K212971 · Guard Medical, Inc. · OKO · Feb 10, 2022 · General, Plastic Surgery
Device Facts
Record ID
K212971
Device Name
NPseal
Applicant
Guard Medical, Inc.
Product Code
OKO · General, Plastic Surgery
Decision Date
Feb 10, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4683
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The NPseal is indicated for patients who would benefit from wound management via application of a negative pressure, particularly as the device may promote wound healing through the removal of small amounts of exudates from closed surgical incisions.
Device Story
NPseal is a single-use, non-powered negative pressure wound therapy (NPWT) system. It consists of an integrated mechanical pump and a multilayer dressing (hydrophilic polyurethane foam pad with a breathable polyurethane film). The user manually actuates the pump to generate and maintain negative pressure in the -75 mmHg to -125 mmHg range. The device is applied over closed surgical incisions to collect and move exudate away from the wound bed via foam absorption and film evaporation. It is intended for up to 6 days of wear time. The system is designed for use on incisions of varying lengths (Small, Medium, Large). By maintaining a negative pressure environment, the device aims to promote wound healing. It is a prescription-use device.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including pressure and exudate handling, benchtop usability, and packaging validation.
Technological Characteristics
Non-powered, integrated mechanical pump system. Materials: Polyurethane film with acrylic adhesive, hydrophilic polyurethane foam pad, thermoplastic elastomer pump. Dimensions: Medium (16x8 cm), Large (21x8 cm). Negative pressure range: -75 to -125 mmHg. Sterilization: Gamma irradiation. Biocompatibility: ISO 10993-1 compliant.
Indications for Use
Indicated for patients requiring wound management via negative pressure to promote healing of closed surgical incisions by removing small amounts of exudate.
Regulatory Classification
Identification
A non-powered suction apparatus device intended for negative pressure wound therapy is a device that is indicated for wound management via application of negative pressure to the wound for removal of fluids, including wound exudate, irrigation fluids, and infectious materials. It is further indicated for management of wounds, burns, flaps, and grafts.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Non-powered Suction Apparatus Device Intended for Negative Pressure Wound Therapy (NPWT).” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
February 10, 2022
Guard Medical Inc % Eric Bannon VP of Regulatory Affairs Alvamed Inc 935 Great Plain Ave. Suite 166 Needham, Massachusetts 02492
Re: K212971
Trade/Device Name: NPseal Regulation Number: 21 CFR 878.4683 Regulation Name: Non-Powered Suction Apparatus Device Intended For Negative Pressure Wound Therapy Regulatory Class: Class II Product Code: OKO Dated: January 7, 2022 Received: January 11, 2022
Dear Eric Bannon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Julie Morabito, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212971
Device Name NPseal
Indications for Use (Describe)
The NPseal is indicated for patients who would management via application of a negative pressure, particularly as the device may promote wound healing through the removal of small amounts of exudates from closed surgical incisions.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Guard Medical. The word "GUARD" is in large, blue, sans-serif font, with the "A" stylized as an upside-down "V". Below the word "GUARD" is the word "MEDICAL", also in blue, sans-serif font, but smaller than the word "GUARD". The logo is simple and modern, and the blue color gives it a sense of trustworthiness and reliability.
### 1.0 TRADITIONAL 510(K) SUMMARY K212971
This Traditional 510(k) for the NPseal is submitted based on the FDA Guidance document "Format for Traditional and Abbreviated 510(k)s: Guidance for Industry and Food and Drug Administration Staff" (issued September 13, 2019).
- 1.1 Name and Address of Sponsor
Guard Medical, Inc. 1221 Brickell Avenue, Suite 900 Miami, FL 33131 USA Phone: +1 (888) 417-3644 Machiel Van der Leest CEO m.vanderleest@guard-medical.com
- 1.2 Correspondent/Primary Contact Person
Eric Bannon Vice President of Regulatory and Clinical Affairs
AlvaMed, Inc. (consultant to Guard Medical) 935 Great Plain Avenue, Unit 166 Needham, MA 02492 USA ebannon@alvamed.com Phone: +1 (781) 710-8243 Fax: +1 (617) 249-0955
- 1.3 Submission Information
| Date Summary Prepared: | September 16, 2021 |
|------------------------|------------------------------------------------------------------|
| Name of Device: | NPseal |
| Common or Usual Name: | Negative Pressure Wound Therapy Non-Powered<br>Suction Apparatus |
| Classification: | Class II |
| Product Code: | OKO (21 CFR 878.4683) |
| Predicate Device: | NPseal (K211130) |
| | |
- 1.4 Device Description
The NPseal Neqative Pressure Advanced System is a single-use device that includes an integrated, mechanical pump system. The NPseal maintains Negative Pressure Wound Therapy (NPWT) in the -75 mmHg to -125 mmHg nominal range.
The NPseal is intended for 6 days of use. Therapy duration of the system may be less than indicated if clinical practice or other factors such as time to wound closure and rate of
{4}------------------------------------------------
exudate, result in earlier removal or need for system change. The NPseal can be replaced only one time for a total maximum wear time of 6 days.
The NPseal Small is intended for surgically closed incisions up to 5cm x 0.5 cm
The NPseal Medium is intended for surgically closed incisions up to 10 cm x 0.5 cm.
The NPseal Large is intended for surgically closed incisions up to 15 cm x 0.5 cm.
- 1.5 Indications for Use
The NPseal is indicated for patients who would benefit from wound management via application of negative pressure, particularly as the device may promote wound healing through the removal of small amounts of exudates from closed surgical incisions.
- 1.6 Comparison of Manufacturer's Cleared Device and Modified Device
## Table 1. Comparison of Subject Device to Cleared Device
| | Subject Devices<br>NPseal Medium<br>NPseal Large | Predicate Device<br>NPseal Small |
|---------------------------|---------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K212971 | K211130 |
| 510k<br>Submitter/Holder | Same as predicate | Guard Medical |
| Product Code | Same as predicate | OKO |
| Regulation No. | Same as predicate | 878.4683 |
| Regulation<br>Description | Same as predicate | Non-Powered suction apparatus<br>device intended for negative pressure<br>wound therapy |
| Common<br>Name | Same as predicate | Negative Pressure Wound Therapy<br>non-powered suction apparatus |
| Indications for Use | Same as predicate | The NPseal is indicated for patients<br>who would benefit from wound<br>management via application of a<br>negative pressure, particularly as the<br>device may promote wound healing<br>through the removal of small amounts<br>of exudates from closed surgical<br>incisions. |
| Wound types | Same as predicate | Closed surgical incisions |
| Incision sizes | NPseal Medium: Up to 10 cm x 0.5 cm<br>NPseal Large: Up to 15 cm x 0.5 cm | Up to 5 cm x 0.5 cm |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Guard Medical. The word "GUARD" is in a large, sans-serif font, with the "U" shaped like a shield. The word "MEDICAL" is in a smaller, sans-serif font and is located to the right of the word "GUARD" and slightly below it. The logo is a light blue color.
3
| | Subject Devices<br>NPseal Medium<br>NPseal Large | Predicate Device<br>NPseal Small |
|---------------------------------------|--------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Single Use | Same as predicate | Yes |
| Negative Pressure<br>Range | Same as predicate | -75 to -125 mmHg (± 17.5 mmHg) |
| Device Technology | Same as predicate | Nonpowered, integrated pump<br>manually actuated to generate<br>negative pressure. Multilayer pad<br>composed of hydrophilic foam with<br>high fluid absorbency and top<br>breathable film designed to collect and<br>move exudate away from the wound<br>bed. |
| Management of<br>Exudates | Same as predicate | Managed by the dressing itself - via<br>combination of absorption into the<br>foam pad and evaporation through the<br>breathable upper film. |
| Materials | Same as predicate | Film: Polyurethane coated with<br>adhesive acrylic,<br>Pad: Hydrophilic polyurethane,<br>Pump: Thermoplastic elastomer |
| Maximum therapy<br>time | Same as predicate | 6 days |
| Maximum number of<br>dressing changes | Same as predicate | 1 dressing change |
| Wear Time per<br>dressing | Same as predicate | Up to 6 days |
| Sterility | Same as predicate | Sterile - Gamma irradiation |
| Biocompatibility | Same as predicate | Complies with ISO 10993-1 |
#### 1.7 Summary of Modification
The dimensions of the current NPseal device cleared in K200305 and K211130 (NPseal Small) is 11 x 8 cm for use on closed surgical incisions up to 5 cm in length and 0.5 cm in width. The corresponding foam pad size is 7.5 x 4.5 cm.
The subject devices are two longer versions of the NPseal Small:
- . The NPseal Medium is 16 x 8 cm with a 12.65 x 4.65 cm foam pad for use on closed surgical incisions up to 10 cm in length and 0.5 cm in width.
{6}------------------------------------------------
- . The NPseal Large is 21 x 8 cm with a 17.65 x 4.65 cm foam pad for use on closed surgical incisions up to 15 cm in length and 0.5 cm in width.
All other dimensions including the foam thickness, as well as the pump system and device materials remain identical to the NPseal Small predicate.
- 1.8 Summary of Functional and Performance Testing
The NPseal dressing length increase for the NPseal Medium and Large was assessed based on risk and the Design Controls in 21 CFR 820.
The following verification and validation testing was performed exclusively on the NPseal Large as it represents the worst-case device size:
- Pressure and Exudate Handling Over Time: Demonstrated that negative pressure could . be maintained, the device was capable of absorbing simulated exudate and the volume of simulated exudate absorbed covered less than 50% of the foam pad.
- Benchtop Usability Testing: Demonstrated that usability requirements including pump . actuation, device removal from packaging, device application and ability to generate negative pressure met acceptance criteria.
- Packaging Validation: Demonstrated that sterile packaging seal was maintained after . undergoing environmental conditioning and simulated distribution.
- Summary of Clinical Testing 1.9
No clinical testing was applicable to this submission.
- 1.10 Conclusion
The NPseal Medium and NPseal Large have been shown to be as safe as the predicate device through bench testing. Size increases made to the device demonstrate substantial equivalence to the predicate device, NPseal Small. Functional and performance testing data demonstrate the substantial equivalence of the NPseal Medium and Large to the NPseal Small (K211130).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.