Encore System

K213159 · Siesta Medical, Inc. · ORY · Dec 22, 2021 · Dental

Device Facts

Record IDK213159
Device NameEncore System
ApplicantSiesta Medical, Inc.
Product CodeORY · Dental
Decision DateDec 22, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.5570
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Encore System is intended for anterior advancement of the tongue base and hyoid suspension. It is indicated for the treatment of obstructive sleep apnea (OSA) and /or snoring.

Device Story

Encore System is a surgical implant kit for anterior tongue base and hyoid suspension. Components include integrated suture passer with #2-0 braided polyester suture, bone screws, bone screw inserters, suspension line lock tool, threading tool, drill bit, and suspension lines (#2 braided polyester or #2 braided polyethylene). Device is used by surgeons to fix soft tissue of tongue base and hyoid bone to the mandible. Procedure aims to reduce airway obstruction in OSA patients. System is non-electronic; no software or algorithms involved. MRI safety labeling updated to allow scanning under specific 1.5T or 3T conditions; drill bit length increased from 5mm to 10mm.

Clinical Evidence

Bench testing only. Performance testing was conducted to verify MRI safety and compatibility of the implantable metallic components under specified magnetic field conditions.

Technological Characteristics

Surgical kit containing bone screws, suspension lines (#2 braided polyester or polyethylene), suture passer, and associated manual surgical tools. Drill bit length is 10 mm. Device is non-electronic, mechanical, and implantable. MRI safety parameters defined for 1.5T and 3T environments.

Indications for Use

Indicated for patients requiring anterior advancement of the tongue base and hyoid suspension for the treatment of obstructive sleep apnea (OSA) and/or snoring.

Regulatory Classification

Identification

Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.

Special Controls

*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. December 22, 2021 Siesta Medical, Inc. Michael Kolber Vice President 101 Church Street Los Gatos, California 95030 Re: K213159 Trade/Device Name: Encore System Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral Devices For Snoring And Intraoral Devices For Snoring And Obstructive Sleep Apnea Regulatory Class: Class II Product Code: ORY Dated: September 25, 2021 Received: September 28, 2021 Dear Michael Kolber: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K213159 Device Name Encore System Indications for Use (Describe) The Encore System is intended for anterior advancement of the tongue base and hyoid suspension. It is indicated for the treatment of obstructive sleep apnea (OSA) and/or snoring. **Type of Use (Select one or both, as applicable)** X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for Siesta Medical. The word "Siesta" is written in a bold, sans-serif font, with a curved line above it. Below "Siesta" is the word "Medical" in a smaller, lighter font. The logo is simple and modern. #### 510(k) Summary ection 5 | 510(k) Number | K213159 | |----------------------------|-----------------------------------------------------------------| | Submitter Name and Address | | | Name | Siesta Medical, Inc. | | Contact | Michael Kolber | | | Vice President, Regulatory Affairs | | Address | 101 Church Street, Suite 3 | | | Los Gatos, CA 95030 | | Telephone | 408-505-6626 | | Fax | 408-399-7600 | | Date Prepared | September 25, 2021 | | General Device Information | | | Product Name | Encore™ System | | Common Name | Intraoral device for snoring and obstructive sleep apnea | | Classification | 21CFR872.5570 | | | Intraoral devices for snoring and intraoral devices for snoring | | | and obstructive sleep apnea. | | Device Class | Class II | | Product Code | ORY | | Predicate Device Manufacturer | Device Name | 510(k) Number | |-------------------------------|---------------|---------------| | Siesta Medical, Inc. | Encore System | K201238 | Device Description The Encore System is designed for anterior tongue base suspension by fixation of the soft tissue of the tongue base to the mandible bone and hyoid bone suspension to the mandible bone using a bone screw and suspension lines. The Encore System consists of 1) an integrated suture passer pre-loaded with size #2-0 braided polyester suture, 2) three (3) bone screws and two (2) bone screw inserters, 3) a suspension line lock tool, 4) a threading tool, and 5) a drill bit. In addition, the following suspension lines are provided depending on the model number: 1) #2 braided polyester suspension line and 2) #2 braided polyethylene suspension line. ## Intended Use (Indications) The Encore System is intended for anterior advancement of the tongue base and hyoid suspension. It is indicated for the treatment of obstructive sleep apnea (OSA) and /or snoring. ## Comparison to the Predicate Device Labeling for the Encore System has been modified with respect to the MRI statement for the Encore Systems packaged with the #2 braided polyester suspension line and the #2 braided polyethylene suspension line. The labeling has changed as follows: ## From: MRI Safety Information: The ENCORE System has not been evaluated for safety and compatibility in the MR environment. It has not been tested for heating, migration or image artifact in the MR environment. The safety of the ENCORE System in the MR environment is unknown. Scanning a patient who has this device may result in patient injury. {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Siesta Medical. The word "Siesta" is in bold, black font, with a gray wave-like line above it. Below "Siesta" is the word "Medical" in a smaller, gray font. The logo is simple and modern. | To: | | |------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------| | MRI Safety Information:<br>A patient implanted with the Encore Bone Screw may be safely scanned under the following<br>conditions. Failure to follow these conditions may result in injury to the patient. | | | Nominal value(s) of Static<br>Magnetic Field [T] | 1.5 T ro 3 T | | Maximum Spatial Field<br>Gradient<br>[T/m and gauss/cm] | 30 T/m (3000 gauss/cm) | | RF Excitation | Circularly Polarized (CP) | | RF Transmit Coil Type | Whole body transmit coil, head RF transmit-receive coil | | Maximum Whole Body SAR<br>[W/kg] | 4.0 W/kg (First Level Operating Mode) | | Limits on Scan Duration | 4.0 W/kg whole body average SAR for 15 minutes of continuous<br>RF (a sequence or back to back series/scan without breaks) | | MR Image Artifact | The presence of this implant may produce an image artifact of<br>12mm | | If information about a specific parameter is not included, there are no conditions associated with | | that parameter. In addition, the drill bit has been lengthened from 5 mm to 10 mm. Summary of Non-Clinical Testing Performance testing was conducted to confirm that the implantable metallic components of the Encore System do not pose a significant risk to a patient in a clinical MRI environment under the conditions presented in the labeling. Statement of Equivalence The Encore System has the same design, indications for use and the same technological characteristics as the predicate device. Based on this and the data provided in this pre-market notification, the subject device and the predicate device have been shown to be substantially equivalent.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...