Silicone double loop ureteral stents: The Silicone double loop ureteral stents are intended for adult and pediatric (children and adolescents) patients for drainage of the upper urinary tract over fistulas or ureteral obstacles and/or for healing of the ureter. These stents may remain implanted for up to 12 months. Pyelostent and Stenostent Silicone double loop ureteral stents: Drainage of the upper urinary tract and/or ureter healing during management of ureteral stenosis: For Pyelostent® Silicone double loop ureteral stents: Partial enlargement of the stent diameter, for localized stenosis of the ureteropelvic junction in adult and pediatric (adolescents) patients. For Stenostent® Silicone double loop ureteral stents: Total enlargement of the stent diameter, for ureteral stenosis in adult and pediatric (children and adolescents) patients. The Pyelostent® and Stenostent® Silicone double loop ureteral sterts may remain implanted for up to 12 months.
Device Story
Implantable silicone double loop ureteral stents; used for drainage of upper urinary tract and ureteral healing; indicated for fistulas, obstacles, or stenosis. Pyelostent and Stenostent models feature partial or total reinforcement along the straight catheter section for crush resistance. Provided as kits with pusher and optional guidewire. Single-use; ethylene oxide sterilized. Implanted by physicians in clinical settings. Stents remain in situ for up to 12 months to maintain ureteral patency and facilitate healing.
Clinical Evidence
No clinical data or animal studies provided. Substantial equivalence supported by bench testing, including visual evaluation, dimensional testing, flowrate (EN 1618:1997), tensile strength (EN 1618:1997), elongation, loop strength, radiopacity, MRI safety, and package integrity.
Technological Characteristics
Silicone double loop ureteral stents; single-use; ethylene oxide sterilized (ISO 11135-2014). Features include reinforced straight catheter sections (Pyelostent/Stenostent) for crush resistance. Mechanical properties tested per EN 1618:1997. MRI compatible.
Indications for Use
Indicated for adult and pediatric patients requiring upper urinary tract drainage due to fistulas, ureteral obstacles, or ureteral stenosis; includes localized ureteropelvic junction stenosis (Pyelostent) and general ureteral stenosis (Stenostent).
Regulatory Classification
Identification
A ureteral stent is a tube-like implanted device that is inserted into the ureter to provide ureteral rigidity and allow the passage of urine. The device may have finger-like protrusions or hooked ends to keep the tube in place. It is used in the treatment of ureteral injuries and ureteral obstruction.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
June 8, 2022
Coloplast Corp. Gayatri Ghadge Principal Regulatory Affairs Specialist 1601 West River Road North Minneapolis, MN 55411
Re: K213185
Trade/Device Name: ImaJin® Silicone double loop ureteral stent kits, ImaJin® Pyelostent® Silicone double loop ureteral stent kits, ImaJin® Stenostent® Silicone double loop ureteral stent kits
Regulation Number: 21 CFR§ 876.4620 Regulation Name: Ureteral Stent Regulatory Class: II Product Code: FAD Dated: May 6, 2022 Received: May 9, 2022
Dear Gayatri Ghadge:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{1}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products): good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Jessica K. Nguven, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
### K213185
Device Name
ImaJin® Silicone double loop ureteral stent kits Imalin® Pyelostent® Silicone double loop ureteral stent kits. ImaJin® Stenostent® Silicone double loop ureteral stent kits
Indications for Use (Describe)
Silicone double loop ureteral stents:
The Silicone double loop ureteral stents are intended for adult and pediatric (children and adolescents) patients for drainage of the upper urinary tract over fistulas or ureteral obstacles and/or for healing of the ureter. These stents may remain implanted for up to 12 months.
Pyelostent and Stenostent Silicone double loop ureteral stents: Drainage of the upper urinary tract and/or ureter healing during management of ureteral stenosis:
For Pyelostent® Silicone double loop ureteral stents:
Partial enlargement of the stent diameter, for localized stenosis of the ureteropelvic junction in adult and pediatric (adolescents) patients.
For Stenostent® Silicone double loop ureteral stents: Total enlargement of the stent diameter, for ureteral stenosis in adult and pediatric (children and adolescents) patients.
The Pyelostent® and Stenostent® Silicone double loop ureteral sterts may remain implanted for up to 12 months.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.4
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{3}------------------------------------------------
#### ર. 510(k) SUMMARY
| I. SUBMITTER<br>510(K) Owner's Name: | Coloplast A/S |
|--------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | Holtedam 1<br>3050 Humlebaek, Denmark |
| Phone/Fax/Email: | Phone: 612-422-3206<br>Email: usggh@coloplast.com |
| Name of Contact Person: | Gayatri Ghadge<br>Principal Regulatory Affairs Specialist |
| Address/Contact: | 1601 West River Road North<br>Minneapolis, MN 55411 |
| Date Prepared: | June 7, 2022 |
| II. DEVICE<br>Device Name: | Imajin® Silicone double loop ureteral stent kits<br>ImaJin® Pyelostent® Silicone double loop ureteral stent kits<br>ImaJin® Stenostent® Silicone double loop ureteral stent kits |
| Common or Usual Name: | Double Loop Ureteral Stents |
| Classification Name:<br>Classification Number:<br>Product Code:<br>Regulatory Class: | Stent, Ureteral<br>21 CFR section 876.4620<br>FAD<br>2 |
#### III. PREDICATE AND REFERENCE DEVICES
| Predicate Device: | Porges™ Silicone Double Loop Ureteral Stents, K013921 |
|-------------------|-----------------------------------------------------------|
| Reference Device: | Silicone Hydro-coated double loop ureteral stent, K180469 |
The predicate and reference devices have not been subject to a design-related recall.
#### IV. DEVICE DESCRIPTION
The ImaJin Silicone double loop ureteral stent kit product family was originally cleared as Porges™ Silicone Double Loop Ureteral Stents in 510(k) K013921. The ImaJin Silicone double loop ureteral stent family consists of three single use, autostatic stents intended to drain the upper urinary tract and allow healing in case of ureteral obstacles or damage to the ureter. The Pyelostent and Stenostent Silicone double loop ureteral stents have partial and complete reinforcement along the straight catheter section respectively to add crush resistance. The devices are provided with a
{4}------------------------------------------------
pusher and in some cases a guidewire. The Silicone double loop ureteral stents are ethylene oxide sterilized, single use, implantable devices.
The current submission is to update the indication for use statement, revise the Instructions for Use, add information on MRI compatibility, modify drainage hole size in some models, modify packaging, and modify the device performance specifications.
#### V. INDICATIONS FOR USE
Silicone double loop ureteral stents:
The Silicone double loop ureteral stents are intended for adult and pediatric (children and adolescents) patients for drainage of the upper urinary tract over fistulas or ureteral obstacles and/or for healing of the ureter. These stents may remain implanted for up to 12 months.
Pyelostent and Stenostent Silicone double loop ureteral stents: Drainage of the upper urinary tract and/or ureter healing during management of ureteral stenosis:
For Pyelostent® Silicone double loop ureteral stents: Partial enlargement of the stent diameter, for localized stenosis of the ureteropelvic junction in adult and pediatric (adolescents) patients.
For Stenostent® Silicone double loop ureteral stents: Total enlargement of the stent diameter, for ureteral stenosis in adult and pediatric (children and adolescents) patients.
The Pyelostent® and Stenostent® Silicone double loop ureteral stents may remain implanted for up to 12 months.
### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
Imajin® Silicone double loop ureteral stent kits, ImaJin® Pyelostent® silicone double loop ureteral stent kits, and ImaJin® Stenostent® silicone double loop ureteral stent kits are currently cleared for the US market under the predicate 510(k) Porges™ Silicone Double Loop Ureteral Stents, K013921. The proposed changes to the previously cleared devices do not alter the performance or intended use of the products. The Imalin Silicone double loop ureteral stent family has equivalent design, function, packaging, sterilization process, material technology, indications for use and operating principles as the predicate device. The differences in indications for use, contraindications, MRI compatibility, packaging, and performance specifications do not significantly alter the device compared to the primary predicate. The modified performance specifications are the same as for the reference device, Silicone Hydro-coated double loop ureteral stent, K180469.
{5}------------------------------------------------
# K213185 Page 3 of 3
#### VII. PERFORMANCE DATA
The following performance data was provided for the proposed changes and to support the substantial equivalence determination.
## Mechanical/Performance Testing
- Visual Evaluation ●
- Dimensional Testing
- o Shaft Diameter and Length
- Guidewire and Pusher Compatibility ●
- Liquid Compatibility
- Flowrate per EN 1618:1997
- Glide on Guidewire (Friction)
- Tensile Strength Testing per EN 1618:1997
- Elongation
- Loop Strength
- Radiopacity
- Shelf Life/Expiration date
- MRI Safety Testing ●
- Package Integrity Testing ●
- Transportation Testing ●
## Sterilization
The Imajin Silicone double loop ureteral stent kits. ImaJin Pyelostent Silicone double loop ureteral stent kits, and ImaJin Stenostent Silicone double loop ureteral stent kits are sterilized using ethylene oxide in a validated cycle in conformance with ISO 11135-2014, demonstrating a sterility assurance level (SAL) of 10-6.
No animal studies or clinical testing were provided to support substantial equivalence between the subject and predicate devices.
## VIII. CONCLUSIONS
Based upon the intended use, technological characteristics, and non-clinical performance data provided, the Imajin Silicone double loop ureteral stent kits, ImaJin Pyelostent Silicone double loop ureteral stent kits, and ImaJin Stenostent Silicone double loop ureteral stent kits have been demonstrated to be substantially equivalent to the predicate, Porges™M Silicone Double Loop Ureteral Stents, cleared under premarket notification number K013921. The changes do not raise new questions of safety or effectiveness and the subject devices are as safe and effective as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.