← Product Code [LZA](/productcode/LZA) · K213048

# Nitrile Patient Examination Gloves (Tested for Use with Chemotherapy Drugs) (K213048)

_Huayuan Medical Technology(Shangqiu) Co., Ltd. · LZA · Dec 18, 2021 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K213048

## Device Facts

- **Applicant:** Huayuan Medical Technology(Shangqiu) Co., Ltd.
- **Product Code:** [LZA](/productcode/LZA.md)
- **Decision Date:** Dec 18, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves to Permeation by Chemotherapy Drugs.

## Device Story

Disposable, non-sterile, ambidextrous nitrile examination gloves; blue color; beaded cuff. Used by healthcare professionals in clinical settings to provide a barrier against contamination between patient and examiner. Tested for permeation resistance against specific chemotherapy drugs per ASTM D6978-05. Output is a physical barrier; healthcare providers use the gloves during patient examinations or chemotherapy drug handling to protect themselves from exposure and prevent patient cross-contamination. Benefits include reduced risk of infection transmission and protection against hazardous drug exposure, excluding specific drugs with low breakthrough times.

## Clinical Evidence

No clinical data. Bench testing only. Performance evaluated per ASTM D6319-19 (physical dimensions, tensile strength, elongation), ASTM D5151-19 (watertightness/holes), ASTM D6124-06 (residual powder), and ASTM D6978-05 (chemotherapy permeation). Biocompatibility confirmed via ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation and sensitization).

## Technological Characteristics

Nitrile rubber material; powder-free; ambidextrous; beaded cuff; blue color. Dimensions: Length ≥220-230mm, Width 70-130mm, Thickness ≥0.05mm. Non-sterile. Complies with ASTM D6319, ASTM D5151, ASTM D6124, and ASTM D6978. Biocompatibility per ISO 10993-5 and ISO 10993-10.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Disposable Powder Free Nitrile Examination Glove, Tested For Use With Chemotherapy Drugs, Disposable Powder Free Nitrile Examination Glove, Tested For Use With Chemotherapy Drugs, Orange Color ([K190860](/device/K190860.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 18, 2021

Huayuan Medical Technology(Shangqiu) Co., Ltd. % Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM.1801,No.161 Lujiazui East Rd.,Pudong Shanghai, Shanghai 200120 China

Re: K213048

Trade/Device Name: Nitrile Patient Examination Gloves (Tested for Use with Chemotherapy Drugs) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC Dated: September 16, 2021 Received: September 22, 2021

Dear Boyle Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Clarence W. Murray, III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known) K213048

#### Device Name

Nitrile Patient Examination Gloves (Tested for Use with Chemotherapy Drugs)

#### Indications for Use (Describe)

A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves to Permeation by Chemotherapy Drugs.

| Chemotherapy Drug          | Concentration          | Breakthrough Detection Time in Minutes |
|----------------------------|------------------------|----------------------------------------|
| Carboplatin                | 10.0 mg/ml(10,000 ppm) | > 240 Minutes                          |
| Carmustine (BCNU)          | 3.3 mg/ml (3,300 ppm)  | 26.5 Minutes                           |
| Cyclophosphamide (Cytoxan) | 20.0 mg/ml(20,000 ppm) | > 240 Minutes                          |
| Doxorubicin                | 2.0 mg/ml(2,000 ppm)   | > 240 Minutes                          |
| Etoposide                  | 20.0 mg/ml(20,000 ppm) | > 240 Minutes                          |
| Fluorouracil               | 50.0 mg/ml(50,000 ppm) | > 240 Minutes                          |
| Methotrexate               | 25.0 mg/ml(25,000 ppm) | > 240 Minutes                          |
| Paclitaxel                 | 6.0 mg/ml(6,000 ppm)   | > 240 Minutes                          |
| ThioTepa                   | 10.0 mg/ml(10,000 ppm) | 48.4 Minutes                           |

Please note that the following drugs have low permeation times: Carmustine (BCNU) 3.3 mg/ml 26.5 Minutes ThioTepa 10.0 mg/ml 48.4 Minutes

Warning: Please do not use with Carmustine (BCNU) and ThioTepa.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary K213048

This summary of 510(k) is being submitted in accordance with 21 CFR 807.92.

#### 1.0 Submitter's Information

Huayuan Medical Technology (Shangqiu) Co., Ltd. Name: Address: In The North Yard of West 1000m Road At The Intersection of Zhuangzhou Avenue and Pingyuan Road, Liangyuan District, Shangqiu City, Henan Province, China Contact: Huamei Wang Date of Preparation: Sept 16, 2021

#### Designated Submission Correspondent

Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Room 1801, No. 161 Lujiazui East Rd., Pudong Shanghai, 200120 China Tel: +86-21-50313932 Email: Info@truthful.com.cn

#### 2.0 Device Information

Nitrile Patient Examination Gloves (Tested for Use with Trade name: Chemotherapy Drugs) Common name: Patient Examination Gloves Classification name: Non-powdered patient examination glove Model(s): XS, S, M, L, XL, XXL

#### 3.0 Classification

Production code: LZA,LZC Regulation number: 21CFR880.6250 Classification: Class I Panel: General Hospital

#### 4.0 Predicate Device Information

Manufacturer: Ever Growth (Vietnam) Co., Ltd. Device: Disposable Powder Free Nitrile Examination Glove, Tested For Use With Chemotherapy Drugs, Disposable Powder Free

{4}------------------------------------------------

Nitrile Examination Glove, Tested For Use With Chemotherapy Drugs, Orange Color

510(k) number: K190860

### 5.0 Device Description

The subject device is single use, disposable gloves intended for medical purposes to be worn on the examiner's hands to prevent contamination between patient and examiner. The gloves are powder-free, ambidextrous with beaded cuff, blue colored, nitrile, and tested for use with chemotherapy drugs. The gloves are offered in six sizes: XS, S, M, L, XL, XXL. The subject device is non-sterile.

#### 6.0 Indication for Use

A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves to Permeation by Chemotherapy Drugs

| Chemotherapy Drug             | Concentration          | Breakthrough Detection<br>Time in Minutes |
|-------------------------------|------------------------|-------------------------------------------|
| Carboplatin                   | 10.0 mg/ml(10,000 ppm) | > 240                                     |
| Carmustine (BCNU)             | 3.3 mg/ml (3,300 ppm)  | 26.5                                      |
| Cyclophosphamide<br>(Cytoxan) | 20.0 mg/ml(20,000 ppm) | > 240                                     |
| Doxorubicin                   | 2.0 mg/ml(2,000 ppm)   | > 240                                     |
| Etoposide                     | 20.0 mg/ml(20,000 ppm) | > 240                                     |
| Fluorouracil                  | 50.0 mg/ml(50,000 ppm) | > 240                                     |
| Methotrexate                  | 25.0 mg/ml(25,000 ppm) | > 240                                     |
| Paclitaxel                    | 6.0 mg/ml(6,000 ppm)   | > 240                                     |
| ThioTepa                      | 10.0 mg/ml(10,000 ppm) | 48.4                                      |

Please note that the following drugs have low permeation times: Carmustine (BCNU) 3.3 mg/ml 26.5 Minutes; Thio Tepa 10.0 mg/ml 48.4 Minutes.

Warning: Please do not use with Carmustine (BCNU) and ThioTepa.

## 7.0 Technological Characteristic Comparison Table

{5}------------------------------------------------

| Item                        | Subject Device<br>(K213048)                                                                                                                                                                                                                                                                                                                                                                                                           | Predicate Device<br>(K190860)                                                                                                                                                                                                                                                                                                                                                                                                         | Remark  |
|-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Product Code                | LZA,LZC                                                                                                                                                                                                                                                                                                                                                                                                                               | LZA,LZC                                                                                                                                                                                                                                                                                                                                                                                                                               | Same    |
| Regulation No.              | 21CFR880.6250                                                                                                                                                                                                                                                                                                                                                                                                                         | 21CFR880.6250                                                                                                                                                                                                                                                                                                                                                                                                                         | Same    |
| Class                       | I                                                                                                                                                                                                                                                                                                                                                                                                                                     | I                                                                                                                                                                                                                                                                                                                                                                                                                                     | Same    |
| Intended Use                | A patient examination<br>gloves is a disposable<br>device intended for medical<br>purpose that is worn on the<br>examiner's hand or<br>fingers to prevent<br>contamination between<br>patient and examiner. In<br>addition, these gloves were<br>tested for use with<br>chemotherapy drugs in<br>accordance with ASTM<br>D6978-05 Standard<br>Practice for Assessment of<br>Medical gloves to<br>Permeation by<br>Chemotherapy Drugs. | A patient examination<br>gloves is a disposable<br>device intended for medical<br>purpose that is worn on the<br>examiner's hand or<br>fingers to prevent<br>contamination between<br>patient and examiner. In<br>addition, these gloves were<br>tested for use with<br>chemotherapy drugs in<br>accordance with ASTM<br>D6978-05 Standard<br>Practice for Assessment of<br>Medical gloves to<br>Permeation by<br>Chemotherapy Drugs. | Same    |
| Powdered or Powered<br>free | Powdered free                                                                                                                                                                                                                                                                                                                                                                                                                         | Powdered free                                                                                                                                                                                                                                                                                                                                                                                                                         | Same    |
| Design Feature              | Ambidextrous                                                                                                                                                                                                                                                                                                                                                                                                                          | Ambidextrous                                                                                                                                                                                                                                                                                                                                                                                                                          | Same    |
| Sterility                   | Non-Sterile                                                                                                                                                                                                                                                                                                                                                                                                                           | Non-Sterile                                                                                                                                                                                                                                                                                                                                                                                                                           | Same    |
| Labeling Information        | Single-use indication,<br>powder free, device color,<br>device name, glove size and<br>quantity,Non-Sterile, a<br>statement of standard<br>ASTM D6978-05<br>compliance and a<br>summary of the testing<br>results.                                                                                                                                                                                                                    | Single-use indication,<br>powder free, device color,<br>device name, glove size<br>and quantity, Non-Sterile, a<br>statement of standard<br>ASTM D6978-05<br>compliance and a<br>summary of the testing<br>results.                                                                                                                                                                                                                   | Same    |
| Dimensions(mm)              | Length:<br>XS:≥220;<br>S/M/L/XL/XXL: ≥230;<br>Width:<br>XS: 70±10;<br>S: 80±10;<br>M: 95±10;<br>L: 110±10;<br>XL: 120±10                                                                                                                                                                                                                                                                                                              | Length:<br>XS/S/M/L/XL: ≥230;<br>Width:<br>XS:70±10;<br>S: 80±10;<br>M: 95±10;<br>L: 110±10;<br>XL: 120±10                                                                                                                                                                                                                                                                                                                            | Similar |

## Table1-General Comparison

{6}------------------------------------------------

| XL: 120±10;<br>XXL: 130±10 |                 |                                                                                                                                                                                  |                                                                                                                                                                                  |                     |            |      |
|----------------------------|-----------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|------------|------|
|                            | Thickness(mm)   | Finger: ≥0.05;<br>Palm: ≥0.05                                                                                                                                                    | Finger: ≥0.05;<br>Palm: ≥0.05                                                                                                                                                    |                     | Same       |      |
|                            | Colorant        | Blue                                                                                                                                                                             | White, Orange                                                                                                                                                                    |                     | Different  |      |
| Physical<br>Properties     | Before<br>Aging | Tensile<br>Strength                                                                                                                                                              | 14MPa, min                                                                                                                                                                       | Tensile Strength    | 14MPa, min | Same |
|                            |                 | Ultimate<br>Elongation                                                                                                                                                           | 500% min                                                                                                                                                                         | Ultimate Elongation | 500% min   | Same |
|                            | After<br>Aging  | Tensile<br>Strength                                                                                                                                                              | 14MPa, min                                                                                                                                                                       | Tensile Strength    | 14MPa, min | Same |
|                            |                 | Ultimate<br>Elongation                                                                                                                                                           | 400%min                                                                                                                                                                          | Ultimate Elongation | 400%min    | Same |
| Freedom from Holes         |                 | Be free from holes when<br>tested in accordance with<br>ASTMD5151 AQL=2.5                                                                                                        | Be free from holes when<br>tested in accordance with<br>ASTMD5151 AQL=2.5                                                                                                        |                     | Same       |      |
| Powder Content             |                 | 0.03 mg per glove, Meet<br>the requirements of<br>ASTM D6124                                                                                                                     | Meet the requirements of<br>ASTM D6124                                                                                                                                           |                     | Similar    |      |
| Biocompatibility           |                 | ISO 10993-10;<br>Under the conditions of<br>the study, not an irritant<br>or a sensitizer<br>ISO 10993-5<br>Under conditions of the<br>study, device extract is<br>not cytotoxic | ISO 10993-10;<br>Under the conditions of<br>the study, not an irritant<br>or a sensitizer<br>ISO 10993-5<br>Under conditions of the<br>study, device extract is<br>not cytotoxic |                     | Same       |      |
|                            |                 |                                                                                                                                                                                  |                                                                                                                                                                                  |                     |            |      |
|                            |                 | Carboplatin 10.0 mg/ml:<br>> 240 Minutes                                                                                                                                         | /                                                                                                                                                                                |                     | Different  |      |
|                            |                 | Carmustine (BCNU) 3.3<br>mg/ml: 26.5 Minutes                                                                                                                                     | Carmustine (BCNU) 3.3<br>mg/ml:<br>White:11.8 Minutes;<br>Orange:31.6Minutes                                                                                                     |                     | Similar    |      |
|                            |                 | /                                                                                                                                                                                | Cisplatin 1.0 mg/ml: >240<br>Minutes                                                                                                                                             |                     | Different  |      |
|                            |                 | Cyclophosphamide<br>(Cytoxan) 20.0 mg/ml:<br>>                                                                                                                                   | Cyclophosphamide<br>(Cytoxan)                                                                                                                                                    |                     | Same       |      |

{7}------------------------------------------------

| Chemotherapy Drugs<br>Tested with Minimum<br>Breakthrough<br>Detection Time as<br>Tested per ASTM D<br>6978 | 240 Minutes<br>/                                             | 20.0 mg/ml: >240 Minutes<br>Dacarbazine (DTIC) 10.0<br>mg/ml:<br>>240 Minutes | Different |
|-------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------|-------------------------------------------------------------------------------|-----------|
|                                                                                                             | Doxorubicin 2.0 mg/ml: ><br>240 Minutes                      | Doxorubicin Hydrochloride<br>2.0<br>mg/ml: >240 Minutes                       | Same      |
|                                                                                                             | Etoposide 20.0 mg/ml: ><br>240 Minutes                       | Etoposide (Toposar) 20.0<br>mg/ml: >240 Minutes                               | Same      |
|                                                                                                             | Fluorouracil 50.0 mg/ml: ><br>240 Minutes                    | Fluorouracil 50.0 mg/ml: ><br>240<br>Minutes                                  | Same      |
|                                                                                                             | Methotrexate<br>25.0<br>mg/ml(25,000 ppm): >240 /<br>Minutes | /                                                                             | Different |
|                                                                                                             | Paclitaxel 6.0 mg/ml: >240<br>Minutes                        | Paclitaxel (Taxol) 6.0<br>mg/ml:<br>>240 Minutes                              | Same      |
|                                                                                                             | ThioTepa 10.0 mg/ml: 48.4<br>Minutes                         | Thio-Tepa 10.0 mg/ml:<br>White:16.9 Minutes;<br>Orange: 72.5 Minutes          | Similar   |

## 8.0 Summary of Non-Clinical Testing

#### Biocompatibility Testing

The biocompatibility evaluation for Nitrile Patient Examination Gloves (Tested for Use with Chemotherapy Drugs) was conducted in accordance with the following standards:

ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests for Irritation And Skin Sensitization.

ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity

#### Performance Testing (Bench)

Physical performance qualities of the proposed device were evaluated per ASTM D6319-10, Standard Specification for Nitrile Examination Gloves for Medical Application.

Permeation testing was conducted to support the addition of the labeling claim:

{8}------------------------------------------------

Tested for use with chemotherapy drugs. In addition, the proposed device was tested according to ASTM D6978-05 (Reapproved 2019), Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs, in which minimum breakthrough times were determined for a wide range of chemotherapy drugs.

In summary, the performance testing of the subject device was conducted to adequately demonstrate the effectiveness of the device in accordance with the relevant test methods cited below:

- ASTM D6124-06 (Reapproved 2017), Standard Test Method for -Residual Powder on Medical Gloves
- -ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves.
- ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application.
- ASTM D6978-05 (Reapproved 2019), Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.

| Test<br>Method | Purpose                                          | Acceptance Criteria                            | Results                                        |
|----------------|--------------------------------------------------|------------------------------------------------|------------------------------------------------|
| ASTM<br>D6319  | Physical<br>Dimensions<br>Test                   | Length(mm):<br>XS: ≥220;                       | Length(mm):<br>> 240/Pass;                     |
|                |                                                  | S/M/L/XL/XXL: ≥230;                            | Width(mm):                                     |
|                |                                                  | Width(mm):<br>XS: 70±10;                       | XS: 80-83/Pass                                 |
|                |                                                  | S: 80±10;                                      | S: 86-90/Pass                                  |
|                |                                                  | M: 95±10;                                      | M: 97-98/ Pass                                 |
|                |                                                  | L: 110±10;                                     | L: 108-112/ Pass                               |
|                |                                                  | XL: 120±10;                                    | XL:115-120/ Pass                               |
|                |                                                  | XXL: 130±10                                    | XXL: 128-130/ Pass                             |
|                |                                                  | Finger: ≥0.05;                                 | Thickness (mm):                                |
|                |                                                  | Palm: ≥0.05                                    | Finger: 0.08-0.10/Pass<br>Palm: 0.08-0.11/Pass |
| ASTM<br>D5151  | Watertightness<br>Test<br>for<br>Detection<br>of | Meet the requirements of ASTM D5151<br>AQL 2.5 | 0/125/Pass                                     |

#### Table 2 - Summary of non-clinical performance testing

{9}------------------------------------------------

|                 | Holes                  |                                             |                        |        |                                                                       |
|-----------------|------------------------|---------------------------------------------|------------------------|--------|-----------------------------------------------------------------------|
| ASTM<br>D6124   | Powder<br>Content      | Meet the requirements of ASTM D6124 < 2.0mg |                        |        | 0.03mg/Pass;                                                          |
| ASTM<br>D412    | Physical<br>properties | Before<br>Aging                             | Tensile<br>Strength    | ≥14MPa | 15.1-16.9MPa/Pass;                                                    |
|                 |                        |                                             | Ultimate<br>Elongation | ≥500%  | 536-559%/Pass;                                                        |
|                 |                        | After<br>Aging                              | Tensile<br>Strength    | ≥14MPa | 16.0-18.0MPa/Pass;                                                    |
|                 |                        |                                             | Ultimate<br>Elongation | ≥400%  | 540-553%/Pass;                                                        |
| ISO<br>10993-5  | Cytotoxicity           | Non- In Vitro Cytotoxicity                  |                        |        | Under conditions of the study, device extract is not cytotoxic. /Pass |
| ISO<br>10993-10 | Irritation             | Non-irritating                              |                        |        | Under the conditions of the study, not an irritant. / Pass            |
| ISO<br>10993-10 | Sensitization          | Non-sensitizing                             |                        |        | Under conditions of the study, not a sensitizer. / Pass               |

## 9.0 _Summary of Clinical Testing

Clinical testing is not needed for this device.

#### 10.0 Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the subject device, Nitrile Patient Examination Gloves (Tested for Use with Chemotherapy Drugs) is as safe, as effective, and performs as well as or better than the legally marketed predicated device under K190860.

---

**Source:** [https://fda-staging.innolitics.com/device/K213048](https://fda-staging.innolitics.com/device/K213048)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).