BTI Interna Dental Implant System UnicCa - Prosthetic Components
K213106 · B.T.I. Biotechnology Institute, Sl. · NHA · Oct 20, 2022 · Dental
Device Facts
Record ID
K213106
Device Name
BTI Interna Dental Implant System UnicCa - Prosthetic Components
Applicant
B.T.I. Biotechnology Institute, Sl.
Product Code
NHA · Dental
Decision Date
Oct 20, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Indications for Use
The BTI Interna Dental Implant System UnicCa® - Prosthetic Components are intended to function in the mandible or maxilla to support single and multiple-unit temporary or definitive restorations on the BTI Interna® Dental Implant System.
Device Story
BTI Interna Dental Implant System UnicCa® - Prosthetic Components consist of various abutments (UNIT transepithelial, Aesthetic Interfaces, Square Aesthetic, Angled Titanium) and screws. These components connect to BTI Interna implant bodies to support single or multiple-unit dental restorations. Abutments are placed by dental professionals in a clinical setting. Angled Titanium Abutments compensate for up to 15° of implant misalignment. The system uses internal engaging/non-engaging connections. Output is a stable prosthetic foundation for restorations, improving patient mastication. Devices are provided sterile via gamma radiation.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical bench testing, including sterilization validation (ISO 11137-1/2, ISO 17665-1), endotoxin testing (USP <85>), packaging validation (ISO 11607-1/2), shelf-life testing (ASTM F88/F1929), biocompatibility (ISO 10993-1), and surface coating characterization (ASTM F1147, ASTM F1160). MRI safety was evaluated via scientific rationale and literature.
Technological Characteristics
Materials: Unalloyed titanium (ASTM F67) with TiN coating; titanium alloy (ASTM F136) screws with WC/C coating. Connection: Internal engaging/non-engaging. Angulation: 0° to 15°. Sterilization: Gamma radiation. Packaging: Thermoform tray with peel-off lid. Connectivity: None (mechanical device).
Indications for Use
Indicated for patients requiring single or multiple-unit temporary or definitive dental restorations in the mandible or maxilla using the BTI Interna® Dental Implant System.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
BTI Interna Narrow/Plus Dental Implant System UnicCa® (K211952)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the symbol of the Department of Health & Human Services on the left and the FDA acronym with the full name of the agency on the right. The FDA part is in blue, with the acronym in a solid blue square and the full name written in a smaller font next to it.
10/20/2022
B.T.I. Biotechnology Institute, SL. Jose Rivero Qualified Person Leonardo Da Vinci 14, Parque Technologico de Alava Minano, Alava 01510 SPAIN
Re: K213106
Trade/Device Name: BTI Interna Dental Implant System UnicCa® - Prosthetic Components Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: September 19, 2022 Received: September 21, 2022
Dear Jose Rivero:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Andrew Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213106
#### Device Name
BTI Interna Dental Implant System UnicCa - Prosthetic Components
Indications for Use (Describe)
The BTI Interna Dental Implant System UnicCa® - Prosthetic Components are intended to function in the mandible or maxilla to support single and multiple-unit temporary or definitive restorations on the BTI Interna® Dental Implant System.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
| X Prescription Use (Part 21 CFR 801 Subpart D) |
| Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510 (K) SUMMARY - K213106
## I. SUBMITTER
## B.T.I. Biotechnology Institute, SL.
Parque Tecnológico de Álava C/Leonardo Da Vinci 14 01510 Miñano (Alava) Spain
## Establishment registration number: 3004417597
Phone: (+34) 945 297 030 Fax: (+34) 945 297 031
Contact Person: Mr. José Ramón Rivero Date Prepared: October 20, 2022
## II. DEVICE
Trade/Proprietary Name: BTI Interna Dental Implant System UnicCa® -Prosthetic Components.
Common or Usual Name: Endosseous Dental Implant Abutment Regulation number: 21 CFR 872.3630 Classification Name: Endosseous Dental Implant Abutments Regulatory Class: II Product Code: NHA
## III. PREDICATE DEVICE
The primary predicate device is the Multi-unit Abutments for CONELOG® from BioHorizons Implant Systems, Inc, which was subject of K203252 (cleared March 16, 2021).
In addition, the following reference devices have been identified for respective abutments/ prosthetic components described within this premarket notification:
- K053355, BTI Interna Dental Implant System O
- K070533, BTI Endosseous Dental Implant Angled Abutments O
- K211952, BTI Interna Narrow/Plus Dental Implant System UnicCa® O
{4}------------------------------------------------
# IV. DEVICE DESCRIPTION
The purpose of this 510(k) is to obtain marketing clearance for expanding the catalogue of abutments and prosthetic components for the Universal Plus and Wide platform of the BTI Interna® Dental Implant System. The premarket notification includes a set of different abutments and screws: UNIT transepithelial abutments, Aesthetic Interfaces, Square Aesthetic Abutments, Angled Titanium Abutments and MULTI-IM transepithelial abutment screws compatible with the cited previous platforms of the BTI Implant System Unicca® in its internal engaging connection version. The implant to abutment connection of all set of abutments range from 4.1mm to 5.5mm. All subject devices are provided sterile via Gamma Radiation. All abutment configurations are not to be used with compatible BTI Interna implant bodies placed at angle (no angular correction), with the exception of the subject Angled Titanium Abutments.
The subject device Transepithelial abutments are straight abutments designed for single (UNIT) and multiple (MULTI-IM) screw-retained restorations. These transepithelial abutments are provided in a range of gingival height from 1.0mm to 4.0mm. All designs have a prosthetic diameter of 5.5mm.
Aesthetic Interfaces are the compatible attachments used to support definitive restorations over transepithelial abutments. Two models are available: Square Aesthetic Interfaces used in single restorations and Aesthetic Interfaces for multiple restorations. Square Aesthetic Interfaces are characterized by an engaging connection to UNIT transepithelial abutments and by four anti rotation lobes placed in prosthetic parts. Aesthetic Interfaces has a conical design with a non-engaging connection to MULTI-IM transepithelial abutments. Prosthetic diameter of aesthetic interfaces can be 5.5mm or 6.5mm, and prosthetic height, 3.5mm or 4.2mm. Aesthetic Interfaces are attached to the transepithelial abutments with their specific retention screws, MULTI-IM transepithelial abutments screws.
The proposed device Square Aesthetic Abutments are straight abutments design for single and multiple restorations in screwed or cemented restorations via traditional casting techniques. The engaging /non-engaging implant abutment connection has a diameter of 4.1mm or 5.5mm, depending on the compatibility with the implant. Square Aesthetic abutments are available in a range of gingival heights from 0.5 to 3.5mm.
Angled Titanium Abutments are premanufactured abutments directly used into the implant in single and multiple cemented restorations. These devices compensate up to 15° misalignment of implants between implant axis and prosthesis axis. Angled Titanium abutments are available in two gingival heights (2 or 4mm) and are compatible with Wide Implant Platform.
All abutments and interfaces are made of unalloyed conforming to ASTM F67 and are coated with titanium nitride (TiN). MULTI-IM transepithelial screws are manufactured in titanium alloy in accordance with ASTM F136, and the surface of the screw thread is coated with tungsten carbide/carbon and chromium (WC/C).
{5}------------------------------------------------
# V. INDICATIONS FOR USE
The BTI Interna Dental Implant System UnicCa® - Prosthetic Components are intended to function in the mandible or maxilla to support single and multiple-unit temporary or definitive restorations on the BTI Interna® Dental Implant System.
#### VI. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE
BTI Interna Dental Implant System UnicCa®- Prosthetic Components is substantially equivalent in terms of indications for use to the primary predicate device K203252. Slight differences in the language of the subject device Indications for Use Statement (IFUS) and reference devices do not affect the intended use for supporting a prosthesis to restore the patient's mastication functions. Minor differences between the IFUS for the subject devices and the primary predicate device include specific device names and compatible implant lines. Minor differences between the subject device IFUS and that of reference devices are related to design features and compatible implant lines. Furthermore, those reference devices mentioned are identified to overcome operating principle, fundamental design, materials, surface coating applied on abutment surface and screw thread - TiN and WC/C, packaging and sterilization to the predicate device.
Table 1 provides a comparison of the indication for use, device features, and other information to demonstrate that the subject device, is substantially equivalent to the primary predicate device and reference devices.
The proposed Transepithelial abutments are substantially equivalent to the transepithelial abutments cleared under K211952. The subject device and reference device K211952 both include straight abutment designs with engaging/non-engaging features, The subject transepithelial abutments and those cleared under K211952 are also substantially equivalent in terms of manufacturing materials, surface coating applied on the transepithelial sleeve and screw, TiN and DLC coating, respectively, in prosthesis attachment (screw-retained), abutment to implant interface (internal connection), packaging (Thermoform tray with peel top lid) and sterilization (gamma irradiation).
The subject device Aesthetic Interfaces and MULTI-IM transepithelial abutment screws are substantially equivalent to the Aesthetic Interfaces and retention screws for Narrow platform cleared under K211952 in terms of design, materials (CP Ti Grade 4 and Ti-6Al-4V), surface coating (TiN and DLC coating) and manufacturing workflow.
Square Aesthetic Abutments design are substantially equivalent to titanium abutment design cleared under K211952 both include straight abutments. The subject device and those titanium abutments cleared under K211952 are also substantially equivalent in terms of manufacturing material (CP Ti Grade 4), in prosthesis attachment (cemented or screw
{6}------------------------------------------------
retained) and abutment to implant interface (internal connection, enganging/non-enganging design). Both devices are also used in single or multiple restorations via casting techniques.
The subject device Angled Titanium Abutment are substantially equivalent to those Angled Titanium Abutments cleared under BTI K070533, same design and same angulation, and the clearance K211952 supports surface coating. . These Angled Titanium Abutments do not present a new worst-case for fatigue testing as compared to the prior clearances for the same system of implants and abutments, so no new fatigue testing was conducted.
The reference device K053355 is for support dimensional differences in the implant platform compatibility, Universal/Plus and/or Wide platform.
Regarding sterilization, the subject devices (aesthetic interfaces, retention screws, square aesthetic abutments and angled titanium abutments) are supplied sterile by gamma radiation in a thermoformed tray with peel-off top lid. This sterilization method and packaging configuration is identical to the transepithelial abutments cleared in K211952. The additional subject abutments which are now provided sterile do not present a new worstcase for sterilization validation.
# VII. NON-CLINICAL TESTING
A series of non-clinical testing has been presented to demonstrate that the proposed device is substantially equivalent to the predicate devices. Tests performed are as follows:
- Sterilization Validation to a sterility assurance level (SAL) of 106 according to ISO . 11137-1, ISO 11137-2, and ISO 17665-1.
- Endotoxin Testing according to ANSI/AAMI ST72 Bacterial endotoxins- Test ● methods, routine monitoring, and alternatives to batch and methods consistent with USP <85> Bacterial Endotoxins Test and the FDA guidance document Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile, issued on January 21, 2016.
- Packaging Validation according to ISO11607-1 and ISO 11607-2. ●
- Shelf-life according to ASTM F88 and F1929 ●
- Biocompatibility testing in conformance with ISO 10993-1.
- Characterization of TiN and WC/C coating surfaces on BTI devices including ● thickness, roughness, SEM images, abrasion testing, adhesion testing by static tensile test according to ASTM F1147, adhesion testing by dynamic shear bonding strength per ASTM F1160.
- Non-clinical worst-case MRI review was performed to evaluate the Subject device ● components in the MRI environment using scientific rationale and published literature (e.g., Woods, Terry O. , Jana G. Delfino and Sunder Rajan, "Assessment of Magnetically Induced Displacement Force and Torque on Metal Allovs Used in Medical Devices." Journal Testing and Evaluation 49.2 (2019):783-795), based on
{7}------------------------------------------------
#### K213106- BTI Interna Dental Implant System UnicCa®- Prosthetic Components 510(K) Summary
the entire system including all variations (all compatible implant bodies, abutments, and fixation screws) and material composition. Rationale addressed parameters per the FDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment', including magnetically induced displacement force and torque.
# VIII. CLINICAL TESTING
Clinical testing is not required.
# IX. CONCLUSIONS
BTI Interna Implant System UnicCa® - Prosthetic Components has same intended use, has similar technological characteristics, are made of identical or similar materials to those clearances identified as primary predicate and reference devices. The subject device and the references devices cover similar range of dimensions, similar packaging, and similar sterilization method. In conclusion the data of the current submission demonstrate that the subject device is substantially equivalent to the predicate and references devices.
{8}------------------------------------------------
| | Subject Device | Primary Predicate Device | Reference Device | Reference Device | Reference Device | | |
|------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|
| | BTI Interna<br>Dental Implant<br>System UnicCa®-<br>Prosthetic<br>Components. | K203252<br>Multi-unit<br>Abutments for<br>CONELOGⓇ | K053355<br>BTI Interna Dental<br>Implant System | K070533<br>Endosseous dental<br>implant angled<br>abutments | K211952<br>BTI Interna Narrow/Plus<br>Dental Implant System<br>UnicCaⓇ | Substantial<br>Equivalence<br>Discussion | |
| Product<br>Classification | Regulation No.: 21<br>CFR 872.3630<br>Product Code:<br>NHA<br>Device Class II | Regulation No.: 21<br>CFR 873.3630<br>Product Code: NHA.<br>Device Class II | Regulation No.: 21 CFR<br>872.3640<br>Product Code: DZE<br>Device Class II | Regulation No.:<br>21 CFR 872.3630<br>Product Code: NHA<br>Device Class II | Regulation No.: 21 CFR<br>872.3640<br>Product Code: DZE<br>Device Class II | Identical to primary<br>predicate device | |
| Indications for Use | The BTI Interna<br>Dental Implant<br>System UnicCa® -<br>Prosthetic<br>Components are<br>intended to function<br>in the mandible or<br>maxilla to support<br>single and multiple-<br>unit temporary or<br>definitive<br>restorations on the<br>BTI Interna®<br>Dental Implant<br>System. | The BioHorizons<br>Multi-unit<br>Abutments for<br>CONELOG® are<br>intended to function<br>in the mandible or<br>maxilla to<br>support single and<br>multiple-unit<br>temporary or<br>definitive<br>restorations on<br>CONELOG dental<br>implants | Dental implant system<br>comprising endosseous<br>titanium implants and<br>prosthetic elements to be<br>attached to the implants, as<br>well as auxiliary elements<br>for surgical and prosthetic<br>procedures.<br>The intended use of the<br>system is the restoration of<br>missing teeth in partially or<br>fully edentulous patients<br>and/or the fixation of<br>overdentures to restore or<br>enhance the chewing<br>capacity of patients. | Endosseous dental<br>implant angled<br>abutments are<br>premanufactured<br>prosthetic components<br>directly connected to the<br>Endosseous dental<br>implant and are<br>intended for use as aids<br>in prosthetic<br>rehabilitation. They can<br>be used in singled and<br>multi-unit restorations<br>where angled correction<br>is required. | The BTI Dental Implant<br>System UnicCa® for oral<br>implant surgery is to be<br>used for the partial or total<br>replacement of<br>teeth in edentulate patients.<br>Once attached to the bone,<br>the implants act as an<br>anchor for various fixed or<br>removable<br>prosthetic solutions that<br>can be used to improve or<br>restore a patient's<br>mastication function.<br>In the case of 5.5 - 6.5mm<br>long UnicCa® implants<br>should be used in a two-<br>stage surgical procedure.<br>These implants are<br>indicated for delayed<br>loading. These implants are<br>indicated only for straight<br>abutments and to support<br>permanently fixed<br>restorations.<br>In the case of Tiny® 3.0<br>UnicCa® implants: These<br>implants shall be used only<br>to replace maxillary lateral<br>incisors and | Equivalent to<br>primary predicate<br>device. | |
| | Subject Device | Primary Predicate<br>Device | Reference Device | Reference Device | Reference Device | Substantial<br>Equivalence<br>Discussion | |
| | BTI Interna<br>Dental Implant<br>System UnicCa®-<br>Prosthetic<br>Components. | K203252 | K053355 | K070533 | K211952 | | |
| | | Multi-unit<br>Abutments for<br>CONELOGⓇ | BTI Interna Dental<br>Implant System | Endosseous dental<br>implant angled<br>abutments | BTI Interna Narrow/Plus<br>Dental Implant System<br>UnicCaⓇ | | |
| Reason for Predicate/ Reference<br>Device | | Indications for Use | Abutment/implant<br>connection design and<br>implant platform<br>compatibility. | Angled<br>abutments Titanium<br>design, material, and usage. | mandibular lateral and<br>central incisors. Immediate<br>loading is recommended<br>when there is good primary<br>stability and an<br>appropriate occlusal load. | | |
| Designs | Transepithelial<br>Abutments,<br>Interfaces, Angled<br>Titanium<br>abutments,<br>Aesthetic<br>Abutments and<br>Screws. | Multi-Unit Straight<br>abutment, Multi-Unit<br>Angled Abutment,<br>Copings, prosthetic<br>screws and<br>temporary cap. | Healing abutments and<br>caps, Temporary abutments,<br>Titanium and Gold<br>Abutments,<br>Transepithelial<br>abutments, Cylinders, and<br>Screws. | Angled Titanium<br>abutments | Healing abutments,<br>Temporary abutments,<br>Titanium Abutments,<br>Transepithelial<br>abutments, Aesthetic<br>Interfaces and Screws | | |
| Restoration | Single & Multiple | Single unit & Multi-<br>unit | Single & Multiple | Single & Multiple | Single & Multiple | Identical to<br>reference devices. | |
| Compatible Implant Platform<br>(Diameter Ø=mm) | Interna Universal/<br>Plus<br>Ø= 4.1<br>Wide Platform<br>Ø = 5.5 | Ø = 3.3- 4.3 | Interna Universal/<br>Plus<br>Ø = 4.1<br>Wide Platform Ø = 5.5 | Interna Universal/Plus<br>Ø =4.1<br>Externa Universal<br>Ø =4.1<br>Externa Wide<br>Ø = 5.5<br>Tiny<br>Ø=3.7<br>Narrow<br>Ø=4.2 | Interna Narrow<br>Ø = 3.5<br>Interna Universal Plus<br>Ø = 4.1 | Identical to<br>reference device<br>K053355 | |
| Implant to Abutment connection | Internal | Internal | Internal | Internal, External | Internal | Equivalent to<br>primary predicate | |
| | | Subject Device | Primary Predicate<br>Device | Reference Device | Reference Device | Reference Device | |
| | | BTI Interna<br>Dental Implant<br>System UnicCa®-<br>Prosthetic<br>Components. | K203252 | K053355 | K070533 | K211952 | Substantial<br>Equivalence<br>Discussion |
| | | | | | | | device and identical<br>to reference devices. |
| | | Engaging/<br>Non-engaging | Engaging | Engaging/<br>Non-engaging | Non-Engaging | Engaging/<br>Non Engaging | Identical<br>to<br>reference<br>device<br>K053355<br>and<br>equivalent<br>to<br>reference<br>device<br>K211952. |
| | Prosthetic<br>Diameter (mm) | 4.1 to 6.5 | 4.8 | 4.1 to 6.5<br>(Selected<br>components<br>design) | 3.7 to 5.5<br>(Selected<br>components<br>design) | 4.1 and 5.5<br>(Selected<br>components<br>design) | Identical<br>to<br>reference<br>device<br>K053355. |
| | Abutment Angulation | 0°, 15° | 0°, 17°, 30° | 0° | 15° | 0° | Identical<br>to<br>reference<br>device<br>K070533. |
| | Gingival Height | Up to 4mm | 2-4mm | Up to 4mm<br>(Selected<br>components<br>design) | Up to 4mm<br>(Selected<br>components<br>design) | Up to 4mm<br>(Selected<br>components design) | Identical to<br>reference devices. |
| Material | Abutments & Aesthetic<br>Interfaces | CP Titanium<br>Grade 4 | Titanium Alloy<br>ASTM F136 | CP Titanium<br>Grade 4 | CP Titanium<br>Grade 4 | CP Titanium<br>Grade 4 | Identical<br>to<br>reference devices.<br>Same<br>manufacturing<br>material. |
| | Screws | Titanium Alloy<br>Ti-6Al-4V | Titanium Alloy<br>ASTM F136 | Titanium CP<br>Grade 4 | | Titanium Alloy<br>Ti-6Al-4V | Identical<br>to<br>reference device<br>K211952.<br>Same<br>manufacturing<br>material. |
| Surface | Abutments & Aesthetic<br>Interfaces | Titanium Nitride | N.A. | - | - | Titanium Nitride | Identical<br>to<br>reference<br>device<br>K211952. |
| Coating | Screws | DLC Coating | N.A. | - | - | DLC Coating<br>(Tungsten Carbide/ Carbon<br>and Chromium) | Same<br>surface<br>coatings. |
| | Subject Device | Primary Predicate Device | Reference Device | Reference Device | Reference Device | | |
| | BTI Interna<br>Dental Implant<br>System UnicCa®-<br>Prosthetic<br>Components. | K203252 | K053355…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.