The NovoFlow Reinforced ureteral stents are intended for patients 12 years of age (40 kg) and over for drainage of the upper urinary tract over fistulas or ureteral obstacles and/or for healing of the ureter. These stents may remain implanted for up to 6 months.
Device Story
Implantable ureteral stent; maintains urine drainage; facilitates ureteral healing. Inserted via retrograde or antegrade surgical technique. Supplied in 7 Fr or 8 Fr diameters; 26-30 cm lengths. Kit includes double loop stent, steerable pusher, and optional hydrophilic guidewire. Used in clinical/surgical settings by physicians. Stent loops at renal and vesical ends secure position. Radiopaque polymeric construction allows visualization. Benefits patient by relieving obstruction and promoting healing. Sterile, single-use device.
Clinical Evidence
No clinical data or animal studies provided. Substantial equivalence supported by bench testing, including visual inspection, dimensional testing, flow rate, tensile strength, loop/curl strength, guidewire compatibility, radiopacity, MRI compatibility, and shelf-life testing per FDA guidance.
Technological Characteristics
Radiopaque polymeric material; double loop design; 7 Fr or 8 Fr diameters; 26-30 cm lengths. Sterile (ethylene oxide, SAL 10^-6). Kit includes steerable pusher and hydrophilic guidewire. No software or electronic components.
Indications for Use
Indicated for patients 12 years of age (40 kg) and older requiring upper urinary tract drainage due to fistulas or ureteral obstacles, or for ureteral healing. Stents may remain implanted for up to 6 months.
Regulatory Classification
Identification
A ureteral stent is a tube-like implanted device that is inserted into the ureter to provide ureteral rigidity and allow the passage of urine. The device may have finger-like protrusions or hooked ends to keep the tube in place. It is used in the treatment of ureteral injuries and ureteral obstruction.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 16, 2022
Coloplast Corp Gayatri Ghadge Principal Regulatory Affairs Specialist 1601 West River Road North Minneapolis. MN 55411
Re: K213186
Trade/Device Name: NovoFlow™ Reinforced Ureteral Stent Regulation Number: 21 CFR§ 876.4620 Regulation Name: Ureteral Stent Regulatory Class: II Product Code: FAD Dated: January 19, 2022 Received: January 20, 2022
## Dear Gayatri Ghadge:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica K. Nguyen, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213186
Device Name
NovoFlow™ Reinforced Ureteral Stent
### Indications for Use (Describe)
The NovoFlow Reinforced ureteral stents are intended for patients 12 years of age (40 kg) and over for drainage of the upper urinary tract over fistulas or ureteral obstacles and/or for healing of the ureter. These stents may remain implanted for up to 6 months.
| Type of Use (Select one or both, as applicable) | | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|--|
| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. | | |
| *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW * | | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the<br>time to review instructions, search existing data sources, gather and maintain the data needed and complete<br>and review the collection of information. Send comments regarding this burden estimate or any other aspect<br>of this information collection, including suggestions for reducing this burden, to: | | |
| Food and Drug Administration<br>Office of Chief Information Officer<br>Paperwork Reduction Act (PRA) Staff<br>PRAStaff(@fda.hhs.gov | Department of Health and Human Services | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of<br>information unless it displays a currently valid OMB number." | | |
Page 1 of 1
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Coloplast logo, which is a blue circle with white lines inside. Next to the logo is the word "Coloplast" in blue, bold letters. Below the logo and word is the address "1601 West River Road North, Minneapolis, MN 55411".
K213186 Page 1 of 3
### 5. 510(k) SUMMARY
| I. SUBMITTER INFORMATION | |
|-----------------------------|---------------------------------------------------------------------------------------|
| 510(K) Owner's Name: | Coloplast A/S |
| Legal Manufacturer Address: | Holtedam 1<br>3050 Humlebaek, Denmark |
| Phone/Fax/Email: | Phone: 612-422-3206<br>Email: usggh@coloplast.com |
| Name of Contact Person: | Gayatri Ghadge<br>Principal Regulatory Affairs Specialist |
| Address/Contact: | 1601 West River Road<br>Minneapolis, MN 55411 |
| Date Prepared: | January 19, 2022 |
| II. DEVICE | |
| Trade or Proprietary Name: | NovoFlow™ Reinforced Ureteral Stent |
| Common or Usual Name: | Ureteral Stents |
| Classification Name: | Stent, Ureteral<br>(21 CFR section 876.4620)<br>Product Code: FAD<br>Device Class: II |
### PREDICATE DEVICE IDENTIFICATION III.
Vortek® Double Loop Ureteral Stents (K180057). Predicate device:
This predicate has not been subject to a design-related recall. No reference devices were used in this submission.
### DEVICE DESCRIPTION IV.
NovoFlow Reinforced Ureteral Stents are implantable ureteral stents used to maintain urine drainage and allow healing of the ureter. They are inserted into the ureter during a surgical procedure using mainly a retrograde technique, although an antegrade approach is also possible.
The stents are supplied in 7 Fr or 8 Fr diameters and lengths from 26 cm through 30 cm, with both tips open (O/O). NovoFlow Reinforced Ureteral Stents are supplied in kits which contain the following components:
- A double loop ureteral stent
Page 5-1
{4}------------------------------------------------
1601 West River Road North Minneapolis, MN 55411
- A steerable pusher, packed separately
Also contained in some kits:
- -An Orchestra® Hydrophilic guidewire, packed separately
The NovoFlow Reinforced Ureteral Stents and accessories included in the kits are supplied sterile via ethylene oxide. The ureteral stent and each accessory are packaged and sterilized separately prior to being combined in the kit.
### V. INDICATIONS FOR USE
The NovoFlow Reinforced ureteral stents are intended for patients 12 years of age (40 kg) and over for drainage of the upper urinary tract over fistulas or ureteral obstacles and/or for healing of the ureter. These stents may remain implanted for up to 6 months.
### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVICE
NovoFlow Reinforced Ureteral Stents are substantially equivalent in performance, design and materials to the Vortek® Double Loop Ureteral Stents, cleared under premarket notification number K180057. All devices are tubes made from the same radiopaque, polymeric materials with loops at both the renal and vesical ends to hold the stent in place. Both devices use guidewires and pushers and are intended to be implanted for up to six months. All the devices are sold as kits with the same accessories intended to facilitate implantation. The main differences between the NovoFlow Reinforced Ureteral Stents and the predicate device are that the subject device is available in only 7 and 8 Fr sizes while the predicate device is available in 4.8 -8 Fr. In addition, some models of the predicate device contain a suture, while the NovoFlow Reinforced Ureteral Stents do not contain a suture.
### VII. PERFORMANCE DATA
The following performance data was provided in support of the substantial equivalence determination.
## Biocompatibility Testing
Biocompatibility testing was conducted based upon ISO 10993-1 (2018): Biological evaluation of medical devices - Part 1: "Evaluation and testing within a risk management process" and FDA Guidance for Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a risk management process" - Guidance for Industry and Food and Drug Administration Staff - June 16, 2016.
{5}------------------------------------------------
1601 West River Road North Minneapolis. MN 55411
## Mechanical Testing
Mechanical testing was completed using the FDA guidance document "Guidance for the content of premarket notifications for ureteral stents" for reference. The following tests were performed:
- Visual Inspection
- Dimensional Testing
- o Diameter
- o Length
- Flow Rate ●
- Elongation/Yield and Tensile Strength ●
- Loop/Curl strength ●
- Guidewire compatibility
- Radiopacity
- Magnetic resonance compatibility ●
- Shelf Life/Expiration Date ●
## Sterilization
The NovoFlow Reinforced Ureteral Stents and accessories are sterilized using ethylene oxide in a validated cycle, demonstrating a sterility assurance level of 10-6.
### Packaging and Distribution
The NovoFlow Reinforced Ureteral Stents were subjected to distribution testing and verification testing to demonstrate that the product and package would be undamaged throughout the product life and maintain the device sterility.
No animal studies or clinical testing were provided to support substantial equivalence between the subject and predicate devices.
## VIII. CONCLUSIONS
The NovoFlow Reinforced Ureteral Stents and related accessories have been demonstrated to be substantially equivalent to the predicate device, Vortek Double Loop Ureteral Stents based on the non-clinical data provided, similar intended use, patient population, materials, biocompatibility, kit composition, and technological characteristics. The differences in technological characteristics do not raise new questions of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.