K213097 · Pain Management Technologies, Inc. · ILO · Jun 10, 2022 · Physical Medicine
Device Facts
Record ID
K213097
Device Name
Armory Motion
Applicant
Pain Management Technologies, Inc.
Product Code
ILO · Physical Medicine
Decision Date
Jun 10, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5720
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Armory Motion is a prescription device that combines cold and compression therapy and is intended to treat postsurgical and acute injuries to reduce edema, swelling, and pain for which cold and compression are indicated. Armory Motion is intended to be used by a licensed health care professional in rehabilitation facilities, athletic training settings, and home settings.
Device Story
Armory Motion is a prescription, AC/battery-powered device providing cold and intermittent pneumatic compression therapy. System components include a control lid with LCD panel, water/air pumps, electronics, an insulated ice/water reservoir, and a dual-bladder joint wrap. The device circulates cold water through one bladder and cyclically inflates/deflates the second bladder with air. Operated via an integrated keypad, the device allows users to select therapy modes as directed by a healthcare professional. It is used in rehabilitation, athletic training, and home settings. The device includes a safety alarm that triggers if air pressure exceeds 100 mmHg, causing the system to stop and deflate. By combining cold and compression, the device aims to reduce edema, swelling, and pain at the treatment site (back, elbow, shoulder, foot, ankle, or knee).
Clinical Evidence
No clinical testing was performed. Substantial equivalence is based on bench testing, risk analysis, and FMEA.
Technological Characteristics
Multimodality software-controlled device; AC/rechargeable Li-ion battery powered. Dual-bladder pneumatic/thermal wrap. Components: water pump, air pump, insulated reservoir. Connectivity: standalone. Standards: ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 14971, IEC 60601-1-6, IEC 62304, ISO 10993-1.
Indications for Use
Indicated for patients with postsurgical or acute injuries requiring cold and compression therapy to reduce edema, swelling, and pain. Intended for use by licensed healthcare professionals or patients in rehabilitation, athletic training, or home settings.
Regulatory Classification
Identification
A water circulating hot or cold pack is a device intended for medical purposes that operates by pumping heated or chilled water through a plastic bag and that provides hot or cold therapy for body surfaces.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 890.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 10, 2022
Pain Management Technologies, Inc. Joshua Lefkovitz President 1760 Wadsworth Rd. Akron, Ohio 44320
Re: K213097
Trade/Device Name: Armory Motion Regulation Number: 21 CFR 890.5720; 21 CFR 890.5650 Regulation Name: Water Circulating Hot Or Cold Pack; Powered inflatable tube massager Regulatory Class: Class II Product Code: ILO, IRP Dated: May 10, 2022 Received: May 18, 2022
Dear Joshua Lefkovitz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K213097
Device Name Armory Motion
Indications for Use (Describe)
The Armory Motion is a prescription device that combines cold and compression therapy and is intended to treat postsurgical and acute injuries to reduce edema, swelling, and pain for which cold and compression are indicated.
Armory Motion is intended to be used by a licensed health care professional in rehabilitation facilities, athletic training settings, and home settings.
| Type of Use (Select one or both, as applicable) | | | |
|-------------------------------------------------|-----------------------------------------------|--|--|
| 7 Prescription Use (Part 21 CED 801 Subnart D) | Over The Counter I lea (21 CED 201 Submort C) | | |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# K213097
Date of Submission prepared: 16 August 2021
#### Submitter l
| Address: | 1760 Wadsworth Rd. Akron, OH 44320, USA |
|------------------------------------|-----------------------------------------|
| Tel: | 614-937-8236 |
| Fax: | 888-304-5454 |
| E-Mail: | josh@paintechnology.com |
| FDA Establishment Registration No: | 1528161 |
| Contact person: | Joshua A. Lefkovitz / President |
## Address of the manufacturing facility:
| Manufacturer: | EasyMed Instruments Co Ltd |
|-------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | 3/F, 5F-6F, Block A, GupoGongmao Building,<br>Fengxin Road, Fengxiang Industrial District,<br>Daliang, 528300 Shunde, Foshan, Guangdong<br>China, |
| FDA Establishment | |
Registration No: 3004049909
#### ll Submitted Device:
| Trade name: | Armory Motion |
|------------------------|-------------------------------------------------------------------------|
| Common name: | Cold and Compression Therapy |
| Classification Name: | Powered Inflatable Tube Massager;<br>Water circulating hot or cold pack |
| Classification Number: | Title 21 CFR 890.5650;<br>Title 21, CFR 890.5720 |
| Product Code: | IRP, ILO |
| Classification Panel: | Neurological and Physical Medicine |
| Regulatory Class: | Class II |
#### Predicate Device: ���
| Trade/Device Name: | Cold and Compression Therapy |
|--------------------|------------------------------|
| Model Number: | Polar Care Wave™ |
| Submitter: | Breg, Inc. |
| 510(k) Number: | K183702 |
| Product Code: | IRP, ILO |
| Type of Use: | Prescription |
{4}------------------------------------------------
Classification: Class II
#### IV DEVICE DESCRIPTION
Armory Motion is a multimodality software-controlled device designed to provide both a muscle massage and a circulatory thermal therapy.
The Armory Motion is an AC Rechargeable Battery powered device designed to provide cold and intermittent Pneumatic Compression to a treatment site on the user. The device includes a lid with LCD panel, pump and electronics, a reservoir, tubing and a single-patient use, non-sterile, pneumatic air compression joint wrap that can also circulate water (dual bladder), to provide cold thermal therapy in addition to the compression. The Joint wrap can be used to treat the back, elbow, shoulder, foot, ankle, and knee.
The Armory Motion device contains several key parts: A cold-compression pad with a connector hose, an insulated container to hold ice and water, and a lid that contains a water pump and air pump, electronics to control the pumps, tubing, and an intuitive keypad to operate the device.
The cold-compression pads are connected to the lid tubing and are applied to the body to deliver cold and intermittent compression therapy. The cold compression pads are made of flexible materials that form two bladders. The dual bladder design allows cold water to continuously flow through one of the bladders to provide cold therapy, and air to be cyclically pumped into and released from a second bladder to provide intermittent pneumatic compression therapy.
The insulated container is filled with ice and water by the user. The water circulates through the cold compression pad to deliver cold therapy to the application site.
The lid which houses the electronics, water pump, and air pump, is sealed and not accessible by the user. The lid also houses the keypad, which controls the device and allows the user to select the therapies as directed by a healthcare professional.
#### Indications for Use of the device >
The Armory Motion is a prescription device that combines cold and compression therapy and is intended to treat post-surgical and acute injuries to reduce edema, swelling, and pain for which cold and compression are indicated.
Armory Motion is intended to be used by a licensed health care professional in rehabilitation facilities, outpatient clinics, athletic training settings, and home settings.
## VI Equivalence Comparison to the Predicate Devices:
The technical characteristics of the applicant device (Model: Armory Motion) are similar to those of the predicate device in design, intended use and function. The predicate device Breg Polar Care Wave (K183702), and the new device Armory Motion is a device that
{5}------------------------------------------------
utilize a pneumatically controlled compression wrap providing intermittent pressure to joints and the lower extremities of a patient.
The Breg Polar Care Wave (K183702) provides both intermittent pneumatically controlled compression and cold thermal therapy.
The table below summarizes the shared and unique technological elements between the Armory Motion and the predicate device. The technology, engineering, and performance for Armory Motion are substantially equivalent to the predicate device.
From the view of safety and effectiveness, the new device Armory Motion uses preset programs that are substantially equivalently to the predicated device. The output characteristics of Armory Motion are similar to those of predicated device, see Table below.
The new device Armory Motion is designed to comply with relevant safety applicable recognized consensus standards; the output pressure is controlled well within the safety and effectiveness ranges specified by relevant FDA guidance. Detailed and strictly controlled testing has been carried out.
Furthermore, Test results, Risk Analysis, and FMEA analysis show that the new device Armory Motion is safe with no hazard.
### As such:
1. the new device (Model: Armory Motion) has the same technological characteristics and intended uses as the predicate devices, and
2. the information submitted to the FDA for the new device (Model: Armory Motion) does not raise new questions about safety or effectiveness and demonstrates with reasonable assurance based on established controls that the device is at least as safe and effective as a legally marketed device.
{6}------------------------------------------------
# Table 1 Substantial Equivalence Comparison Table
| 510(K)No.: | (to be assigned) | Primary Predicate Device<br>K183702 | Conclusion |
|--------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|
| Device name: | Armory Motion and<br>Armory Force | Polar Care Wave | |
| Submitter: | Pain Management Technologies | Breg, Inc. | |
| Indications for use: | The Armory Motion is a prescription device<br>that combines cold and compression<br>therapy and is intended to treat<br>post-surgical and acute injuries to reduce<br>edema, swelling, and pain for which cold<br>and compression are indicated.<br><br>Armory Motion is intended to be used by<br>licensed health care professional<br>rehabilitation facilities, outpatient clinics,<br>athletic training settings, and home settings. | The Breg is a prescription device<br>intended to treat post-surgical and acute<br>injuries to reduce edema, swelling, and<br>pain where cold and compression are<br>indicated. It is intended to be used by or<br>on the order of licensed healthcare<br>professionals in hospitals, outpatient<br>clinics, athletic training settings, or home<br>settings. | Substantially equivalent |
| OTC use | Prescription | Prescription | Substantially Equivalent. |
| Classification<br>Number | 21 CFR 820.5650<br>21 CFR 890.5720 | 21 CFR 820.5650<br>21 CFR 890.5720 | Substantially Equivalent. |
| Classification Name | Powered Inflatable Tube Massager;<br>Water Circulating hot or cold pack | Powered Inflatable Tube Massager;<br>Water Circulating hot or cold pack | Substantially Equivalent. |
| Regulatory Class | II | II | Substantially Equivalent. |
| 510(K)No.: | (to be assigned) | Primary Predicate Device<br>K183702 | Conclusion |
| Device name: | Armory Motion and<br>Armory Force | Polar Care Wave | |
| Product Code | IRP, ILO | IRP, ILO | Substantially Equivalent. |
| Classification Panel | Neurological and Physical Medicine; | Neurological and Physical Medicine; | Substantially Equivalent. |
| Therapy | Cold and Compression<br>Work together or independently | Cold and Compression<br>Work together or independently | Substantially Equivalent. |
| Therapy modes | Manually adjustable mode - allows the user<br>to adjust cold and two compression settings. | Manually adjustable mode - allows the<br>user to adjust cold and two compression<br>settings. | Substantially Equivalent. |
| Compression Setting | Available in two levels Low (0-25 mm Hg)<br>Regular (0-50 mm Hg) | Available in two levels Low (0-25 mm Hg)<br>Regular (0-50 mm Hg) | Substantially Equivalent |
| Types of Pads | Various anatomical pads: Knee,<br>Shoulder, Back, calf, Universal/joint,<br>Foot/Ankle | Various anatomical pads: Knee,<br>Shoulder, Back, Hip, Universal,<br>Foot/Ankle | Substantially Equivalent |
| Therapy Temperature Range | 45°F-60°F Cold | 45°F-60°F | Substantially Equivalent |
| Operating fluid and<br>cooling agent | Tap water and Ice | Tap water and Ice | Substantially Equivalent |
| Single User and<br>sterility | Single user and non-sterile | Single user and non-sterile | Substantially Equivalent |
| 510(K)No.: | (to be assigned) | Primary Predicate Device<br>K183702 | Conclusion |
| Device name: | Armory Motion and<br>Armory Force | Polar Care Wave | |
| Line Voltage /<br>Frequency | Rechargeable battery and<br>100-240 VAC 50/60 HZ | 100-240 VAC 50/60 HZ | Substantially Equivalent. |
| Treatment times | 15, 20, 30, 45; compression stops at 30<br>min max | Continuous, until turn-off byuser | Substantially Equivalent |
| Cooling on/off<br>times | 15, 20, 30, 45 | Continuous, until turn-off byuser | Substantially Equivalent |
| Pressure Mode | Two compression modes and one cold<br>therapy mode | Three | Substantially Equivalent |
| Power Supply | 3.7V rechargeable Li-ion Polymer Battery<br>Power Adapter | Power Adapter | Substantially Equivalent. |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
Both the new device Armory Motion and the predicate device Breg Polar Care Wave (K183702) provide continuous intermittent pneumatic massage using inflatable garments. The new device Armory Motion and predicate device have the same indications for use, both operate within the similar parameters and performance specifications. The new device Armory Motion and predicate device use the similar inflation pressures and cycle times.
Furthermore, a subtle difference between the new device Armory Motion and the predicate device is in their respective alarm function. The new device has a malfunction overpressure safety alarm: any time the pressure in the air chamber exceeds 100mmHq. it will stop working and deflate immediately. The device panel shall sleep with the red warning sign on and flash for 20 seconds.
Substantial Equivalence None of the performance or technological differences between the Armory Motion Device and the predicate device raise any new issues of safety and effectivenes.
### CONCLUSION:
The data, and information provided in this submission, support the conclusion that the Armory Motion Safety features are Substantially Equivalent to those of Breg Polar Care Wave (K183702). No special safety features have been shown to be necessary for home use without instruction.
{10}------------------------------------------------
| FDA recognition No. | Standard Title |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 19-4 | ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012,<br>C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text) Medical<br>electrical equipment - Part 1: General requirements for basic safety<br>and essential performance (IEC 60601-1:2005, MOD) |
| 19-8 | IEC 60601-1-2 Edition 4.0 2014-02<br>Medical electrical equipment - Part 1-2: General requirements for<br>basic safety and essential performance - Collateral Standard:<br>Electromagnetic disturbances - Requirements and tests |
| 19-14 | IEC 60601-1-11 Edition 2.0 2015-01<br>Medical electrical equipment - Part 1-11: General requirements for<br>basic safety and essential performance - Collateral Standard:<br>Requirements for medical electrical equipment and medical<br>electrical systems used in the home healthcare environment. |
| 5-40 | ISO 14971: Second Edition 2007-03-01 Medical devices-<br>Application of Risk Management To Medical Devices. |
| 5-89 | IEC 60601-1-6 Edition 3.1 2013-10<br>Medical electrical equipment - Part 1-6: General requirements for<br>basic safety and essential performance - Collateral standard:<br>Usability |
| 13-79 | IEC 62304 Edition 1.1 2015-06 CONSOLIDATED<br>VERSION Medical device software - Software life cycle processes |
| 2-220 | ANSI AAMI ISO 10993-1:2009/(R)2013<br>Biological evaluation of medical devices - Part 1: Evaluation and<br>testing within a risk management process |
#### VII Performance Tests:
#### V Clinical Testing
No formal clinical testing was performed on the Armory Motion. The clinical testing is not applicable in this submission.
#### IX Conclusion:
- � The new device Armory Motion has the same technological characteristics and intended uses as the predicate deivce Breg Polar Care Wave (K183702); and
- � The labelling of the new device Armory Motion is concordant with the predicate device and FDA requirements; and
- � The information submitted to the FDA for the new device Armory Motion does not raise new questions about safety or effectiveness and demonstrates with reasonable assurance based on established controls that the device is at least as safe and effective as a legally marketed device.
Therefore, the new device Armory Motion is substantially equivalent to the predicate device and differences between the devices do not raise any new questions about safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.