Surgical Mask

K212997 · Guangdong Zhizhen Biological Medical Co., Ltd. · FXX · Jan 14, 2022 · General, Plastic Surgery

Device Facts

Record IDK212997
Device NameSurgical Mask
ApplicantGuangdong Zhizhen Biological Medical Co., Ltd.
Product CodeFXX · General, Plastic Surgery
Decision DateJan 14, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4040
Device ClassClass 2

Indications for Use

The surgical mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device, provided sterile.

Device Story

Surgical mask; three-layer, plane-bandage design with ear straps and nose clip. Protects patient and healthcare personnel from microorganisms, body fluids, and particulates. Used in clinical settings for infection control. Single-use, disposable, provided sterile via EO sterilization. Available in two sizes (175x95mm, 145x95mm) and two colors (Blue, Black).

Clinical Evidence

No clinical data. Bench testing only. Performance verified against ASTM F2100-19 standards, including fluid resistance (ASTM F1862), bacterial filtration efficiency (ASTM F2101), particulate filtration efficiency (ASTM F2299), differential pressure (EN 14683), and flammability (16 CFR 1610). Biocompatibility confirmed via ISO 10993-5 and ISO 10993-10. Sterilization residuals and package integrity verified per ISO 10993-7 and ASTM standards.

Technological Characteristics

Three-layer non-woven construction: outer/inner layers (polypropylene), middle layer (polypropylene melt-blown). Ear straps: polyamide/polyurethane. Nose clip: polyethylene terephthalate/iron. Dimensions: 175x95mm or 145x95mm. Sterilization: Ethylene Oxide (EO) to SAL 10^-6. Complies with ASTM F2100-19 Level 1 requirements.

Indications for Use

Indicated for use by healthcare personnel and patients to reduce exposure to microorganisms, body fluids, and particulate material during infection control practices.

Regulatory Classification

Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible. (ii) Analysis and nonclinical testing must: (A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and (B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device. (iii) NIOSH approved under its regulation. (2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters. January 14, 2022 Guangdong Zhizhen Biological Medical Co., Ltd % Diana Hong General Manager Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai, 200120 China Re: K212997 Trade/Device Name: Surgical Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical apparel Regulatory Class: Class II Product Code: FXX Dated: October 11, 2021 Received: October 15, 2021 Dear Diana Hong: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Clarence W. Murray, III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use 510(k) Number (if known) K212997 Device Name Surgical Mask The surgical mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These masks are in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device, provided sterile. Serial number: ZZ-YYKZ-01, size 175x95mm, Blue ZZ-YYKZ-02, size 145x95mm, Blue ZZ-YYKZ-03.size 175x95mm, Black ZZ-YYKZ-04, size 145x95mm, Black Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### 510(k) Summary This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92. The assigned 510(k) Number: K212997 - Date of Preparation: 01/12/2022 1. - 2. Sponsor Identification ### Guangdong ZhiZhen Biological Medicine Co., Ltd. 3rd and 4th floors of Building 2 & 1ª and 3nd Building 3, No.5 South Street, Datian First Team, Minzhu Village, Tanbu Town, Huadu District Guangzhou Guangdong, China Establishment Registration Number: 3015457451 Contact Person: Caiyun Zhou Position: Manager of Regulations Department Tel: +86-20-37713902 Fax: +86-20-37713902 Email: gfpeng@zhizhenmedic.com - 3. Designated Submission Correspondent Ms. Diana Hong (Primary Contact Person) Ms. Tingting Su (Alternative Contact Person) Mid-Link Consulting Co., Ltd. P.O. Box 120-119, Shanghai, 200120, China Tel: +86-21-22815850, Fax: +360-925-3199 Email: info@mid-link.net {4}------------------------------------------------ #### 4. Identification of Proposed Device Trade Name: Surgical Mask Common Name: Surgical mask Regulatory Information Classification Name: Mask, Surgical; Classification: II; Product Code: FXX; Regulation Number: 21CFR 878.4040; Review Panel: General Hospital; Indication for use: The surgical mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device, provided sterile. Serial number: ZZ-YYKZ-01, size 175x95mm, Blue ZZ-YYKZ-02, size 145x95mm, Blue ZZ-YYKZ-03, size 175x95mm, Black ZZ-YYKZ-04, size 145x95mm, Black Device Description: The proposed device, Surgical Mask is a three-layer, plane bandage type mask with ear strap and nose clip. The ear straps are held in place over the users' mouth and nose by two elastic ear straps welded to the facemask. The proposed device includes masks in two colors and two sizes, as shown in Table 1. The device is single use and provided sterile. | Serial number | Length (mm) | Width (mm) | Color | |---------------|-------------|------------|-------| | ZZ-YYKZ-01 | 175 | 95 | Blue | | ZZ-YYKZ-02 | 145 | 95 | Blue | | ZZ-YYKZ-03 | 175 | 95 | Black | | ZZ-YYKZ-04 | 145 | 95 | Black | Table 1 Product Model #### ર. Identification of Predicate Device ### Predicate Device: 510(k) Number: K202905 Product Name: Disposable Face Mask {5}------------------------------------------------ ### Technological Characteristic Comparison Table 6. | | | Table 1 Comparison of Surgical Mask | | | |--|--|-------------------------------------|--|--| | | | | | | | ITEM | Proposed Device | Predicate Device | Remark | | |---------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|--| | Product Code | FXX | FXX | same | | | Regulation No. | 21 CFR 878.4040 | 21 CFR 878.4040 | same | | | Class | II | II | same | | | Indication for Use | The surgical mask is intended to be worn to<br>protect both the patient and healthcare<br>personnel from transfer of microorganisms,<br>body fluids and particulate material. These<br>masks are intended for use in infection<br>control practices to reduce the potential<br>exposure to blood and body fluids. This is a<br>single use, disposable device, provided<br>sterile.<br>Serial number: ZZ-YYKZ-01, size<br>175x95mm, Blue<br>ZZ-YYKZ-02, size 145x95mm, Blue<br>ZZ-YYKZ-03, size 175x95mm, Black<br>ZZ-YYKZ-04, size 145x95mm, Black | The Disposable Face Mask is<br>intended to be worn to protect<br>both the patient and healthcare<br>personnel from transfer of<br>microorganisms, body fluids, and<br>particulate material. These face<br>masks are intended for use in<br>infection control practices to<br>reduce the potential exposure to<br>blood and body fluids. This is a<br>single-use, disposable device,<br>provided non-sterile. | different | | | Configuration | ear strap<br>nose clip<br>mask body | ear strap<br>nose clip<br>mask body | same | | | Mask color | Blue, Black | Blue | different | | | Dimension | 17.5×9.5 cm<br>14.5×9.5 cm | 14.5×9.5cm,<br>17.0×9.5cm,<br>17.5×9.5cm | different | | | Level | Level I | Level I | same | | | Fluid Resistance | Pass at 80mmHg | Pass at 80mmHg | same | | | Particulate efficiency<br>level | Blue: Passed at 99.14%<br>Black: Passed at 99.14 % | ≥95% | different | | | Bacterial filtration<br>level | Blue: Passed at 99.60%<br>Black: Passed at 99.61 % | ≥95% | different | | | Differential pressure | <5 mmH2O/cm² | <5mmH2O/cm² | same | | | Flammability | Class 1 | Class 1 | same | | | Label/Labeling | Complied with 21 CFR part 801 | Complied with 21 CFR part 801 | same | | | Patient Contacting Material | | | | | | Ear strap | Polyamide and Polyurethane | Spandex and Polyester | different | | | Nose clip | Polyethylene terephthalate and Iron | HDPE and Iron | different | | | Mask<br>body | Outer<br>material | Polypropylene nonwoven fabric | PP meltblown non-woven cloth | | | | Middle<br>material | Polypropylene melt-blown cloth | PP spunbond non-woven cloth | | | | Inner<br>material | Polypropylene nonwoven fabric | PP spunbond non-woven cloth | | | Biocompatibility | | | | | | Cytotoxicity | No Cytotoxicity | No Cytotoxicity | same | | | Sensitization | No Sensitization | No Sensitization | same | | | Irritation | No Irritation | No Irritation | same | | | Sterilization | | | | | | Method | EO sterilized | Non-sterile | different | | | SAL | 10-6 | / | | | {6}------------------------------------------------ Different Analysis 1- Indication for Use The proposed device and predicate devices have essentially same indications for use. Different Analysis 2- Color The color of the proposed device is different from the predicate device. Different Analysis 3- Dimension The dimension for the proposed device is similar to the predicate device, Different Analysis 4-Particulate efficiency level The test result for particulate efficiency for the proposed device is different from predicate device. Different Analysis 5-Bacterial filtration level The test result for bacteria efficiency for the proposed device is different from predicate device. Different Analysis 6-Patient Contacting Material The patient contact material for the propose device is different from predicate device. Different Analysis 7-Sterilezation The final product status of the proposed device is different from predicate device, {7}------------------------------------------------ #### 7. Non-Clinical Test Conclusion Non clinical tests were conducted to verify that the proposed device met all design specifications and the acceptance criteria in the standard and test methodology. The test results demonstrated that the proposed device complies with the following standards: - > ISO 10993-1:2018 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process - > ISO 10993-5:2009 Biological evaluation of medical device- Part 5: Tests for in vitro cytotoxicity - > ISO 10993-7:2008 Biological Evaluation of Medical Device- Part 7: Ethylene Oxide Sterilization Residuals - > ISO 10993-10:2010 Biological evaluation of medical device- Part 10: Tests for irritation and skin sensitization - > ASTM F88/F88M-15, Standard Test Method for Seal Strength of Flexible Barrier Materials. (Sterility) - > ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical Package by Dye Penetration - > ASTM F1886 / F1886M-16, Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection - > ASTM F1862/F1862M-17 Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at a Known Velocity) - > ASTM F2299/F2299M-03 (2017) Standard Test Method for Determining the Initial Efficiency of Material Used in medical Face Masks to Penetration by Particulates using Latex Spheres - > ASTM F2101-19 Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) of Medical Face Mask Materials, Using a Biological Aerosol of Staphylococcus aureus - > 16 CFR 1610 Standard for the Flammability of Clothing Textiles Corrections - > EN14683 Medical face masks-Requirements and test methods {8}------------------------------------------------ The results of performance tests, biocompatibility test, conducted on the Surgical Mask demonstrate that the device met the specification and the acceptance criteria. A summary of the tests performed is provided in the table below: | Test | Purpose | Acceptance<br>Criteria per ASTM<br>F2100-19 | Results<br>(Statistics of three lots,<br>32 per lot) | | |---------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|------------------------------------------------------|------------------------------------| | | | | Blue | Black | | Fluid Resistance<br>(ASTM F1862) | Verify the fluid resistance of the<br>proposed device can meet the<br>requirements for Level 1<br>specified in ASTM F2100-19. | Pass at 80 mmHg | 96 out of 96<br>pass at 80<br>mmHg | 96 out of 96<br>pass at 80<br>mmHg | | Bacterial filtration<br>efficiency (BFE)<br>(ASTM F2101) | Verify the bacterial filtration<br>efficiency of the proposed device<br>can meet the requirements for<br>Level 1<br>specified in ASTM F2100-19. | ≥95% | Passed at<br>99.60% | Passed at<br>99.61% | | Particulate<br>filtration<br>efficiency (PFE)<br>(ASTM F2299) | Verify the particulate filtration<br>efficiency of the proposed device<br>can meet the requirements for<br>Level 1<br>specified in ASTM F2100-19. | ≥95% | Passed at<br>99.14% | Passed at<br>99.14% | | Differential<br>pressure<br>(EN 14683) | Verify the differential pressure of<br>the proposed device can meet the<br>requirements for Level 1<br>specified in ASTM F2100-19. | <5.0<br>mmH2O/cm² | <5.0<br>mmH2O/cm² | <5.0<br>mmH2O/cm² | | Flammability<br>(16 CFR 1610) | Verify the flammability of the<br>proposed device can meet the<br>requirements for Level 1<br>specified in ASTM F2100-19. | Class 1 | Class 1 | Class 1 | | Table 2 Performance Testing | | |-----------------------------|--| |-----------------------------|--| ## Table 3 Biocompatibility Testing | Test | Purpose | Acceptance Criteria | Result | |----------------------------------------|--------------------------------------------------------------|---------------------------------------------------------------|--------| | In vitro Cytotoxicity<br>(ISO 10993-5) | Verify that the proposed device extract is<br>non-cytotoxic. | The extract is non-cytotoxic<br>under the testing conditions. | Pass | {9}------------------------------------------------ | Skin Irritation<br>(ISO 10993-10) | Verify that the proposed device extract is non-irritating. | The polar and non-polar extracts are non-irritating under the testing conditions. | Pass | |--------------------------------------|-------------------------------------------------------------|------------------------------------------------------------------------------------|------| | Skin Sensitization<br>(ISO 10993-10) | Verify that the proposed device extract is non-sensitizing. | The polar and non-polar extracts are non-sensitizing under the testing conditions. | Pass | # Table 4 Sterilization | Test | Purpose | Acceptance Criteria | Result | |-----------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------| | EO ECH residuals<br>(ISO 10993-7) | Verify the EO ECH residuals<br>of the proposed device can<br>meet the requirements of ISO<br>10993-7. | The average daily dose of EO to<br>patient shall not exceed 4mg.<br>The average daily dose of ECH<br>to patient shall not exceed 9mg. | Pass | | Bacterial Endotoxin<br>(USP <85>) | Verify the bacterial endotoxin<br>of the proposed device can<br>meet the requirements of USP<br><85>. | Bacterial Endotoxin Levels<br>were below the level of 20<br>EU/device | Pass | | Package Integrity<br>(ASTM F1886 /<br>F1886M-16, ASTM<br>F88/F88M-15,<br>ASTM F1929-15) | Verify the package integrity of<br>the proposed device can meet<br>the requirements of ASTM<br>F1886 / F1886M-16, ASTM<br>F88/F88M-15,<br>ASTM<br>F1929-15. | There should be no channel on the<br>package and no damage on device.<br>The maximum seal strength<br>should be no less than<br>2.0N/15mm.<br>There should be no dye<br>penetration on the package | Pass | - 8. Clinical Test Conclusion No clinical study is included in this submission. {10}------------------------------------------------ #### 9. Conclusion The conclusion drawn from the nonclinical tests demonstrates that the subject device in the 510(k) submission, the Surgical Masks are as safe and effective, and performs as well as or better than the legally marketed predicate device cleared under K202905.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...