This video-bronchoscope is intended to provide an optical display of the pulmonary tract using a monitor (Screeni®) and to be used with endotherapy accessories and instruments. This video-bronchoscope is designed for use exclusively in a hospital environment.
Device Story
System comprises single-use video-bronchoscope (Broncoflex Agile/Vortex) and reusable display monitor (Screeni). Bronchoscope captures pulmonary tract images via distal tip sensor; transmits data to Screeni monitor for real-time visualization. Used in hospital settings by clinicians for airway examination and procedures requiring suction or endotherapy instruments. Screeni features touch interface, internal storage, and HDMI output for secondary display. System aids clinical decision-making by providing direct visual access to airways. Benefits include reduced cross-contamination risk due to single-use bronchoscope design.
Indicated for adult patients requiring optical visualization of the pulmonary tract and bronchial tree, and for use with endotherapy accessories and instruments. Designed for hospital use only.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
Predicate Devices
Broncoflex Agile, Broncoflex Vortex and Screeni (K202180)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 16, 2022
Axess Vision Technology Marie-Hélène Bacheley Regulatory Affairs and Ouality Assurance Manager Zone de la Liodière 6. rue de la Flottière Joué-lès-Tours, Indre-et-Loire 37300 France
Re: K212886
Trade/Device Name: Broncoflex Agile, Broncoflex Vortex, Screeni Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (flexible or rigid) and accessories Regulatory Class: Class II Product Code: EOQ Dated: October 29, 2021 Received: November 4, 2021
Dear Marie-Hélène Bacheley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known) K212886
Device Name
Broncoflex® Agile, Broncoflex® Vortex and Screeni®
Indications for Use (Describe) Broncoflex® Agile and Broncoflex® Vortex
This video-bronchoscope is intended to provide an optical display of the pulmonary tract using a monitor (Screeni®) and to be used with endotherapy accessories and instruments.
This video-bronchoscope is designed for use exclusively in a hospital environment.
Endotherapy accessories and instruments
The effective length of an endoscopic instrument should be at least 30 cm greater than the effective length of the endoscope.
| Model | Minimum compatible<br>endotracheal tube size | Minimum compatible dual<br>lumen endo-bronchial tube size | Maximum size of<br>endotherapy instruments |
|-------------------|----------------------------------------------|-----------------------------------------------------------|--------------------------------------------|
| Broncoflex Agile | $\geq 5.0$ mm | $\geq 35$ Fr. | $\leq 1.2$ mm |
| Broncoflex Vortex | $\geq 6.0$ mm | $\geq 41$ Fr. | $\leq 2.6$ mm |
Patient Population
The Broncoflex can only be used on adult patients.
#### Screeni®
This device is designed for use exclusively in a hospital environment, in combination with an Axess Vision Broncoflex® video-bronchoscope.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
# Broncoflex® Vortex, Broncoflex® Agile and Screeni® (Flexible Bronchoscope System)
## K212886
### 1. Submitter:
| Submitter's Name: | Axess Vision Technology | |
|------------------------|--------------------------------------------------------------------------------|------------------------------------------------|
| Address: | Zone de la Liodière<br>6 rue de la flottière<br>37300 Joué-lès-Tours<br>France | |
| Contact Person: | Marie-Hélène BACHELEY | Regulatory Affairs & Quality Assurance Manager |
| Address: | Zone de la Liodière<br>6 rue de la flottière<br>37300 Joué-lès-Tours | |
| Phone Number: | +33 (0)2 47 34 32 96 | |
| Fax Number: | NA | |
| Date Summary Prepared: | 20 August 2021 | |
#### 2. Device:
| Name of Device: | Broncoflex® Agile; Product Reference: 20030001<br>Broncoflex® Vortex; Product Reference: 10030001<br>Screeni®; Product Reference: 30030001 |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| Common or Usual Name: | Flexible Bronchoscope |
| Classification Name: | Bronchoscope (Flexible or Rigid) and accessories |
| Regulation Number: | 21 CFR 874.4680 |
| Device Class: | II |
| Product Code: | EOQ |
| Device Panel: | Ear Nose & Throat |
#### 3. Predicate Devices:
Broncoflex® Agile, Broncoflex® Vortex and Screeni® (Broncoflex® Agile or Vortex + Screeni® = the system) are substantially equivalent to the legally marketed predicate devices:
| Manufacturer: | Axess Vision Technology |
|----------------|-----------------------------------------------------|
| Trade Name: | Broncoflex® Agile<br>Broncoflex® Vortex<br>Screeni® |
| 510(K) Number: | K202180 |
This predicate has not been subject to a design-related recall.
{5}------------------------------------------------
#### 4. Device Description:
Broncoflex Agile and Broncoflex Vortex are a single-use video-bronchoscope which is part of a system made up of the endoscope (Broncoflex) and its reusable display monitor (Screeni). The videobronchoscope is intended to provide an optical display of the pulmonary tract using a monitor (Screeni) and to be used with endotherapy accessories and instruments. The applied part of the medical device is the whole of the flexible tubing containing the insertion tube, the articulated section and the distal tip (BF type applied part: BF). Broncoflex is sterile and supplied ready to use in its sealed packaging (sterilization method ethylene oxide). The Screeni is a reusable nonsterile medical device and is a video processor with integrated touch interface designed to display live imaging data captured by Broncoflex Agile and Broncoflex Vortex. The consumables (Broncoflex) are supplied in sets of 5 in a carton. Inside this box, each sterilized consumable is placed in a single sterile pouch. Screeni is supplied in a carton with all the accessories (quick mounting bracket, knurled screw, power supply, EU/US/UK/AU adaptors and video cable (HDMI/DVI)). Broncoflex Agile and Broncoflex Vortex are the same devices except for some dimensional aspects such as external diameter of insertion tube, distal tip outer diameter and working channel inner diameter. Indeed, Broncoflex Agile has an applied part with a smaller diameter than the Broncoflex Vortex.
#### 5. Indications for Use:
#### 1. Broncoflex Agile and Vortex
This video-bronchoscope is intended to prove an optical display of pulmonary tract using a monitor (Screeni®) and to be used with endotherapy accessories and instruments.
This video-bronchoscope is designed for use exclusively in a hospital environment.
#### Endotherapy accessories and instruments
The effective length of an endoscopic instrument should be at least 30 cm greater than the effective length of the endoscope.
| Model | Minimum compatible<br>endotracheal tube size | Minimum compatible<br>dual lumen endo-<br>bronchial tube size | Maximum size of<br>endotherapy instruments |
|-------------------|----------------------------------------------|---------------------------------------------------------------|--------------------------------------------|
| Broncoflex Agile | $\ge$ 5.0 mm | $\ge$ 35 Fr. | $\le$ 1.2 mm |
| Broncoflex Vortex | $\ge$ 6.0 mm | $\ge$ 41 Fr. | $\le$ 2.6 mm |
#### Patient Population
The Broncoflex can only be used on adult patients.
#### 2. Screeni
This device is designed for use exclusively in a hospital environment, in combination with an Axess Vision Broncoflex® video-bronchoscope.
#### 6. Comparison of technological characteristics with the predicate device:
#### 1. Broncoflex Agile and Vortex
Broncoflex Agile and Broncoflex Vortex are technologically substantially equivalent to the predicate devices. The subject devices do not raise any new issues of safety or effectiveness based on the similarities to the predicate devices.
{6}------------------------------------------------
| Type | Broncoflex® Vortex<br>Subject device | Broncoflex® Agile<br>Subject device | Broncoflex® Vortex<br>Predicate device | Broncoflex® Agile<br>Predicate device |
|-------------------------------------------|--------------------------------------|-------------------------------------|----------------------------------------|---------------------------------------|
| Product reference | 10030001 | 20030001 | 10030001 | 20030001 |
| Field of vision direction | 0° | 0° | 0° | 0° |
| Field angle | 87,5° | 87,5° | 87,5° | 87,5° |
| Field depth | 5 - 50 mm | 5 - 50 mm | 5 - 50 mm | 5 - 50 mm |
| High/low<br>deflection<br>angle | 200°/200° | 220°/220° | 200°/200° | 220°/220° |
| External<br>diameter<br>of insertion tube | 5.4 mm | 3.6 mm | 5.4 mm | 3.6 mm |
| Distal<br>end<br>outer<br>diameter | 5.6 mm | 3.9 mm | 5.6 mm | 3.9 mm |
| Operator channel inner<br>diameter | 2.8 mm | 1.4 mm | 2.8 mm | 1.4 mm |
| Working length | 605 mm | 605 mm | 605 mm | 605 mm |
| Lighting system | 2 LEDs | 2 LEDs | 2 LEDs | 2 LEDs |
| Image resolution | 400x400 | 400x400 | 400x400 | 400x400 |
| Sterilisation | Ethylene oxide | Ethylene oxide | Ethylene oxide | Ethylene oxide |
The Broncoflex Vortex and Broncoflex Agile are the same as the predicate devices in the following areas:
- Bronchoscope used to reach the target organs, tissues and subsystems (pulmonary tract) .
- Device inserted through oral or nasal route .
- Viewing of the upper airways and of the bronchial tree
- Performance of other procedures such as as performing examinations requiring suction of . secretions or the use of endotherapy accessories or instruments designed for use in combination with a bronchoscope
- . Single-use devices provided sterile
- . Field of vision direction
- . Lighting system (2 LEDs)
- . Image resolution
- Sterile with ETO .
#### 2. Screeni
Screeni is technologically substantially equivalent to the predicate device.
| Type | Screeni®<br>Subject device | Screeni®<br>Predicate device |
|---------------------------------------------|----------------------------------------------------------------|----------------------------------------------------------------|
| Product reference | 30030001 | 30030001 |
| ELECTRICAL CHARACTERISTICS | | |
| Power requirement | 100-240V AC / 50-60Hz / 0.6A | 100 - 240V AC; 50-60Hz; 0.6A |
| Power output | 15 VDC / 2A | 18V DC; 1.67A |
| Type of protection against<br>electrocution | Class 2 | Class 2 |
| Battery type | Lithium-Ion battery (11,25V -<br>2950mAh or 10,8V – 3350mAh) | Lithium-Ion battery (11,25V -<br>2950mAh or 10,8V – 3350mAh) |
| Battery Operation | At least 3 hours (for a new and fully<br>charged battery) | At least 3 hours (for a new and fully<br>charged battery) |
| MECHANICAL CHARACTERISTICS | | |
| Dimensions | L: 300 mm x H: 200 mm x D: 110 mm<br>(mounting bracket folded) | L: 300 mm x H: 200 mm x D: 110 mm<br>(mounting bracket folded) |
| Weight | 1.8 kg (with mounting bracket) | 1.8 kg (with mounting bracket) |
| Mounting interface | VESA 75 mm | VESA 75 mm |
Table 2: Screeni and aViewTechnical Characteristics
{7}------------------------------------------------
| TOUCH SCREEN | | |
|---------------------------------------|-------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|
| Resolution | 1280x800 | 800x480 |
| Display type | 10.1-inch TFT LCD | 10.1-inch TFT LCD |
| MISCELLANEOUS CHARACTERISTICS | | |
| IP ProtectionClassification<br>System | IP30 | IP30 |
| Exported file formats | Photos: .JPG<br>Videos: .AVI (h264 compression) | Photos: .JPG<br>Videos: .AVI (h264 compression) |
| TECHNICAL CHARACTERISTICS | | |
| Internal storage capacity | 16 Gb (can store up to 14 h of video or<br>more than 100,000 photos) | 16 Gb (can store up to 14 h of video or<br>more than 100,000 photos) |
| Connections | Type A (for USB key connection only) | Type A (for USB key connection only) |
| | Endoscope port | Endoscope port |
| | DC 15V / 2A input | DC 15V / 2A input |
| | HDMI port: video output to be used with a<br>DVI compatible monitor only and with the<br>HDMI/DVI cable supplied. | NA |
The Screeni is the same as the predicate device in the following areas:
- Viewing of the upper airways and of the bronchial tree ●
- Correct image orientation ●
- . Touch screen
- . Battery time/operation (3 hours)
- Battery type (Lithium-ion) ●
- . Storage capacity
- USB, endoscope and power out (electrical power) connection .
The Screeni is different from the predicate device in the following area:
· HDMI port for use with a DVI compatible monitor and the provided HDMI/DVI cable to duplicate the live image from the endoscope to a secondary monitor
This difference has no impact on performance, safety nor effectiveness of the devices. ณ
#### 7. Non-Clinical Performance data:
As part of demonstrating safety and effectiveness of Flexible Bronchoscope System (= Broncoflex® Agile or Broncoflex® Vortex + Screen®) and in showing substantial equivalence to the predicate devices that are subject to this 510(k) submission, Axess Vision Technology completed a number of non-clinical performance tests. The Flexible Bronchoscope System meets all the requirements for overall design, sterilization, biocompatibility, sterile packaging integrity, shelf life, transportation, electromagnetic compatibility and electrical safety results confirming that the design output meets the design inputs and specifications for the device.
The Flexible Bronchoscope System passed all the testing in accordance with internal requirements, national standards, and international standards shown below to support substantial equivalence of the subject device:
- . Electrical safety testing per IEC 60601-1:2012 and IEC 60601-2-18:2009
- . Electromagnetic Compatibility (EMC) testing per IEC 60601-1-2:2014 / A1:2020 (Ed. 4.1)
- Software verification and validation testing per IEC 62304:2006/A1 & FDA Guidance
#### 8. Conclusion:
Based on the indications for use, principle of operation, overall technological characteristics and the performance data provided, it is concluded that the functionality and the intended use of the Broncoflex® and Screeni® system is equivalent to the cited predicate devices (= Broncoflex® and Screeni® devices), and is therefore determined to be substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.