K212897 · Shenzhen Century Dongyuan Technology CO , Ltd. · OHT · Nov 4, 2021 · General, Plastic Surgery
Device Facts
Record ID
K212897
Device Name
IPL Hair Removal Device, model: AI01
Applicant
Shenzhen Century Dongyuan Technology CO , Ltd.
Product Code
OHT · General, Plastic Surgery
Decision Date
Nov 4, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The IPL Hair Removal Device (Model: AI01) is intended for the removal of unwanted body hair.
Device Story
Hand-held IPL device for permanent reduction of unwanted body hair; utilizes Xenon Arc Flashlamp to deliver thermal energy to hair follicles; operates via external power adapter; includes skin sensor to ensure full contact before light pulse emission; activated by finger switch; intended for home use; provides non-invasive hair growth reduction; includes goggles for eye protection.
Clinical Evidence
No clinical data. Bench testing only, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), photobiological safety (IEC 62471), and biocompatibility (ISO 10993-5/10).
Technological Characteristics
Hand-held IPL device; Xenon Arc Flashlamp source; wavelength 510-1100nm; fluence 3.8-5.2 J/cm²; pulse duration 3ms; PC material construction; external power supply; skin contact sensor; IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-57, IEC 62471, ISO 10993-5, ISO 10993-10 compliance; software developed per IEC 62304.
Indications for Use
Indicated for the removal of unwanted body hair in adult patients.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 4, 2021
Shenzhen Century Dongyuan Technology Co Ltd. % You Yijie Manager Qimmiq Medical Consulting Service Co., Ltd. Rm.1711, Building K. No.101 Science Ave International Creative Valley Guangzhou, Guangdong 510663 China
Re: K212897
Trade/Device Name: IPL Hair Removal Device, model: AI01 Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: ONF Dated: August 27, 2021 Received: September 10, 2021
Dear You Yijie:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K212897
Device Name IPL Hair Removal Device, model: AI01
#### Indications for Use (Describe)
The IPL Hair Removal Device (Model: AI01) is intended for the removal of unwanted body hair.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 5: 510(k) Summary
### 1. Submitter's Information
#### Establishment Registration Information
Name: Shenzhen Century Dongyuan Technology Co Ltd. Address: Floor 2, Building B1, Gonghe Fourth Industrial Zone, Shajing Street, Bao'an District, Shenzhen City, Guangdong Province China Contact Person of applicant Name: Zhang Benrong Address: Floor 2, Building B1, Gonghe Fourth Industrial Zone, Shajing Street, Bao'an District, Shenzhen City, Guangdong Province China TEL: +86 18565833539 Email: aijofi01@163.com
#### Contact Person of the Submission:
Name: Yijie You Address: RM.1711, Building K, NO.101 Science Ave International Creative Valley Development Zone, Guangzhou China Floor 2, Building B1, Gonghe Fourth Industrial Zone, Shajing Street, Bao'an District, Shenzhen City, Guangdong Province China TEL: +86 020-8224 5821 FAX: +86 020-8224 5821 Email: Jet.you@qimmiq-med.com
Date prepared: Oct. 20, 2021
#### 2. Device Information
| Trade Name: | IPL Hair Removal Device, model: AI01 |
|-----------------------|----------------------------------------------------------|
| Classification name: | Laser Surgical Instrument For Use In General And Plastic |
| | Surgery And In Dermatology |
| Common or Usual Name: | Powered Light Based Non-Laser Surgical Instrument With |
| | Thermal Effect |
| Review panel: | General & Plastic Surgery |
| Product code: | ONF |
| Regulation Class: | II |
| Regulation Number: | 878.4810 |
{4}------------------------------------------------
### 3. Predicate Device Information
| 510(k) submitter/holder: | Touchbeauty Beauty & Health (Shenzhen) Co., Ltd. |
|--------------------------|--------------------------------------------------|
| 510(K) Number: | K183217 |
| Trade Name: | IPL Hair Removal Device |
| Model: | TB-1755 |
| Classification name: | Laser Surgical Instrument For Use In General And |
| | Plastic Surgery And In Dermatology |
| Review panel: | General & Plastic Surgery |
| Product code: | OHT, ONF |
| Regulation Class: | II |
| Regulation Number: | 878.4810 |
## 4. Device description
IPL Hair Removal Device, model: Al01 is a device for the permanent reduction of hair growth based on Intense Pulsed Light (IPL). It works below the skin's surface
and does not involve any cutting or pulling, reducing hair growth with minimal pain. The device is only powered by the external power adapter and its IPL emission activation is by finger switch. The device contains a Xenon Arc Flashlamp and a skin sensor to detect appropriate skin contact. If the device is not properly applied to the treatment area (in full contact with the skin), the device cannot emit the treatment light pulses. The device incorporates Intense Pulse Light (IPL) technology. The purpose of the light is to heat the root where the hair grows.
Device includes a main unit, an adaptor, and accessories, accessories include goggle.
#### Principle of operation:
The device works by use of thermal energy of Intense Pulse Light (IPL) to kill hair follicles for hair removal. The device generally targets skin (including dermis, epidermis), underlying soft tissue and hair follicles.
### 5. Indications for Use
The IPL Hair Removal Device (Model: Al01) is intended for the removal of unwanted body hair.
### 6. Summary of technological characteristics of device
#### compared to the predicate devices (K183217)
{5}------------------------------------------------
# K212897
| SE<br>Comparisons | Subject device<br>(IPL Hair Removal Device,<br>model: AI01) | Predicate device<br>(IPL Hair Removal Device, Model:<br>TB-1755) | Discussion of<br>difference |
|--------------------------------------------------------|----------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------|
| 510K Number | / | K183217 | / |
| Classification | 21CFR 878.4810 | 21CFR 878.4810 | Same |
| Product Code | ONF | OHT, ONF | Same |
| FDA Class | II | II | Same |
| Indications for Use | The IPL Hair Removal Device (Model:<br>AI01) is intended for the removal of<br>unwanted body hair. | The IPL Hair Removal Device (Model:<br>TB-1755) is an Over the Counter<br>device intended for the removal of<br>unwanted body hair. | Same<br>(Discussion is<br>indicated in<br>D1) |
| Model | AI01 | TB-1755 | / |
| Environment of Use | Home use | Home use | Same |
| Design | Hand-hold | Hand-hold | Same |
| Patient Population | Adult | Adult | Same |
| Material of Patient<br>contact<br>components | PC | ABS | Different<br>(Discussion is<br>indicated in<br>D2) |
| Biocompability<br>testing | Meets ISO 10993- 5<br>ISO 10993-10 | Meets ISO 10993- 5<br>ISO 10993-10 | Same |
| Single Patient,<br>multi-use | Yes | Yes | Same |
| Patient Interface | Buttons | Buttons | Same |
| Technology | Intense Pulse Light (IPL) | Intense Pulse Light (IPL) | Same |
| Dimensions | 140*203*56 | 182*72.4*69.2mm | Different<br>(Discussion is<br>indicated in<br>D3) |
| Power source | an external power supply | an external power supply | Same |
| Light source | Xenon Arc Flashlamp | Xenon Arc Flashlamp | Same |
| Wavelength | 510nm~1100nm | 510nm~1100nm | Same |
| Spot Size | 3.1 cm² | 3.1 cm² | Same |
| Max. Fluence<br>(J/cm²) | 3.8-5.2 J/cm² | 3.8-5.2 J/cm² | Same |
| Pulse duration | 3 milliseconds | 3 milliseconds | Same |
| Output energy | 12-16 J | 12-16 J | Same |
| Pulsing<br>Control | Finger switch | Finger switch | Same |
| Output<br>Channel | One channel | One channel | Same |
| Delivery | Direct Illumination to Tissue | Direct Illumination to Tissue | Same |
| Software | Yes | Yes | Same |
| Control | | | |
| Electrical<br>safety, EMC,<br>Biological<br>Evaluation | IEC 60601-1 | IEC 60601-1 | Same |
| | IEC 60601-1-2 | IEC 60601-1-2 | |
| | IEC 60601-1-11 | IEC 60601-1-11 | |
| | IEC 60601-2-57 | IEC 60601-2-57 | |
| | IEC 62471 | IEC 62471 | |
| | ISO 10993-5 | ISO 10993-5 | |
| | ISO 10993-10 | ISO 10993-10 | |
{6}------------------------------------------------
# K212897
#### The discussion of differences exist between the subject and predicate devices is listed in following:
- D1: The subject device has same indication with equivalent device regarding remove unwanted hair, the difference is that the equivalent device is for OTC use and the subject device is for Prescription Use, this difference does not raise new questions of safety.
- D2: The subject device has been validated for cytotoxicity per ISO 10993- 5 and Irritation as well as Sensitization per ISO 10993-10 with positive results, therefore, the material difference of subject device with Predicate device TB-1755 (K183217) do not raise new questions of safety and effectiveness.
- D3: The difference of dimensions will not affect the safety and effectiveness.
### 7. Discussion of Non-Clinical Tests Performed for Safety and
#### effectiveness are as follows
The recognized consensus standards for safety of medical electrical equipment: ANSI AAMI ES60601-1, IEC 60601-1-11 for safety, IEC 60601-1-2 for electromagnetic compatibility, IEC 60601-2-57:2011 for performance and IEC 62304 for software verification, and ISO 10993-5:2009 for Cytotoxicity endpoints, ISO 10993-10:2010 for Sensitization and Irritation endpoints are complied. See below table for details:
| Standards | Standards Name |
|-----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ANSI AAMI ES60601-<br>1:2005/(R)2012 and<br>A1:2012 | Medical Electrical Equipment - Part 1: General Requirements<br>For Basic Safety And Essential Performance |
| IEC 60601-1-2: 2014 | Medical Electrical Equipment -- Part 1-2: General Requirements<br>For Basic Safety And Essential Performance -- Collateral<br>Standard: Electromagnetic Disturbances -- Requirements And<br>Tests |
| IEC 60601-1-11: 2015 | Medical Electrical Equipment - Part 1-2: General Requirements<br>For Basic Safety And Essential Performance - Collateral |
{7}------------------------------------------------
| | Standard: Electromagnetic Disturbances - Requirements And<br>Tests |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ISO 60601-2-57: 2011 | Particular requirements for the basic safety and essential<br>performance of non-laser light source equipment intended for<br>therapeutic, diagnostic, monitoring and cosmetic/aesthetic use |
| IEC 62471: 2006 | Photobiological safety of lamps and lamp systems |
| ISO 10993-5:2009 | Biological evaluation of medical devices - Part 5: Tests for in<br>vitro cytotoxicity |
| ISO 10993-10:2010 | Biological evaluation of medical devices - Part 10: Tests for<br>irritation and skin sensitization |
| IEC 62304:2006+A1:2015 | Medical device software - Software life cycle processes |
Software verification and validation was performed for the subject device in accordance with Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices - Guidance for Industry and FDA Staff, May 2005.
## 8. Discussion of Clinical Accuracy Testing Performed
There was no clinical testing performed.
### 9. Conclusions
Based on performance testing, comparison and analysis, the subject device IPL Hair Removal Device, model Al01 is as safe, as effective, and performs as well as the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.