K213262 · STERIS Corporation · QKM · Jan 7, 2022 · General Hospital
Device Facts
Record ID
K213262
Device Name
CELERITY HP Chemical Indicator
Applicant
STERIS Corporation
Product Code
QKM · General Hospital
Decision Date
Jan 7, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2800
Device Class
Class 2
Indications for Use
The Celerity™ HP Chemical Indicator is a vaporized hydrogen peroxide multivariable chemical indicator. It is designed for routine monitoring of the following cycles: · Lumen, Non Lumen, Flexible, Fast Non Lumen or Fast sterilization cycle of a V-PRO® Low Temperature Sterilization System, or • Standard, Advanced, Express, Flex or Duo cycles of an ASP STERRAD® System, including those systems with ALLClear Technology.
Device Story
Celerity HP Chemical Indicator is a multivariable chemical indicator strip for monitoring vaporized hydrogen peroxide (VHP) sterilization cycles. Device consists of reactive indicator ink printed on a polypropylene strip. When exposed to VHP sterilization cycles, ink changes color from magenta to orange/yellow or lighter yellow if critical variables (time, temperature, sterilant concentration) are achieved. Used in clinical settings to verify sterilization cycle efficacy. Healthcare providers visually inspect the indicator post-cycle; color change confirms cycle completion. Provides quality assurance for sterilized instruments, ensuring patient safety by confirming sterilization parameters were met.
Clinical Evidence
Bench testing only. Performance verified against ANSI/AAMI/ISO 11140-1:2014 requirements for Type 4 VHP chemical indicators. Testing confirmed reaction to critical variables (time, temperature, concentration) and endpoint color stability up to 15 months. Simulated use testing (leveraged from K192020) confirmed performance in commercial sterilizers with worst-case loads, showing complete color change in full cycles and incomplete change in abbreviated cycles.
Technological Characteristics
Chemical indicator strip; polypropylene substrate; proprietary VHP-reactive ink. Multivariable indicator (Type 4 per ISO 11140-1:2014). Dimensions/form factor: strip. No energy source or software. Sterilization monitoring principle: chemical reaction to VHP exposure (4.7 mg/L, 29 seconds, 50°C).
Indications for Use
Indicated for routine monitoring of vaporized hydrogen peroxide sterilization cycles in V-PRO® Low Temperature Sterilization Systems and ASP STERRAD® Systems (including those with ALLClear Technology).
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
Predicate Devices
3M™ Attest™ Vaporized Hydrogen Peroxide Tri-Metric Chemical Indicator 1348/1348E (K203284)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
January 7, 2022
Steris Corporation Anthony Piotrkowski Director, Regulatory Affairs 5960 Heisley Rd Mentor, Ohio 44060
Re: K213262
Trade/Device Name: CELERITY HP Chemical Indicator Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: QKM Dated: December 7, 2021 Received: December 8, 2021
Dear Anthony Piotrkowski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
542 of the Act); 21 CFR 1000-1050.
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K213262
Device Name Celerity™ HP Chemical Indicator
### Indications for Use (Describe)
The Celerity™ HP Chemical Indicator is a vaporized hydrogen peroxide multivariable chemical indicator. It is designed for routine monitoring of the following cycles:
· Lumen, Non Lumen, Flexible, Fast Non Lumen or Fast sterilization cycle of a V-PRO® Low Temperature Sterilization System, or
• Standard, Advanced, Express, Flex or Duo cycles of an ASP STERRAD® System, including those systems with ALLClear Technology.
Type of Use (Select one or both, as applicable)Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# STI
# 510(k) Summary For CELERITY HP Chemical Indicator (CI)
## Sponsor Facility
STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No: (440) 639-4459
## Manufacturing Facility
Albert Browne Ltd., a subsidiary of STERIS Corporation Rayns Way Watermead Business Park Syston Leicester LE7 1PF UNITED KINGDOM
| Contact: | Anthony Piotrkowski |
|----------|------------------------------|
| | Director, Regulatory Affairs |
| | Telephone: (440) 392-7437 |
| | Fax No: (440) 357-9198 |
| | Tony_piotrkowski@steris.com |
| Submission Date: | January 6, 2022 |
|------------------|-----------------|
|------------------|-----------------|
K213262 510K number:
STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600
{4}------------------------------------------------
## STERIS Traditional 510(k) PREMARKET NOTIFICATION K213262 CELERITY HP Chemical Indicator (CI)
#### 1. Predicate Device
| Trade Name: | 3M™ Attest™ Vaporized Hydrogen Peroxide Tri-Metric<br>Chemical Indicator 1348/1348E |
|--------------------------|-------------------------------------------------------------------------------------|
| Common/Usual Name: | Chemical Indicator Classification: |
| | Class II |
| Classification Name: | Physical/chemical sterilization processindicator |
| 510(k) Submitter/Holder: | 3M Corporation |
| 510(k) Number: | K203284 |
#### 2. Device Description
The Celerity™ HP Chemical Indicator is a chemical indicator strip consisting of indicator ink containing the reactive chemicals printed on one end of a polypropylene strip. If the critical variables are achieved, the color of the indicator ink changes from magenta to orange/yellow or lighter yellow when exposed to the V-PRO® Low Temperature Sterilization System cycles or ASP STERRAD® System cycles. The indicator is validated to function as a multiple variable indicator with increased resistance characteristics similar to ISO 11140-1:2014 end points for a Type 4 Vaporized Hydrogen Peroxide (VHP) Chemical Indicator (CI). No changes have been made to the device other than additional testing and updating the labeling for multivariable indicator claims.
#### 3. Indications for Use
The Celerity™ HP Chemical Indicator is a vaporized hydrogen peroxide multivariable chemical indicator. It is designed for routine monitoring of the following cycles:
- · Lumen, Non Lumen, Flexible, Fast Non Lumen or Fast sterilization cycle of a V-PRO® Low Temperature Sterilization System, or
- Standard, Advanced, Express, Flex or Duo cycles of an ASP STERRAD® . System, including those systems with ALLClear Technology.
#### 4. Technological Characteristics
The proposed and predicate devices are chemical indicators in accordance with ISO 11140-1:2014 for use in monitoring Vaporized Hydrogen Peroxide sterilization cycles. The mechanism of action and endpoint are similar and when exposed to the defined processing conditions, the proposed and predicate devices exhibit a visible color change.
{5}------------------------------------------------
## STERIS Traditional 510(k) PREMARKET NOTIFICATION K213262 CELERITY HP Chemical Indicator (CI)
| Feature | Proposed: Celerity™ HP | Predicate: K203284, 3M Attest | Comparison |
|------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Chemical Indicator | Tri-Metric Chemical Indicator | |
| Intended Use /<br>Indications for<br>Use | The Celerity™ HP Chemical<br>Indicator is a vaporized hydrogen<br>peroxide multivariable chemical<br>indicator. It is designed for<br>routine monitoring of the<br>following cycles:<br>• Lumen, Non Lumen,<br>Flexible, Fast Non Lumen or<br>Fast sterilization cycle of a V-<br>PRO® Low Temperature<br>Sterilization System, or<br>• Standard, Advanced, Express,<br>Flex or Duo cycles of an ASP<br>STERRAD® System, including<br>those systems with ALLClear<br>Technology. | Use the 3MTM Attest™ Vaporized<br>Hydrogen Peroxide Tri-Metric Chemical<br>Indicator 1348/1348E for pack control<br>monitoring of the following hydrogen<br>peroxide sterilization sterilizers and<br>cycles: STERRAD® 100S System,<br>STERRAD® NX® System (Standard and<br>Advanced cycles), STERRAD® NX®<br>System with AllClear™ Technology<br>(Standard and Advanced cycles),<br>STERRAD® 100NX® System (Standard,<br>Flex, Express and Duo cycles)<br>STERRAD® 100NX® System with<br>AllClear™ Technology (Standard, Flex,<br>Express and Duo cycles) vaporized<br>hydrogen peroxide sterilizers and<br>STERIS® V-PRO® 1 (Lumen cycle),<br>STERIS® V-PRO® 1 Plus (Lumen and<br>Non-Lumen cycles) and STERIS® V-<br>PRO® maX Low Temperature<br>Sterilization System (Lumen, Non-<br>Lumen and Flexible cycles), STERIS®<br>V-PRO® 60 Temperature Sterilization<br>System (Lumen, Non-Lumen and<br>Flexible cycles), STERIS® V-PRO®<br>maX 2 Temperature Sterilization System<br>(Lumen, Non-Lumen, Flexible, and Fast<br>Non-lumen cycles), STERIS® V- PRO®<br>s2 Low Temperature Sterilization<br>System (Lumen, Non- Lumen, Flexible,<br>and Fast cycles) | Both predicate and subject<br>indicators are used for<br>monitoring the same<br>STERRAD and V-PRO<br>sterilization cycles. |
| Device design | Indicator Ink printed<br>onto polypropylene. | The 3MTM Attest™ Vaporized<br>Hydrogen Peroxide Tri-Metric<br>Chemical Indicator 1348/1348E is a<br>chemical indicator consisting of a non-<br>cellulose based coated indicator strip<br>sensitive to vaporized hydrogen<br>peroxide, contained in a film laminate | Both proposed and predicate<br>devices have a VH2O2<br>reactive ink on a non-<br>cellulosic substrate that<br>demonstrates a color change<br>when exposed to VH2O2. |
| Indicator agent | Proprietary | Proprietary | Both proposed and<br>predicate devices have a<br>VH2O2 reactive ink that<br>demonstrates a color<br>change when exposed to<br>VH2O2. |
| Critical<br>parameters | 4.7 mg/L - VH2O2<br>29 second - exposure time<br>50 °C - exposure Temperature | 5.1 mg/L - VH2O2<br>1 minute - exposure time<br>50 °C - exposure Temperature | Both proposed and<br>predicate devices only<br>react when all the variables<br>are reached. |
| Color change | Magenta to orange/yellow or<br>lighter yellow | Blue to pink with a moving dye front | Both proposed and<br>predicate devices use color<br>changes that are the same<br>as other cleared devices. |
| Endpoint<br>stability | 15 months | At least one month (4 weeks) | Both proposed and predicate<br>devices include labeling to<br>indicate endpoint stability. |
# Table 4-1. Summary of CI Physical Description and Technological Properties
{6}------------------------------------------------
## STERIS Traditional 510(k) PREMARKET NOTIFICATION K213262 CELERITY HP Chemical Indicator (CI)
The predicate and proposed devices are similar in terms of basic scientific technology and indications for use. Testing is included in the submission to demonstrate that the proposed CI meets the criteria of a multivariable indicator and to extend the endpoint color stability. The multivariable indicator claims change the product code from JOJ to QKM. Both product codes are under regulation number 21 CFR 880.2800.
#### .. Performance Testing
Performance testing was conducted to verify that the proposed Celerity CI meets the requirements for multivariable chemical indicator using the Type 4 vaporized hydrogen peroxide sterilization indicator requirements defined in ANSI/AAMI/ISO 11140-1:2014. In addition, ongoing endpoint color stability testing has demonstrated stability throughout 15 months, so the labeling has been updated to reflect this.
| Testing | Purpose | Acceptance Criteria | Study Result |
|---------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|
| Multivariable CI<br>Performance<br>Testing | Demonstrates that the indicator<br>reacts to all critical variables | Pass testing per Table 8<br>of ISO 11140-1:2014 | Pass. All CI show a<br>complete color change |
| | (time, temperature and sterilant<br>concentration) of the<br>sterilization process. | Fail testing per Table 8<br>of ISO 11140-1:2014 | Pass. All CI show an<br>incomplete color change |
| Endpoint color<br>stability | Demonstrates that the<br>processed indicator color does<br>not significantly change over<br>the labeled time period. | Color remains stable | Pass. Color stable up to<br>15 months |
| Simulated Use<br>Testing<br>(Leveraged from<br>K192020) | Demonstrates that the indicator<br>performs as expected in a<br>commercial sterilizer with<br>worst-case load. | Complete color change in<br>a full cycle.<br>Incomplete color change<br>in an abbreviated cycle. | Pass.<br>Complete color change in<br>a full cycle.<br>Incomplete color change<br>in an abbreviated cycle. |
Table 5-1. Verification Results Summary
The results of the performed testing demonstrate that the Celerity™ HP Chemical Indicator performs as intended.
#### 6. Conclusion
The conclusion drawn from the non-clinical test demonstrate that the subject device is as safe, as effective and perform as well as or better that the legally marketed predicate device (K203284, Class II as per 21 CFR 880.2800, product code QKM).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.