K213166 · Kaneka Medical America, LLC · QEZ · Oct 26, 2021 · Cardiovascular
Device Facts
Record ID
K213166
Device Name
Thrombuster II Aspiration Catheter
Applicant
Kaneka Medical America, LLC
Product Code
QEZ · Cardiovascular
Decision Date
Oct 26, 2021
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.5150
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Thrombuster II Aspiration Catheter is indicated for the removal of fresh, soft emboli and thrombi from vessels in the coronary and peripheral system. The Thrombuster II is not intended for use in the cerebral vasculature.
Device Story
Rapid exchange, dual-lumen aspiration catheter for embolectomy/thrombectomy. Input: manual negative pressure generated by an external lock syringe. Operation: catheter inserted into coronary or peripheral vessels over a 0.014-inch guidewire; negative pressure applied via syringe to aspirate thrombi through the distal tip. Output: physical removal of emboli/thrombi. Used in clinical settings (PCI/peripheral intervention) by physicians. Benefits: revascularization via thrombus removal. Includes core wire for kink resistance, syringe kit, and accessories (flushing needle, filter).
Clinical Evidence
Bench testing only. No clinical data provided. Performance verified via trackability, dimensional verification, and aspiration performance testing. Biocompatibility relies on predicate device data as materials remain unchanged.
Technological Characteristics
Rapid-exchange dual-lumen catheter; 1400 mm working length; compatible with 0.014" guidewires and 6/7 Fr guiding catheters. Features hydrophilic coating and radiopaque ring marker. Aspiration via external syringe-generated negative pressure. Non-powered, mechanical device.
Indications for Use
Indicated for removal of fresh, soft emboli and thrombi from coronary and peripheral vasculature. Contraindicated for use in cerebral vasculature.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
October 26, 2021
Kaneka Medical America LLC % Takeaki Miyata Manager, Regulatory Affairs Team Kaneka Corporation 1-12-32 Akasaka Minato-ku, Tokyo 107-6028 Japan
Re: K213166
Trade/Device Name: Thrombuster II Aspiration Catheter Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: QEZ Dated: September 27, 2021 Received: September 28, 2021
Dear Takeaki Miyata:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K213166
Device Name Thrombuster II Aspiration Catheter
Indications for Use (Describe)
The Thrombuster II Aspiration Catheter is indicated for the removal and thrombi from vessels in the coronary and peripheral system. The Thrombuster II is not in the cerebral vasculature.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
#### K213166
# Thrombuster II Aspiration Catheter
#### 510(k) Submitter
Kaneka Medical America LLC 623 Fifth Avenue, New York NY, 10022, The United States Contact Person: Audra Bogucki Telephone: 917-628-9870 Email: Audra.boguchi@kaneka.com
#### Official Correspondent
Takeaki Miyata Manager, Regulatory Affairs Team KANEKA CORPORATION 1-12-32, Akasaka, Minato-ku, Tokyo 107-6028, Japan Phone: +81-3-5574-8023 Email: Takeaki.Miyata@kaneka.co.jp
#### Date Prepared: October 21, 2021
## Subject Device Name:
| Trade Name | Thrombuster II Aspiration Catheter |
|----------------------|----------------------------------------------------------------|
| Common or usual name | Aspiration Thrombectomy Catheter |
| Classification name | Embolectomy Catheter [21 CFR 870.5150; product code is<br>QEZ] |
| Class | II |
| Classification Panel | Cardiovascular (74) |
## Predicate Device:
- Xpress-Way RX [K101839 (Kaneka Corporation)]
{4}------------------------------------------------
The predicate device have not been subject to a design-related recall.
## Device Description:
Thrombuster II Aspiration Catheter [Thrombuster II] is a rapid exchange type of dual lumen catheter for embolectomy. The Thrombuster II has been commercially available in EU region, Asia including Japan and China, Oceania and Middle east countries.
The main body consists of a distal shaft, proximal shaft, ring marker, hub, and strain relief. The catheter working length is 1400 mm. This product includes a core wire to resist kink during delivery in clinical usage, syringe kit (extension tube, stopcock, and lock syringe) for connecting the catheter main body and aspiration of fresh, soft emboli and thrombi, and accessories (flushing needle and filter). After insertion of the catheter into a patient body, this product aspirates and removes emboli or thrombi from a target lesion by transmitting negative pressure generated by connected lock syringe to the distal end of the catheter.
The maximum compatible diameter of a guidewire used together with Thrombuster II in a procedure is 0.014 inches. Additionally, guiding catheters with a diameter of 6 or 7 Fr have been deemed to be compatible with Thrombuster II.
#### Indications for Use
The Thrombuster II Aspiration Catheter is indicated for the removal of fresh, soft emboli and thrombi from vessels in the coronary and peripheral system. The Thrombuster II is not intended for use in the cerebral vasculature.
#### Comparison of Indications for Use to Predicate Device
The Thrombuster II has the same intended use (removal of thrombus form the peripheral and/or coronary vasculature through aspiration) as the predicate device. The indications for use of the Thrombsuter II is also identical to that of the predicate device.
## Comparison of Technological Characteristics to Predicate Device
Embolectomy or thrombectomy by aspiration is the technological principle for both Thrombuster II and the predicate device. These devices are used in Percutaneous Coronary Intervention (PCI) or peripheral intervention for the purpose of revascularization.
The subject device and predicate device are based on the following same technological elements:
- . Fundamental catheter design - rapid exchange type of dual lumen catheter
- . Catheter components - proximal shaft, distal shaft, wire lumen tube with radiopaque ring marker, hub, and hydrophilic coating
- . Product configuration - catheter main body, core wire, syringe kit, and accessories
{5}------------------------------------------------
- . Principle of operation - aspiration of emboli and thrombi from catheter tip due to negative pressure created by connected syringe
- . Dimensional specification - catheter working length, outer diameters of distal shaft and proximal shaft
- . Concomitant devices - a guidewire of 0.014 inches and a guiding catheter of 6Fr or 7Fr
There are following minor technological differences between Thrombuster II and the predicate device:
- . Aspiration opening shape at the distal end of catheter
- . Number of lock syringes
- . Filter shape
These differences do not affect common intended use, indications for use, fundamental scientific technology and performance specification in these devices.
## Performance Testing
To reduce risks identified in our risk analysis of the differences, and to demonstrate substantial equivalence of Thrombuster II to the predicate device, following tests were conducted using well-established methods:
- . Trackability test
- . Dimensional verification
- . Aspiration performance test
The results from these tests demonstrate that the technological characteristics and performance of the Thrombuster II are substantially equivalent to the predicate device.
## Biocompatibility:
The modifications do not affect the nature of body contact and do not change raw materials for the predicate device. Therefore, in the design control activities for the modifications, the original testing on the predicate device applies to the Thrombuster II.
## Conclusions:
The Thrombuster II met all the predetermined acceptance criteria of design verification and validation as specified by applicable standards, guidance, test protocols and customer inputs. The Thrombuster II is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.