K213266 · Lespine Innovations, LLC · PEK · Dec 3, 2021 · Orthopedic
Device Facts
Record ID
K213266
Device Name
Inspan ScrewLES Fusion System
Applicant
Lespine Innovations, LLC
Product Code
PEK · Orthopedic
Decision Date
Dec 3, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3050
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K213266 · Dec 3, 2021
Inspan ScrewLES Fusion System
Lespine Innovations, LLC
Published clinical literature (Chin et al., 2020) reporting on a retrospective review of prospectively collected clinical data
The study evaluated clinical outcomes (VAS back pain and ODI) in 56 patients treated with the device for degenerative disc disease and lumbar spinal stenosis, demonstrating statistically significant improvements at 24 months post-operatively.
Retrospective review; Lumbar spinal stenosis; Degenerative disc disease; VAS back pain; ODI
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Chin et al. (2020); Retrospective review of prospectively collected data; Follow-up/Duration: 24 months
Patients with degenerative disc disease and lumbar spinal stenosis; Sample Size: 56
Not applicable for this study
VAS back pain and ODI
Indications for Use
The Inspan ScrewLES Fusion System is a posterior non-pedicle supplemental fixation system intended for use in the noncervical spine (T1-S1). It is intended for plate fixation/attachment to the spinous process for the purpose of achieving supplemental fusion for the following indications: spondylolisthesis, trauma (fracture or dislocation), tumor, degenerative disc disease (defined as discogenic pain with degeneration of the disc confirmed by history and radios), or lumbar spinal stenosis. The device is intended for use with bone graft material and is not intended for stand-alone use.
Device Story
Inspan ScrewLES Fusion System is a posterior non-pedicle spinal fixation device; provides supplemental stabilization to spinous processes to facilitate fusion. Components include plates of varying lengths/hub diameters, set screws, and implantation instruments. Plates feature integrated spikes to prevent rotation; set screws secure the construct. Used in non-cervical spine (T1-S1) by surgeons; requires bone graft material. Device is not for stand-alone use. Clinical benefit includes stabilization of spinal segments to address pain and degeneration associated with spondylolisthesis, trauma, tumor, or stenosis.
Clinical Evidence
Retrospective review of prospectively collected data (Chin et al., 2020) involving 56 patients treated for degenerative disc disease with lumbar spinal stenosis. Results showed statistically significant improvements in VAS back pain and ODI scores at 24 months post-operatively.
Technological Characteristics
Posterior spinal fixation orthosis; non-pedicle spinous process plate system. Materials: Titanium alloy (Ti-6Al-4V ELI) conforming to ASTM F 136. Components: Plates with integrated spikes, set screws. Supplied non-sterile; single-use. Mechanical fixation principle.
Indications for Use
Indicated for patients requiring supplemental posterior non-pedicle spinal fixation in the non-cervical spine (T1-S1) for spondylolisthesis, trauma (fracture/dislocation), tumor, degenerative disc disease, or lumbar spinal stenosis. Not for stand-alone use; must be used with bone graft.
Regulatory Classification
Identification
A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.
Predicate Devices
Vega™ SPAN™ Spinous Process Plate System (K102020)
Submission Summary (Full Text)
{0}------------------------------------------------
December 3, 2021
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
LESspine Innovations % Mr. Justin Eggleton VP, Spine Regulatory Affairs MCRA, LLC 803 7th Street NW Washington, District of Columbia 20141
Re: K213266
Trade/Device Name: Inspan® ScrewLES® Fusion System Regulation Number: 21 CFR 888.3050 Regulation Name: Spinal Interlaminal Fixation Orthosis Regulatory Class: Class II Product Code: PEK Dated: September 30, 2021 Received: September 30, 2021
Dear Mr. Eggleton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K213266
Device Name Inspan® ScrewLES® Fusion System
#### Indications for Use (Describe)
The Inspan ScrewLES Fusion System is a posterior non-pedicle supplemental fixation system intended for use in the noncervical spine (T1-S1). It is intended for plate fixation/attachment to the spinous process for the purpose of achieving supplemental fusion for the following indications: spondylolisthesis, trauma (fracture or dislocation), tumor, degenerative disc disease (defined as discogenic pain with degeneration of the disc confirmed by history and radios), or lumbar spinal stenosis. The device is intended for use with bone graft material and is not intended for stand-alone use.
Type of Use (Select one or both, as applicable)
| <span>☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|-------------------------------------------------------------|
| <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Device Trade Name: | Inspan® ScrewLES® Fusion |
|--------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer: | LESspine Innovations<br>350 Main Street, 2nd Floor<br>Malden, MA 02148, USA |
| Contact: | John Sullivan<br>VP, Global Quality & Regulatory<br>LESspine Innovations<br>350 Main Street, 2nd Floor<br>Malden, MA 02148, USA<br>Email: johnsullivan@kicventures.com |
| Date Prepared: | November 24, 2021 |
| Classifications: | Interlaminal spinal fixation orthosis (21 CFR §888.3050) |
| Class: | II |
| Product Code: | PEK |
## Indications for Use:
The Inspan ScrewLES Fusion System is a posterior non-pedicle supplemental fixation system intended for use in the non-cervical spine (T1-S1). It is intended for plate fixation/attachment to the spinous process for the purpose of achieving supplemental fusion for the following indications: spondylolisthesis, trauma (fracture or dislocation), tumor, degenerative disc disease (defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies), or lumbar spinal stenosis. The device is intended for use with bone graft material and is not intended for stand-alone use.
## Device Description:
The Inspan ScrewLES Fusion System, previously cleared (under K102020) and marketed as, "Vega™ SPANTM Spinous Process Plate System", consists of varying length and hub diameters, set screws, and instruments required for implantation. Spikes are present on the sides of the plate that interface with the spinous process to restrain the plate from rotating postoperatively. The device is offered with varying lengths and hub diameters to accommodate anatomical needs. Set screws are used to secure the two sides of the device into the final compressed and implanted construct.
System components are supplied non-sterile, are single use and are fabricated from titanium alloy (Ti-6AI-4V ELI) that conforms to ASTM F 136.
{4}------------------------------------------------
## Primary Predicate Device:
The subject device, Inspan ScrewLES Fusion System has been demonstrated to be substantially equivalent to the primary predicate device.
| Subject Device | Primary Predicate | | |
|-------------------------------|-------------------------------------------------|-----------------------|----------|
| | Device Name | Manufacturer | K-Number |
| Inspan ScrewLES Fusion System | VegaTM SPANTM Spinous<br>Process Plate System * | SpineFrontier, Inc.** | K102020 |
#### Table 1: Primary Predicate Device
*Since the original 510(k) clearance, Vega Span has been rebranded as Inspan ScrewLES Fusion System ** Since the original 510(k) clearance, SpineFrontier is now operating as, "LESspine, LLC."
#### Non-Clinical Performance Testing Summary:
Pre-clinical testing data submitted, referenced or relied upon to demonstrate substantial equivalence includes biocompatibility, sterilization, and non-clinical testing data.
## Clinical Performance Testing Summary:
A retrospective review of prospectively collected data was performed by Chin et al. (2020). In total, 56 patients received the subject device for degenerative disc disease with lumbar spinal stenosis where the authors reported statistically significant improvements in patients receiving the subject device for lumbar spinal stenosis at 24 months postoperatively for VAS back pain and ODI.
#### Substantial Equivalence:
The subject device was demonstrated to be substantially equivalent to the predicate devices cited in the passage above with respect to indications, design, materials, function, manufacturing, and performance.
## Conclusion:
The subject device is substantially equivalent to the cited predicate device with respect to indications for use, design, function, materials, and performance. Additionally, the provided clinical data related to the subject device, supports this traditional 510(k).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.