# Rapid (K213165)

_Ischemaview, Inc. · LLZ · Feb 8, 2022 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K213165

## Device Facts

- **Applicant:** Ischemaview, Inc.
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Feb 8, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** AI/ML, Software as a Medical Device, Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K213165 · Feb 8, 2022 | Rapid | Ischemaview, Inc. | Routine clinical imaging data (axial image slices) from multiple clinical sites | Retrospective clinical imaging data was used to train and validate an AI/ML module designed to detect motion artifacts in non-contrast CT (NCCT) images. | AI/ML; NCCT; Motion Artifacts; Retrospective clinical data |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| NCCT Motion Artifact AI/ML Module Training and Validation; Retrospective analysis of clinical imaging data | Patients undergoing NCCT scans at multiple clinical sites; Sample Size: Training: 23,066 slices; Validation: 5,906 slices; Test: 3,262 slices; Independent Validation: 619 cases; Number of Sites: Multiple (Siemens, GE, Toshiba, Philips, and Neurologica) | Not applicable for this study | Sensitivity, Specificity, and AUC for motion artifact detection |

## Indications for Use

Rapid is an image processing software package to be used by trained professionals, including but not limited to physicians and medical technicians. The software runs on a standard off-the-shelf computer or a virtual platform, such as VMware, and can be used to perform image viewing, processing and analysis of images. Data and images are acquired through DICOM compliant imaging devices. Rapid provides both viewing and analysis capabilities for functional and dynamic imaging datasets acquired with CT Perfusion (CTP), CT Angiography (CTA), and MRI including a Diffusion Weighted MRI (DWI) Module and a Dynamic Analysis Module (dynamic contrast-enhanced imaging data for MRI and CT). The CT analysis includes NCCT maps showing areas of hypodense and hyperdense tissue. The DWI Module is used to visualize local water diffusion properties from the analysis of diffusion - weighted MRI data. The Dynamic Analysis Module is used for visualization and analysis of dynamic imaging data, showing properties of changes in contrast over time. This functionality includes calculation of parameters related to tissue flow (perfusion) and tissue blood volume. Rapid CT-Perfusion and Rapid MR-Perfusion can be used by physicians to aid in the selection of acute stroke patients (with known occlusion of the intracranial internal carotid artery or proximal middle cerebral artery). Instructions for the use of contrast agents for this indication can be found in Appendix A of the User's Manual. Additional information for safe and effective drug use is available in the product-specific iodinated CT and gadolinium-based MR contrast drug labeling. In addition to the Rapid imaging criteria, patients must meet the clinical requirements for thrombectomy, as assessed by the physician, and have none of the following contraindications or exclusions: Bolus Quality: absent or inadequate bolus. Patient Motion: excessive motion leading to artifacts that make the scan technically inadequate. Presence of hemorrhage.

## Device Story

Rapid is a DICOM-compliant image processing software (SaMD) for CT and MRI data; operates on off-the-shelf hardware or virtual platforms (VMware). It automates processing tasks for stroke workup; includes modules for CTP, CTA, DWI, and dynamic contrast-enhanced analysis. Features an AI/ML-based NCCT motion artifact filter that provides textual overlays on images. Used by physicians/technicians in clinical settings to visualize tissue perfusion, diffusion, and density (hypodense/hyperdense tissue). Outputs include quantified parameters (CBF, CBV, MTT, Tmax) and thresholded maps to aid in thrombectomy patient selection. Integrates into hospital networks via DICOM; supports secure VPN/SSH connections. Benefits include streamlined, consistent image analysis, reducing clinician fatigue and aiding rapid clinical decision-making for acute stroke.

## Clinical Evidence

Bench testing only. NCCT Motion Artifact AI/ML module validated on 23,066 training slices and 3,262 test images. Independent validation (N=619) showed AUC=0.96 (95% CI: 0.94-0.97), Sensitivity=0.91 (95% CI: 0.83-0.95), and Specificity=0.86 (95% CI: 0.83-0.89). Performance validation used phantoms and case data to confirm accurate representation of processing parameters.

## Technological Characteristics

DICOM-compliant PACS software; runs on standard off-the-shelf Linux-based servers or virtualized environments. Connectivity via VPN (RFC2401/4301) or SSH. Materials: N/A (software). Standards: EN ISO 14971:2019, IEC 62304:2016, IEC 62366:2015, NEMA PS 3.1-3.20. Algorithm: Mixed traditional rule-based processing and AI/ML-based motion artifact filter.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- qER-Quant ([K211222](/device/K211222.md))
- iSchemaView Rapid ([K182130](/device/K182130.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

March 15, 2022

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

iSchemaView Inc. James Rosa SVP Regulatory and Quality 433 Park Point Drive, Suite 220 Golden, Colorado 80401

Re: K213165

Trade/Device Name: Rapid Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: QIH, LLZ

Dear James Rosa:

The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated 02/08/2022. Specifically, FDA is updating this SE Letter to change the company name from "ischema View Inc." to "iSchemaView Inc." as an administrative correction.

Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact CAPT Patrick Hintz, MSIH, CIH, USPHS, OHT7: Office of In Vitro Diagnostics and Radiological Health, 301-796-6927, Patrick.Hintz(@fda.hhs.gov.

Sincerely,

Michael D. O'Hara

For

CAPT Patrick Hintz, MSIH, CIH, USPHS Chief Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".

iSchema View Inc. % James Rosa SVP Regulatory and Quality 433 Park Point Drive, Suite 220 GOLDEN CO 80401

Re: K213165

February 8, 2022

Trade/Device Name: Rapid Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: QIH, LLZ Dated: January 4, 2022 Received: January 5, 2022

Dear James Rosa:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

{2}------------------------------------------------

https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica Lamb, Ph.D. Assistant Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Indications for Use

510(k) Number (if known)

#### K213165

Device Name Rapid

#### Indications for Use (Describe)

Rapid is an image processing software package to be used by trained professionals, including but not limited to physicians and medical technicians. The software runs on a standard off-the-shelf computer or as VMware, and can be used to perform image viewing, processing and analysis of images. Data and images are acquired through DICOM compliant imaging devices.

Rapid provides both viewing and analysis capabilities for functional and dynamic imaging datasets acquired with CT, CT Perfusion (CTP), CT Angiography (CTA), and MRI including a Diffusion Weighted MRI (DWI) Module and a Dynamic Analysis Module (dynamic contrast-enhanced imaging data for MRI and CT).

The CT analysis includes NCCT maps showing areas of hypodense and hyperdense tissue.

The DWI Module is used to visualize local water diffusion properties from the analysis of diffusion - weighted MRI data.

The Dynamic Analysis Module is used for visualization and analysis of dynamic imaging data, showing properties of changes in contrast over time. This functionality includes calculation of parameters related to tissue flow (perfusion) and tissue blood volume.

Rapid CT-Perfusion and Rapid MR-Perfusion can be used by physicians to aid in the selection of acute patients (with known ocuusion of the intracranial internal carotid artery or proximal middle cerebral artery)

Instructions for the use of contrast agents for this in Appendix A of the User's Manual. Additional information for safe and effective drug use is available in the product-specific iodinated CT and gadolinium-based MR contrast drug labeling.

In addition to the Rapid imaging criteria, patients must requirements for thrombectomy, as assessed by the physician, and have none of the following contraindications or exclusions:

- Bolus Quality: absent or inadequate bolus.
- Patient Motion: excessive motion leading to artifacts that make the scan technically inadequate .
- . Presence of hemorrhage

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------|---------------------------------------------|
|----------------------------------------------|---------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

iSchemaView - Traditional 510(k) Rapid 510(k) Summary

#### 510(k) Summary

#### iSchemaView, Inc.'s Rapid

This document contains the 510(k) summary for the iSchemaView Rapid. The content of this summary is based on the requirements of 21 CFR Section 807.92(c).

#### Applicant Name and Address:

| Name:             | iSchemaView, Inc.                                                |
|-------------------|------------------------------------------------------------------|
| Address:          | 1120 Washington St., Suite 200<br>Golden, CO 80401               |
| Official Contact: | Jim Rosa<br>Phone: (303) 704-3374<br>Email: rosa@ischemaview.com |

#### Summary Preparation Date: `September 24, 2021

#### Device Name and Classification:

| Trade Name:           | iSchemaView Rapid                              |
|-----------------------|------------------------------------------------|
| Common Name:          | PACS – Picture Archiving Communications System |
| Classification:       | II                                             |
| Product Code:         | Primary: QIH, Secondary: LLZ                   |
| Regulation No:        | 21 C.F.R. §892.2050                            |
| Classification Panel: | Radiology Devices                              |

#### Predicate Devices:

The iSchemaView Rapid is claimed to be substantially equivalent to the following legally marketed predicate devices:

Primary: qER-Quant (K211222) Secondary: iSchemaView Rapid (K182130)

## Previous Related FDA Submission:

iSchemaView Rapid (K121447) iSchemaView Rapid (K172477) iSchemaView Rapid (K182130)

#### Device Description:

Rapid is a software package that provides for the visualization and study of changes in tissue using digital images captured by diagnostic imaging systems including CT (Computed

{5}------------------------------------------------

#### 510(k) Summary

Tomography) and MRI (Magnetic Image Resonance), as an aid to physician diagnosis. Rapid can be installed on a customer's Server or it can be accessed online as a virtual system. It provides viewing, quantification, analysis and reporting capabilities.

Rapid works with the following types of (DICOM compliant) medical image data:

- CT (Computed Tomography) ●
- MRI(Magnetic Image Resonance)

Rapid acquires (DICOM compliant) medical image data from the following sources:

- . DICOM file
- DICOM CD-R
- Network using DICOM protocol ●

Rapid provides tools for performing the following types of analysis:

- selection of acute stroke patients for endovascular thrombectomy ●
- volumetry of thresholded maps
- time intensity plots for dynamic time courses ●
- measurement of mismatch between labeled volumes on co-registered image ● volumes
- large vessel density

Rapid is a Software as a Medical Device (SaMD) consisting of one or more Rapid Servers (dedicated or virtual). The Rapid Server is an image processing engine that connects to a hospital LAN, or inside the Hospital Firewall. It can be a dedicated Rapid Server or a VM Rapid appliance, which is a virtualized Rapid Server that runs on a dedicated server.

Rapid is designed to streamline medical image processing tasks that are time consuming and fatiguing in routine patient workup. Once Rapid is installed it operates with minimal user interaction. Once the CT (NCCT. CT, CTA) or MR (MR, MRA) data are acquired, the CT or MRI console operator selects Rapid as the target for the DICOM images, and then the operator selects which study/series data to be sent to Rapid. Based on the type of incoming DICOM data, Rapid will identify the data set scanning modality and determine the suitable processing module. The Rapid platform is a central control unit which coordinates the execution image processing modules which support various analysis methods used in clinical practice today:

- Rapid CTP/MRP, DWI, Dynamic Analysis (Original: K121447, Updated with K172477; ● and K182130):
- Rapid CTA (K172477); ●
- Rapid ASPECTS(K190395); ●
- Rapid ICH (K193087);
- . Rapid LVO (K200941);

The iSchemaView Server is a dedicated server that provides a central repository for Rapid data. All iSchemaView Server data is stored on encrypted hard disks. It also provides a user interface for accessing Rapid data. It connects to a firewalled Data Center Network and

{6}------------------------------------------------

510(k) Summarv

has its own firewall for additional cyber/data security. The iSchemaView Server connects to one or more Rapid Servers via WAN. Available types of connection include VPN (Virtual Private Network - RFC2401 and RFC4301 Standards) Tunnel and SSH (Secure Shell).

## Indications for Use:

Rapid is an image processing software package to be used by trained professionals, including but not limited to physicians and medical technicians. The software runs on a standard off-the-shelf computer or a virtual platform, such as VMware, and can be used to perform image viewing, processing and analysis of images. Data and images are acquired through DICOM compliant imaging devices.

Rapid provides both viewing and analysis capabilities for functional and dynamic imaging datasets acquired with CT Perfusion (CTP), CT Angiography (CTA), and MRI including a Diffusion Weighted MRI (DWI) Module and a Dynamic Analysis Module (dynamic contrast-enhanced imaging data for MRI and CT).

The CT analysis includes NCCT maps showing areas of hypodense and hyperdense tissue.

The DWI Module is used to visualize local water diffusion properties from the analysis of diffusion - weighted MRI data.

The Dynamic Analysis Module is used for visualization and analysis of dynamic imaging data, showing properties of changes in contrast over time. This functionality includes calculation of parameters related to tissue flow (perfusion) and tissue blood volume.

Rapid CT-Perfusion and Rapid MR-Perfusion can be used by physicians to aid in the selection of acute stroke patients (with known occlusion of the intracranial internal carotid artery or proximal middle cerebral artery)

Instructions for the use of contrast agents for this indication can be found in Appendix A of the User's Manual. Additional information for safe and effective drug use is available in the product-specific iodinated CT and gadolinium-based MR contrast drug labeling.

In addition to the Rapid imaging criteria, patients must meet the clinical requirements for thrombectomy, as assessed by the physician, and have none of the following contraindications or exclusions:

- Bolus Quality: absent or inadequate bolus. ●
- . Patient Motion: excessive motion leading to artifacts that make the scan technically inadequate
- Presence of hemorrhage ●

{7}------------------------------------------------

510(k) Summary

## Technological Characteristics:

Rapid performs the following functions:

- processes DICOM images from multiple sources to provide visualization of changes . of tissue perfusion, diffusion and change.
- receives DICOM images from external DICOM image providers (modalities (CT/MRI . Scanners), PACS and Workstations) and sends DICOM images to external image consumers.
- . processes requests, statuses and results, and references therein, which are stored in a searchable database
- processing status is available through a web browser using HTTP, HTML and PHP. ●
- . for NCCT images, a motion filter (AI/ML) is employed which provides a textual overlay on an image suspected of having motion artifacts, without distorting the original image
- can send summary results to the user over email. For this, Rapid generally connects to ● the infrastructure of the medical partner (e.g., the hospital). In particular, Rapid uses a SMTP protocol with security extensions to provide secure communications.

Rapid is available in the following configurations:

- Standard Rapid, which is installed directly on a customer's Linux-based server and . integrated with medical image processing software such as commercial PACS.
- . Virtual Rapid, wherein the user accesses Rapid online and uses it to process DICOM images otherwise available on his/her computer.

Rapid is a DICOM-compliant PACS software that provides comprehensive functionality to transfer, process, and display modality specific imaging data. Rapid runs on standard "offthe- shelf computer and networking hardware. Rapid is entirely independent from CT, MRI, or independent PACS platforms. It supports secure VPN (Virtual Private Network) networking or encapsulated Secure Shell (SSH), and seamlessly integrates into an existing radiological data network.

NCCT Motion Artifact AI/ML Module Performance:

Training was performed on 23066 (Pos:1021, Neg:12877) axial image slices from multiple sites, training validation included 5906 (pos: 422.neg: 5484) with a test set of 3262 (pos:2914, neg:348) images. Slice thickness ranged from 1.2-6.0 mm; The optimal performance for the final engineering solution showing an optimal AUC = 0.95, Sensitivity=0.95, Specificity=0.96. Samples were obtained from Siemens, GE, Toshiba, Philips, and Neurologica.

For final independent validation, an N=619 was used with ground truth established by 3 experienced truthers. Testing was performed indevendent of the development group to avoid bias. The primary endpoint was passed (weak artifact = 0) with Sensitivity = 0.91(0.83,0.95) and Specificity = 0.86(0.83,0/89) with AUC = 0.96(0.94,0.97). The cases were split Male:55%, Female 45% with an age range or 32-88 years. The samples were primarily from Siemens with GE mixed.

{8}------------------------------------------------

510(k) Summary

#### Clinical Characteristics:

The primary users of Rapid software are medical imaging professionals who analyze tissue using CT or MRI images. The images generated by Rapid provide additional diagnostic information, which is derived from the temporal/diffusion/density features of the native CT or MRI images.

Rapid CT Perfusion and Rapid MRI can be used by physicians to select acute stroke patients for endovascular thrombectomy. The recommended selection criteria are listed in the table below. Patients must meet the clinical requirements for thrombectomy as assessed by the physician.

## Performance Standards:

Rapid has been developed in conformance with the following standards, as applicable:

| EN ISO 14971:2019  | Application of Risk Management to Medical Devices       |
|--------------------|---------------------------------------------------------|
| IEC 62304:2016     | Medical device software – Software lifecycle processes  |
| IEC 62366:2015     | Application of Usability Engineering to Medical Devices |
| NEMA PS 3.1 - 3.20 | Digital Imaging and Communications in Medicine (DICOM)  |

## Performance Data:

Rapid complies with DICOM (Digital Imaging and Communications in Medicine) - Developed by the American College of Radiology and the National Electrical Manufacturers Association. NEMA PS 3.1 - 3.20.

Additionally, iSchemaView conducted extensive performance validation testing and software verification and validation testing of the Rapid system. This performance validation testing demonstrated that the Rapid system provides accurate representation of key processing parameters under a range of clinically relevant parameters and perturbations associated with the intended use of the software. Software performance, validation and verification testing demonstrated that the Rapid system met all design requirements and specifications.

#### Prescriptive Statement:

Caution: Federal law restricts this device to sale by or on the order of a physician.

## Safety & Effectiveness:

Rapid has been designed, verified and validated in compliance with 21 CFR, Part 820.30 requirements. The device has been designed to meet the requirements associated with EN ISO 14971:2019 (risk management). The Rapid System performance has been validated through the use of phantoms and case data.

## Substantial Equivalence:

Rapid is as safe and effective as the previously cleared Rapid (K182130) with an extension of two parameters similar to the hyperdensity defined in qER-Quant (K211222). Rapid has the same intended use and similar indications, technological characteristics and principles of operation as its predicate devices. Rapid raises no new issues of safety or effectiveness compared to qER-Quant (K211222) or Rapid (K182130), as demonstrated by the testing

{9}------------------------------------------------

510(k) Summary

conducted with Rapid that confirms the software reliably processes and supports analysis of CT and MRI medical images for tissue evaluation. Thus, the Rapid software is substantially equivalent. The claims have been expanded to include the use of Rapid to show areas of hypodensity and hyperdensity and NCCT Motion Suspicion (AI/ML).

{10}------------------------------------------------

# 510(k) Summary

Substantial Equivalence Discussion:

| Parameter                               | Rapid (K182130) – Secondary                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | qER-Quant (K211222) - Primary                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Rapid                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|-----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code                            | LLZ                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | QIH                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | QIH, LLZ                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Regulation                              | 21 CFR §892.2050                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 21 CFR §892.2050                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 21 CFR §892.2050                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Intended Use/ Indications<br>for<br>Use | iSchemaView's Rapid is an image<br>processing software package to be used<br>by trained professionals, including but<br>not limited to physicians and medical<br>technicians. The software runs on a<br>standard off-the-shelf computer or a<br>virtual platform, such as VMware, and<br>can be used to perform image viewing,<br>processing and analysis of images.<br>Data and images are acquired through<br>DICOM compliant imaging devices.<br><br>The iSchemaView Rapid provides both<br>viewing and analysis capabilities for<br>functional and dynamic imaging datasets<br>acquired with CT Perfusion, CT<br>Angiography, and MRI including a<br>Diffusion Weighted MRI (DWI) Module<br>and a Dynamic Analysis Module<br>(dynamic contrast-enhanced imaging<br>data for MRI and CT).<br><br>The DWI Module is used to visualize<br>local water diffusion properties from<br>the analysis of diffusion - weighted<br>MRI data.<br><br>The Dynamic Analysis Module is used<br>for visualization and analysis of dynamic<br>imaging data, showing properties of<br>changes in contrast over time. This<br>functionality includes calculation of<br>parameters related to tissue flow<br>(perfusion) and tissue blood volume. | The qER-Quant device is intended<br>for automatic labeling, visualization<br>and quantification of segmentable<br>brain structures from a set of Non-<br>Contrast head CT (NCCT) images.<br>The software is intended to automate<br>the current manual process of<br>identifying, labeling and quantifying<br>the volume of segmentable brain<br>structures identified on NCCT<br>images.<br><br>qER-Quant provides volumes from<br>NCCT images acquired at a single<br>time point and provides a table with<br>comparative analysis for two or more<br>images that were acquired on the<br>same scanner with the same image<br>acquisition protocol for the same<br>individual at multiple time points.<br><br>The qER-Quant software is indicated<br>for use in the analysis of the<br>following structures: Intracranial<br>Hyperdensities, Lateral Ventricles<br>and Midline Shift. | Rapid is an image processing software<br>package to be used by trained<br>professionals, including but not limited<br>to physicians and medical technicians.<br>The software runs on a standard off-the-<br>shelf computer or a virtual platform,<br>such as VMware, and can be used to<br>perform image viewing, processing and<br>analysis of images. Data and images are<br>acquired through DICOM compliant<br>imaging devices.<br><br>Rapid provides both viewing and<br>analysis capabilities for functional and<br>dynamic imaging datasets acquired with<br>CT, CT Perfusion (CTP), CT<br>Angiography (CTA), and MRI including<br>a Diffusion Weighted MRI (DWI)<br>Module and a Dynamic Analysis Module<br>(dynamic contrast-enhanced imaging<br>data for MRI and CT).<br><br>The CT analysis includes NCCT maps<br>showing areas of hypodense and<br>hyperdense tissue.<br><br>The DWI Module is used to visualize<br>local water diffusion properties from the<br>analysis of diffusion - weighted MRI<br>data.<br><br>The Dynamic Analysis Module is used<br>for visualization and analysis of dynamic<br>imaging data, showing properties of<br>changes in contrast over time. This<br>functionality includes calculation of |
|                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|                                         | Perfusion can be used by physicians to<br>aid in the selection of acute stroke<br>patients (with known occlusion of the<br>intracranial internal carotid artery or<br>proximal middle cerebral artery)<br><br>Instructions for the use of contrast agents<br>for this indication can be found in<br>Appendix A of the User's Manual.<br>Additional information for safe and<br>effective drug use is available in the<br>product-specific iodinated CT and<br>gadolinium-based MR contrast drug<br>labeling.<br><br>In addition to the Rapid imaging criteria,<br>patients must meet the clinical<br>requirements for thrombectomy, as<br>assessed by the physician, and have none<br>of the following contraindications or<br>exclusions.<br><br>Contraindications/Exclusions:<br>Bolus Quality: absent or<br>inadequate bolus. Patient Motion: excessive motion<br>leading to artifacts that make the<br>scan technically inadequate Presence of Hemorrhage                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | parameters related to tissue flow<br>(perfusion) and tissue blood volume.<br><br>Rapid CT-Perfusion and Rapid MR-<br>Perfusion can be used by physicians to aid in<br>the selection of acute stroke patients (with<br>known occlusion of the intracranial internal<br>carotid artery or proximal middle cerebral<br>artery)<br><br>Instructions for the use of contrast agents for<br>this indication can be found in Appendix A<br>of the User's Manual. Additional<br>information for safe and effective drug use is<br>available in the product-specific iodinated<br>CT and gadolinium-based MR contrast drug<br>labeling.<br><br>In addition to the Rapid imaging criteria,<br>patients must meet the clinical requirements<br>for thrombectomy, as assessed by the<br>physician, and have none of the following<br>contraindications or exclusions.<br><br>Contraindications/Exclusions:<br>Bolus Quality: absent or inadequate<br>bolus. Patient Motion: excessive motion<br>leading to artifacts that make the<br>scan technically inadequate Presence of hemorrhage                                                                                                                                                              |
|                                         | PACS Functionality                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Basic PACS Functions                    | Software package which interfaces to a<br>PACS or allows viewing within the<br>application                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Viewing through user PACS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Computer Platform                       | Standard off-the-shelf Hardware: On-<br>Premise                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Standard off-the-shelf Hardware: On-<br>Premise and Secure Cloud                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Standard off-the-shelf Hardware: On-<br>Premise                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Software                                | Traditional Coding                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | AI/ML                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Mixed Traditional and AI/ML(NCCT<br>Motion Filter)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| DICOM Compliance                        | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Functional Overview                     | Rapid is a software package that<br>provides for the visualization and<br>study of changes of tissue in digital<br>images captured by CT and MRI.<br>Rapid provides viewing and<br>quantification.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Data/Image Types                        | Computed Tomography (CT) via<br>DICOM Format                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
|                                         | Magnetic Image Resonance (MRI) via<br>DICOM Format                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Not supported                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Supported                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                                         | Acquisition and Modalities Features                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| MRI                                     | Diffusion Weighted Image (DWI)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Not supported                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Supported                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                                         | Dynamic Analysis tissue flow<br>(perfusion) and tissue blood volume…

---

**Source:** [https://fda-staging.innolitics.com/device/K213165](https://fda-staging.innolitics.com/device/K213165)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).