K213037 · Digital Diagnostics, Inc. · PIB · Jun 17, 2022 · Ophthalmic
Device Facts
Record ID
K213037
Device Name
IDx-DR v2.3
Applicant
Digital Diagnostics, Inc.
Product Code
PIB · Ophthalmic
Decision Date
Jun 17, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.1100
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K213037 · Jun 17, 2022
IDx-DR v2.3
Digital Diagnostics, Inc.
Retrospective analysis of pivotal clinical study data (Abràmoff et al. 2018)
The sponsor performed a retrospective analysis of 892 participants from a previously conducted pivotal study to compare the performance of the modified IDx-DR v2.3 algorithm against the predicate IDx-DR v2.0.
Pivotal study (Abràmoff et al. Digital Medicine 2018;1:39): 850 participants
—
Indications for Use
IDx-DR is indicated for use by healthcare providers to automatically detect more than mild diabetic retimopathy (mtmDR) in adults diagnosed with diabetes who have not been previously diagnosed with diabetic retinopathy. IDx-DR is indicated for use with the Topcon NW400.
Device Story
IDx-DR is an autonomous AI-based system for detecting more than mild diabetic retinopathy (mtmDR). Input: two macula- and disc-centered color fundus images per eye (45° field of view) captured via Topcon NW400 fundus camera. Operation: IDx-DR Client (local computer) transmits images to IDx-DR Service (secure datacenter); IDx-DR Analysis System processes images to determine exam quality and presence/absence of mtmDR. Output: results (mtmDR detected/not detected/insufficient quality) displayed on IDx-DR Client. Used in clinical settings by healthcare providers. Benefits: automated screening for diabetic retinopathy in patients without prior diagnosis, facilitating timely referral to eye care professionals.
Clinical Evidence
Retrospective study using 850 evaluable participants from the pivotal study (Abràmoff et al. 2018). Primary endpoints: sensitivity, specificity, and diagnosability. Final submission sensitivity: 87.69% (v2.3) vs 87.37% (v2.0); specificity: 90.07% (v2.3) vs 89.53% (v2.0). Diagnosability of final submission: 95.18% (v2.3) vs 96.35% (v2.0). Results demonstrate substantial equivalence.
Technological Characteristics
AI-based software as a medical device (SaMD). Inputs: 45° color fundus images (min 22 pixels/degree). Architecture: client-server model; IDx-DR Client (local), IDx-DR Service (web server/database), IDx-DR Analysis System (server-side). Connectivity: networked to secure datacenter. Software: v2.3.0 (Analysis), v1.2.0 (Service), v3.5.0 (Client).
Indications for Use
Indicated for adults diagnosed with diabetes without prior diabetic retinopathy diagnosis. Used by healthcare providers to automatically detect more than mild diabetic retinopathy (mtmDR).
Regulatory Classification
Identification
A retinal diagnostic software device is a prescription software device that incorporates an adaptive algorithm to evaluate ophthalmic images for diagnostic screening to identify retinal diseases or conditions.
Special Controls
In combination with the general controls of the FD&C Act, the retinal diagnostic software device is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) Software verification and validation documentation, based on a comprehensive hazard analysis, must fulfill the following:
(i) Software documentation must provide a full characterization of technical parameters of the software, including algorithm(s).
(ii) Software documentation must describe the expected impact of applicable image acquisition hardware characteristics on performance and associated minimum specifications.
(iii) Software documentation must include a cybersecurity vulnerability and management process to assure software functionality.
(iv) Software documentation must include mitigation measures to manage failure of any subsystem components with respect to incorrect patient reports and operator failures.
(2) Clinical performance data supporting the indications for use must be provided, including the following:
(i) Clinical performance testing must evaluate sensitivity, specificity, positive predictive value, and negative predictive value for each endpoint reported for the indicated disease or condition across the range of available device outcomes.
(ii) Clinical performance testing must evaluate performance under anticipated conditions of use.
(iii) Statistical methods must include the following:
(A) Where multiple samples from the same patient are used, statistical analysis must not assume statistical independence without adequate justification.
(B) Statistical analysis must provide confidence intervals for each performance metric.
(iv) Clinical data must evaluate the variability in output performance due to both the user and the image acquisition device used.
(3) A training program with instructions on how to acquire and process quality images must be provided.
(4) Human factors validation testing that evaluates the effect of the training program on user performance must be provided.
(5) A protocol must be developed that describes the level of change in device technical specifications that could significantly affect the safety or effectiveness of the device.
(6) Labeling must include:
(i) Instructions for use, including a description of how to obtain quality images and how device performance is affected by user interaction and user training;
(ii) The type of imaging data used, what the device outputs to the user, and whether the output is qualitative or quantitative;
(iii) Warnings regarding image acquisition factors that affect image quality;
(iv) Warnings regarding interpretation of the provided outcomes, including:
(A) A warning that the device is not to be used to screen for the presence of diseases or conditions beyond its indicated uses;
(B) A warning that the device provides a screening diagnosis only and that it is critical that the patient be advised to receive followup care; and
(C) A warning that the device does not treat the screened disease;
(v) A summary of the clinical performance of the device for each output, with confidence intervals; and
(vi) A summary of the clinical performance testing conducted with the device, including a description of the patient population and clinical environment under which it was evaluated.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the date June 17, 2022. The text is written in a clear, sans-serif font. The date is formatted with the month, day, and year. The text is centered and takes up most of the image.
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Digital Diagnostics Inc. % Kelliann Payne Partner Hogan Lovells US LLP 1735 Market St., Floor 23 Philadelphia, Pennsylvania 19103
Re: K213037
Trade/Device Name: IDx-DR v2.3 Regulation Number: 21 CFR 886.1100 Regulation Name: Retinal Diagnostic Software Device Regulatory Class: Class II Product Code: PIB Dated: September 21, 2021 Received: May 17, 2022
Dear Kelliann Payne:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Elvin Ng Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213037
Device Name IDx-DR v2.3
Indications for Use (Describe)
IDx-DR is indicated for use by healthcare providers to automatically detect more than mild diabetic retimopathy (mtmDR) in adults diagnosed with diabetes who have not been previously diagnosed with diabetic retinopathy. IDx-DR is indicated for use with the Topcon NW400.
| <b>Type of Use (Select one or both, as applicable)</b> |
|--------------------------------------------------------|
|--------------------------------------------------------|
| <div style="display:inline-block"><span style="font-size: 20px; vertical-align: middle;">☒</span> Research Use (21 CFR 201.320(b))</div> | <div style="display:inline-block"><span style="font-size: 20px; vertical-align: middle;">☐</span> Over-The-Counter Use (21 CFR 201.66)</div> |
|------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary K213037
#### I. Submitter
Digital Diagnostics Inc. 2300 Oakdale Blvd. Coralville, IA 52241 Phone: 319-248-5620
Contact Person: Ashley Miller Date Prepared: June 14, 2022
#### II. Device
Name of Device: IDx-DR v2.3 Common or Usual Name: Diabetic Retinopathy Detection Device Classification Name: Retinal diagnostic software device Regulatory Class: II Regulation: 21 CFR 886.1100 Product Code: PIB
#### III. Device Description
The IDx-DR device is an autonomous, artificial intelligence (AI)-based system for the automated detection of more than mild diabetic retinopathy (mtmDR). It consists of several component parts (see Figure 1 below).
The component parts of IDx-DR are summarized as follows:
- IDx-DR Analysis: Software that analyzes the patient's images and determines ● exam quality and the presence/absence of mtmDR.
- . IDx-DR Client: A software application component running on a computer, usually connected to the fundus camera, at the user site. Using this software, the user can transfer images to IDx-DR Analysis via IDx-DR Service and receive results back.
- . IDx-DR Service: IDx-DR Service comprises a general exam analysis service delivery software package. IDx-DR Service contains a webserver front-end that securely handles incoming requests, a database that stores user information, and a logging system that records information about each transaction through IDx-DR Service. IDx-DR Service is also primarily responsible for device cybersecurity.
The Topcon NW400 fundus camera is attached to a computer, where IDx-DR Client is installed. Guided by the IDx-DR Client, end-users acquire two fundus images per eye to
{4}------------------------------------------------
be dispatched to IDx-DR Service. IDx-DR Service is installed on a server hosted at a secure datacenter. From IDx-DR Service, images are transferred to IDx-DR Analysis System. No information other than the fundus images is required to perform the analysis. IDx-DR Analysis System, which runs on dedicated servers hosted in the same secure datacenter as IDx-DR Service, processes the fundus images and returns information on the exam quality and the presence or absence of more than mild diabetic retinopathy (mtmDR) to IDx-DR Service. IDx-DR Service then transports the results to the IDx-DR Client that displays them to the user.
Image /page/4/Figure/2 description: This image shows a diagram of the IDx secure server system. The diagram shows the flow of data from the patient and camera to the IDx-DR client on the customer's computer. The exam images are then sent to the IDx web service on the IDx secure servers, where they are analyzed by the IDx-DR analysis software. The results are then sent back to the IDx-DR client on the customer's computer.
Figure 1: IDx-DR Components
#### IV. Indications for Use
IDx-DR is indicated for use by healthcare providers to automatically detect more than mild diabetic retinopathy (mtmDR) in adults diagnosed with diabetes who have not been previously diagnosed with diabetic retinopathy. IDx-DR is indicated for use with the Topcon NW400.
#### V. Predicate Device
IDx-DR, Diabetic Retinopathy Detection Device, K203629 This predicate has not been subject to a design-related recall.
No reference devices were used in this submission.
#### VI. Purpose of Submission
The purpose of this 510(k) submission is to modify the following: the image quality classifier of the IDx-DR Analysis System, from version 2.1.1 to 2.3.0; update IDx-DR Service from version 1.1.2 to 1.2.0; and update IDx-DR Client from v3.3.0 to v3.5.0.
{5}------------------------------------------------
#### VII. Comparison of Technological Characteristics with the Predicate Device
The main technological principle for both the subject and predicate devices is Al-based technology to analyze specific pathologic features from fundus retinal images. The subject device has the same intended use and indications for use as those cleared under K203629.
The major technological differences that exist between each component of the subject and predicate devices are described below.
# IDx-DR Service
- Updated to look up the IDx-DR Analysis System version and sends it to IDx-DR Client for display on the user interface.
# IDx-DR Client
- Updated to provide clear language of "exam completed" after repeated imaging . attempts have resulted in an insufficient image quality output. The result output and report will remain unchanged, indicating exam quality was insufficient and a diagnostic result is not provided.
- . Updated to visually communicate image quality feedback by adding a green checkmark or a red "X" to images on the submission feedback screen.
## IDx-DR Analysis System
- . The image quality classifier of the subject device was replaced with a new image quality classifier.
- Updated to improve the speed of the device. .
Table 1 provides a comparison between the technical characteristics and indications for use of the subject and predicate devices.
| | Subject Device<br>IDx-DR v2.3 | Predicate Device<br>IDx-DR v2.0, K203629 | Discussion |
|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|
| Component<br>Software<br>Versions | IDx-DR Client v3.5.0<br>IDx-DR Analysis System v2.3.0<br>IDx-DR Service v1.2.0 | IDx-DR Client v3.2.0<br>IDx-DR Analysis System v2.1.1<br>IDx-DR Service v1.1.2 | See above for the<br>major<br>technological<br>differences<br>between each<br>component of the<br>subject and<br>predicate device. |
| Technological<br>Principle | Artificial intelligence<br>software as a medical device. | Artificial intelligence<br>software as a medical device. | Equivalent |
| | Subject Device<br>IDx-DR v2.3 | Predicate Device<br>IDx-DR v2.0, K203629 | Discussion |
| Indications for Use | For use by healthcare<br>providers to automatically<br>detect more than mild diabetic<br>retinopathy in adults<br>diagnosed with diabetes who<br>have not been previously<br>diagnosed with diabetic<br>retinopathy. | For use by healthcare<br>providers to automatically<br>detect more than mild diabetic<br>retinopathy in adults<br>diagnosed with diabetes who<br>have not been previously<br>diagnosed with diabetic<br>retinopathy. | Equivalent |
| Indicated<br>Camera | Topcon NW400 fundus<br>camera | Topcon NW400 fundus<br>camera | Equivalent |
| Inputs | Macula- and disc-centered<br>color fundus images with 45°<br>field of view, 2 per eye. | Macula- and disc-centered<br>color fundus images with 45°<br>field of view, 2 per eye. | Equivalent |
| Outputs | Detection of diabetic<br>retinopathy and referral<br>decision:<br>• mtmDR detected: Refer to<br>an eye care professional<br>• mtmDR not detected:<br>Rescreen in 12 months<br>• Insufficient image quality | Detection of diabetic<br>retinopathy and referral<br>decision:<br>• mtmDR detected: Refer to<br>an eye care professional<br>• mtmDR not detected:<br>Rescreen in 12 months<br>• Insufficient image quality | Equivalent |
| Architecture | User facing client software<br>transfers images to and<br>receives results from analysis<br>software through a web<br>server. | User facing client software<br>transfers images to and<br>receives results from analysis<br>software through a web<br>server. | Equivalent |
| Workflow | The graphical user interface<br>includes on-screen prompts to<br>guide the user through the<br>image acquisition workflow<br>one image at a time and<br>submission of the exam | The graphical user interface<br>includes on-screen prompts to<br>guide the user through the<br>image acquisition workflow<br>one image at a time and<br>submission of the exam. | Equivalent |
## Table 1: Comparison of the Subject and Predicate Device
{6}------------------------------------------------
## VIII. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
# A. Summary of Non-clinical Studies
IDx-DR was identified as having a major level of concern as defined in the FDA guidance document Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. The software documentation includes:
{7}------------------------------------------------
- 1. Software/Firmware Description
- 2. Device Hazard Analysis
- 3. Software Requirement Specifications
- 4. Architecture Design Chart
- 5. Software Design Specifications
- 6. Traceability
- 7. Software Development Environment Description
- 8. Verification and Validation Documentation
- 9. Revision Level History
- 10. Unresolved Anomalies
- 11. Cybersecurity
A comprehensive risk analysis was performed on IDx-DR with identification and detailed description of the hazards, their causes and severity, as well as acceptable methods for control of the identified hazards. A description of acceptable verification and validation activities, at the unit, integration, and system level, including test protocols with pass/fail criteria and a report of the results, was provided. The expected impact of various hardware features on performance was assessed and minimum specifications for acceptable images for analysis were specified.
The cybersecurity considerations of data confidentiality, data integrity, data availability, denial of service attacks, and malware were adequately addressed utilizing platform controls, application controls, and procedure controls, and evidence was provided for the intended performance of the controls. Risks related to failure of various software components and their potential impact on patient reports and operator failures were also adequately addressed in the risk analysis. This software documentation information provided sufficient evidence of safe and effective software performance.
A full characterization of the technical parameters of all of the components of the software, including a description of the algorithms that analyzes the patient's images to determine exam quality and the diagnostic screening of diabetic retinopathy, has been provided. IDx-DR requires one optic disc-centered image and one macula centered image from a fundus camera with at least 22 pixels per degree on the retina. So, a 1000 pixel field of view diameter for a 45 degree field of view image.
The IDx-DR artificial intelligence device design has the ability to perform analysis on the specific disease features that are important to a retina specialist for diagnostic screening of diabetic retinopathy. Future algorithm improvements will be made under a consistent medically relevant framework. A protocol was provided to mitigate the risk of algorithm changes leading to changes in the device technical specifications, which would lead to changes in false positive or false negative results. These changes could significantly affect clinical functionality or performance specifications directly associated with the intended use of the device. The protocol specifies the level of change in device
{8}------------------------------------------------
specifications that could significantly affect the safety or effectiveness of the device, triggering the requirement for a new 510(k) premarket notification submission before commercial introduction. The protocol incorporates a risk management approach and other approaches provided in the FDA guidance document Deciding When to Submit a 510(k) for a Software Change to an Existing Device: Guidance for Industry and FDA Staff in development, validation, and execution of the device changes.
### Usability
Usability validation testing was performed under simulated-use to assess the user interface (IDx-DR Client) of the subject device. The testing was performed in an environment equivalent to the intended use environment of IDx-DR with subjects that had no prior experience using the IDx-DR Client. The critical task for the IDx-DR system is the ability to capture four images of sufficient quality. The purpose of the usability validation test plan was to demonstrate that the intended image capture workflow and training methodology can successfully be used by the intended operators to capture four retinal images. The results of the usability validation study indicated that no existing critical tasks were impacted by the modification and no new critical tasks were introduced, and demonstrated that previously untrained camera operators could capture four retinal images of sufficient quality following the imaging protocol and using the indicated camera system and standardized training and operating materials.
### B. Summary of Clinical Performance Testing
A retrospective study was conducted to validate the clinical performance of the modified IDx-DR ("IDx-DR v2.3). Data previously collected from the pivotal study of the predicate ("IDx-DR v2.0"; Abràmoff et al. Digital Medicine 2018;1:39) was analyzed to evaluate the performance of the upgraded Analysis component. The three co-primary endpoints are sensitivity, specificity, and "diagnosability" (the proportion of evaluated participants for whom IDx-DR returns a diagnostic result). The secondary endpoints are positive predictive value (PPV) and negative predictive value (NPV).
Data from the 892 participants evaluated during the pivotal study were used to evaluate the modified algorithm; of these, images from 850 participants were available for analysis and diagnosable by the clinical reference standard, thus were evaluable for performance. Of the "first submission" images (i.e., first images taken and without pharmacologic pupil dilation), IDx-DR v2.3 was able to analyze images from 552 of the 850 participants (64.9%) and IDx-DR v2.0 was able to analyze images from 533 of the 850 subjects (62.7%). Of the "final submission" images (after following the as-needed pharmacologic pupil dilation image acquisition protocol), IDx-DR v2.3 was able to analyze images from 809 of the 850 participants (95.2%) and IDx-DR v2.0 was able to analyze images from 819 of the 850 participants (96.4%).
{9}------------------------------------------------
Table 2 presents the diagnosability for participants who were diagnosable by both the subject and predicate devices based on images from the "first" and "final submission" for each participant from the pivotal study.
| Table 2: Diagnosability Results for the Subject and Predicate Devices Based on First | | |
|--------------------------------------------------------------------------------------|---------------------------------------|---------------------------------|
| and Final Submissions | | |
| Characteristic | Pivotal Study Device<br>(IDx-DR v2.0) | Subject Device<br>(IDx-DR v2.3) |
| Diagnosability of first<br>submission images<br>Point Estimate | 62.71% (533/850) | 64.94% (552/850) |
| 95% Confidence Interval1 | (59.40%, 65.89%) | (61.67%, 68.08%) |
| Diagnosability of final<br>submission images<br>Point Estimate | 96.35% (819/850) | 95.18% (809/850) |
| 95% Confidence Interval2 | (94.86%, 97.51%) | (93.51%, 96.52%) |
1 Calculated using the modified Wald method
2 Calculated using an exact binomial model
Table 3 presents the sensitivity and specificity for participants who were diagnosable by both the subject and predicate devices based on images from the "final submission" for each participant from the pivotal study.
## Table 3: Performance Results for the Subject and Predicate Devices Based on Final Submissions
| Characteristic | Pivotal Study Device<br>(IDx-DR v2.0) | Subject Device<br>(IDx-DR v2.3) |
|--------------------------------|---------------------------------------|---------------------------------|
| Sensitivity*<br>Point Estimate | 87.37% (173/198) | 87.69% (171/195) |
| 95% Confidence Interval¹ | (81.93%, 91.66%) | (82.24%, 91.95%) |
| Specificity*<br>Point Estimate | 89.53% (556/621) | 90.07% (553/614) |
| 95% Confidence Interval¹ | (86.85%, 91.83%) | (87.42%, 92.32%) |
¹ Calculated using an exact binomial model
*Excludes exam quality insufficient images, 31 for IDx-DR v2.0, 41 for IDx-DR v2.3
Table 4 presents the PPV and NPV for participants who were diagnosable by both the subject and predicate devices based on images from the "final submission" for each subject from the pivotal study.
{10}------------------------------------------------
| Characteristic | Pivotal Study Device<br>(IDx-DR v2.0) | Subject Device<br>(IDx-DR v2.3) |
|---------------------------------------------|---------------------------------------|---------------------------------|
| Positive Predictive Value<br>Point Estimate | 72.69% (173/238) | 73.71% (171/232) |
| 95% Confidence Interval1 | (66.56%, 78.25%) | (67.55%, 79.25%) |
| Negative Predictive Value<br>Point Estimate | 95.70% (556/581) | 95.84% (553/577) |
| 95% Confidence Interval1 | (93.71%, 97.20%) | (93.87%, 97.32%) |
### Table 4: Secondary Performance for the Subject and Predicate Devices Based on Final Submissions
1 Calculated using an exact binomial model
Additional analyses were performed to include submissions that were non-diagnosable by the respective version of IDx-DR but were diagnosable by the reference standard.
Table 5 presents the worst-case imputations for IDx-DR v2.0 and IDx-DR v2.3 based on images from the final submission for each participant from the pivotal study, wherein the non-diagnosable submissions are assumed to have IDx-DR results disagreeing with the reference standard.
| Characteristic | Pivotal Study Device<br>(IDx-DR v2.0) | Subject Device<br>(IDx-DR v2.3) |
|---------------------------|---------------------------------------|---------------------------------|
| Sensitivity | | |
| Point Estimate | 85.22% (173/203) | 84.24% (171/203) |
| 95% Confidence Interval1 | (79.58%, 89.80%) | (78.48%, 88.96%) |
| Specificity | | |
| Point Estimate | 85.94% (556/647) | 85.47% (553/647) |
| 95% Confidence Interval1 | (83.02%, 88.52%) | (82.52%, 88.10%) |
| Positive Predictive Value | | |
| Point Estimate | 65.53% (173/264) | 64.53% (171/265) |
| 95% Confidence Interval1 | (59.46%, 71.25%) | (58.44%, 70.29%) |
| Negative Predictive Value | | |
| Point Estimate | 94.88% (556/586) | 94.53% (553/585) |
| 95% Confidence Interval1 | (92.77%, 96.52%) | (92.37%, 96.23%) |
Table 5: Worst-Case Final Submission Performance by IDx-DR Device Version
1 Calculated using an exact binomial model
Table 6 presents the best-case imputations for IDx-DR v2.0 and IDx-DR v2.3 based on images from the final submission for each participant from the pivotal study, wherein the non-diagnosable submissions are assumed to have IDx-DR results agreeing with the reference standard.
{11}------------------------------------------------
| Characteristic | Pivotal Trial Device<br>(IDx-DR v2.0) | Subject Device<br>(IDx-DR v2.3) |
|---------------------------|---------------------------------------|---------------------------------|
| Sensitivity | | |
| Point Estimate | 87.68% (178/203) | 88.18% (179/203) |
| 95% Confidence Interval1 | (82.36%, 91.87%) | (82.92%, 92.28%) |
| Specificity | | |
| Point Estimate | 89.95% (582/647) | 90.57% (586/647) |
| 95% Confidence Interval1 | (87.37%, 92.16%) | (88.05%, 92.71%) |
| Positive Predictive Value | | |
| Point Estimate | 73.25% (178/243) | 74.58% (179/240) |
| 95% Confidence Interval1 | (67.22%, 78.71%) | (68.58%, 79.97%) |
| Negative Predictive Value | | |
| Point Estimate | 95.88% (582/607) | 96.07% (586/610) |
| 95% Confidence Interval1 | (93.98%, 97.32%) | (94.20%, 97.46%) |
Table 6: Best-Case Final Submission Performance by IDx-DR Device Version
¹Calculated using an exact binomial model
The results of the clinical study support a determination of substantial equivalence between IDx-DR v2.3 and IDx-DR v2.0.
#### IX. Conclusions
The modified IDx-DR device is substantially equivalent to the predicate IDx-DR device cleared under K203629. The modifications do not raise new questions of safety and effectiveness of the device. The subject and predicate devices have the same indications for use, technological characteristics, and performance specifications.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.