CMFlex™ is indicated for filling and/or augmenting maxillofacial, mandibular, and intraoral osseous defects. Indications include: - Intrabony periodontal osseous defects - Furcation defects - Bony defects or bony deficiencies of the alveolar ridge - Intraoral, maxillofacial, and mandibular augmentation - Bony defects of the upper or lower jaw - Filling of tooth extraction sites - Sinus elevation grafting
Device Story
CMFlex™ is a synthetic, osteoconductive bone grafting material provided as a sterile, single-use block. It is fabricated via extrusion-based 3D-printing of liquid inks, resulting in a porous structure composed of synthetic hydroxyapatite powder bound by a biodegradable polylactide-co-glycolide (PLG) polymer. Surgeons trim the blocks to fit specific patient bone defects in maxillofacial, mandibular, or intraoral sites. The device is not intended for immediate load-bearing; it functions by providing a scaffold that remodels over time, being replaced by new bone tissue. The device benefits patients by facilitating bone regeneration in osseous defects. It is used in clinical settings by surgeons who manually place the graft into the defect site.
Clinical Evidence
Non-clinical performance testing included chemical composition (TGA), crystallinity (XRD, ASTM F2024), morphology (SEM), porosity (BET), mechanical strength (ASTM D695), pH (ASTM F1635), and biocompatibility (ISO 10993). Preclinical performance was evaluated in a canine one-wall dental defect model at 4, 8, and 12 weeks using radiography, microCT, and histomorphometry. Results showed CMFlex™ maintained ridge dimensions and exhibited bone-forming capacity and inflammatory response comparable to the predicate device.
Technological Characteristics
Synthetic bone graft composed of 90% hydroxyapatite and 10% PLG. Fabricated via extrusion-based 3D-printing. Porosity >80%, pore size 0.20-0.27 mm. Sterile, single-use block form. Standards: ASTM F2024-10R21 (crystallinity), ASTM D695-15 (compressive strength), ASTM F1635-16 (pH), ISO 10993 (biocompatibility), ISO 11737-1 (bioburden), ISO 11138-1 (sterilization).
Indications for Use
Indicated for filling and/or augmenting maxillofacial, mandibular, and intraoral osseous defects, including periodontal defects, furcation defects, alveolar ridge deficiencies, tooth extraction sites, and sinus elevation grafting.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Dimension Inx Corp. Ramille Shah Chief Science Officer 3440 S Dearborn St. Suite 142S Chicago, Illinois 60616
Re: K213260
Trade/Device Name: CMFlex™ Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: Class II Product Code: LYC Dated: December 19, 2022 Received: December 19, 2022
Dear Ramille Shah:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Andrew I. Steen -S
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K21360
Device Name CMFlex(TM)
#### Indications for Use (Describe)
CMFlex™ is indicated for filling and/or augmenting maxillofacial, mandibular, and intraoral osseous defects. Indications include: 22552
- · Intrabony periodontal osseous defects
- · Furcation defects
- · Bony defects or bony deficiencies of the alveolar ridge
- · Intraoral, maxillofacial, and mandibular augmentation
- · Bony defects of the upper or lower jaw
- · Filling of tooth extraction sites
- · Sinus elevation grafting
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K213260 CMFlex™
#### 510(K) Summary
DEVICE TRADE NAME CMFlex™
#### MANUFACTURER
Dimension Inx 3440 S. Dearborn St., Suite 142S Chicago, IL 60616 Phone: (312) 235-3510
## CONTACT
Ramille Shah, Ph.D. ramilleshah(@dimensioninx.com
#### Consultant:
Mehdi Kazemzadeh-Narbat, PhD, PMP, CQA Associate Director, Regulatory Affairs, MCRA, LLC Office: (202) 552-6011 mkazemzadeh(@mcra.com
#### DATE PREPARED
December 30, 2022
#### CLASSIFICATION
Bone Grafting Material, Synthetic (21 CFR 872.3930)
## PRODUCT CODE LYC
#### PRIMARY PREDICATE
Trade name: OsteoScaf™ Common name: Bone Grafting Material, Synthetic 510(k) number: K101827
#### REFERENCE DEVICE
Trade name: Easy-Graft® Common name: Bone Grafting Material, Synthetic 510(k) number: K131385
#### INDICATIONS FOR USE
CMFlex™ is indicated for filling and/or augmenting maxillofacial, mandibular, and intraoral osseous defects. Indications include:
- -Intrabony periodontal osseous defects
- Furcation defects
- Bony defects or bony deficiencies of the alveolar ridge -
- -Intraoral, maxillofacial, and mandibular augmentation
- Bony defects of the upper or lower jaw -
- -Filling of tooth extraction sites
- -Sinus elevation grafting
{4}------------------------------------------------
## DEVICE DESCRIPTION
CMFlex™ is a synthetic bone grafting material provided in block form of varying sizes (See Table 1.1) that can be easily trimmed or cut by the surgeon to fit the patient's bone defect(s). CMFlex™ is composed of majority synthetic hydroxyapatite powder bound by minority biodegradable polylactide-co-glycolide. CMFlex™ is fabricated via extrusion-based 3D-printing of liquid inks into regular porous structures. The combined macroscopic 3D-printed porosity, microporosity within the printed struts, and micron-sized hydroxyapatite particles gives CMFlex unique microstructural and physical properties. CMFlex™ is an osteoconductive, highly absorbent, and flexible bone graft that can be used in defects where new bone is needed. Although it is not intended for immediate load bearing applications, the implant remodels over time and is replaced by new bone tissue, functioning in the same manner as the predicate device. The blocks are provided sterile and are intended for single use. There are no accessories associated with CMFlex™.
# CMFLEXTM BLOCK SIZES AND VOLUMES
| Product | Dimensions (mm) | Approximate<br>Volume |
|---------------------|-------------------|-----------------------|
| CMFlex™ Block, 1cc | <br>15<br>15<br>5 | 1 cc |
| CMFlex™ Block, 6cc | <br>38<br>38<br>4 | 6 cc |
| CMFlex™ Block, 13cc | <br>38<br>38<br>9 | 13 cc |
Table 1.1. CMFlex™ Block Sizes and Volumes
# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
| | Table 1.2. CMFlex™ as Compared to Predicate and Reference Devices |
|--|-------------------------------------------------------------------|
|--|-------------------------------------------------------------------|
| | Subject Device | Primary Predicate<br>Device | Secondary Predicate<br>Device | Comparison |
|--------------------------|---------------------|-------------------------------------------------------|-------------------------------|----------------------------|
| Trade Name | CMFlex™ | OsteoScaf™ | Easy-Graft® | N/A |
| Manufacturer | Dimension Inx Corp. | Texas Innovative<br>Medical Devices<br>(DBA) Skeletal | Guidor® | N/A |
| 510(k) Number | K213260 | K101827 | K131385 | N/A |
| Product Code | | LYC | | Same product code |
| Device Class | | Class II | | Same device class |
| Device<br>Classification | | Bone grafting material, synthetic | | Same device classification |
{5}------------------------------------------------
# Dimension Inx
| Indications for Use | Indicated for filling<br>and/or augmenting<br>maxillofacial,<br>mandibular, and<br>intraoral osseous<br>defects. Indications<br>may include:<br>- Intrabony<br>periodontal osseous<br>defects<br>- Furcation defects<br>- Bony defects or bony<br>deficiencies of the<br>alveolar ridge<br>- Intraoral,<br>maxillofacial, and<br>mandibular<br>augmentation<br>- Bony defects of the<br>upper or lower jaw<br>- Filling of tooth<br>extraction sites<br>- Sinus elevation<br>grafting | Indicated for filling<br>and/or augmenting<br>intraoral/maxillofacial<br>osseous defects, such<br>as intrabony<br>periodontal osseous<br>defects, furcation<br>defects, augmentation<br>of bony defects of the<br>alveolar ridge, filling<br>of tooth extraction<br>sites, and sinus<br>elevation grafting. | Indicated for the<br>treatment of intraoral /<br>maxillofacial osseous<br>defects. Dental and<br>maxillo-facial<br>indications may<br>include: extraction<br>defects (alveolar ridge<br>preservation),<br>periodontal defects,<br>pen-implant defects,<br>augmentation of<br>deficient alveolar crest,<br>sinus floor<br>augmentation, defects<br>after surgical<br>extractions, defects<br>after removal of bony<br>cysts, defects after root<br>resection or<br>apicoectomy, defects<br>after removal of<br>autologous bone. | Intended use includes<br>elements of the predicate<br>devices, supported by<br>performance testing in canines |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
{6}------------------------------------------------
| Prescription/OTC | Prescription | | | All require prescription |
|------------------------|-----------------------------------------|-----------------------------------------|---------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| Composition | 90% HA<br>10% PLG | 78% CP<br>22% PLG | 99% β-TCP<br>1% PLG<br>N-methyl-2-<br>pyrrolidone | All devices comprised of a<br>majority calcium phosphate<br>component combined with a<br>degradable polymer<br>component |
| Physical<br>Morphology | Interconnecting pore structure | | | Similar pore morphology |
| Porosity (Nominal) | 85% | 80% | 53% | Similar porosity to primary<br>predicate |
| Pore Size | 0.20 - 0.27 mm | 0.20 mm | 0.23 mm | All devices have similar pore<br>sizes |
| Form | Block | Particulate, Cylinder,<br>Block | Granule | Similar form factor to primary<br>predicate |
| Crystallinity | 96% | Not Available | 92% | Similar crystallinity to<br>secondary predicate |
| Resorption1 | Resorbable | Resorbable | Resorbable | Similar resorption<br>characteristics to primary<br>predicate |
| Sterility | Sterile; One-time<br>single patient use | Sterile; One-time<br>single patient use | Sterile; One-time<br>single patient use | All devices are sterile and for<br>single use only |
1Based on composition
# Device Intended Use and Descriptions
CMFlex™, OsteoScaf™, and Easy-Graft® are all sterile, porous, ceramic-polymer composite single-use bone grafting materials intended for use in the filling, augmenting, or treating intraoral/maxillofacial defects. All three are class II devices and fall under the same product code, LYC.
# Device Compositions and Resorption
CMFlex™, OsteoScaf™ and Easy-Graft® are comprised of discrete micro-scale, spherical calcium phosphate ceramic particles (90. 78. 99 wt.%, respectively) and the biodegradable polyester binder, poly(lactide-co-glycolide), or PLG, (10, 22, 1 wt.%, respectively). The phase of the calcium phosphate component of CMFlex™ is primarily synthetic hydroxyapatite, with minor amounts of beta tricalcium phosphate (B-TCP), while the phases of the calcium phosphate components of OsteoScaf™ and Easy-Graft® are minority hydroxyapatite, with majority di/tetra calcium phosphate and ß-TCP, respectively. The calcium phosphate components of CMFlex™ and Easy-Graft® are primarily crystalline (>90%) and the particles comprising CMFlex™ are 23 um in diameter on average, while those of OsteoScaf™ are 27.5 um on average. The difference in composition among the devices primarily affect the rate of resorption of the devices, but do not affect the substantial equivalanceof the devices for their intended use.
{7}------------------------------------------------
# Device Porositv
Like CMFlex™, OsteoScaf™, and Easy-Graft® possess interconnected micro-and macro-pores formed by overlapping fibers or bound calcium phosphate particles. While CMFlex™ and OsteoScaf™ have total porosity exceeding 80 vol. %, Easy-Graft® had a measured porosity of 53.4 vol.%; however, this value does not include the interconnected porosity within the larger calcium phosphate particles that comprise the majority of Easy-Graft® (which could not be accurately measured). Thus, the actual total porosity of Easy-Graft® may be significantly greater than 53.4 vol.%. and more similar to CMFlex™ and OsteoScaf™ (both >80% porosity with 0.25-1.20 mm pore size). Like CMFlex™, OsteoScaf™ and Easy-Graft® are flexible or moldable, absorb fluid, and are bioresorbable.
# Device Form Factors and Volumes
CMFlex™ and OsteoScaf™ are comparable in form factors and sizing: CMFlex™ is available in blocks of varying volumes, while OsteoScaf™ is available in blocks, cylinders, and particulates. Like the various OsteoScaf™ form factors, the various CMFlex™ block volume options are provided as a convenience to the surgeon. Both CMFlex™ and OsteoScaf™ may be further trimmed to ensure an appropriate, custom fit to the patient's defect. Easy-Graft® is available in multiple volumes, but rather than solid form factors, it comes as a kit of PLG coated calcium phosphate particles that are mixed with solvent prior to injection into boney defect. All CMFlex™ volumes share a single "indications for use" statement, which is common to the "indications for use" statement provided for the OsteoScaf™ and Easy-Graft® predicate devices.
# PERFORMANCE DATA
The following non-clinical and preclinical tests were performed to support a demonstration of substantial equivalence:
| Purpose | Method | Conclusions<br>(based on testing results and published information) |
|------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Chemical composition | Thermogravimetric analysis | The subject and predicate devices are composed of a majority calcium phosphate component and minority biodegradable polyester component. |
| Crystallinity (of calcium phosphate component) | X-ray diffraction, ASTM F2024-10R21 | The subject and predicate devices have calcium phosphate crystallinities >90%. |
| Calcium phosphate particle morphology | Scanning electron microscopy | The subject and predicate devices possess spherical calcium phosphate particles. |
| Pore structure | Scanning electron microscopy | The subject and predicate devices have interconnected pores around 0.2mm and primary pores around 1mm. |
| Porosity | Porosimetry (BET analysis) | The subject and predicate devices have porosities greater than 50%. |
| Mechanical strength (compressive) | Compressive mechanical testing, ASTM D695-15 | The devices provide adequate handling properties and strength to ensure dimensional integrity when handled by the surgeon and delivered to bony application sites; devices have similar compressive properties. |
| pH in phosphate buffered saline | ASTM F1635-16 | The subject and predicate device have similar pH. |
| Biocompatibility | ISO 10993 series of standards | Devices passed cytotoxicity, sensitization, irritation, material-mediated pyrogenicity, genotoxicity, and systemic toxicity testing according to ISO 10993. |
| Performance in animal<br>model | Canine critical sized<br>defect model and ISO<br>10993-6 | CMFlexTM was evaluated for performance and<br>biocompatibility as compared to the predicate device in a<br>one-wall dental defect canine model at 4, 8, and 12 weeks.<br>Assessments included radiography, microCT, and<br>histomorphometric analysis. Endpoints included new bone<br>growth, residual implant material, preservation of ridge<br>(height, width, depth), changes to surrounding bone, and<br>any significant adverse findings (i.e., exuberant necrosis,<br>proliferative granulation tissue/fibrosis, extensive<br>inflammation, or evidence of infection). Results<br>demonstrated that CMFlexTM was able to maintain ridge<br>height, width, and depth and showed similar bone forming<br>capacity and steady state inflammatory response<br>accompanying the absorption<br>process when compared to the predicate device. |
| Sterilization Validation | Pre-Sterilization<br>Bioburden (ISO 11737-<br>1); LAL Endotoxin<br>(USP<85>); Chamber<br>Mapping; Residuals<br>testing; Bioburden<br>Recovery, Inactivation<br>Kinetics (ISO 11138-1);<br>Performance<br>Qualification (ISO<br>11138-1:2017, ISO<br>11138-7:2019, ISO<br>14937:2009) | CMFlexTM passed all acceptance criteria for validating the<br>end sterilization method. |
| Shelf Life | Device properties after<br>accelerated and real time<br>aging: Compressive<br>mechanical testing,<br>polymer molecular<br>weight, ceramic<br>composition and<br>crystallinity (ASTM<br>F2024:2021), and<br>microstructure (SEM) | CMFlexTM had no significant mechanical property changes<br>or clinically meaningful compositional (ceramic and<br>polymer component) and microstructural changes after<br>accelerated (12-month simulation) and real-time aging (6<br>months and 12 months). |
| Package Integrity Testing<br>(Initial and aged<br>samples) | ASTM F2096; ASTM<br>F88 | CMFlexTM packaging passed all package integrity testing<br>as is and after accelerated (12-month simulation) and real-<br>time aging (6 months and 12 months). |
Table 1.3 Substantial Equivalence Performance Testing Summary
{8}------------------------------------------------
# CONCLUSIONS
CMFlex™ was compared with the identified predicate devices based on the indications for use, composition, porosity, mechanical strength, biocompatibility testing, and performance studies. CMF lex™ has similar compositional components (e.g., a combination of a biodegradable polymer and calcium phosphate particles), microstructural properties (percent porosity and pore size), pH, and compressive mechanical characteristics. Furthermore, CMFlex™ passed all biocompatibility testing per ISO 10993, and head-to-head comparison with the predicate in a canine critically sized dental defect model showed similar bone forming capacity associated with device remodeling and osteoconduction over time. Based on these studies, the subject device, CMFlex™ is found to be substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.