K212958 · Viewray, Incorporated · IYE · Dec 14, 2021 · Radiology
Device Facts
Record ID
K212958
Device Name
MRIdian Linac System
Applicant
Viewray, Incorporated
Product Code
IYE · Radiology
Decision Date
Dec 14, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The MRIdian Linac system, with magnetic resonance imaging capabilities, is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated.
Device Story
MRIdian Linac is an image-guided radiation therapy (IGRT) system combining a 6 MV linear accelerator with a 0.345 Tesla superconducting MRI. It integrates three subsystems: Treatment Planning and Delivery System (TPDS), Magnetic Resonance Imaging System (MRIS), and Radiation Therapy Delivery System (RDS). The device provides real-time MRI guidance during radiation delivery to target lesions, tumors, and conditions. It is operated by clinical staff in a radiation oncology setting. The system uses MRI inputs to visualize anatomy in real-time, allowing clinicians to adjust radiation delivery for precise targeting. This enables stereotactic radiosurgery and precision radiotherapy, potentially improving treatment accuracy and sparing healthy tissue. Modifications in this version include a redesigned Treatment Delivery subsystem, updated MR control system software (Siemens Avanto Dot), and support for an updated head receive coil.
Clinical Evidence
No clinical or animal testing was performed. Substantial equivalence is supported by bench testing and verification/validation activities demonstrating conformance to design specifications and relevant safety standards.
Technological Characteristics
6 MV linear accelerator; 0.345 Tesla superconducting MRI; 138-leaf MLC (0.83 cm nominal width, 0.415 cm effective width); 90 cm isocenter distance; 600 MU/min dose rate. Patient surface complies with ISO 10993-1. Software includes MR control system (Siemens Avanto Dot). Connectivity/interoperability via standard medical imaging protocols. Standards: IEC 60601-1, IEC 60601-2-1, IEC 60601-2-33, IEC 62304, IEC 62366-1.
Indications for Use
Indicated for patients requiring stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}------------------------------------------------
December 14, 2021
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
ViewRay, Incorporated % Mr. Sean Delaney Senior Advisor, Continuous Improvement & Development Strategy 815 E. Middlefield Road MOUNTAIN VIEW CA 94043
Re: K212958
Trade/Device Name: MRIdian Linac System Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: Class II Product Code: IYE, LNH Dated: September 14, 2021 Received: September 16, 2021
Dear Mr. Delaney:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting
{1}------------------------------------------------
combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
, for
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
K212958
Device Name MRIdian Linac System
Indications for Use (Describe)
The MRIdian Linac system, with magnetic resonance imaging capabilities, is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated.
| Type of Use (Select one or both, as applicable) | | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| <table><tr><td><div> <input checked="true" type="checkbox"/> Research Use (Part 21 CFR 601 Subpart D) </div></td></tr><tr><td><div> <input type="checkbox"/> Compassionate Use (21 CFR 601 Subpart E) </div></td></tr></table> | <div> <input checked="true" type="checkbox"/> Research Use (Part 21 CFR 601 Subpart D) </div> | <div> <input type="checkbox"/> Compassionate Use (21 CFR 601 Subpart E) </div> |
| <div> <input checked="true" type="checkbox"/> Research Use (Part 21 CFR 601 Subpart D) </div> | | |
| <div> <input type="checkbox"/> Compassionate Use (21 CFR 601 Subpart E) </div> | | |
> Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the ViewRay logo. The logo consists of a green sphere with white lines running through it, followed by the word "VIEWRAY" in bold, black letters. Below the word "VIEWRAY" is the tagline "Visibly Different" in a smaller, lighter font.
## Section 6: 510(k) Summary
The information provided for the modified MRIdian Linac system, following the format of 21 CFR 807.92.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the ViewRay logo. The logo consists of a green sphere with white lines and a blue dot on the upper left side. To the right of the sphere is the word "VIEWRAY" in bold, black letters, and below that is the phrase "Visibly Different" in a smaller, lighter font.
### Premarket Notification 510(k) Summary
As Required by 21 CFR 807.92
| 510(k) Number: | K212958 |
|---------------------------|--------------------------------------------------------------------------------|
| Product Name: | MRIdian Linac System |
| Date Prepared: | September 13, 2021 |
| Submitter: | ViewRay Incorporated<br>815 E. Middlefield Road<br>Mountain View, CA 94043 USA |
| Primary Contact Person: | Sean Delaney<br>(650) 252-0969<br>sdelaney@viewray.com |
| Secondary Contact Person: | Elizabeth Osuna<br>(408) 431-1046<br>eosuna@viewray.com |
| Common or Usual Name: | Medical Linear Accelerator |
| Regulatory Name: | 21 CFR 892.5050 |
| Classification Name: | Medical charged-particle radiation therapy<br>system |
| Product Code: | IYE, LNH |
| Primary Predicate Device: | MRIdian Linac System<br>K181989 |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image is a logo for ViewRay. The logo consists of a green sphere with white lines running across it, and the word "VIEWRAY" in bold black letters. Below the word "VIEWRAY" is the tagline "Visibly Different" in a smaller, non-bold font. The logo is clean and modern, and the green color suggests a focus on health or technology.
### Device Description
The MRIdian Linac system is an image-quided radiation therapy (IGRT) system that uses a 6 MV linear accelerator radiotherapy system to deliver ionizing radiation while using a magnetic resonance imaging system for image guidance in real time. The MRIdian consists of the Treatment Planning and Delivery System (TPDS), Magnetic Resonance Imaging System (MRIS) and the Radiation Therapy Delivery System (RDS). These three subsystems are designed to operate concurrently for accurate targeted administration of radiation therapy.
### Intended Use
The MRIdian Linac system, with magnetic resonance imaging capabilities, is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated.
### Indication for Use
The MRIdian Linac system, with magnetic resonance imaging capabilities, is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated.
#### Significant Differences
- The Treatment Planning and Delivery System modified with a redesigned . Treatment Delivery subsystem (TDS) and the same Treatment Planning subsystem (TPS) modified only to support the new TDS implementation;
- The Magnetic Resonance Imaging System (MRIS) which has been updated to . support an end-of-life replacement of the MR control system software (Siemens Avanto Dot) and include additional support for an updated head (receive) coil; and
- The Radiation Therapy Delivery System (RDS) modified only to support the TDS ● software interfaces changes.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the ViewRay logo. The logo consists of a green sphere with white lines inside, followed by the text "VIEWRAY" in black, with the words "Visibly Different" underneath. The text is separated by a horizontal line.
# Summary of Technological Characteristics
| Device<br>Characteristic | Predicate Device | Subject Device | Comparison |
|------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| Radiation Source | | | |
| Nominal Beam<br>Energy | 6 MV | 6 MV | Same |
| Isocenter Distance | 90 cm | 90 cm | Same |
| Nominal Dose Rate | 600 MU/min | 600 MU/min | Same |
| Collimation | | | |
| Radiation Field<br>Sizes | 2 mm x 4.15 mm<br>to 27.4 x 24.1 cm<br>projected by<br>isocenter | 2 mm x 4.15 mm<br>to 27.4 x 24.1 cm<br>projected by<br>isocenter | Same |
| Primary<br>Collimation Method | 138-leaf MLC | 138-leaf MLC | Same |
| Resolution at<br>isocenter | Each leaf projects<br>with a nominal<br>width of 0.83 cm<br>at 90 cm SAD. The<br>double- stack MLC<br>uses an offset<br>between the top<br>and bottom stack<br>to achieve a<br>nominal effective<br>leaf width of 0.415<br>cm at 90 cm SAD. | Each leaf projects<br>with a nominal<br>width of 0.83 cm<br>at 90 cm SAD. The<br>double- stack MLC<br>uses an offset<br>between the top<br>and bottom stack<br>to achieve a<br>nominal effective<br>leaf width of 0.415<br>cm at 90 cm SAD. | Same |
| Collimation<br>Transition Time | MLC leaf speed of<br>4 cm/s at isocenter | MLC leaf speed of<br>4 cm/s at isocenter | Same |
| Imaging | | | |
| Energy/Type | 0.345 Tesla<br>superconducting<br>MRI | 0.345 Tesla<br>superconducting<br>MRI | Same |
| Device<br>Characteristic | Predicate Device | Subject Device | Comparison |
| Field of View Size | 50 cm diameter<br>spherical volume<br>(DSV) | 50 cm diameter<br>spherical volume<br>(DSV) | Same |
| Contrast<br>Resolution | Complies with IEC<br>60601-2-33 | Complies with IEC<br>60601-2-33 | Same |
| Spatial Resolution | Volume scans:<br>Min 0.075 x 0.075<br>x 0.15 | Volume scans:<br>Min 0.075 x 0.075<br>x 0.075 | Volume scans:<br>Minimum<br>longitudinal<br>resolution from<br>0.15 to 0.75. |
| | Max 0.3 x 0.3 x<br>0.3 | Max 0.3 x 0.3 x<br>0.3 | |
| | Ciné scans:<br>Slice thickness 5,<br>7, or 10 mm | Ciné scans:<br>Slice thickness 5,<br>7, or 10 mm | Ciné scans: |
| | One plane: Sagittal | One, two, or three<br>planes: Axial,<br>sagittal, or coronal | |
| | Pixel size: | Pixel size: | Pixel size: |
| | 0.24 x 0.24 | 0.13 x 0.13; head | 0.13 x 0.13; head |
| | 0.35 x 0.35 | 0.20 x 0.20; body | 0.20 x 0.20; body |
| | | 0.24 x 0.24; body<br>0.35 x 0.35; body | |
| Dose | None | None | Same |
| Physical<br>Geometry | | | |
| Bore size | The system has a<br>diameter of 70 cm<br>nominal | The system has a<br>diameter of 70 cm<br>nominal | Same |
| Room Dimensions | The MRIdian Linac<br>minimum vault<br>size with finishes is | The MRIdian Linac<br>minimum vault<br>size with finishes is | Same |
| Device<br>Characteristic | Predicate Device | Subject Device | Comparison |
| | 19' 2" wide by x<br>24' 8" long [5.9 m<br>x 6.93 m]. The<br>minimum finished<br>ceiling height over<br>the footprint of<br>system is 9' 6"<br>[2.9 m] to allow<br>for clearance of<br>system covers<br>which are 9' 4" | 19' 2" wide by x<br>24' 8" long [5.9 m<br>x 6.93 m]. The<br>minimum finished<br>ceiling height over<br>the footprint of<br>system is 9' 6"<br>[2.9 m] to allow<br>for clearance of<br>system covers<br>which are 9' 4" | |
| Patient Couch -<br>Degrees of<br>Freedom | 3D shifts and<br>shifted values are<br>within ± 0.1 cm in<br>each dimension | 3D shifts and<br>shifted values are<br>within ± 0.1 cm in<br>each dimension | Same |
| Patient Surface -<br>Biocompatibility | Complies with ISO<br>10993-1 | Complies with ISO<br>10993-1 | Same |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the ViewRay logo. The logo consists of a green sphere with white lines and a blue dot on the top left. To the right of the sphere is the word "VIEWRAY" in black, with the words "Visibly Different" in a smaller font below.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the ViewRay logo. The logo consists of a green sphere with white lines running across it on the left side. To the right of the sphere is the word "VIEWRay" in black, with the words "Visibly Different" in a smaller font below it.
### Performance Data
The MRIdian Linac System was tested according to the FDA Quality System Regulation (21 CFR §820). Test results demonstrate that the device conforms to design specifications and meets the needs of the intended users, including assuring risk mitigations were implemented and functioned properly. Software testing was performed and documented as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."
### Standards
The MRIdian Linac System conforms to the following regulatory standards including FDA recognized standards and references additional standards as applicable.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image contains the ViewRay logo. The logo consists of a green sphere with white lines inside, followed by the word "VIEWRAY" in bold, black letters. Below the word "VIEWRAY" is the tagline "Visibly Different" in a smaller, lighter font.
| Rec. # | Standards<br>Developing<br>Organization | Standard<br>Designation<br>Number and<br>Date | Title of Standard |
|--------|---------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 19-4 | ANSI AAMI | ES60601-<br>1:2005/(R)20<br>12 and<br>A1:2012,<br>C1:2009/(R)2<br>012 and<br>A2:2010/(R)2<br>012<br>(Consolidated<br>Text) | Medical electrical equipment - Part 1: General<br>requirements for basic safety and essential<br>performance (IEC 60601-1:2005, MOD) |
| 19-8 | ANSI AAMI<br>IEC | 60601-1-<br>2:2014 | Medical electrical equipment -- Part 1-2: General<br>requirements for basic safety and essential<br>performance -- Collateral Standard: Electromagnetic<br>disturbances -- Requirements and tests |
| 19-8 | EC | 60601-1-2<br>Edition 4.0<br>2014-02 | Medical electrical equipment - Part 1-2: General<br>requirements for basic safety and essential<br>performance - Collateral Standard: Electromagnetic<br>disturbances - Requirements and tests |
| 12-285 | IEC | 60601-2-1<br>Edition 3.0 | Medical electrical equipment - Part 2-1: Particular<br>requirements for the basic safety and essential<br>performance of electron accelerators in the range 1<br>MeV to 50 MeV |
| 12-295 | IEC | 60601-2-33<br>Ed. 3.2<br>b:2015 | Medical electrical equipment - Part 2-33: Particular<br>requirements for the basic safety and essential<br>performance of magnetic resonance equipment for<br>medical diagnosis |
| 12-267 | IEC | 61217 Edition<br>2.0 2011-12 | Radiotherapy equipment - Coordinates, movements,<br>and scales |
| 12-217 | IEC | 62083 Edition<br>2.0 2009-09 | Medical electrical equipment - Requirements for the<br>safety of radiotherapy treatment planning systems |
| 13-79 | ANSI AAMI<br>IEC | 62304:2006/A<br>1:2016 | Medical device software - Software life cycle<br>processes [Including Amendment 1 (2016)] |
| 13-79 | IEC | 62304 Edition<br>1.1 2015-06<br>CONSOLIDAT<br>ED VERSION | Medical device software - Software life cycle<br>processes |
| 5-114 | ANSI AAMI<br>IEC | 62366-1:2015 | Medical devices - Part 1: Application of usability<br>engineering to medical devices |
| Rec. # | Standards<br>Developing<br>Organizatio<br>n | Standard<br>Designation<br>Number and<br>Date | Title of Standard |
| | IEC | 62366-1<br>Edition 1.0<br>2015-02 | Medical devices - Part 1: Application of usability<br>engineering to medical devices [Including<br>CORRIGENDUM 1 (2016)] |
| 5-89 | IEC | 60601-1-6<br>Edition 3.1<br>2013-10 | Medical electrical equipment - Part 1-6: General<br>requirements for basic safety and essential<br>performance - Collateral standard: Usability |
| 12-253 | IEC | 60976 Edition<br>2.0 2007-10 | Medical electrical equipment - Medical electron<br>accelerators - Functional performance characteristics |
| 2-258 | ANSI AAMI<br>ISO | 10993-1:<br>2018 | Biological evaluation of medical devices - Part 1:<br>Evaluation and testing within a risk management<br>process |
| | ISO | 10993-1 Fifth<br>edition 2018-<br>08 | Biological evaluation of medical devices - Part 1:<br>Evaluation and testing within a risk management<br>process |
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the ViewRay logo. The logo consists of a green circle with white lines inside, and the word "VIEWRAY" in black letters to the right of the circle. Below the word "VIEWRAY" is the phrase "Visibly Different" in a smaller font. The logo is simple and modern, and the colors are eye-catching.
## Bench Testing
The performance data demonstrate that the MRIdian Linac System is as safe and effective and performs as well as the predicate device.
## Clinical Testing
No animal or clinical tests are being submitted to establish substantial equivalence with the predicate device.
### Conclusion
The MRIdian Linac System is substantially equivalent to the predicate device. The intended use and indications for use are the same. The major technological characteristics are substantially equivalent to the predicate device, and the differences do not raise new questions of safety and effectiveness. The results of testing as well as conformance to relevant safety standards demonstrate that MRIdian Linac System meets the safety and performance criteria and is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.