V-HEPES PLUS is intended for oocyte collection, intracytoplasmic sperm injection (ICSI) fertilization, and oocyte and embryo washing and handling procedures. V-HEPES PLUS is also intended for transfer of embryos to the uterine cavity.
Device Story
V-HEPES PLUS is a reproductive medium used in assisted reproduction technology (ART) procedures. It serves as a solution for the collection, washing, handling, and fertilization of oocytes and embryos, as well as for the transfer of embryos into the uterine cavity. The device is provided as a sterile, aseptically filtered liquid in 100 mL PETG bottles. It is intended for use by trained clinicians in a laboratory or clinical setting. The medium supports the viability of oocytes and embryos during manipulation outside the body. By maintaining appropriate physiological conditions (pH, osmolality, and chemical composition), the device facilitates successful ART procedures, potentially improving outcomes for patients undergoing fertility treatments.
Clinical Evidence
Bench testing only. Biocompatibility testing (cytotoxicity, sensitization, vaginal irritation) per ISO 10993. Sterile filtration and aseptic fill validation per ISO 13408. Shelf-life validation per ASTM F1980. Mouse Embryo Assay (MEA) demonstrated ≥80% development to expanded blastocysts at 96 hours following 120-minute exposure. Endotoxin levels <0.25 EU/mL.
Technological Characteristics
Aqueous reproductive medium containing salts, amino acids, vitamins, glucose, HEPES buffer, and Human Serum Albumin (HSA). Provided in 100 mL PETG bottles. Sterilized via aseptic filtration. Shelf-life 12 months at 2-8°C. pH 7.0-7.4; osmolality 257-273 mOsm/kg. Biocompatibility testing per ISO 10993-1; shelf-life testing per ASTM F1980-16; transportation testing per ASTM D4169-16.
Indications for Use
Indicated for oocyte collection, ICSI fertilization, oocyte and embryo washing/handling, and embryo transfer to the uterine cavity.
Regulatory Classification
Identification
Reproductive media and supplement are products that are used for assisted reproduction procedures. Media include liquid and powder versions of various substances that come in direct physical contact with human gametes or embryos (including water, acid solutions used to treat gametes or embryos, rinsing solutions, sperm separation media, supplements, or oil used to cover the media) for the purposes of preparation, maintenance, transfer or storage. Supplements are specific reagents added to media to enhance specific properties of the media (e.g., proteins, sera, antibiotics, etc.).
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). The device, when it is phosphate-buffered saline used for washing, and short-term handling and manipulation of gametes and embryos; culture oil used as an overlay for culture media containing gametes and embryos; and water for assisted reproduction applications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 14, 2022
Vitromed Langenfeld c/o Greg Holland Sr. Partner Regulatory Specialists, Inc. 3722 Ave. Sausalito Irvine, CA 92606
Re: K213293
> Trade/Device Name: V-HEPES PLUS Regulation Number: 21 CFR 884.6180 Regulation Name: Reproductive Media and Supplements Regulatory Class: II Product Code: MQL Dated: August 15, 2022 Received: August 16, 2022
Dear Greg Holland:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213293
Device Name V-HEPES PLUS
Indications for Use (Describe)
V-HEPES PLUS is intended for oocyte collection, intracytoplasmic sperm injection (ICSI) fertilization, and oocyte and embryo washing and handling procedures. V-HEPES PLUS is also intended for transfer of embryos to the uterine cavity.
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.qov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary K213293
| 510(k) Owner | VITROMED GmbH |
|--------------|------------------------------|
| | Raiffeisenstr. 15a |
| | 40764 Langenfeld |
| | Germany |
| | Ines Diener |
| | Phone: +49 2173-20041-30 |
| | Facsimile: +49 2173-20041-58 |
Submission Correspondent Greg Holland Regulatory Specialists, Inc. 3722 Ave. Sausalito Irvine, CA 92606 Phone: 949.262.0411 Fax: 949.552.2821 Email: greg@requlatoryspecialists.com
Date Prepared
September 9, 2022
Trade Name V-HEPES PLUS Common Name Assisted Reproduction Media Requlation Name Reproductive Media and Supplements Regulation Number 21 CFR 884.6180 Class Class II Product Code MQL (Media, Reproductive) K142991 Predicate Device LifeGlobal Group, LLC LifeGlobal Global Total W/HEPES W/HSA
The predicate device has not been subject to a design-related recall.
### Device Description
V-HEPES PLUS is a medium for oocyte collection, intracytoplasmic sperm injection (ICSI) fertilization, and oocyte and embryo washing and handling procedures. V-HEPES PLUS is also intended for transfer of embryos to the uterine cavity. The medium is aseptically filtered and provided in a volume of 100 mL in pre-sterilized PETG bottles. V-HEPES PLUS has a shelf-life of 12 months when stored at 2-8°C and can be used for up to seven days after bottle opening. Additional information on the formulation and specifications of V-HEPES PLUS are provided in the Comparison of the Subject and Predicate Device Intended Use and Technological Characteristics section of this summary.
{4}------------------------------------------------
## Indications for Use
V-HEPES PLUS is intended for oocyte collection, intracytoplasmic sperm injection (ICSI) fertilization, and oocyte and embryo washing and handling procedures. V-HEPES PLUS is also intended for transfer of embryos to the uterine cavity.
## Comparison of the Subject and Predicate Device Intended Use and Technological Characteristics
A comparison of the intended use and technological features of the subject and predicate devices are described in the table below:
| | K213293 | K142991 | Comparison |
|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | V-HEPES PLUS | LifeGlobal Global Total | |
| | | w/HEPES w/HSA | |
| Indications for<br>Use | V-HEPES PLUS is intended<br>for oocyte collection,<br>intracytoplasmic sperm<br>injection (ICSI) fertilization,<br>and oocyte and embryo<br>washing and handling<br>procedures. V-HEPES<br>PLUS is also intended for<br>transfer of embryos to the<br>uterine cavity. | Oocyte and embryo<br>washing, manipulation,<br>fertilization by<br>intracytoplasmic sperm<br>injection (ICSI), and<br>embryo transfer. | There are differences in<br>the indications for use<br>statements for the<br>subject and predicate<br>device, including use of<br>the subject device for<br>collection of aspirated<br>oocytes; however, the<br>intended uses of the<br>subject and predicate<br>devices are the same. |
| Conditions for<br>Use | Prescription Use Only | Prescription Use Only | Same |
| Formulation | Water; Sodium Bicarbonate;<br>Calcium Lactate; Potassium<br>Chloride; Sodium Chloride;<br>Magnesium Sulfate;<br>Potassium Phosphate;<br>EDTA; Phenol Red, Sodium;<br>Taurine; Sodium Citrate; D-<br>Calcium Pantothenate;<br>Alanyl-Glutamine; Glucose;<br>Sodium Pyruvate;<br>Gentamicin Sulfate; L-<br>Aspartic Acid; L-Glutamic<br>acid; Glycine; L-Proline; L-<br>Asparagine; L-Serine; L-<br>Arginine Hydrochloride; L-<br>Cystine; L-Histidine<br>Hydrochloride; L-Isoleucine;<br>L-Leucine; L-Lysine<br>Hydrochloride: L-<br>Methionine; L-<br>Phenylalanine; L-Threonine;<br>L-Tryptophan; L-Tyrosine; L-<br>Valine; HEPES; Human<br>Serum Albumin (HSA) | Sodium Chloride;<br>Potassium Chloride;<br>Calcium Chloride;<br>Potassium Phosphate;<br>Magnesium Sulfate;<br>Sodium Bicarbonate;<br>Glucose; Sodium Lactate;<br>Sodium Pyruvate; L-<br>Arginine; L-Cystine; L-<br>Histidine; L-Isoleucine; L-<br>Leucine; L-Lysine; L-<br>Methionine; L-<br>Phenylalanine; L-<br>Threonine; L-Tryptophan;<br>L-Tyrosine; L-Valine; L-<br>Alanine; L-Asparagine; L-<br>Aspartic Acid; L-Glutamic<br>Acid; Glycine; L-Proline; L-<br>Serine: Glycyl-L-<br>glutamine; EDTA; Phenol<br>Red; Gentamicin Sulfate;<br>Water; Human Serum<br>Albumin (HSA); HEPES | Different: The<br>formulations of the<br>subject and predicate<br>devices include the<br>same types of chemical<br>constituents; however,<br>the formulations are not<br>the identical.<br>Differences in device<br>formulations do not<br>raise different<br>questions of safety and<br>effectiveness (S&E). |
| | K213293<br>V-HEPES PLUS | K142991<br>LifeGlobal Global Total<br>w/HEPES w/HSA | Comparison |
| Sterility | No growth | Sterile (SAL 10-3) | Different: The subject<br>and predicate devices<br>are both provided<br>sterile; however, their<br>specifications are<br>different. This<br>difference in sterility<br>specification does not<br>raise a different<br>question of S&E. |
| pH | 7.0-7.4 | 7.2-7.4 | Different: The subject<br>device has a lower pH<br>range than the<br>predicate device. This<br>difference in pH range<br>does not raise different<br>questions of S&E. |
| Osmolality<br>(mOsm/kg) | 257-273 | 260-270 | Similar |
| Mouse Embryo<br>Assay (MEA) | 1-Cell MEA:≥80%<br>developed to expanded<br>blastocysts at 96 h after<br>120-minute exposure | 1-Cell MEA: ≥80%<br>developed to expanded<br>blastocysts at 96 h after<br>one-hour exposure | Different: The duration<br>of exposure of the<br>subject and predicate<br>devices to mouse<br>embryos are not the<br>same. This difference<br>in MEA method due to<br>expected worst-case<br>exposure of media to<br>oocytes/embryos<br>during use does not<br>raise different<br>questions of S&E. |
| Endotoxin<br>(EU/mL) | <0.25 | ≤0.5 | Different: The subject<br>device has a lower<br>endotoxin level than the<br>predicate device. This<br>difference in endotoxin<br>specifications does not<br>raise different<br>questions of S&E. |
| Sterilization | Aseptic filtration | Aseptic filtration | Same |
| Storage<br>Condition | 2-8°C | 2-8°C | Same |
| Volume | 100 mL | Not known | Different: The volume<br>of the predicate device<br>is not known; however,<br>differences in device<br>volume do not raise<br>different questions of<br>S&E. |
| Shelf-Life | 12 months | 10 weeks | Different: The subject<br>device has a longer |
| K213293<br>V-HEPES PLUS | K142991<br>LifeGlobal Global Total<br>w/HEPES w/HSA | Comparison | |
| | | shelf-life than the<br>predicate device.<br>Differences in shelf-life<br>do not raise different<br>questions of S&E. | |
{5}------------------------------------------------
{6}------------------------------------------------
As shown in the table above, there are differences in the indications for use statements and technological features of the subject and predicate devices. However, as stated in the table, the differences in indications for use do not represent a new intended use and the differences in technological features do not raise different questions of safety and effectiveness.
## Summary of Non-Clinical Performance Testing
The following studies have been performed to support substantial equivalence to the predicate device:
- Biocompatibility testing was conducted in support of the subject device that will have direct contact with the patient during embryo transfer procedures. Testing was conducted in accordance with the 2020 FDA guidance Use of International Standard ISO 10993-1, Biological Evaluation of Medical Devices - Part 1: Evaluation and testing within a risk management process. Testing included:
- -Cytotoxicity per ISO 10993-5:2009
- Sensitization per ISO 10993-10:2010 -
- Vaginal Irritation per ISO 10993-23:2021 -
The testing demonstrated the device formulation to be non-cytotoxic, nonsensitizing, and non-irritating.
- Sterile filtration and aseptic fill validation, per ISO 13408-1:2008 and ISO 13408-● 2:2018.
- . Shelf-life testing was conducted to support the 12-month shelf-life for the subject device through demonstration that the product specifications (shown below) were met at time 0 and after accelerated aging in accordance with ASTM F1980-16:
- -Appearance: Pink rose color, no precipitates
- pH: 7.0—7.4 -
- Osmolality: 257-273 mOsm/kg -
- Endotoxin, per USP <85>: < 0.25 EU/mL -
- MEA testing, in accordance with the 2021 FDA quidance Mouse Embryo Assay for Assisted Reproduction Technology Devices: One-cell system: ≥80% embryos
{7}------------------------------------------------
developed to expanded blastocyst at 96 hours after 120-minute exposure to the subject device.
- Sterility, per USP <71>: No growth -
- Transportation testing per ASTM D4169-16 .
## Conclusions
The results of the performance testing described above demonstrate that V-HEPES PLUS is as safe and effective as the predicate device and support a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.