← Product Code [HRS](/productcode/HRS) · K212979

# CalcShift™ Displacement Calcaneal Osteotomy System (K212979)

_Nextremity Solutions, Inc. · HRS · May 20, 2022 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K212979

## Device Facts

- **Applicant:** Nextremity Solutions, Inc.
- **Product Code:** [HRS](/productcode/HRS.md)
- **Decision Date:** May 20, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Nextremity Solutions CalcShift™ Displacement Calcaneal Osteotomy System is indicated for osteotomies, nonunions, malunions, revisions, fusions, and reconstruction of the calcaneus requiring a medial or lateral displacement osteotomy.

## Device Story

CalcShift Displacement Calcaneal Osteotomy System is a metallic bone fixation system consisting of plates and screws. Used by surgeons in clinical settings for calcaneal procedures; system provides mechanical stabilization of bone segments following osteotomy. Implants include 1.6mm thick plates with 4 screw holes and various 3.5mm locking/non-locking screws and 5.0mm cannulated compression/stability screws. Accessory instruments facilitate site preparation and implant insertion. System provides rigid fixation to maintain bone alignment during healing; benefits patient by supporting structural reconstruction and bone fusion.

## Clinical Evidence

No clinical data. Substantial equivalence demonstrated via non-clinical bench testing: endotoxin testing (LAL method per AAMI ST72/USP 161/85), axial pullout strength, torque to failure, insertion/removal torque (ASTM F543-17), and static/dynamic 4-point bend tests (ASTM F382).

## Technological Characteristics

Metallic bone fixation system; plates and screws manufactured from Ti-6Al-4V ELI (ASTM F136). Plate dimensions: 1.6mm thick, 19.75mm long, 11.5mm wide. Screws: 3.5mm diameter (locking/non-locking) and 5.0mm diameter (cannulated). Provided sterile. Mechanical testing per ASTM F543-17 and ASTM F382.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Wright Medical Technology, Ortholoc 3Di Midfoot/Flatfoot System ([K121651](/device/K121651.md))
- Nextremity Solutions, Stratum Foot Plating System ([K182201](/device/K182201.md))
- Zimmer Biomet, TiMAX™ Cannulated Screw System ([K140891](/device/K140891.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the date May 20, 2022. The date is written in a clear, legible font. The month is spelled out, followed by the day and year, separated by commas.

Image /page/0/Picture/10 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the symbol of the Department of Health & Human Services on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The text is in blue, with "FDA" in a larger font size than the rest of the text.

Nextremity Solutions, Inc. Elise Fox Quality and Regulatory Specialist 1195 Polk Drive Warsaw, Indiana 46582

Re: K212979

Trade/Device Name: CalcShift™ Displacement Calcaneal Osteotomy System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS. HWC Dated: April 14, 2022 Received: April 15, 2022

Dear Elise Fox:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known) K212979

Device Name

CalcShift™ Displacement Calcaneal Osteotomy System

Indications for Use (Describe)

The Nextremity Solutions CalcShift™ Displacement Calcaneal Osteotomy System is indicated for osteotomies, nonunions, malunions, revisions, fusions, and reconstruction of the calcaneus requiring a medial or lateral displacement osteotomy.

Type of Use (Select one or both, as applicable)

|X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

| 510(k) Summary    |                                                                                                                      |
|-------------------|----------------------------------------------------------------------------------------------------------------------|
| Prepared:         | May 17, 2022                                                                                                         |
| Submitter:        | Nextremity Solutions, Inc.<br>1195 Polk Drive<br>Warsaw, IN 46582                                                    |
| Contact:          | Elise Fox<br>Quality and Regulatory Specialist<br>elise.fox@medartis.com<br>Phone: 574-376-2062<br>FAX: 574-966-1396 |
| Proprietary Name: | CalcShift™ Displacement Calcaneal Osteotomy System                                                                   |
| Common Name:      | Bone Plate and Screw System                                                                                          |
| Classifications:  | 21 CFR §888.3030: Single/Multiple Component Metallic Bone<br>Fixation Appliances And Accessories; Class II (Primary) |
|                   | 21 CFR §888.3040: Smooth or Threaded Metallic Bone Fixation<br>Fastener; Class II                                    |
| Product Codes:    | 87 / HRS, 87 / HWC                                                                                                   |

#### Substantially Equivalent Devices:

- Wright Medical Technology, Ortholoc 3Di Midfoot/Flatfoot System, K121651- Primary . Predicate
- Nextremity Solutions, Stratum Foot Plating System, K182201-Additional Predicate .
- Zimmer Biomet, TiMAX™ Cannulated Screw System (Cleared as Biomet Cannulated . Screw System), K140891-Additional Predicate

#### Device Description:

The CalcShift Displacement Calcaneal Osteotomy System is a foot and ankle plate and screw system consisting of a plate and screw implants. The plate is 1.6mm thick, 19.75mm long, and 11.5mm wide. The plate has 4 screw holes. The system provides locking and non-locking screws that are 3.5mm in diameter and range in length from 10mm to 70mm. The system includes cannulated compression screws and cannulated stability screws that are 5.0mm in diameter and range from 30mm to 100mm in length. The plate and screws are provided sterile. The plates and screws are manufactured from Ti-6Al-4V ELI conforming to ASTM F136.

{4}------------------------------------------------

The system is provided with a set of accessory instruments designed for preparation of the implant site and insertion of the implants into bone.

#### Indications for Use:

The Nextremity Solutions CalcShift Displacement Calcaneal Osteotomy System is indicated for osteotomies, non-unions, malunions, revisions, and reconstruction of the calcaneus requiring a medial or lateral displacement osteotomy.

## Summary of Technologies/Substantial Equivalence:

The CalcShift Displacement Calcaneal Osteotomy System is substantially equivalent to the predicate devices regarding the intended use and indications, material, design, sizes, and mechanical properties. Differences between the subject device system and the predicate device systems do not raise new types of safety and effectiveness questions.

### Non-Clinical Testing:

Endotoxin testing was performed using the Limulus Amebocyte Lysate (LAL) method according to AAMI ST72, USP 161 and USP 85. Results met the Endotoxin limit of ≤20 EU per device.

To evaluate the strength of the CalcShift Displacement Calcaneal Osteotomy System, axial pullout strength, torque to failure tests, and insertion/removal torque were performed on worst case screws in accordance with ASTM F543-17. Static 4-point bend tests and dynamic 4-point bend tests were performed on the CalcShift plate according to ASTM F382. These tests confirmed that the strength of the CalcShift Displacement Calcaneal Osteotomy System is substantially equivalent to predicate devices with similar indications and is adequate for its intended use.

# Clinical Testing:

Clinical testing was not necessary to demonstrate substantial equivalence of the CalcShift Displacement Calcaneal Osteotomy System to the predicate device.

#### Conclusions/Substantial Equivalence:

The CalcShift Displacement Calcaneal Osteotomy System is substantially equivalent to the predicate devices regarding its intended use, material, design, sizes, and mechanical properties. Differences between the subject device system and the predicate device systems do not raise different types of safety and effectiveness questions.

---

**Source:** [https://fda-staging.innolitics.com/device/K212979](https://fda-staging.innolitics.com/device/K212979)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
