The ExamVue Duo software is designed for use by radiologists and radiology technicians for the acquisition of digital x-ray images. It interfaces with 3rd party digital x-ray detectors or CR scanners and manufacturer supplied software for the acquisition and storage of digital x-ray images. The ExamVue Duo software then provides a user interface for the viewing, annotating, and other workstation functions. ExamVue Duo software includes the ability to receive patient information and send x-ray images to remote destinations using the DICOM 3.0 protocol.
Device Story
ExamVue Duo is digital X-ray acquisition software; interfaces with 3rd-party detectors/generators to acquire, process, store, and view digital X-ray images. Operated by radiologists/technicians in clinical environments; supports DICOM 3.0 for image transfer/worklist management. Provides tools for annotation, measurement, auto-stitching, and optional bone suppression/specialty measurements (chiropractic/podiatry). Output displayed on workstation for clinical evaluation; assists in diagnostic decision-making by providing enhanced/processed images.
Clinical Evidence
Bench testing only. Provided direct comparison of functions between ExamVue Duo and predicate device. No clinical data presented.
Technological Characteristics
Software-based digital X-ray acquisition system. Runs on Windows 10; requires Intel Core i5 or higher, 8GB RAM, 500GB HDD, 1GBit network. Interfaces with 3rd-party detectors/generators via DICOM 3.0. Features include auto-stitching, bone suppression, and specialty measurement tools. Software version 1.0.
Indications for Use
Indicated for acquisition, processing, storage, and viewing of digital radiology studies in adult and pediatric subjects. Used in general imaging, including podiatry, orthopedic, and other specialties, and in mobile x-ray systems. Not indicated for mammography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
February 18, 2022
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left, there is a seal with an eagle-like figure. To the right of the seal, there is a blue square with the letters 'FDA' in white. Next to the blue square, the words 'U.S. FOOD & DRUG' are written in blue, with the word 'ADMINISTRATION' written in a smaller font size below.
JPI Healthcare Co, Ltd. % William Little Senior Product Manager JPI Healthcare Solutions. Inc 52 Newtown Plaza PLAINVIEW NY 11803
Re: K213057
Trade/Device Name: ExamVue Duo Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: KPR Dated: February 8, 2022 Received: February 10, 2022
Dear William Little:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for
combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laurel Burk Assistant Director Diagnostic X-ray Systems Team Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213057
Device Name ExamVue Duo
Indications for Use (Describe)
Exam Vue Duo is a software for the acquisition, processing, storage and viewing of digital radiology studies. Exam Vue Duo is intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects. Exam Vue Duo is indicated for use in general imaging including podiatry, orthopedic, and other specialties, and in mobile x-ray systems.
Exam Vue Duo is not indicated for use in mammography.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | <span> X </span> |
| Over-The-Counter Use (21 CFR 801 Subpart C) | <span> </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K213057 510 (k) Summary
## April 12, 2021
### 1. Company and Correspondant Making the Submission:
Name: JPI Healthcare Co,. LTD Address: 608-ho, 28, Digital-ro 33-gil, Guro-gu, 08377 Seoul, Korea Telephone: +82-2-2108 - 2580 Fax: +82-2-2108-1180 Contact: Lee Jang Won Website: http://www.jpi.co.kr/
## 2. Identification of Device
| Classification Name: | Stationary x-ray system |
|-------------------------|------------------------------------|
| Common Name: | Digital X-ray Acquisition Software |
| Trade/Proprietary Name: | ExamVue Duo |
#### 3. Predicate Device
| Manufacturer: | JPI Healthcare Solutions, Inc |
|----------------------|------------------------------------|
| Device: | ExamVue DR |
| 510(k) Number: | K142930 |
| Classification Name: | Stationary x-ray system |
| Common Name: | Digital X-ray Acquisition Software |
| Regulatory Number: | 21 CFR 892.1680 |
| Regulatory Class: | II |
| Product Code: | 90 KPR |
#### Product Classification Names and Citations 4.
| Regulatory Number: | 21 CFR 892.1680 |
|--------------------|-----------------|
| Regulatory Class: | II |
| Product Code: | 90 KPR |
#### 5. Description:
The ExamVue Duo software is designed for use by radiologists and radiology technicians for the acquisition of digital x-ray images. It interfaces with 3rd party digital x-ray detectors and (optionally) generators and manufacturer supplied software for the acquisition and storage of digital x-ray studies. The ExamVue Duo software then provides a user interface for the viewing, annotating, and other workstation functions.
{4}------------------------------------------------
Exam Vue DR includes the ability to receive patient information and send studies to remote destinations using the DICOM 3.0 protocol.
- 6. Indication for use
ExamVue Duo is a software for the acquisition, processing, storage and viewing of digital radiology studies. ExamVue Duo is intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects.
ExamVue Duo is indicated for use in general imaging, specialist imaging including podiatry, orthopedic, and other specialties, and in mobile x-ray systems.
ExamVue Duo is not indicated for use in mammography.
- 7. Comparison with Predicate Device:
JPI Healthcare Co., Ltd. believes that the ExamVue Duo software is substantially equivalent to the ExamVue DR software registered by JPI Healthcare Co. Ltd The ExamVue Duo software and the predicate device both -Provide a user interface for the registration, acquisition, and evaluation of x-ray studies. -Perform the functions of image transfer, image acquisition, image processing, and maintaining a patient database. -Use the DICOM 3.0 standard for medical imaging -Are intended for installation on Windows operating systems for use in a medical environment. -Interface with and process images from multiple models of hardware.
ExamVue Duo and the predicate device share the same essential functions of image acquisition, transfer, and processing; however they have different user interfaces and ExamVue Duo has some additional registration, annotation and imaging processing features. We believe this does not represent a substantial difference between the two devices, as the user interface presents the same essential data and supports similar workflow as the predicate device.
# 8. Safety, EMC and Performance Data
Bench tests reports and clinical data have been provided, detailing the direct comparison of functions between the ExamVue Duo and predicate device.
- 9. Conclusions:
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification JPI Healthcare Co., Ltd. concludes that the ExamVue PACS is safe and effective and substantially equivalent to predicate devices as described herein.
A detailed comparison supporting this conclusion can be found in Exhibit 1, Substantial Equivalence Chart.
{5}------------------------------------------------
# EXHIBIT 1
SUBSTANTIAL EQUIVALENCE CHART
{6}------------------------------------------------
| Characteristics | Proposed<br>ExamVue Duo | Predicate 1<br>ExamVue DR |
|-------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | JPI Healthcare Solutions, Inc | JPI Healthcare Solutions, Inc |
| 510(k) Number | | K142930 |
| Intended use. | The ExamVue Duo software is<br>designed for use by radiologists<br>and radiology technicians for the<br>acquisition of digital x-ray images. It<br>interfaces with 3rd party digital x-ray<br>detectors or CR scanners and<br>manufacturer supplied software for<br>the acquisition and storage of digital<br>x-ray images. The ExamVue<br>Duo software then provides a user<br>interface for the viewing,<br>annotating, and other workstation<br>functions. ExamVue Duo software<br>includes the ability to receive<br>patient information and send x-ray<br>images to remote destinations<br>using the DICOM 3.0 protocol. | The ExamVue DR software is<br>designed for use by radiologists<br>and radiology technicians for<br>the acquisition of digital x-ray<br>images. It interfaces with 3rd<br>party digital x-ray detectors or<br>CR scanners and manufacturer<br>supplied software for the<br>acquisition and storage of digital<br>x-ray images. The ExamVue DR<br>software then provides a user<br>interface for the viewing,<br>annotating, and other<br>workstation functions. ExamVue<br>DR includes the ability to<br>receive patient information and<br>send x-ray images to remote<br>destinations using the DICOM<br>3.0 protocol. |
| Performance<br>Standard | 21 CFR 820<br>21 CFR 830 | SAME |
| Processor | Intel Core i5 or higher | Intel Core i5 or higher |
| RAM | 8GB or higher | Min. 4GB |
| Hard Disk | Min. 500GByte | Min. 500GByte |
| Network | 1GBit | 1GBit |
| Operating<br>System<br>Requirements | Windows 10 | Windows 7, Windows 8 or<br>Windows 10 |
| Resolution | Min. 1366 x 768 | 1920 x 1080 |
| Image<br>Transfer | DICOM 3.0 Standard | Same |
| Image<br>Acquisition | Yes | Yes |
| Image<br>Processing | Yes | Yes |
| Windowing | Yes | Yes |
| Image<br>Formatting | Yes (1x1, 1x2, 2x1, 2x2, 3x3) | Yes (1x1, 1x2, 2x1, 2x2, 3x3) |
| Annotations<br>and<br>Measurements | Yes | Yes |
| Image<br>Rotation | Yes | Yes |
| Zoom | Yes | Yes |
| Patient<br>Database | Yes | Yes |
| DICOM<br>Worklist | Yes | Yes |
| DICOM Store | Yes | Yes |
| DICOM Print | Yes | Yes |
| Detector<br>Specific | No | No |
| X-ray<br>Generator<br>Control | Yes | Yes |
| SW Ver. | ExamVue Duo – 1.0 | ExamVue DR - 1.0 |
| Added<br>Functions | - Auto Stitching<br>- Bone Suppression (option)<br>- Chiropractic measurement features (option)<br>- Podiatry measurement tools (option) | |
{7}------------------------------------------------
{8}------------------------------------------------
# Summary
- The pre / post image processing algorithm is the same for ExamVue DR and ExamVue Duo. -
- The measurement features for Chiropractic and Podiatry in ExamVue Duo are additional features. -However, they are similar to the measurement function of length and angle already incorporated in ExamVue DR (Ver 1.0). For the user convenience, a specific region of interest for chiropractic and podiatry was designated as a dedicated in ExamVue Duo (Ver 1.0).
- -In Conclusion, the intended use and environment of the device is the same as the predicate devices, with only minor differences in features that are not integral to the function of the device. The differences do not introduce a fundamentally new scientific technology.
The intended use and environment of the device is the same as the predicate devices, with only minor differences in features that are not integral to the function of the device. For this reason, we believe it is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.