The INHANCE SHOULDER SYSTEM with the humeral stemless anchor is intended for use in anatomic total shoulder replacement procedures to address the following: - Osteoarthritis - Post-traumatic arthrosis - Focal avascular necrosis of the humeral head - Previous surgeries of the shoulder that do not compromise the fixation The INHANCE SHOULDER SYSTEM with a humeral stem is intended for use in anatomic total or hemi-shoulder replacement procedures to address the following: - Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis. - Rheumatoid arthritis. - Revision where other devices or treatments have failed. - Correction of functional deformity. - Fractures of the humeral head (with Short Humeral Stems) - Fractures of the proximal humerus (with Standard or Long Stems) - Difficult clinical management problems where other methods of treatment may not be suitable or may be inadequate.
Device Story
The INHANCE Hybrid Anatomic Glenoid Implant is a prosthetic component for anatomic total shoulder arthroplasty. It consists of a cross-linked, Vitamin E-stabilized UHMWPE articulating surface and a central Ti-6Al-4V titanium alloy fixation post. The device is implanted by an orthopedic surgeon in a clinical setting. The glenoid component is intended for cemented fixation, while the humeral components (stemmed or stemless) are used for cementless or cemented fixation. The device articulates with the humeral head to restore joint function, address degenerative disease, and correct deformities. It benefits patients by providing a stable, wear-resistant joint surface, reducing pain, and improving shoulder mobility.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including Range of Motion (ASTM F1378), biocompatibility (ISO 10993-1), glenoid wear rate, torsional resistance (ASTM F2028), axial pull-out, Ti6Al4V post disassembly, and MRI compatibility (ASTM F2052, F2213, F2182, F2119).
Technological Characteristics
Materials: Cross-linked Vitamin E UHMWPE (ASTM F2695) and additively manufactured Ti-6Al-4V titanium alloy. Design: Five sizes (21.5mm to 31.5mm). Fixation: Cemented glenoid, cementless/cemented humeral components. Sterilization: Validated via VDmax method (ISO 11137-1/2) to SAL 10^-6. Packaging: ISO 11607-1/2 compliant with 5-year shelf life.
Indications for Use
Indicated for patients requiring anatomic total or hemi-shoulder replacement due to osteoarthritis, post-traumatic arthrosis, focal avascular necrosis, rheumatoid arthritis, functional deformity, or proximal humeral fractures. Contraindicated for cases where fixation is compromised or other methods are suitable.
Regulatory Classification
Identification
A shoulder joint metal/polymer/metal nonconstrained or semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits movement in one or more planes. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral component made of alloys such as cobalt-chromium-molybdenum (Co-Cr-Mo) and titanium-aluminum-vanadium (Ti-6Al-4V) alloys, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, or a combination of an articulating ultra-high molecular weight bearing surface fixed in a metal shell made of alloys such as Co-Cr-Mo and Ti-6Al-4V. The humeral component and glenoid backing have a porous coating made of, in the case of Co-Cr-Mo components, beads of the same alloy or commercially pure titanium powder, and in the case of Ti-6Al-4V components, beads or fibers of commercially pure titanium or Ti-6Al-4V alloy, or commercially pure titanium powder. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. This generic type of device is designed to achieve biological fixation to bone without the use of bone cement.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance: Shoulder Joint Metal/Polymer/Metal Nonconstrained or Semi-Constrained Porous-Coated Uncemented Prosthesis.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Depuy Ireland UC % Russ Parrott Chief Technology Officer Ignite Orthopedics LLC 700 Park Avenue Suite F Winona Lake, Indiana 46590 June 8. 2022
Re: K212933
Trade/Device Name: INHANCETM Hybrid Anatomic Glenoid Implant Regulation Number: 21 CFR 888.3670 Regulation Name: Shoulder joint metal/polymer/metal nonconstrained or semi-constrained porouscoated uncemented prosthesis Regulatory Class: Class II Product Code: MBF, KWT, KWS, PAO, HSD Dated: May 9, 2022 Received: May 10, 2022
Dear Russ Parrott:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For Jiping Chen, M.D., Ph.D., M.P.H. Acting Division Director DHT6A:Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known) K212933
#### Device Name
INHANCE™ Hybrid Anatomic Glenoid Implant
Indications for Use (Describe) Anatomic Total Shoulder or Hemi-Shoulder
The INHANCE SHOULDER SYSTEM with the humeral stemless anchor is intended for use in anatomic total shoulder replacement procedures to address the following:
- · Osteoarthritis
- · Post-traumatic arthrosis
- Focal avascular necrosis of the humeral head
- · Previous surgeries of the shoulder that do not compromise the fixation
The INHANCE SHOULDER SYSTEM with a humeral stem is intended for use in anatomic total or hemi-shoulder replacement procedures to address the following:
- · Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis.
- · Rheumatoid arthritis.
- · Revision where other devices or treatments have failed.
- · Correction of functional deformity.
- · Fractures of the humeral head (with Short Humeral Stems)
- · Fractures of the proximal humerus (with Standard or Long Stems)
- · Difficult clinical management problems where other methods of treatment may not be inadequate.
#### Fixation Methods
The humeral stems are intended or cementless use. The humeral stemless anchor is intended for cementless use. The glenoid is intended for cemented use only.
Type of Use (Select one or both, as applicable)
| <div> <input checked="true" type="checkbox"/> <span>For Scientific Research/STEM Education Purposes</span> </div> |
|-------------------------------------------------------------------------------------------------------------------|
| <div> <input type="checkbox"/> <span>For Other Health/STEM Workforce Purposes</span> </div> |
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| 510(k) Summary | |
|--------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Prepared: | June 6, 2022 |
| Submitter: | DePuy Ireland UC<br>Loughbeg<br>Ringaskiddy<br>CO. CORK Munster, IE |
| Contact: | Russ Parrott<br>Chief Technology Officer<br>Phone: 574.527.2864<br>russ.parrott@igniteorthopedics.com |
| Proprietary Name: | INHANCE™ Hybrid Anatomic Glenoid Implant |
| Common Name: | Shoulder Arthroplasty System |
| Classification: | Shoulder Joint, Metal/Polymer/Metal, Non-Constrained or Semi-Constrained, Porous Coated, Uncemented Prosthesis (21 CFR §888.3670); Class II<br>Prosthesis, Shoulder, Non-Constrained, Metal/Polymer, Cemented (21 CFR §888.3650); Class II<br>Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer, Cemented (21 CFR §888.3660); Class II<br>Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer + Additive, Cemented (21 CER §888.3660); Class II<br>Prosthesis, Shoulder, Hemi-, Humeral, Metallic Uncemented (21 CER §888.3690); Class II |
| Product Codes: | MBF, KWT, KWS, PAO, HSD |
| Predicate Devices: | K202716 - Ignite Anatomic Shoulder System; Ignite Orthopedics LLC<br>K060694 - Modular Hybrid Glenoid, Biomet Manufacturing Corp |
{4}------------------------------------------------
# Reference Devices: K052472 - DePuy Global Shoulder Crosslink Glenoid, DePuy Orthopedics, Inc.
K212737 - INHANCE Reverse Shoulder System, DePuy Ireland UC
# Device Description:
The INHANCE™ Hybrid Anatomic Glenoid Implants are offered in five sizes: X-Small (21.5mm), Small (24.0mm), Medium (26.5mm), Large (29.0mm), and X-Large (31.5mm). The Glenoid Implants consist of a Cross-linked, Vitamin E Ultra High Molecular Weight Polyethylene (Cross-linked, VE UHMWPE) articulation surface and fixation ring along with an additively manufactured Ti-6Al-4V (titanium alloy) central fixation post.
The INHANCE™ Hybrid Anatomic Glenoid Implants have a lateral surface that is concave and designed to articulate with the Humeral Heads from the Ignite Anatomic Stemmed and Stemless Shoulder Systems that are indicated for use in total shoulder arthroplasty.
### Intended Use/Indications for Use:
#### Anatomic Total Shoulder or Hemi-Shoulder
The INHANCE SHOULDER SYSTEM with the humeral stemless anchor is intended for use in anatomic total shoulder replacement procedures to address the following:
- · Osteoarthritis
- Post-traumatic arthrosis
- Focal avascular necrosis of the humeral head
- Previous surgeries of the shoulder that do not compromise the fixation
The INHANCE SHOULDER SYSTEM with a humeral stem is intended for use in anatomic total or hemi-shoulder replacement procedures to address the following:
- · Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis.
- Rheumatoid arthritis.
- · Revision where other devices or treatments have failed.
- Correction of functional deformity.
- · Fractures of the humeral head (with Short Humeral Stems)
{5}------------------------------------------------
- Fractures of the proximal humerus (with Standard or Long Stems)
• Difficult clinical management problems where other methods of treatment may not be suitable or may be inadequate.
#### Fixation Methods
The humeral stems are intended for cemented or cementless use. The humeral stemless anchor is intended for cementless use. The glenoid is intended for cemented use only.
# Summary of Technologies/Substantial Equivalence:
The INHANCE™ Hybrid Anatomic Glenoid Implants are substantially equivalent to the predicate device in terms of their intended use and indications, material, design, sizes, and mechanical properties. While the predicate device is composed entirely of cross-linked VE UHMWPE machined from ram extruded bar stock, the cross-linked VE UHMWPE body of the INHANCE Hybrid Glenoid Implants are direct compression molded into an additively manufactured Ti-6Al-4V central porous post. Differences between the subject device system and the predicate device system do not raise different questions of safety and effectiveness.
### Non-Clinical Testing:
The INHANCE™ Hybrid Anatomic Glenoid Implants underwent non-clinical testing and analyses to support a determination of substantial equivalence to the predicate device. The following were completed:
# Range of Motion (RoM) Evaluation
An evaluation was conducted to ensure the RoM of the worst-case subject device components meet established specifications per ASTM F1378. The RoM targets were met.
#### Biocompatibility Assessments
The contact classification for the subject device is Implant, Bone/Tissue with permanent contact (>30 days). A Biocompatibility Assessment was completed and provided per ISO 10993-1 and FDA Guidance Document Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process. The subject device was found to be biocompatible.
# Porous Structure Characterization
The porous structure used for the subject device is identical to the porous structure that was applied to the implants cleared under K202716 and K203108.
{6}------------------------------------------------
# Characterization of AO-HXLPE
The antioxidant highly crosslinked UHMWPE (AO-HXLPE) was fully characterized and conforms to ASTM F2695. The Vitamin E Ultra High Molecular Weight Polyethylene (Crosslinked, VE UHMWPE) material used for the INHANCE Hybrid Anatomic Glenoids is identical in base resin, blending concentration of antioxidant, and crosslinking irradiation dose to the VE UHMWPE material that was used on the devices cleared under K202716.
### Evaluation of Glenoid Wear Rate
The direct compression molded articulating surfaces on the INHANCE Hybrid Anatomic Glenoid Implants do not present a new worst case for wear testing when compared to articulating surfaces of the same geometry and material that are machined from ram extruded bar stock.
### Glenoid Loosening Testing
The INHANCE™ Hybrid Anatomic Glenoids were tested according to ASTM F2028. The acceptance criteria were met.
### Glenoid Torsional Resistance Testing
The INHANCE™ Hybrid Anatomic Glenoids were subjected to static torsional resistance testing. The acceptance criteria were met.
#### Axial Pull-out Testing
The INHANCE™ Hybrid Anatomic Glenoids were subjected to axial pullout testing. The acceptance criteria were met.
# Ti6Al4V Post Disassembly Evaluation
A mechanical evaluation was conducted to study the risk of dissociation of the Ti6Al4V Post and the VE XLPE Glenoid body. The acceptance criteria were met.
# MRI Compatibility
Quantitative data was obtained to inform Magnetic Resonance Imaging (MRI) Conditional Labeling through the following evaluations:
- Force: Static Magnetic Field Induced Displacement Force per ASTM F2052-15 ●
- Torque: Static Magnetic Field Induced Torque per ASTM F2213-17 ●
- Heating: Radiofrequency field (RF) induced heating per ASTM F2182-19e2 .
- Image Quality: Susceptibility induced image artifacts per ASTM F2119-07 ●
{7}------------------------------------------------
#### Shelf Life Evaluation
A shelf life evaluation per ISO 11607-1 and ISO 11607-2 was completed on the packaging materials that make up the sterile barrier. A five-year shelf life was established based on the resultant data.
### Sterilization Validation
Sterilization validation was completed using the VDmax method specified in ISO 11137-1 and ISO 11137-2. The Sterility Assurance Level (SAL) was found to be 10-6.
# Clinical Testing:
Clinical testing was not necessary to demonstrate substantial equivalence of the INHANCE™ Hybrid Anatomic Glenoid Implants to the predicate device.
### Conclusion:
A comparison of the subject and predicate devices, including comparison of the intended use, technological characteristics, and non-clinical testing results has demonstrated that the subject device has a safety and effectiveness profile equivalent to that of the predicate device. Thus, the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.