K213104 · Auxein Medical Private Limited · MBI · Jul 29, 2022 · Orthopedic
Device Facts
Record ID
K213104
Device Name
Auxilock Titanium Screw-In Suture Anchor
Applicant
Auxein Medical Private Limited
Product Code
MBI · Orthopedic
Decision Date
Jul 29, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Auxilock Titanium Screw-In Suture Anchor is intended to be used with #2 Biobraid suture with white, blue, black colors for use for the reattachment of soft tissue to bone for the following indications: Shoulder: Bankart lesion repairs, SLAP lesion repairs, Acromiolavicular separation repairs, Rotator cuff tear repairs, Capsular shift or Capsutolabral reconstructions, Biceps tenodesis, and Deltoid Repairs, Anterior Shoulder Instability Repair. Foot and Ankle: Hallux valgus reconstruction, Medial or Lateral instability repairs/reconstructions, Achilles tendon repair/reconstructions, Mid-foot reconstruction, Metatarsal ligament/tendon repairs/reconstructions, Bunionectomy. Elbow, Wrist and Hand: Scapholunate ligament reconstruction, Ulnar or radial collateral ligament reconstructions, Lateral epicondylitis repair, Biceps tendon reattachment, Tennis elbow repair. Knee: Extra-capsular repairs: Medial collateral ligament repair. Lateral ligament repair. Posterior oblique ligament. Iliotibial band tenodesis, Patellar tendon repair - Vastus medialis obliquous advancement. Hip: Capsular repair, Acetabular labral repair.
Device Story
Threaded, tapered titanium alloy fastener; used for reattaching soft tissue to bone; self-tapping design; available in 2.8mm, 3.5mm, 5.0mm, and 6.5mm diameters; incorporates cancellous and cortical threads for pull-out strength. Used in orthopedic surgeries (shoulder, knee, hip, foot/ankle, hand/wrist); operated by surgeons; provided preassembled on an inserter with #2 Biobraid suture. Drill guide and bit included for accurate placement; minimizes anchor slippage/breakage. Device provides mechanical fixation to facilitate healing; sold sterile for single use.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included insertion tests, pull-out tensile static tests, and pull-out tensile fatigue tests compared against predicate devices. Biocompatibility evaluated per ISO 10993-3.
Technological Characteristics
Material: Titanium alloy (Ti-6Al-4V, ASTM F136); Suture: UHMWPE (ASTM F2848-17). Threaded, tapered fastener design. Sterilization: Ethylene Oxide (ISO 11135). Single-use, sterile. No software or electronic components.
Indications for Use
Indicated for adult patients requiring soft tissue-to-bone reattachment in shoulder, foot, ankle, elbow, wrist, hand, knee, and hip procedures, including ligament/tendon repairs and reconstructions.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
TwinFix 2.8mm, 3.5mm and BioRaptor 2.9mm Suture Anchors (K053344)
Submission Summary (Full Text)
{0}------------------------------------------------
July 29, 2022
Image /page/0/Picture/1 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left and features a stylized eagle. The FDA logo is on the right and features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Auxein Medical Private Limited Rahul Luthra Director Plot No. 168, 169, 170 Phase-IV, Sector 57, Kundli Industrial Area Sonipat, Haryana 131028 India
## Re: K213104
Trade/Device Name: Auxilock Titanium Screw-In Suture Anchor Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: MBI Dated: June 22, 2022 Received: June 27, 2022
## Dear Rahul Luthra:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For.
Laura C. Rose, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known) K213104
Device Name Auxilock Titanium Screw-In Suture Anchor
#### Indications for Use (Describe)
Auxilock Titanium Screw-In Suture Anchor:
The Auxilock Titanium Screw-In Suture Anchor is intended to be used with #2 Biobraid suture with white, blue, black colors for use for the reattachment of soft tissue to bone for the following indications:
Shoulder:
Bankart lesion repairs, SLAP lesion repairs, Acromiolavicular separation repairs, Rotator cuff tear repairs, Capsular shift or Capsutolabral reconstructions, Biceps tenodesis, and Deltoid Repairs, Anterior Shoulder Instability Repair.
Foot and Ankle:
Hallux valgus reconstruction, Medial or Lateral instability repairs/reconstructions, Achilles tendon repair/reconstructions, Mid-foot reconstruction, Metatarsal ligament/tendon repairs/reconstructions, Bunionectomy.
Elbow, Wrist and Hand:
Scapholunate ligament reconstruction, Unar or radial collateral ligament reconstructions, Lateral epicondylitis repair, Biceps tendon reattachment, Tennis elbow repair.
#### Knee:
Extra-capsular repairs: Medial collateral ligament repair. Lateral ligament repair. Posterior oblique ligament. Iliotibial band tenodesis, Patellar tendon repair - Vastus medialis obliquous advancement.
#### Hip:
Capsular repair, Acetabular labral repair.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Auxein. The logo consists of a blue circle with a stylized "A" inside of it, followed by the word "Auxein" in blue, sans-serif font. A registered trademark symbol is located to the right of the word.
## 510(k) Summary
# General Company Information as required by 807.92 (a)
| Submitter's Name: | Auxein Medical Private Limited |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | Auxein Medical Private Limited<br>Plot No. 168-169-170, Phase-4, Kundli Industrial Area, HSIIDC,<br>Sector-57, Sonepat-131028, Haryana, India |
| Contact Person Name: | Mr. Rahul Luthra |
| Title: | Director |
| Phone Number: | +91 9560557733 |
| Dated: | 29.07.2022 |
#### Person Responsible for Regulatory Compliance
| Name: | Mr. Mohit Kumar |
|-----------|-----------------------|
| Title: | Sr. Research Engineer |
| Email Id: | m.kumar@auxein.com |
| Dated: | 29.07.2022 |
# Proprietary Name:
Auxilock Titanium Screw-In Suture Anchor
#### Common or Usual Name:
Soft Tissue Fixation Device
# Classification Name:
Fastener, fixation, non-degradable, soft tissue
## Product Code: MBI
Device Class: II
Review Panel: Orthopedic
# Regulation Number:
21 CFR 888.3040
| Primary Product<br>Code | Classification Name | Common Name | Regulation Number |
|-------------------------|-----------------------------------------------------|--------------------------------|-------------------|
| MBI | Fastener, fixation, non-<br>degradable, soft tissue | Soft Tissue Fixation<br>Device | 21 CFR 888.3040 |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Auxein. The logo consists of a blue circle with a stylized "A" inside of it, followed by the word "Auxein" in blue, sans-serif font. A small "R" in a circle is located to the upper right of the word "Auxein".
Following are the predicate device 510(k) with which we are declaring substantial equivalence: The following is the range of variants covered with their corresponding predicate devices.
#### Primary Predicate:
| 510(k) Number | K152566 |
|---------------|--------------------------------|
| Applicant | Smith & Nephew |
| Common Name | Suture Anchor |
| Device Name | PEBA Anchor/Suture Combination |
#### Secondary Predicate:
| 510(k) Number | K053344 |
|---------------|---------------------------------------------------------|
| Applicant | Smith & Nephew |
| Common Name | Suture Anchor |
| Device Name | TwinFix 2.8mm, 3.5mm and BioRaptor 2.9mm Suture Anchors |
## Device Description:
The Titanium Screw-In Suture Anchor is a threaded, tapered fastener for use in attachment of soft tissue to bone. The self-tapping AUXILOCK® Titanium screw-in suture anchor is mostly used as a medial row anchor in rotator cuff repair surgery, Bankart or SLAP surgeries. It is recommended for use in small and large-joint repairs without the need for tapping or drilling. AUXILOCK® Titanium screw-in anchors are available in diameter of 2.8mm, 3.5mm, 5.0mm and 6.5mm. The anchor design incorporates a cancellous thread with a tiny core diameter to maximize pull-out strength in cancellous or osteoporotic bone. The AUXILOCK® Titanium screw-in anchor is designed for ultimate mechanical properties (pull-out strength, tensile strength, etc.) and ease-of-use. The anchor design incorporates a cortical thread with a small core diameter to maximize pull-out strength in a cortical or hard bone. The drill guide and drill bit are provided in the instrument set for the accurate placement of the anchor. It in turn minimizes the anchor slippage or breakage during the surgery.
All of the Auxein's Non-Absorbable Suture Anchors are provided sterile, for single use only. All Non-Absorbable Suture Anchors are preassembled onto an inserter, sized appropriately to accommodate the indicated procedures. Auxein's Non-Absorbable Suture Anchors come in various configurations, including: with attached non-absorbable suture(s). In certain configurations, the Non-Absorbable Suture Anchors are packaged with a guide and awl. All configurations of the Non-Absorbable Suture Anchors are identical to the identified predicate devices.
The device is made from a Titanium alloy, Ti-6Al-4V (ASTM F136) as per grade 5.
Note: The #2 Biobraid Suture is not to be used as a stand-alone. It should be used only with the assembled implant.
These implants are sold in sterile conditions (Ethylene Oxide Sterilization). The system is indicated for use in adult patients only. All implants are for single use only.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Auxein. The logo consists of a blue circle with a stylized "A" inside of it, followed by the word "Auxein" in blue, sans-serif font. The logo is simple and modern, and the blue color gives it a sense of trustworthiness and reliability.
#### Indications for Use:
#### Auxilock Titanium Screw-In Suture Anchor:
The Auxilock Titanium Screw-In Suture Anchor is intended to be used with #2 Biobraid suture with white, blue, black colors for use for the reattachment of soft tissue to bone for the following indications:
#### Shoulder
Bankart lesion repairs, SLAP lesion repairs, Acromiolavicular separation repairs, Rotator cuff tear repairs, Capsular shift or Capsutolabral reconstructions, Biceps tenodesis, and Deltoid Repairs, Anterior Shoulder Instability Repair.
#### Foot and Ankle
Hallux valgus reconstruction, Medial or Lateral instability repairs/reconstructions, Achilles tendon repair/reconstructions, Mid-foot reconstruction, Metatarsal ligament/tendon repairs/reconstructions, Bunionectomy.
## Elbow, Wrist and Hand
Scapholunate ligament reconstruction, Ulnar or radial collateral ligament reconstructions, Lateral epicondylitis repair, Biceps tendon reattachment, Tennis elbow repair.
#### Knee
Extra-capsular repairs: Medial collateral ligament repair, Lateral ligament repair, Posterior oblique ligament, Iliotibial band tenodesis, Patellar tendon repair - Vastus medialis obliquous advancement.
#### Hip
Capsular repair, Acetabular labral repair.
## Summary of Technological Characteristics as compared to the predicate devices
A comparison between the Auxilock Titanium Screw-In Suture Anchor and predicate devices has been performed which has resulted in demonstration of similarities in dimensional and performance criteria.
Following is the summary of parameters in which the comparison has been verified:
| S.No. | Characterist<br>ics | Subject Device (Auxein) | Predicate Devices | Remarks |
|-------|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|
| 1. | Product Code | MBI | K152566, PEBA<br>Anchor/Suture Combination,<br>(Primary Predicate)<br><br>K053344, TwinFix 2.8mm,<br>3.5mm Suture Anchors<br>(Secondary Predicate) | Identical |
| | | | | as<br>predicate<br>device. |
| 2. | Regulation<br>Number | 21 CFR 888.3040 | 21 CFR 888.3040 | Identical<br>as<br>predicate<br>device. |
| 3. | Regulatory<br>Class | Class II | Class II | Identical<br>as<br>predicate<br>device. |
| 4. | Indications<br>for use | <b>Auxilock Titanium<br/>Screw-In Suture Anchor</b><br>The Auxilock Titanium Screw-In Suture Anchor is<br>intended to be used with #2 Biobraid suture with white,<br>blue, black colors for use for the reattachment of soft<br>tissue to bone for the following indications:<br><b>Shoulder:</b><br>Bankart lesion repairs, SLAP lesion repairs,<br>Acromiolavicular separation repairs, Rotator<br>cuff tear repairs, Capsular shift or Capsutolabral<br>reconstructions, Biceps tenodesis, and Deltoid<br>Repairs, Anterior Shoulder Instability Repair.<br><b>Foot and Ankle:</b><br>Hallux valgus reconstruction, Medial or<br>Lateral instability repairs/reconstructions,<br>Achilles tendon repair/reconstructions,<br>Mid-foot reconstruction, Metatarsal ligament/tendon repairs/reconstructions | The PeBA Series<br>Anchor/Suture Combination<br>(Primary Predicate) is intended<br>for the fixation of surgical<br>suture material for the<br>following indications:<br><b>Shoulder:</b><br>Bankart lesion repairs, SLAP<br>lesion repairs, Acromio-<br>clavicular separation repairs,<br>Rotator cuff tear repairs,<br>Capsular shift or capsulolabral<br>reconstructions, Biceps<br>tenodesis, Deltoid repairs.<br><b>Foot and Ankle:</b><br>Hallux Valgus repairs, Medial<br>or lateral instability repairs,<br>Achilles tendon<br>repairs/reconstructions, Midfoot<br>reconstructions, Metatarsal<br>ligament/tendon<br>repairs/reconstructions<br><b>Elbow, Wrist, and Hand:</b><br>Scapholunate ligament<br>reconstructions, Ulnar or radial<br>collateral ligament<br>reconstructions, Tennis elbow<br>repair, Biceps tendon<br>reattachment. | Identical<br>as<br>predicate<br>devices. |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image features the logo for Auxein. The logo consists of a stylized letter "A" inside of a circle, followed by the word "Auxein" in a sans-serif font. The logo and text are both in a light blue color.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for Auxein. The logo consists of a blue circle with a stylized "A" inside of it, followed by the word "Auxein" in blue, sans-serif font. A small "R" in a circle is located to the right of the word "Auxein" indicating that the logo is a registered trademark.
| | Bunionectomy. | <b>Knee:</b><br>Extra-capsular repairs, medial<br>collateral ligament, lateral<br>collateral ligament, posterior<br>oblique ligament, Iliotibial band<br>tenodesis, Patellar realignment<br>and tendon, repairs, including<br>vastus medialis, obliquous<br>advancement. |
|--|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | <b>Elbow, Wrist and Hand:</b><br>Scapholunate ligament<br>reconstruction, Ulnar or<br>radial collateral ligament<br>reconstructions, Lateral<br>epicondylitis repair, Biceps<br>tendon reattachment,<br>Tennis elbow repair. | The K053344, TwinFix 2.8mm,<br>3.5mm Suture Anchors<br>(Secondary Predicate) is<br>intended for use for the<br>reattachment of soft tissue to<br>bone for the following<br>indications: |
| | <b>Knee:</b><br>Extra-capsular repairs:<br>Medial collateral ligament<br>repair, Lateral collateral<br>ligament repair, Posterior<br>oblique ligament, Iliotibial<br>band tenodesis, Patellar<br>tendon repair - Vastus<br>medialis obliquous<br>advancement. | <b>Shoulder</b><br>Bankart lesion repairs, SLAP<br>lesion repairs,<br>Acromiolavicular separation<br>repairs, Rotator cuff tear<br>repairs, Capsular shift or<br>Capsutolabral reconstructions,<br>Biceps tenodesis, and Deltoid<br>Repairs, Anterior Shoulder<br>Instability Repair. |
| | <b>Hip:</b><br>Capsular repair, Acetabular<br>labral repair. | <b>Foot and Ankle</b><br>Hallux valgus reconstruction,<br>Medial or Lateral instability<br>repairs/reconstructions,<br>Achilles tendon<br>repair/reconstructions, Mid-foot<br>reconstruction, Metatarsal<br>ligament/tendon<br>repairs/reconstructions,<br>Bunionectomy. |
| | | <b>Elbow, Wrist and Hand</b><br>Scapholunate ligament<br>reconstruction, Ulnar or radial<br>collateral ligament |
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for Auxein. The logo consists of a blue circle with a blue triangle inside of it, followed by the word "Auxein" in blue font. There is a registered trademark symbol to the right of the word.
| | | | reconstructions, Lateral<br>epicondylitis repair, Biceps<br>tendon reattachment, Tennis<br>elbow repair. | |
|-----|-----------------------------|------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|
| | | | Knee<br>Extra-capsular repairs: Medial<br>collateral ligament repair,<br>Lateral collateral ligament<br>repair, Posterior oblique<br>ligament, Iliotibial band<br>tenodesis, Patellar tendon repair<br>- Vastus medialis obliquous<br>advancement. | |
| | | | Hip<br>Capsular repair, Acetabular<br>labral repair. | |
| 5. | Material | Titanium Alloy as per<br>ASTM F136 and<br>UHMWPE Suture as per<br>ASTM F2848-17. | Titanium Alloy as per ASTM<br>F136 and and UHMWPE<br>Suture as per ASTM F2848-17. | Conforms<br>to Identical<br>standard as<br>predicate<br>device. |
| 6. | Performance<br>Standards | Mechanical Testing<br>Protocol AMPL-MT-PRT-<br>002, Rev. 01. was used in<br>the scoped Device. | Mechanical Testing Protocol<br>AMPL-MT-PRT-002, Rev. 01.<br>was used in the predicate<br>Device. | Identical<br>as<br>predicate<br>device. |
| 7. | Sterilization | Ethylene Oxide<br>Sterilization Method was<br>used in the subject device. | Ethylene Oxide Sterilization<br>Method was used in the<br>predicate devices. | Identical in<br>predicate<br>device. |
| 8. | Dimensional<br>Verification | Same dimension as of Predicate device. | | Identical<br>as<br>predicate<br>device. |
| 9. | Shelf-life | 5 Years | 5 Years | Identical<br>as<br>predicate<br>device. |
| 10. | Single<br>Use/Reuse | Single Use | Single Use | Identical<br>as<br>predicate<br>device. |
| 11. | Operating Principle | Secure tendon and ligament reconstructions with increased boney ingrowth at the anchor site with minimal invasive technique. | Secure tendon and ligament reconstructions with increased boney ingrowth at the anchor site with minimal invasive technique. | Identical as predicate device. |
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for Auxein. The logo consists of a blue circle with a stylized "A" inside of it, followed by the word "Auxein" in blue, with a registered trademark symbol next to it. The logo is simple and modern, and the blue color gives it a sense of trust and reliability.
## Discussion on the non-clinical testing performed
Following are the applicable product standards considered for non-clinical standards:
- Material Standards. ●
- . Biocompatibility Standards
- Performance Standards.
- Sterilization, shelf-life and packaging for sterile product.
- . Bacterial Endotoxin
#### Non-Clinical Test Summary:
Bench tests were conducted to verify that the subject device met all design specifications. The test results demonstrated that the subject device complies with the following standards:
#### Material Standards:
The material standards are the essential part to be complied with first, as it is the basis of manufacturing metallic surgical implants.
We have complied with the following material standards:
- ASTM F136/ISO 5832-3: Standard specification for wrought Titanium-6 Vanadium . ELI (Extra low interstitial) Alloy for surgical implant applications.
- · ASTM F2848-17: Standard Specification for Medical-Grade Ultra-High Molecular Weight Polyethylene Yarns.
- ASTM F899-20: Standard Specification for Wrought Stainless Steels for surgical instruments.
Note: We have used Grade 304 of Stainless steel (SS 304) Material for instruments as per ASTM F899-20, UHMWPE for Suture as per ASTM F2848-17 and Titanium Alloy (Ti- 6Al-4V) Grade 5 for Titanium Implants.
#### Summary of Biocompatibility
The device in its final finished form has been evaluated for biocompatibility according to ISO 10993-3.
#### Conclusion of Mechanical performance
The following are the mechanical tests that have been performed on the Subject device (i.e. Auxilock Titanium Screw-In Suture Anchor) and Predicate device (i.e. Smith & Nephew's Twinfix Suture Anchor):
- 1. Insertion Test
- 2. Pull-Out Tensile Static Test and Pull-Out Tensile Fatigue Test.
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the logo for Auxein. The logo consists of a blue circle with a white "A" inside of it, followed by the word "Auxein" in blue. There is a registered trademark symbol next to the "n" in "Auxein".
# Sterilization, shelf-life and packaging for sterile product
## Sterilization: ETO Sterilization
The ETO sterilization has been performed to sterilize this medical device. EO penetrates the packaging, making contact with all accessible surfaces of the product to deliver the required sterility assurance level (SAL).
Trace levels of EO and ethylene chlorohydrin (ECH) may remain on products after an EO sterilization process. To detect these traces, EO residual test was done. ISO 10993-7 outlines the specific limits of EO and ECH that must not be exceeded in order to ensure product and patient safety.
- ISO 11135: 2014, Sterilization of health-care products Ethylene oxide Requirements for the ● development, validation and routine control of a sterilization process for medical devices.
- ISO 17665-1:2006, Sterilization of health care products Moist heat Part 1: Requirements for ● the development, validation and routine control of a sterilization process for medical devices.
- ISO/TS 17665-2:2009, Sterilization of health care products - Moist heat - Part 2: Guidance on the application of ISO 17665-1.
- . ISO/TS 17665-3:2013 (en), Sterilization of health care products - Moist heat - Part 3: Guidance on the designation of a medical device to a product family and processing category for steam sterilization.
- . ISO 11140-1:2014, Sterilization of health care products — Chemical indicators — Part 1: General requirements.
- . ISO 11737-1:2018 Sterilization of medical devices - Microbiological methods- Part 1: Estimation of population of microorganisms on products.
- ISO 11737-2:2009 Sterilization of medical devices - Microbiological methods- Part 2: Tests of sterility performed in the validation of a sterilization process.
## Packaging of the Product: Tyvek Packaging
The integrity of the final package is maintained at least for the claimed shelf-life of the medical device.
The tyvek pouch is used for the packaging of Auxilock Titanium Screw-In Suture Anchor. The double sterile barrier is used for the packaging.
We have followed the below standards for packaging of the device:
- · ISO 11607-1:2006/AMD1:2014 Packaging for terminally sterilized medical devices part 1: requirements for materials, sterile barrier systems and packaging system.
- . ISO 11607-2:2006/AMD1:2014 Packaging for terminally sterilized medical devices - Part 2: Validation requirements for forming, sealing and assembly processes.
## Shelf-life: 5 years
The stability study has been done to determine the shelf life.
We have followed the below standards for performing shelf-life of the device:
- ASTM F1980:2016 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices.
- ASTM F88/F88M:2015 Standard test method for seal strength of flexible barrier materials.
- ASTM F1929:2015 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging ●
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the logo for Auxein. The logo consists of a blue circle with a white triangle inside, followed by the word "Auxein" in blue, stylized font. There is a small "R" in a circle to the right of the word "Auxein".
by Dye Penetration.
#### Safety in MRI Not Evaluated
The Auxilock Titanium Screw-In Suture Anchor has not been evaluated for safety in the MR environment. It has not been tested for heating or unwanted movement in the MR environment. The safety of Auxilock Titanium Screw-In Suture Anchor in the MR environment is unknown. Performing an MR exam on a person who has this medical device may result in injury or device malfunction.
#### Bacterial Endotoxin
- USP <85> Bacterial Endotoxin Test. .
- USP <161> Medical Devices-Bacterial Endotoxin and Pyrogen Tests.
Bacterial Endotoxin Test (BET): We perform Bacterial Endotoxin Test of every sterile batch products by using acceptance criteria as endotoxin testing limit must be <20EU/Device (as the products are general medical devices for implantation). The testing is done by using Limulus amebocyte lysate (LAL) test. The testing procedure is performed as per the standards USP 85 and ANSI/AAMI ST72:2019.
## Conclusion:
There are no significant differences between the subject device and the predicate devices that would adversely affect the use of the product. It is substantially equivalent to these devices in design, function, materials, and operational principles as internal fixation components. From the data available we can justify that the Auxilock Titanium Screw-In Suture Anchor is as safe, and as effective and performs the same indications for use as that of already marketed predicate devices identified in 510(k) summary.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.