Ivenix Infusion System, LVP Epidural Administraion Set NRFit Connector
K213089 · Ivenix, Inc. · FPA · Nov 10, 2021 · General Hospital
Device Facts
Record ID
K213089
Device Name
Ivenix Infusion System, LVP Epidural Administraion Set NRFit Connector
Applicant
Ivenix, Inc.
Product Code
FPA · General Hospital
Decision Date
Nov 10, 2021
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Ivenix LVP Epidural Administration Set with NRFit Connector (SET-0030-1) is intended for the controlled administration of pharmaceutical drugs or other therapies via the epidural pathway. The Ivenix LVP Epidural Administration Set is intended only for use with the Ivenix Infusion System in a hospital or outpatient.
Device Story
The Ivenix LVP Epidural Administration Set is a sterile fluid delivery component for the Ivenix Infusion System. It consists of a primary inlet tubing with a vented spike, a pumping cassette, and outlet tubing terminating in an ISO 80369-6:2013 compliant NRFit connector. The device is operated by clinicians in hospital or outpatient environments. It interfaces with the Ivenix Large Volume Pump (LVP) to enable controlled fluid delivery; it is not intended for gravity infusion. The NRFit connector prevents misconnections with non-epidural access devices, enhancing patient safety by ensuring the sterile fluid path is maintained during epidural therapy.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including ISO 80369-6:2016 compliance, connector bond strength (ISO 8536-4:2019), flow accuracy, shelf life, and microbial ingress testing.
Technological Characteristics
Materials: PE-lined PVC tubing, Terlux 2802 ABS connector. Dimensions: 29 +/- 1 in upstream, 73 +/- 2 in downstream. Priming volume: 10 mL. Connector: NRFit (ISO 80369-6:2016). Energy source: Powered by Ivenix LVP. Sterilization: Sterile fluid path.
Indications for Use
Indicated for controlled administration of pharmaceutical drugs or other therapies via epidural pathway in hospital or outpatient settings for adult, pediatric, and neonate patients.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
Predicate Devices
Ivenix Infusion System, Epidural Administration set with luer connector (SET-0016-1) (K183311)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right of this symbol, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" above the word "ADMINISTRATION".
Ivenix, Inc. John Sokolowski VP, Regulatory Affairs 50 High St North Andover, Massachusetts 01845
Re: K213089
Trade/Device Name: Ivenix Infusion System, LVP Epidural Administraion Set NRFit Connector Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: FPA Dated: October 7, 2021 Received: October 8, 2021
### Dear John Sokolowski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Carolyn Dorgan Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213089
Device Name
Ivenix LVP Epidural Administration Set with NRFit Connector
#### Indications for Use (Describe)
The Ivenix LVP Epidural Administration Set with NRFit Connector (SET-0030-1) is intended for the controlled administration of pharmaceutical drugs or other therapies via the epidural pathway. The Ivenix LVP Epidural Administration Set is intended only for use with the Ivenix Infusion System in a hospital or outpatient.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;"> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span style="font-family: Arial, sans-serif;"> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY K213089
| Manufacturer's Name: | Ivenix, Inc.<br>50 High Street, North Andover, MA 01845 |
|-------------------------|-------------------------------------------------------------------------------------------------|
| Corresponding Official: | John J. Sokolowski<br>Vice President, Regulatory Affairs |
| Telephone Number: | (978) 775-8050 |
| Fax Number: | (978) 775-8052 |
| E-mail Address: | john.sokolowski@ivenix.com |
| Trade Name: | Ivenix LVP Epidural Administration Set with NRFit Connector, SET-<br>0030-1 |
| Common or Usual Name: | Epidural infusion administration set |
| Regulation Name: | Set, administration, intravascular |
| Regulation Number: | 21 CFR 880.5440 |
| Product Codes: | FPA |
| Device Class: | Class II |
| Predicate Device: | K183311 Ivenix Infusion System, Epidural Administration set with luer<br>connector (SET-0016-1) |
### Device Description
The Ivenix LVP Epidural Administration Set with NRFit connector is comprised of a primary channel inlet tubing with a vented cap spike, a cassette which contains the fluid pumping mechanism and serves as the interface to the Ivenix LVP, and outlet tubing for connection to patient access. The end of the outlet tubing is fitted with a NRFit connector that is compliant to ISO 80369-6:2013.
The Ivenix LVP Epidural Administration Set with NRFit Connector provides a sterile fluid path for the delivery of pharmaceutical drugs and other therapies via the epidural pathway, provides a means for a clinician to aseptically connect to the fluid source and to the patient access site, and interfaces with the LVP to enable measurement and control of the fluid delivery.
The Ivenix LVP Epidural Administration Set with NRFit Connector is intended for use within the Ivenix Infusion System for controlled fluid delivery; it is not recommended for gravity infusion.
### Indications for Use
The Ivenix LVP Epidural Administration Set with NRFit Connector (SET-0030-1) is intended for the controlled administration of pharmaceutical drugs or other patient therapies via the epidural pathway. The Ivenix LVP Epidural Administration Set is intended only for use with the Ivenix Infusion System in a hospital or outpatient care environment.
{4}------------------------------------------------
# Substantial Equivalence Discussion
# Intended Use Comparison
The table below includes a comparison of the intended use between the subject device and those of the predicate device:
| Characteristic | Subject Device<br>Ivenix Infusion System,<br>Microbore single inlet, single<br>outlet, for administration of<br>epidural fluids with NRFit®<br>Connector (SET-0030-1) | Predicate Device<br>Ivenix Infusion System<br>K183311, Microbore<br>single inlet, single outlet,<br>for administration of<br>epidural fluids with luer<br>connector (SET-0016-1) | Discussion |
|--------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The Ivenix Infusion<br>System is indicated for use<br>in a hospital and in<br>outpatient care<br>environments for the<br>controlled administration<br>of fluids through clinically<br>accepted routes of<br>administration:<br>intravenous, intra-arterial,<br>epidural, and subcutaneous<br>to adults, pediatric and<br>neonate patients.<br>Administered fluids may<br>be pharmaceutical drugs,<br>red blood cells, platelets,<br>plasma, and other mixtures<br>required for patient<br>therapy. | The Ivenix Infusion<br>System is indicated for<br>use in a hospital and in<br>outpatient care<br>environments for the<br>controlled administration<br>of fluids through<br>clinically accepted routes<br>of administration:<br>intravenous, intra-<br>arterial, epidural, and<br>subcutaneous to adults,<br>pediatric and neonate<br>patients. Administered<br>fluids may be<br>pharmaceutical drugs,<br>red blood cells, platelets,<br>plasma, and other<br>mixtures required for<br>patient therapy. | No differences<br>between the devices. |
| Prescription Only or<br>Over the Counter | Prescription Only | Prescription Only | No differences<br>between the devices. |
| Intended Population | Adult, pediatric, neonate | Adult, pediatric, neonate | No differences<br>between the devices. |
| Environment of Use | Hospital or out-patient setting | Hospital or out-patient<br>setting | No differences<br>between the devices. |
| Item | Subject Device - Epidural<br>Administration Set with<br>NRFit connector (SET-0030-<br>1) | Predicate Device - Epidural<br>Administration Set with<br>luer connector (SET-0016-<br>1) | Discussion |
| Single inlet | Primary line with vented spike<br>and cap, no secondary inlet. | Primary line with vented<br>spike and cap, no secondary<br>inlet. | No differences between the<br>devices. |
| Flow<br>dial/regulator | The Flow Dial integrates with<br>the LVP to maintain flow<br>accuracy. | The Flow Dial integrates<br>with the LVP to maintain<br>flow accuracy. | No differences between the<br>devices. |
| Connector | The connector that attaches<br>the downstream outlet tubing<br>to the patient access device<br>has an NRFit connector,<br>complaint to ISO 80369-<br>6:2016 with cap. | The connector that attaches<br>the distal line to the patient<br>access device has a luer lock<br>with cap. | The subject device has a<br>connector that meets the<br>requirements of a small bore<br>connector, per ISO 80369-<br>6:2016 |
| Outlet Tubing /<br>Distal Line<br>(Patient Line) | Runs from the cassette to the<br>patient. | Runs from the cassette to the<br>patient. | No differences between the<br>devices. |
| Upstream tubing | PE lined PVC with yellow<br>stripe, length 29 +/- 1 in. | PE lined PVC with yellow<br>stripe, length 29 +/- 1 in. | No differences between the<br>devices |
{5}------------------------------------------------
### Discussions of differences in Indications for Use statement
The indications for use statements are identical for both subject and predicate devices.
### Discussions of differences in intended population
The intended population for the subject device is identical to the predicate device.
### Discussions of differences in environment of use There are no differences in the environment of use.
### Technological Characteristics
The Ivenix LVP Epidural Administration Set with NRFit Connector is a modification of the Ivenix LVP Epidural Administration Set with Luer Connector (SET-0016-1). These sets have equivalent technological characteristics, as described below, and support substantial equivalence.
The table below includes a comparison of the technological characteristics between the new device and those of the predicate device:
{6}------------------------------------------------
| Item | Subject Device - Epidural<br>Administration Set with<br>NRFit connector (SET-0030-<br>1) | Predicate Device - Epidural<br>Administration Set with<br>luer connector (SET-0016-<br>1) | Discussion |
|-------------------------------------------------------------|------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|---------------------------------------|
| Downstream<br>tubing | PE lined PVC with yellow<br>stripe, length 73 +/- 2 in. | PE lined PVC with yellow<br>stripe, length 73 +/- 2 in. | No differences between the<br>devices |
| Priming volume | 10 mL | 10 mL | No differences between the<br>devices |
| Patient access<br>connector<br>materials of<br>construction | Terlux 2802 ABS | Terlux 2802 ABS | No differences between the<br>devices |
### Discussions of differences in technological characteristics
The technological characteristics of the subject and predicate devices are identical except for the dimensional characteristics of the patient access connector. The subject device meets the dimensional specifications of ISO 80369-6:2016.
### Performance Testing
The following bench testing was performed and reviewed to support the substantial equivalence of the subject device:
| Conformance to ISO<br>80369-6:2013 | The NRFit connector was tested against the requirements of ISO-80369-6:2016, using the<br>methods described in ISO-80369-20:2015, by an accredited test laboratory. All requirements<br>were met. |
|-----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Connector bond strength | The epidural administration set with NRFit met the pull strength requirements of ISO 8536-<br>4:2019. |
| Flow accuracy and shelf<br>life testing | The epidural administration set with NRFit connector met all functional requirements of the<br>Ivenix epidural set with luer connector. |
| Microbial Ingress | The sterile fluid path of the administration set was challenged, under simulated worst case<br>conditions, at the patient access device, with four strains of bacteria (2 gram positive, 2 gram<br>negative). Test results showed no growth of microorganism in any test article, demonstrating<br>the integrity of the NRFit connector to maintain the sterile fluid path under simulated<br>conditions of use. |
| Biocompatibility | The polymer material of the NRFit connector is identical to the standard luer connector of the<br>predicate device in its final finished form in formulation, processing, sterilization, and<br>geometry (with the exception of the geometry necessary to meet the requirements of ISO<br>80369-6:2016), and no other chemicals have been added (e.g., plasticizers, fillers, additives,<br>cleaning agents, mold release agents). Therefore, no additional biocompatibility testing was<br>conducted |
{7}------------------------------------------------
### Clinical Tests
Clinical evaluation is not required for this submission to support substantial equivalence. Similarly, the predicate devices did not undergo clinical evaluation to support substantial equivalence.
### Conclusions
The differences between the predicate and the subject device do not raise any new or different questions of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.