Disposable Vinyl/Nitrile Blend Medical Examination Gloves

K212899 · Bytech(Dongtai) Co., Ltd. · LYZ · Dec 11, 2021 · General Hospital

Device Facts

Record IDK212899
Device NameDisposable Vinyl/Nitrile Blend Medical Examination Gloves
ApplicantBytech(Dongtai) Co., Ltd.
Product CodeLYZ · General Hospital
Decision DateDec 11, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

The Disposable Vinyl/Nitrile Blend Medical Examination Gloves is intended for medical purposes that is worn on the examiners hands to prevent contamination between patient and examiner.

Device Story

Disposable vinyl/nitrile blend medical examination gloves; non-sterile; ambidextrous; blue color. Used in medical settings by clinicians to prevent cross-contamination between patient and examiner. Device acts as a physical barrier. Performance verified through bench testing for physical dimensions, tensile strength, elongation, and freedom from holes. No clinical data included.

Clinical Evidence

No clinical data. Bench testing only. Testing included biocompatibility (ISO 10993-5, -10, -11), powder residue (ASTM D6124), freedom from holes (ASTM D5151), and physical properties/dimensions (ASTM D5250, ASTM D412). All tests met acceptance criteria.

Technological Characteristics

Vinyl/nitrile blend material; non-sterile; ambidextrous; powder-free. Dimensions and physical properties comply with ASTM D5250, ASTM D5151, and ASTM D6124. Biocompatibility per ISO 10993-5, 10993-10, and 10993-11.

Indications for Use

Indicated for use as a disposable medical glove worn on the examiner's hands to prevent contamination between patient and examiner.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. December 11, 2021 Bytech(Dongtai) Co., Ltd. Boyle Wang Manager Shanghai Truthful information Technology Co., Ltd. Room608,No.738,Shangcheng Rd.,Pudong Shanghai, Shanghai 200120 China Re: K212899 Trade/Device Name: Disposable Vinyl/Nitrile Blend Medical Examination Gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LYZ Dated: August 30, 2021 Received: September 13, 2021 Dear Boyle Wang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For Clarence W. Murray, III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### Indications for Use 510(k) Number (if known) K212899 Device Name Disposable Vinyl/Nitrile Blend Medical Examination Gloves Indications for Use (Describe) The Disposable Vinyl/Nitrile Blend Medical Examination Gloves is intended for medical purposes that is worn on the examiners hands to prevent contamination between patient and examiner. Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # K212899 510(k) Summary This summary of 510(k) is being submitted in accordance with 21 CFR 807.92. #### 1.0 Submitter's Information Name: Bytech (Dongtai) Co., Ltd. Address: XinYuan Industrial Park,TangYang Town, DongTai City, Jiangsu Province, China Phone Number: Contact: Mr. Wang Cheng Date of Preparation: 2021.08.30 ## Designated Submission Correspondent Mr. Boyle Wanq Shanghai Truthful Information Technology Co., Ltd. Room 608, No. 738 Shangcheng Rd., Pudong Shanghai, 200120 China Tel: +86-21-50313932 Email: Info@truthful.com.cn #### 2.0 Device information Trade name: Disposable Vinyl/Nitrile Blend Medical Examination Gloves Common name: Disposable Vinyl/Nitrile Blend Medical Examination Gloves Classification name: Non-powdered Patient Examination Glove Model(s): S, M, L, XL # 3.0 Classification Production code: LYZ Regulation number: 21CFR880.6250 Classification: Class I Panel: General Hospital # 4.0 Predicate device information Tangshan Hongyun Healthcare Products Co.,Ltd. Manufacturer: Device: Disposable Vinyl Nitrile Synthetic Gloves Powder Free 510(k) number: K211262 {4}------------------------------------------------ ## 5.0 Indication for Use The Disposable Vinyl/Nitrile Blend Medical Examination Gloves is intended for medical purposes that is worn on the examiners hands to prevent contamination between patient and examiner. #### 6.0 Device description The subject device is powder free vinyl patient examination gloves. The subject device is blue. The design of subject device is addressing the standards as ASTM D6124, ASTM D5151, and ASTM D5250. The subject device is non-sterile. #### 7.0 Summary comparing technological characteristics with predicate device | Item | Subject device | Predicated device | Comparison | |--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------| | 510(k) number | K212899 | K211262 | / | | Product Code | LYZ | LYZ | Same | | Regulation No. | 21CFR880.6250 | 21CFR880.6250 | Same | | Class | I | I | Same | | Intended Use | The Disposable<br>Vinyl/Nitrile Blend<br>Medical Examination<br>Gloves is a disposable<br>device intended for<br>medical purposes that is<br>worn on the examiner's<br>hands to prevent<br>contamination between<br>patient and examiner. | The Disposable Vinyl<br>Nitrile Synthetic Gloves<br>Powder Free is a<br>disposable device<br>intended for medical<br>purposes that is worn<br>on the examiner's<br>hands to prevent<br>contamination between<br>patient and examiner. | Same | | Powdered or Powered free | Powdered free | Powdered free | Same | | Design Feature | ambidextrous | ambidextrous | Same | | Labeling Information | Single use, powder free,<br>device color, device<br>name, glove size and<br>quantity, Vinyl<br>Examination Gloves,<br>Non-Sterile | Single use, powder<br>free, device color,<br>device name, glove size<br>and quantity, Vinyl<br>Examination Gloves,<br>Non-Sterile | Same | #### Table1-General Comparison #### Table2 Device Dimensions Comparison | Predicate | Designation | Size | Tolerance | |-----------|-------------|------|-----------| | | | | | {5}------------------------------------------------ | Device(K211262) | | S | M | L | XL | | |-----------------|----------------|------|-----|-----|-----------|---------| | | Length, mm | 230 | 230 | 230 | 230 | min | | | Width, mm | 85 | 95 | 105 | 115 | $\pm$ 5 | | | Thickness, mm: | | | | | | | | Finger | 0.05 | | | | min | | | Palm | 0.05 | | | | min | | Subject Device | Designation | Size | | | Tolerance | | | | | S | M | L | XL | | | | Length, mm | 230 | 230 | 230 | 230 | min | | | Width, mm | 85 | 95 | 105 | 115 | $\pm$ 5 | | | Thickness, mm: | | | | | | | | Finger | 0.05 | | | | min | | | Palm | 0.05 | | | | min | | Remark | Same | | | | | | #### Table3 Performance Comparison | Item | | Subject device | Predicated device | Comparison | | |------------------------|----------------|----------------------------------------------------------------------|----------------------------------------------------------------------|------------|------| | Colorant | | Blue | Blue | SAME | | | Physical<br>Properties | Before Aging | Tensile<br>Strength | SAME | 11MPa, min | SAME | | | | Ultimate<br>Elongation | SAME | 300%min | SAME | | | After<br>Aging | Tensile<br>Strength | SAME | 11MPa, min | SAME | | | | Ultimate<br>Elongation | SAME | 300%min | SAME | | | | Comply with ASTM D5250 | Comply with ASTM D5250 | SAME | | | Freedom from Holes | | Be free from holes when tested in accordance with ASTM D5151 AQL=2.5 | Be free from holes when tested in accordance with ASTM D5151 AQL=2.5 | SAME | | | Powder Content | | 0.15-0.18 mg per glove | 0.07 mg per glove. Meet the requirements of ASTM D6124 | SIMILAR | | Analysis: The powder content is different with that of the predicate, but they all meet the requirements of ASTM D5250 ASTM D6124 - 06(2017). #### Table4 Safety Comparison | Item | Subject device | Predicated device | Comparison | |------------------|------------------------------------------------------------------|----------------------------|------------| | Material | Vinyl | Vinyl | SAME | | Biocompatibility | Irritation<br>Under the conditions of the study, not an irritant | Comply with<br>ISO10993-10 | SAME | {6}------------------------------------------------ | | Sensitization | Under conditions of the study, not a sensitizer. | | | |--------------------|-------------------|-----------------------------------------------------------------------------------------------------------------|--------------------------------------------------|---------| | | Cytotoxicity | Under the conditions of the study, the device is potentially cytotoxic | Comply with ISO10993-5 | SIMILAR | | | Systemic toxicity | Under the conditions of the study, the device does not elicit a systemic toxicity response in the model animal. | Complies with ISO 10993-11 Third edition 2017-09 | | | Label and Labeling | | Meet FDA's Requirement | Meet FDA's Requirement | SAME | Analysis: The material of subject device are different with that of the predicate device, but biocompatibility test has been performed on subject device and the test result can meet the requirements of ISO 10993 standards. #### 8.0 Summary of Non-Clinical Performance Testing The following performance data has been provided to demonstrate that the subject device meets the acceptance criteria in the standard. | Table 5 Summary of Non-Clinical Performance Testing | | |-----------------------------------------------------|--| |-----------------------------------------------------|--| | No. | Name of the Test<br>Methodology / Standard | Purpose | Acceptance Criteria | Results | |-----|--------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 1 | ISO 10993-5 | Cytotoxicity | Non-cytotoxic | Under conditions of the study, did<br>not show potential toxicity to<br>L-929 cells. | | 2 | ISO 10993-10 | Irritation | Non-irritating | Under the conditions of the study,<br>not an irritant. Pass | | 3 | ISO 10993-10 | Sensitization | Non-sensitizing | Under conditions of the study, not<br>a sensitizer. Pass | | 4 | ISO 10993-11 | Systemic toxicity | Non-Systemic toxic | Under the conditions of the study,<br>the device does not elicit a<br>systemic toxicity response in the<br>model animal. Pass | | 5 | ASTM D6124-06 | This standard is<br>designed to<br>determine the<br>amount ofresidual<br>powder (or<br>filter-retained<br>mass) found | powder residue limit of<br>2.0 mg | 0.15mg -0.18 mg /glove | | | | | | | | 6 | ASTM<br>D5151-06(Reapproved2<br>015) | on medical gloves<br>This test method<br>covers the<br>detection of holes<br>in<br>medical gloves. | Samples number: 125 gloves<br>AQL: 2.5 (ISO 2859)<br>Criterion ≤7 gloves<br>for water leakage | no glove water leakage found | | 7 | ASTM D5250 | Physical<br>Dimensions Test | Sterility: no need<br>Freedom from holes:<br>pl. Refer to No. 5 in<br>table 5<br>Dimensions:<br>Length(mm):≥230;<br>Width(mm):<br>S: 85±5;<br>M: 95±5;<br>L: 105±5;<br>XL: 115±5;<br><br>Thickness (mm):<br>Finger: ≥0.05 Palm:<br>≥0.05 | N.A.<br>Please refer to No. 5 in table 5<br><b>Lot no. 1:</b><br>Dimensions:<br>S: width: 83-88 mm<br>Length 231-238 mm (one data<br>227mm, the number of<br>non-conforming is 1, acceptable)<br>M: width 92-96 mm<br>Length 233-236 mm<br>L: width 102-106 mm<br>Length 246-253 mm<br>XL: width 111-116 mm<br>Length 242-245 mm<br>Thickness:<br>Finger 0.10-0.11 mm<br>Palm 0.07-0.08 mm<br><br><b>Lot no.:2</b><br>Dimensions:<br>S: width: 82-85 mm<br>Length 232-235 mm<br>M: width 92-96 mm<br>Length 232-236 mm<br>L: width 103-106 mm<br>Length 243-249 mm<br>XL: width 111-116 mm<br>Length 237-243 mm<br>Thickness:<br>Finger 0.11-0.12 mm<br>Palm 0.07-0.08 mm<br><br><b>Lot no.:3</b><br>Dimensions:<br>S: width: 82-87 mm | {7}------------------------------------------------ {8}------------------------------------------------ | | | | | | | M: width 93-96mm<br>Length 230-239 mm | |---|-----------|---------------------|----------------|------------------------|------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | | | | | | L: width 103-107mm<br>Length 243-249 mm | | | | | | | | XL: width 112-116 mm<br>Length 244-249 mm | | | | | | | | Thickness: | | | | | | | | Finger 0.11 mm | | | | | | | | Palm 0.07-0.08 mm | | | | | | | | Pass | | 8 | ASTM D412 | Physical properties | Befor<br>Aging | Tensile<br>Strength | ≥<br>11MPa | Lot 1: 12.7-24 MPa<br>Lot 2: 15.1-25.1 MPa<br>Lot 3: 14.5-23.5 MPa<br>Pass | | | | | | Ultimate<br>Elongation | ≥<br>300% | Lot 1: 300.08-397.467 %<br>Lot 2: 300.547-380.29 %<br>(one data 228.080, the number<br>of non-conforming is 1,<br>acceptable)<br>Lot 3: 303.365-389.716 %<br>Pass | | | | | After<br>Aging | Tensile<br>Strength | ≥<br>11MPa | Lot 1: 14.2-22.4 MPa<br>Lot 2: 16.0-23.8 MPa<br>Lot 3: 14.4-23.3 MPa<br>Pass | | | | | | Ultimate<br>Elongation | ≥<br>300% | Lot 1: 301.276-394.245 %<br>Lot 2: 300.79-399.170 %<br>Lot 3: 307.674-399.758 %<br>Pass | #### 9. Summary of Clinical Performance Test No clinical study is included in this submission. #### 10.0 Conclusion The conclusions drawn from the nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicated device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...