K212875 · Philips Medical Systems Nederland B.V. · JAK · May 13, 2022 · Radiology
Device Facts
Record ID
K212875
Device Name
Spectral CT on Rails
Applicant
Philips Medical Systems Nederland B.V.
Product Code
JAK · Radiology
Decision Date
May 13, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Indications for Use
The Spectral CT on Rails is a Computed Tomography X-Ray System intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories. The Spectral CT on Rails system acquires one CT dataset – composed of data from a higher-energy detected x-ray spectrum and a lower- energy detected x-ray spectra may be used to analyze the differences in the energy dependence of the attenuation coefficient of different materials. This allows for the generation of images at energies selected from the available spectrum and to provide information about the chemical composition of the body materials and/or contrast agents. Additionally, materials analysis provides for the quantification and graphical display of attenuation, material density, and effective atomic number. This information may be used by a trained healthcare professional as a diagnostic tool for the visualization and analysis of anatomical and pathological structures in patients of all ages, and to be used for diagnostic imaging in radiology, interventional radiology, and cardiology and in oncology. The system is also intended to be used for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer*. The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society.
Device Story
Spectral CT on Rails is a whole-body CT system featuring a continuously rotating gantry that slides horizontally on rails over a stationary patient support. Input: X-ray transmission data acquired at different angles and planes using a dual-layer detector (NanoPanel Prism). Operation: System acquires high- and low-energy spectra to analyze material attenuation coefficients; enables material decomposition, density, and effective atomic number quantification. Output: Cross-sectional images, material analysis, and graphical displays. Used in radiology, interventional radiology, cardiology, and oncology settings. Operated by trained healthcare professionals. Enhancements include interventional control features, passive dose alerts, and workflow integrations with angio systems. Benefits: Provides advanced material characterization and diagnostic visualization for clinical decision-making.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical performance testing, including system-level verification, validation, and image quality performance testing (CT number, uniformity, noise, spatial resolution, low contrast resolution, and spectral accuracy) compared against the primary predicate.
Technological Characteristics
Whole-body CT system with sliding gantry on rails. Detectors: Dual-layer solid-state yttrium-based scintillator (GOS + Photodiode). Energy: 120kW generator, 80-140 kVp settings. Connectivity: DICOM 3.0. Software: Windows 10-based, includes spectral reconstruction and viewer. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-44, IEC 62304, ISO 10993-1, ISO 14971.
Indications for Use
Indicated for diagnostic imaging of anatomical and pathological structures in patients of all ages in radiology, interventional radiology, cardiology, and oncology. Also indicated for low-dose CT lung cancer screening for early detection of lung nodules in accordance with approved protocols.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Philips CT Big Bore Sliding Gantry Configuration (K181797)
Submission Summary (Full Text)
{0}------------------------------------------------
May 13, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.
Philips Medical Systems Nederland B.V. % Maya Tolchinsky Regulatory Affairs Specialist Veenpluis 6 Best. 5684 PC NETHERLANDS
Re: K212875
Trade/Device Name: Spectral CT on Rails Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: Class II Product Code: JAK Dated: April 12, 2022 Received: April 15, 2022
#### Dear Maya Tolchinsky:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laurel Burk, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT 8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K212875
Device Name Spectral CT on Rails
#### Indications for Use (Describe)
The Spectral CT on Rails is a Computed Tomography X-Ray System intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories. The Spectral CT on Rails system acquires one CT dataset – composed of data from a higher-energy detected x-ray spectrum and a lower- energy detected x-ray spectra may be used to analyze the differences in the energy dependence of the attenuation coefficient of different materials. This allows for the generation of images at energies selected from the available spectrum and to provide information about the chemical composition of the body materials and/or contrast agents. Additionally, materials analysis provides for the quantification and graphical display of attenuation, material density, and effective atomic number.
This information may be used by a trained healthcare professional as a diagnostic tool for the visualization and analysis of anatomical and pathological structures in patients of all ages, and to be used for diagnostic imaging in radiology, interventional radiology, and cardiology and in oncology.
The system is also intended to be used for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer*.
The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society.
*Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl. J Med 2011; 365:395-409) and subsequent literature, for further information.
Type of Use (Select one or both, as applicable)
| <div> <span> <span style="font-size:16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|---------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px;">□</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### K212875
PHILIPS BG Diagnostic Imaging
#### SECTION 5 510(K) SUMMARY
HA300-006-05 Revision: 01 Status: Approved Page: 1 of 21
#### CT/AMI
### SECTION 5 510(K) SUMMARY
## 510(k) Summary
This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR §807.92.
Date Prepared: September 2, 2021
| Manufacturer: | Philips Medical Systems Nederland B.V.<br>Veenpluis 6,<br>5684 PC BEST<br>The Netherlands<br>Establishment Registration Number: 3015777306 | |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------|
| Primary<br>Contact<br>Person: | Maya Tolchinsky<br>Regulatory Affairs Specialist<br>Philips Medical Systems Technologies Ltd.<br>Phone: 972-52-8598223<br>E-mail: maya.tolchinsky@philips.com | |
| Secondary<br>Contact<br>Person: | Ilana Ben-Moshe<br>Regulatory Affairs Haifa Site Lead<br>Philips Healthcare<br>Phone: 972 (52) 5233496<br>E-mail: Ilana.Ben-Moshe@philips.com | |
| Device: | Trade Name: | Spectral CT on Rails |
| | Common name: | Computed Tomography x-Ray System |
| | Classification Name: | Computed Tomography x-Ray System |
| | Classification Regulation: | 21CFR 892.1750 |
| | Classification Panel: | Radiology |
| | Device Class: | II |
| | Primary Product Code: | JAK |
| | Secondary Product Code: | Not Applicable |
| Primary<br>Predicate<br>Device: | Trade Name: | Spectral CT |
| | Manufacturer: | Philips Medical Systems Nederland B.V. |
| | 510(k) Clearance: | K203020 |
| | Classification Name: | Computed Tomography x-Ray System |
| | Classification Regulation: | 21CFR §892.1750 |
| | Classification Panel: | Radiology |
| | Device Class: | Class II |
| | Product Code: | JAK |
| Secondary<br>Predicate<br>Device: | Trade Name: | Philips CT Big Bore Sliding Gantry Configuration |
| | Manufacturer: | Philips Medical Systems Nederland B.V. |
| | 510(k) Clearance: | K181797 |
| | Classification Name: | Computed Tomography x-Ray System |
| | Classification Regulation: | 21CFR §892.1750 |
| | Classification Panel: | Radiology |
| | Device Class: | Class II |
| | Product Code: | JAK |
{4}------------------------------------------------
| PHILIPS<br>BG Diagnostic Imaging<br>CT/AMI | HA300-006-05<br>Revision: 01<br>Status: Approved<br>Page: 2 of 21 |
|--------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| SECTION 5 510(K) SUMMARY | |
| Device Description: | The proposed Spectral CT on Rails System is a whole-body computed tomography (CT) X-Ray System featuring a continuously rotating x-ray tube and detectors gantry and multi-slice capability. The acquired x-ray transmission data is reconstructed by computer into cross-sectional images of the body taken at different angles and planes. The proposed Spectral CT on Rails System consists of three main components: a rotating gantry that slides on a carriage in the horizontal direction, stationary patient support and an operator console for scan control and image reconstruction. On the gantry, the main active components are the X-ray HV power supply, the X-ray tube and the detection system. The fundamental design and characteristics of the main components used in the proposed Spectral CT on Rails System, are identical to the cleared to market primary predicate device, Spectral CT System (K203020). The proposed Spectral CT on Rails System consists of main components that are similar to the cleared for market primary predicate device, Spectral CT (K203020) Gantry. The Gantry consists of the following main internal units: Stator - a fixed mechanical frame that carries HW and SW Rotor - A rotating circular stiff frame that is mounted in and supported by the stator. X-Ray Tube (XRT) and its power Generator, and the upper beam mechanism – fixed to the Rotor frame Rails - the rails system includes a carriage which the gantry sits on so that it may be moved back and forth on the rails horizontally relative to a stationary patient support that the patient lays on. The moving gantry functionality has previously been cleared in, Philips CT Big Bore Sliding Gantry Configuration (K181797), secondary predicate device. Data Measurement System (DMS) – a detectors array, fixed to the rotor in front of the XRT. |
Console - A computer and display that interfaces between the system and the user. Common Image Reconstruction Unit (CIRS) – a dedicated powerful image reconstruction system
In addition to the above components and the operating software, the system includes: Workstation hardware and software for data acquisition and image display, manipulation, storage, and filming; as well as post-processing into views other than the original axial images. Spectral Reconstruction System Spectral CT Viewer.
- The Spectral CT on Rails is a Computed Tomography X-Ray System intended to produce cross-Indications sectional images of the body by computer reconstruction of x-ray transmission data taken at for Use: different angles and planes. This device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
The Spectral CT on Rails system acquires one CT dataset - composed of data from a higherenergy detected x-ray spectrum and a lower- energy detected x-ray spectrum. The two spectra may be used to analyze the differences in the energy dependence of the attenuation coefficient of different materials. This allows for the generation of images at energies selected from the available spectrum and to provide information about the chemical composition of the body materials and/or contrast agents. Additionally, materials analysis provides for the quantification and graphical display of attenuation, material density, and effective atomic number.
This information may be used by a trained healthcare professional as a diagnostic tool for the visualization and analysis of anatomical and pathological structures in patients of all ages, and to be used for diagnostic imaging in radiology, interventional radiology, and cardiology and in oncology.
{5}------------------------------------------------
#### CT/AMI
The system is also intended to be used for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer*.
The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society.
*Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
{6}------------------------------------------------
| PHILIPS<br>BG Diagnostic Imaging | SECTION 5 510(K) SUMMARY | HA300-006-05<br>Revision: 01 |
|----------------------------------|--------------------------|-----------------------------------|
| CT/AMI | | Status: Approved<br>Page: 4 of 21 |
The main components (detection system, the reconstruction and the x-ray system) that are used in the proposed Spectral CT on Rails Technological System have the same fundamental design characteristics and are based on the seared to market primary predicate Characteristics device, Spectral CT System (K203020).
> The design fundamental scientific technology of both the proposed Spectral CT on Rails System and the cleared to market primary predicate device, Spectral CT System (K203020) are the same. The design changes (e.g. patient support, horizontal motion range) do not change the fundamental scientific technology of the proposed Spectral CT on Rails System.
The following table lists the technological characteristics differences for the proposed Spectral CT on Rails
| Table 05-1 | |
|---------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Modifications made to the proposed device, Spectral CT on Rails as compared to<br>the Primary predicate device, Spectral CT (K203020) | |
| Design feature | Description |
| Couch | The proposed Spectral CT on Rails is not installed with a couch that<br>can be controlled by the CT system. |
| Horizontal<br>Movement | The Proposed Spectral CT on Rails moves the rotating gantry<br>towards the patient who is positioned on a couch that is stationary in<br>the horizontal axis; whereas the primary predicate system the<br>rotating gantry was stationary and the couch was mobile in the<br>horizontal axis. |
| Horizontal Motion<br>Range | The Proposed Spectral CT on Rails has a larger range of horizontal<br>motion that allows the user to move the Gantry away from couch to<br>a parking position |
| Interventional<br>Features | The proposed Spectral CT on Rails has enhanced interventional and<br>interventional control features |
{7}------------------------------------------------
| | Table 05-2. Technological Characteristics Comparison | | | |
|----------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Device<br>Feature | Primary Predicate Device:<br>Spectral CT (K203020) | Proposed Device: Spectral CT on<br>Rails | Conclusion | |
| Application | Head, Body and Cardiac | Head, Body and Cardiac | Identical, therefore substantially<br>equivalent. | |
| Scan Regime | Continuous Rotation | Continuous Rotation | Identical, therefore substantially<br>equivalent. | |
| Scan Field of View | Up to 500 mm | Up to 500 mm | Identical, therefore substantially<br>equivalent. | |
| Scan modes | Surview<br>Axial-after-Axial<br>Dynamic Scan<br>Helical Scan | Surview<br>Axial-after-Axial<br>Dynamic Scan<br>Helical Scan | Identical, therefore substantially<br>equivalent. | |
| Spatial Resolution | 16 lp/cm max (high mode)<br>13 lp/cm max (standard mode) | 16 lp/cm max (high mode)<br>13 lp/cm max (standard mode) | Identical, therefore substantially<br>equivalent. | |
| Low Contrast<br>Resolution (20 cm<br>Catphan phantom) | 4 mm @ 0.3% @ 25 mGy<br>CTDIvol | 4 mm @ 0.3% @ 25 mGy CTDIvol | Identical, therefore substantially<br>equivalent. | |
| Minimum Scan Time | 0.18 sec for 240° rotation,<br>0.27 sec for 360° rotation | 0.18 sec for 240° rotation,<br>0.27 sec for 360° rotation | Identical, therefore substantially<br>equivalent. | |
| Number of Slices | Up to 128 slices of 0.625 mm | Up to 128 slices of 0.625 mm | Identical, therefore substantially<br>equivalent. | |
| Scan Coverage | Scanner Center of Rotation<br>(COR) is up to 80 mm | Scanner Center of Rotation (COR) is<br>up to 80 mm | Identical, therefore substantially<br>equivalent. | |
| Noise in Standard<br>Mode (as measured<br>on 21.6 cm water-<br>equivalent) | 0.27% at 27 mGy | 0.27% at 27 mGy | Identical, therefore substantially<br>equivalent. | |
| Image Matrix | Up to 1024 x 1024 | Up to 1024 x 1024 | Identical, therefore substantially<br>equivalent. | |
| Table 05-2. Technological Characteristics Comparison | | | | |
| Device<br>Feature | Primary Predicate Device:<br>Spectral CT (K203020) | Proposed Device: Spectral CT on<br>Rails | Conclusion | |
| Display (Pixels) | 1024 x 1280 | 1024 x 1280 | Identical, therefore substantially<br>equivalent. | |
| Communication | Compliance with DICOM 3.0 | Compliance with DICOM 3.0 | Identical, therefore substantially<br>equivalent. | |
| Detectors | NanoPanel Prism | NanoPanel Prism | Identical, therefore substantially<br>equivalent. | |
| Type | NanoPanel Prism | NanoPanel Prism | Identical, therefore substantially<br>equivalent. | |
| Material | Solid-state yttrium-based<br>scintillator, GOS + Photodiode | Solid-state yttrium-based scintillator,<br>GOS + Photodiode | Identical, therefore substantially<br>equivalent. | |
| DMS Detector | 8 cm - Dual-Layer scintillator, up<br>to 128 detector rows | 8 cm - Dual-Layer scintillator, up to<br>128 detector rows | Identical, therefore substantially<br>equivalent. | |
| DMS structure | Spherical DMS structure | Spherical DMS structure | Identical, therefore substantially<br>equivalent. | |
| Detector structure | 0.625 mm and various<br>combinations, such as:<br>2x0.625, 16x0.625, 32x0.625,<br>64x0.625, 96x0.625, 112x0.625,<br>128x0.625 mm | 0.625 mm and various combinations,<br>such as:<br>2x0.625, 16x0.625, 32x0.625,<br>64x0.625, 96x0.625, 112x0.625,<br>128x0.625 mm. | Identical, therefore substantially<br>equivalent. | |
| Collimation | 2x0.625, 16x0.625, 32x0.625,<br>64x0.625, 96x0.625, 112x0.625,<br>128x0.625 mm | 2x0.625, 16x0.625, 32x0.625,<br>64x0.625, 96x0.625, 112x0.625,<br>128x0.625 mm. | Identical, therefore substantially<br>equivalent. | |
| Slice thickness | Various slice thickness options<br>available in the range of 0.67 - 10<br>mm for helical mode and 0.625 –<br>20 for axial mode. | Various slice thickness options<br>available in the range of 0.67 - 10 mm<br>for helical mode and 0.625 - 20 for<br>axial mode. | Identical, therefore substantially<br>equivalent. | |
| Scan field of view | Up to 500 mm | Up to 500 mm | Identical, therefore substantially<br>equivalent. | |
| Table 05-2. Technological Characteristics Comparison | | | | |
| Device<br>Feature | Primary Predicate Device:<br>Spectral CT (K203020) | Proposed Device: Spectral CT on<br>Rails | Conclusion | |
| Gantry | | | | |
| Gantry rotation<br>speed | 0.27 sec -1.5 sec (360° rotation)<br>0.18 sec, 0.2 sec (240° rotation) | 0.27 sec -1.5 sec (360° rotation)<br>0.18 sec, 0.2 sec (240° rotation) | Identical, therefore substantially<br>equivalent. | |
| Bore size | 800 mm | 800 mm | Identical, therefore substantially<br>equivalent. | |
| Operator Controls<br>located on Gantry | Touch Panel Controls | Touch Panel Controls | Identical, therefore substantially<br>equivalent. | |
| Eclipse Collimation | A-Plane | A-Plane | Identical, therefore substantially<br>equivalent. | |
| Generator and Tube | |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.