| Depth of cure | A0 - 3.86(0.15) mm<br>UO - 3.43(0.11) mm | 2.60(0.02) mm | Both<br>acceptance<br>criteria | met |
| Filler particle<br>size distribution | Silica filler 16nm<br>Glass filler 4 $ \mu $ m | Silica filler<br>- non-agglomerated/non-<br>aggregated : 20 nm<br>Zirconia filler<br>-non-agglomerated/non-<br>aggregated : 4 to11 nm<br>zirconia/silica cluster filler<br>- aggregated : comprised of<br>20 nm silica and 4 to 11 nm<br>zirconia particles | Both<br>acceptance<br>criteria | met |
| Surface<br>hardness | 43.86(3.02) KHN | 78.664(0.68) KHN | Both<br>acceptance<br>criteria | met |
| Radio-opacity | 3.2 mmAl | 2.1(0.0) mmAl | Both<br>acceptance<br>criteria | met |
| Water sorption | 18.48(0.84) $\mu$ g/mm3 | 33.1(2.1) $\mu$ g/mm3 | Both<br>acceptance<br>criteria | met |
| Solubility | 0.82(0.76) $\mu$ g/mm3 | 1.0(0.7) $\mu$ g/mm3 | Both<br>acceptance<br>criteria | met |
| Curing time | 20 sec | Dentin/Enamel/Translucent<br>shade(Increment depth<br>2.0mm) : 20 sec<br>Dentin, A6B, B5B shade<br>(Increment depth 1.5mm) : | Both<br>acceptance<br>criteria | met |
| | | 40 sec | | |
| Intensity<br>for curing | 400mW/cm²<br>(Halogen or LED) | 400 mW/cm2<br>(Halogen or LED) | Equivalent | |
| Wavelength for<br>curing | 400-500 nm (Halogen or<br>LED) | 400-500 nm (Halogen or<br>LED) | Equivalent | |
| Sterile | Non-sterile | Non-sterile | Equivalent | |
| Shelf Life | 3 years | 3 years | Equivalent | |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
## DENU Flow Resin
| Descriptive<br>Information | New device | Predicate device | Discuss/Justify<br>the Differences | |
|----------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------|-----|
| 510(k) Number | New | K100235 | - | |
| Trade Name | DENU Flow Resin | FiltekTM Supreme Ultra<br>Flowable Restorative | - | |
| Manufacturer | HDI, Inc. | 3M ESPE | - | |
| Common Name | Tooth shade resin material | Tooth shade resin material | Equivalent | |
| Device Class | 2 | 2 | Equivalent | |
| Product Code | EBF | EBF | Equivalent | |
| Regulation<br>Number | 21 CFR 872.3690 | 21 CFR 872.3690 | Equivalent | |
| Device<br>Description | DENU Flow Resin is light-cured flowable composite resin which can be used in anterior and posterior teeth. This is classified into type 1, class 2 group 1 according to 4, ISO 4049:2019. | Filtek Supreme Ultra Flowable Restorative, is a low viscosity, visible-light activated, flowable nanocomposite. | Equivalent | |
| Intended<br>Use(including<br>the indications<br>for use) | - Class III restorations<br>- Class V restorations<br>- Small Class I restorations (non stress-bearing<br>restorations)<br>- Pit and fissure sealing in molars and premolars<br>- Repair of small defects in esthetic indirect inlays<br>- Base/liner under direct restorations | -Class III and V restorations<br>- Restoration of minimally invasive cavity preparations (including small, non stress-bearing occiusal restorations)<br>- Base/liner under direct restorations<br>- Repair of small defects in esthetic indirect restorations<br>- Pit and fissure sealant<br>- Undercut blockout<br>- Repair of resin and acrylic temporary materials | Equivalent | |
| Intended user | Dental professional | Dental professional | Equivalent | |
| Composition of<br>Materials | BISGMA<br>UDMA<br>TEGDMA<br>Ba glass<br>Silane<br>Camphorquinone<br>Ethyl 4-dimethylaminobenzoate<br>Titanium(IV) oxide<br>Yellow ferric oxide<br>Iron(III) oxide<br>Iron(II,III) oxide | BISGMA<br>Substituted Demethacrylate<br>TEGDMA<br>Silane treated ceramic<br>Silane Treated Silica<br>Ytterbium Fluoride (Ybf3)<br>Reacted Polycaprolactone Polymer<br>Diphenyliodonium<br>Hexafluorophosphate | Similar | |
| Applicable<br>standards | ISO 4049<br>ISO 10993 | ISO 4049<br>ISO 10993 | Equivalent | |
| Compressive<br>strength | 277.4(29.4) MPa | 317.82 (17.20) MPa | Both<br>acceptance<br>criteria | met |
| Flexural<br>strength | 105.24(6.74) MPa | 120.96 (18.64) MPa | Both<br>acceptance<br>criteria | met |
| Elastic modulus | 8772.3(558.1) MPa | 6815.80 (924.00) MPa | Both<br>acceptance<br>criteria | met |
| Depth of cure | A0 - 2.93(0.05) mm<br>UO - 2.66(0.05) mm | 2.837(0.13) mm | Both<br>acceptance<br>criteria | met |
| Filler particle<br>size distribution | Silica filler 16nm<br>glass filler 4 μm | ytterbium trifluoride filler<br>-0.1 to 5.0 microns<br>Silica filler<br>-non-agglomerated/non-<br>aggregated surface<br>modified : 20 nm, 75 nm<br>zirconia/silica cluster filler<br>- surface modified<br>aggregated(comprised of<br>20 nm silica and 4 to 11 nm<br>zirconia particles) : 0.6 to<br>10 microns | Both<br>acceptance<br>criteria | met |
| Surface<br>hardness | 26.02(1.83) KHN | 45.124(0.16) KHN | Both<br>acceptance<br>criteria | met |
| Radio-opacity | 2.7 mmAl | 1.70(0.05) mmAl | Both<br>acceptance<br>criteria | met |
| Water sorption | 25.24(0.88) µg/mm3 | High translucency - 24.87<br>(2.23) µg/mm3<br>Medium translucency -<br>24.48 (1.99) µg/mm3<br>Low translucency - 28.32<br>(1.23) µg/mm3 | Both<br>acceptance<br>criteria | met |
| Solubility | 1.52(0.56) µg/mm3 | High translucency - 3.22<br>(0.20) µg/mm3<br>Medium translucency -<br>6.29 (0.23) µg/mm3<br>Low translucency - 9.77<br>(0.69) µg/mm3 | Both<br>acceptance<br>criteria | met |
| Curing time | 20 sec | Curing time :<br>Opaque(Increment depth<br>1.5mm) - 40 sec<br>All other shade(Increment<br>depth 2.0mm) - 20 sec | Both<br>acceptance<br>criteria | met |
| Intensity for<br>curing | 400mW/cm² (Halogen or<br>LED) | 400 mW/cm2 (Halogen or<br>LED) | Equivalent | |
| Wavelength for<br>curing | 400-500 nm (Halogen or<br>LED) | 400-500 nm (Halogen or<br>LED) | Equivalent | |
| Sterile | Non-sterile | Non-sterile | Equivalent | |
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
# 8. SUBSTANTIAL EQUIVALENCE DISCUSSION
DENU Composite Resin/DENU Flow Resin have the same Indications for Use and the principle of operations as the predicate devices. They are intended to perform as lightcured resin which met the requirement according to ISO 4049. They demonstrate similar physical properties and biocompatibilities with comparable performance specifications to the predicate devices.
{12}------------------------------------------------
The chemical compositions might be slightly different from the predicate devices, however subject devices and predicate devices use same resin matrix based on BISGMA, UDMA and TEGDMA. Other compositions such as filler, photoinitiator, and pigment operate under the same principle of operations.
The bench and biocompatibility testing performed demonstrates that any differences in their technological characteristics do not raise any new questions as to safety and effectiveness. Therefore, it is concluded that DENU Composite Resin/DENU Flow Resin is substantially equivalent to the predicate devices. Hence, its equivalent is acceptable.
#### 9. CONCLUSION
HDI Inc. believes that DENU Composite Resin and DENU Flow Resin are substantially equivalent to the legally marketed predicate device. They do not introduce new indications for use, has similar technological characteristics and do not introduce any new safety or effectiveness concerns.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.