← Product Code [QNQ](/productcode/QNQ) · K213010

# FEELject LDV (Low dead volume) syringe (K213010)

_Feeltech Co., Ltd. · QNQ · May 31, 2023 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K213010

## Device Facts

- **Applicant:** Feeltech Co., Ltd.
- **Product Code:** [QNQ](/productcode/QNQ.md)
- **Decision Date:** May 31, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5860
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

FEELject LDV(Low dead volume) Syringe is intended for use to inject fluid into or withdraw fluids from the parts of body below the surface of the skin.

## Device Story

FEELject LDV is a single-use, manually operated disposable syringe with a permanently attached hypodermic needle. Device consists of a calibrated plastic barrel, plunger, rubber piston, and needle. Operates by manual movement of the plunger within the barrel to draw in or expel fluids through the needle. Designed as a low dead volume (LDV) syringe (≤ 0.03 mL) to minimize fluid loss. Used by healthcare professionals or patients in clinical or home settings for subcutaneous, intramuscular, or intravenous fluid administration/withdrawal. Output is the delivery or extraction of fluid. Benefits include reduced medication waste due to low dead space design.

## Clinical Evidence

Bench testing only. No clinical data. Testing included biocompatibility (ISO 10993 series: cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, pyrogenicity, hemolysis), sterility (ISO 11737-2), EO residuals (ISO 10993-7), and mechanical performance (ISO 7886-1, ISO 7864, ISO 9626, USP 788). All tests passed.

## Technological Characteristics

Materials: Polypropylene (barrel, plunger, hub, cap), Rubber (piston), SUS304 (needle). Manual operation. Single-use. E.O. gas sterilization (SAL 10^-6). Low dead volume (≤ 0.03 mL). 1 mL capacity. Complies with ISO 7886-1, ISO 7864, ISO 9626, and ISO 10993 standards.

## Regulatory Identification

A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.

## Predicate Devices

- Syringe ([K192551](/device/K192551.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 31, 2023

Feeltech Co., Ltd. % Peter Chung President Plus Global 300. Atwood Pittsburgh, Pennsylvania 15213

Re: K213010

Trade/Device Name: FEELject LDV (Low dead volume) syringe Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: QNQ, FMI Dated: April 28, 2023 Received: May 1, 2023

Dear Peter Chung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Alan Guth

CAPT Alan M. Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K213010

Device Name FEELject LDV(Low dead volume) Syringe

Indications for Use (Describe)

FEELject LDV(Low dead volume) Syringe is intended for use to inject fluid into or withdraw fluids from the parts of body below the surface of the skin.

Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

### K213010 - 510(k) Summarv

#### [as required by 21 CFR 807.92(c)]

#### 1. Applicant

- Company: Feeltech Co., Ltd. 1)
- Address: 3, 4 Floor, Standard Factory 2-dong, 15, Jayumuyeok2-gil, Gunsan-si, Jeollabuk-do, Korea 2)
- Tel : +82-63-468-6626~8 3)
- Fax : +82-63-468-6623 4)
- Prepared date : May 18, 2023 ર)
- 6) Contact person : Peter Chung, 412-512-8802
- 7) Contact person address : 300 Atwood Street, Pittsburgh, PA, 15213, USA

#### 2. Subject Device Information

- 1) Trade name: FEELject LDV(Low dead volume) Syringe
- Common name : Disposal syringe with needle 2)
- 3) Classification name : Piston syringe
- 4) Product code : QNQ, FMI
- Regulation number : 21 CFR 880.5860 ર)
- Class of device : Class II 6)
- Panel : General hospital 7)

# 3. The legally marketed device to which we are claiming equivalence

Syringe, K192551, Jiangsu Caina Medical Co., Ltd.

#### 4. Device description

A Disposable Syringe with Needle is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a hypodermic single lumen needle which is permanently attached. The device is used to inject fluids into, or withdraw fluids from, the body. It is made of plastic and silicone materials that allows for smooth plunger movement, and manually operated. This is a single-use device. This product is packed by sterile paper and sterilized by E.O gas.

#### 5. Intended Use:

This product is intended for use to inject fluids into or withdraw fluids from parts of the body below the surface of the skin.

#### 6. Comparison Table

| Table 1. Comparison table of proposed to predicate device |              |                                                                                                                               |                                                                                                                              |                 |
|-----------------------------------------------------------|--------------|-------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|-----------------|
|                                                           | Manufacturer | FeelTech Co., Ltd.                                                                                                            | Jiangsu Caina Medical Co., Ltd.                                                                                              | Remark          |
|                                                           | Item         | Proposed device                                                                                                               | Predicate device                                                                                                             |                 |
|                                                           | 510(K) No.   | K213010                                                                                                                       | K192551                                                                                                                      | N/A             |
|                                                           | Product      | FEELject LDV Syringe                                                                                                          | Syringe with permanently<br>attached needle                                                                                  |                 |
|                                                           | Intended use | This product is intended for use to inject<br>fluids into or withdraw from parts of the<br>body below the surface of the skin | This device is intended for use to inject<br>fluids into or withdraw from parts of the<br>body below the surface of the skin | Same            |
|                                                           | Components   | Barrel, Plunger, Plunger rod<br>Needle, needle cap                                                                            | Barrel, Plunger, Plunger rod<br>Needle, needle cap                                                                           | Same            |
|                                                           | Nozzle type  | Permanently attached                                                                                                          | Permanently attached<br>Luer-slip/Luer-lock                                                                                  | Different<br>#1 |
| Material                                                  | Barrel       | Polypropylene (PP)                                                                                                            | Polypropylene (PP)                                                                                                           |                 |
|                                                           | Plunger      | Polypropylene (PP)                                                                                                            | Polypropylene (PP)                                                                                                           |                 |
|                                                           | Piston       | Rubber                                                                                                                        | Rubber                                                                                                                       | Same            |
|                                                           | Needle hub   | Polypropylene (PP)                                                                                                            | Polypropylene (PP)                                                                                                           |                 |
|                                                           | Needle       | SUS304                                                                                                                        | SUS304                                                                                                                       |                 |
|                                                           | Needle cap   | Polypropylene (PP)                                                                                                            | Polypropylene (PP)                                                                                                           |                 |
|                                                           | Capacity     | 1ml                                                                                                                           | 0.3, 0.5, 1ml                                                                                                                | Different<br>#2 |
|                                                           | Dead space   | Low dead volume (≤ 0.03 mL)                                                                                                   | Unknown                                                                                                                      | Different<br>#3 |

## Table 1. Comparison table of proposed to predicate device

{4}------------------------------------------------

| Principle of<br>operation   | Manual                                                                                                                                                                                                                        | Manual                                                                                                                                                                                                                        | Same      |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Syringe<br>Performances¹    | Complies with ISO 7886-1 : 2017 Sterile<br>hypodermic syringes for single use - Part<br>1 : Syringes for manual use                                                                                                           | Complies with ISO 7886-1 : 2017 Sterile<br>hypodermic syringes for single use - Part<br>1 : Syringes for manual use                                                                                                           | Same      |
| Needle Length               | 25.0, 25.4, 38.1mm                                                                                                                                                                                                            | 8, 10, 13 16, 20, 25mm                                                                                                                                                                                                        |           |
| Gauge                       | 23, 25G                                                                                                                                                                                                                       | 21, 23, 25, 26, 27, 28, 29, 30, 31G                                                                                                                                                                                           | Different |
| Needle Tip<br>configuration | Bevel (11°±2°)                                                                                                                                                                                                                | Bevel (11°±2°, 15°±2°)                                                                                                                                                                                                        | #4        |
| Needle wall type            | TW                                                                                                                                                                                                                            | RW, TW                                                                                                                                                                                                                        |           |
| Needle<br>performance¹      | ISO 7864 : 2016 Sterile hypodermic needles<br>for single use - Requirements and test<br>methods<br>ISO 9626 : 2016 Stainless steel needle<br>tubing for the manufacture of medical<br>devices - Requirements and test methods | ISO 7864 : 2016 Sterile hypodermic needles<br>for single use - Requirements and test<br>methods<br>ISO 9626 : 2016 Stainless steel needle<br>tubing for the manufacture of medical<br>devices - Requirements and test methods | Same      |
|                             | Conforms to the                                                                                                                                                                                                               | Conforms to the                                                                                                                                                                                                               |           |
| Bio-<br>Compatibility¹      | requirements of ISO 10993<br>series standards.<br>Cytotoxicity<br>Acute systemic toxicity<br>Pyrogenicity<br>Sensitization<br>Irritation<br>Hemolysis<br>Intracutaneous reactivity<br>Bacterial Endotoxins                    | requirements of ISO 10993<br>series standards.<br>Cytotoxicity<br>Acute systemic toxicity<br>Pyrogenicity<br>Sensitization<br>Irritation<br>Hemolysis<br>Intracutaneous reactivity<br>Bacterial Endotoxins                    | Same      |
| Principle of<br>operation   | The plunger of syringe can be pulled and<br>pushed along inside the barrel, allowing the<br>syringe to take in and expel the fluids<br>through the connector to the patient.                                                  | The plunger of syringe can be pulled and<br>pushed along inside the barrel, allowing the<br>syringe to take in and expel the fluids<br>through the connector to the patient.                                                  | Same      |
| Sterilization<br>method     | E.O Gas sterilization                                                                                                                                                                                                         | E.O Gas Sterilization                                                                                                                                                                                                         | Same      |
| SAL                         | 10-6                                                                                                                                                                                                                          | 10-6                                                                                                                                                                                                                          | Same      |
| Endotoxin limit¹            | 20 EU / Device                                                                                                                                                                                                                | 20 EU / Device                                                                                                                                                                                                                | Same      |
| Label/Labeling              | Complied with 21 CFR part 801                                                                                                                                                                                                 | Complied with 21 CFR part 801                                                                                                                                                                                                 | Same      |

* Note 1: Refer to the following section (8. Performance data) for the detail descriptions of conducted testing result.

#### 7. Equivalence Discussion

#### 1) Different #1 – Syringe nozzle type

Subject device only has permanently attached nozzle type. However, predicate device has not only permanently attached nozzle, but also luer-lock and luer-slip. While this configuration is different from each other, the subject device's nozzle configuration is included to the predicate device. Therefore, the differences on nozzle type does not raise new questions of safety and effectiveness.

- Different #2 Capacity 2)
Subject device has only one capacity configuration. (1 ml) Although predicate device has 3 configuration for the capacity (0.3, 0.5, 1 ml) this difference does not affect substantial equivalence because it is covered within the predicate submission.

#### 3) Different #3 - Dead space

Subject device is intended to provide low dead volume (LDV), minimizing the loss of fluid left inside the syringe. Although predicate device does not have this aspect, subject device is better than its requirements for the dead space. In accordance with product code 'QNQ', this device has proved with related performance testing that it conforms to the FDA's requirements. Therefore, the differences between predicate and subject device does not affect the device's effectiveness and safety.

- Different #4 Needle configurations (Needle length, gauge, tip configuration, wall type) The needle configurations (Needle length, gauge, tip configuration, wall type) for proposed devices is different from the predicate devices. However, the predicate device covers the range of needle dimensions.

{5}------------------------------------------------

Therefore, the differences on needle length and gauge does not affect safety and effectiveness.

#### 8. Performance data evaluation

1) Performance testing - Bench

Bench tests were performed. Bench testing included biocompatibility, mechanical testing, sterility testing including EO residues. The tests demonstrated that the device performs in a substantially equivalent manner to the predicate device. The following bench testing is performed to demonstrate the functionality is substantially equivalent.

| Requirement - Test (ISO 7886-1)                          | Result |
|----------------------------------------------------------|--------|
| Limits for extractable metals                            | Pass   |
| General                                                  | Pass   |
| Limits for acidity or alkalinity                         | Pass   |
| Conical fitting                                          | Pass   |
| Position of nozzle on end of barrel                      | Pass   |
| Tolerance on graduated capacity                          | Pass   |
| Scale                                                    | Pass   |
| Numbering of scales                                      | Pass   |
| Overall length of scale to nominal capacity line         | Pass   |
| Position of scale                                        | Pass   |
| Barrel flanges                                           | Pass   |
| Plunger stopper/plunger assembly                         | Pass   |
| Dead space                                               | Pass   |
| Freedom from air and liquid leakage past plunger stopper | Pass   |
| Force to operate the piston                              | Pass   |
| Fit of plunger stopper/plunger in barrel                 | Pass   |
| Unit packaging and self-contained syringe units          | Pass   |
| Quantity of silicone oil                                 | Pass   |

| Requirement - Test (ISO 7864)             | Result |
|-------------------------------------------|--------|
| Cleanliness                               | Pass   |
| Tolerances on length                      | Pass   |
| Needle Point, needles with sharp tip only | Pass   |
| Bond between hub and needle tube          | Pass   |
| Patency lumen                             | Pass   |

| Requirement – Test (ISO 9626) | Result |
|-------------------------------|--------|
| Stiffness                     | Pass   |
| Resistance to breakage        | Pass   |
| Resistance to corrosion       | Pass   |

| Requirement - Test (USP 788)  | Result |
|-------------------------------|--------|
| Particulate matter injections | Pass   |

#### 2) Biocompatibility

| # | Test item                         | Test method / Test criteria                                                                                | Result |
|---|-----------------------------------|------------------------------------------------------------------------------------------------------------|--------|
| 1 | Cytotoxicity                      | ISO 10993-5 Tests for in vitro cytotoxicity                                                                | Pass   |
| 2 | Skin<br>Sensitization<br>Test     | ISO 10993-10 irritation and skin sensitization                                                             | Pass   |
| 3 | Intracutaneous<br>Reactivity Test | ISO 10993-10 Test for irritation and skin sensitization, maximization test<br>for delayed hypersensitivity | Pass   |
| 4 | Acute Systemic<br>Toxicity Test   | ISO 10993-11 Test for systemic toxicity - Acute Systemic Toxicity                                          | Pass   |
| 5 | Pyrogen Test                      | ISO 10993-11 Tests for systemic toxicity, Annex(F) Information on<br>material-mediated pyrogens.           | Pass   |
| 6 | Hemolysis Test                    | ISO 10993-4 Selection of tests for interactions with blood                                                 | Pass   |

3) Sterility and LAL test

| # | Test item      | Test standard                                                          | Result |
|---|----------------|------------------------------------------------------------------------|--------|
| 1 | LAL test       | USP39 <85>, Bacterial Endotoxins Test (Unit : EU/Device)               | Pass   |
| 2 | Sterility test | According to ISO 11737-2                                               | Pass   |
| 3 | E.O Residual   | Under the conditions of ISO 10993-7:2008, Ethylene oxide sterilization | Pass   |

{6}------------------------------------------------

|  | test | residuals, the test articles should meet the test requirements. |  |  |
|--|------|-----------------------------------------------------------------|--|--|
|--|------|-----------------------------------------------------------------|--|--|

Above conducted performance tests demonstrated that this device is performs in a substantially equivalent manner to the predicate device.

#### 9. Conclusion

The device has completed testing to show that the device meets its intended use and demonstrates substantial equivalence to the predicate device, K192551. Therefore, it is concluded that the subject device, FEELject LDV (Low dead volume) Syringe is substantially equivalent to the legally marketed predicate device, K192551.

---

**Source:** [https://fda-staging.innolitics.com/device/K213010](https://fda-staging.innolitics.com/device/K213010)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
