K212867 · Nanchang Kanghua Health Materials Co., Ltd. · FXX · Dec 11, 2021 · General, Plastic Surgery
Device Facts
Record ID
K212867
Device Name
Medical Surgical Mask
Applicant
Nanchang Kanghua Health Materials Co., Ltd.
Product Code
FXX · General, Plastic Surgery
Decision Date
Dec 11, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4040
Device Class
Class 2
Indications for Use
The medical surgical mask are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.
Device Story
Pleated 3-layer surgical mask; inner/outer layers spun-bond non-woven fabric; middle layer melt-blown non-woven filter; malleable aluminum nose piece; spandex elastic ear loops. Used in clinical settings for infection control; worn by healthcare personnel and patients. Provides physical barrier against blood, body fluids, and particulates. Single-use, disposable, non-sterile. Protects wearer and patient from cross-contamination.
Clinical Evidence
No clinical data. Bench testing only. Performance verified against ASTM F2100 Level 2 requirements: BFE (ASTM F2101) >98%, PFE (ASTM F2299) >98%, Differential Pressure (EN 14683) <6.0 mm H2O/cm2, Synthetic Blood Penetration (ASTM F1862) 32/32 pass at 120 mmHg, and Flammability (16 CFR 1610) Class 1. Biocompatibility confirmed per ISO 10993-5 and ISO 10993-10.
Technological Characteristics
3-layer pleated mask; spun-bond/melt-blown/spun-bond non-woven polypropylene construction. Malleable aluminum nose piece; spandex ear loops. Dimensions: 175x95 mm. ASTM F2100 Level 2 compliant. Non-sterile. No electronic components or software.
Indications for Use
Indicated for use by patients and healthcare personnel to reduce exposure to microorganisms, body fluids, and particulate material during infection control practices.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 11, 2021
Nanchang Kanghua Health Materials Co., Ltd % Doris Chen Regulatory Affairs Staff Shanghai Jiushun Enterprise Management Technology Service Co., Ltd. Room 1502,BaoAn Building,No.800 Dongfang Road Shanghai, 200122 China
Re: K212867
Trade/Device Name: Medical Surgical Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: August 25, 2021 Received: September 8, 2021
Dear Doris Chen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212867
Device Name Medical Surgical Mask
### Indications for Use (Describe)
The medical surgical mask are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K212867
This summary of 510(k) is being submitted in accordance with the requirements of 21 CFR 807.92.
#### Applicant 1.
Sponsor Name: Nanchang Kanghua Health Materials Co., Ltd
Address: Moonlight Road,Medical Device Technology Park, Jinxian County,Nanchang
City,Jiangxi Province, 331700,China.
- � Contact Person (including title): Jiamin Wang (Manager)
- ◆ Phone:+86-791-85621388
#### 2. Submission Correspondent
Contact Person: Doris Chen
Shanghai Jiushun Enterprise Management Technology Service Co., Ltd. Address: Room 1502,BaoAn
Buiding,No.800 Dongfang Road,Shanghai,China. Tel: +86-21-50931939
Email: doris-chen@isosh.com
#### 3. Subject Device Information
| Type of 510(k): | Traditional |
|----------------------|-----------------------|
| Common Name: | Medical Face Mask |
| Trade Name: | Medical Surgical Mask |
| Classification Name: | Mask,Surgical |
| Review Panel: | General Hospital |
| Product Code: | FXX |
| Regulation Number: | 21 CFR 878.4040 |
| Regulation Class: | II |
{4}------------------------------------------------
#### Predicate Device Information 4.
### Predicate Device
| Sponsor: | Xiantao Rayxin Medical Products Co., Ltd. |
|--------------------|-------------------------------------------|
| Common Name: | Surgical Face Mask |
| Trade Name: | Disposable Surgical Face Mask |
| 510(k) number: | K153496 |
| Review Panel: | General Hospital |
| Product Code: | FXX |
| Regulation Number: | 21 CFR 878.4040 |
| Regulation Class: | II |
#### 5. Device Description
The medical surgical mask is pleated three-layer mask with ear loops and nose piece. The inner and outer layers are made of spun-bond non-woven fabric. The middle layer is made of melt blown non-woven fabric. only the outer layers' color is blue (colorant: Pigment Blue 15:3 /CAS number: 147-14-8), which is held to cover the users' mouth and nose by two spandex elastic bands ultrasonic welded to the medical surgical mask. The elastic ear loops are not made with natural rubber latex. The nose piece contained in the medical surgical mask is in the middle layer of medical surgical mask to allow the user to fit the medical surgical mask around their noses, which is made of malleable aluminum wire coated with plastic materials. The dimensions of each medical surgical mask are length 175±5 mm and width 95±5 mm, The dimensions of nose piece is length 110±20 mm, and the ear loop is length 165±10 mm. The medical surgical mask are sold non-sterile and are intended to be single use, disposable devices.
#### 6. Intended Use / Indications for Use
The medical surgical mask are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These masks
{5}------------------------------------------------
are intended for use in infection control practices to reduce the potential exposure to blood and
body fluids. This is a single use, disposable device(s), provided non-sterile.
#### Comparison with predicate device 7.
Table 1 General Comparison
| Elements of<br>Comparison | Subject Device | Predicate Device | Comparison | | |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------|----------------|
| Manufacturer | Nanchang Kanghua Health<br>Materials Co., Ltd | Xiantao Rayxin Medical<br>Products Co., Ltd. | -- | | |
| Product Name | Medical Surgical Mask | Disposable Surgical Face<br>Masks | -- | | |
| K Number | K212867 | K153496 | -- | | |
| Product Code | FXX | FXX | Same | | |
| Regulation<br>Number | 21 CFR 878.4040 | 21 CFR 878.4040 | Same | | |
| Intended use/<br>Indications for Use | The medical surgical mask are<br>intended to be worn to protect<br>both the patient and healthcare<br>personnel from transfer of<br>microorganisms, body fluids<br>and particulate material. These<br>masks are intended for use in<br>infection control practices to<br>reduce the potential exposure<br>to blood and body fluids. This is<br>a single use, disposable<br>device(s), provided non-sterile. | The Disposable Surgical<br>Face Masks are intended to<br>be worn to protect both the<br>patient and healthcare<br>personnel from transfer of<br>microorganisms, body<br>fluids and particulate<br>material. These face masks<br>are intended for use in<br>infection control practices to<br>reduce the potential<br>exposure to blood and body<br>fluids. This is a single use,<br>disposable device(s),<br>provided nonsterile. | Same | | |
| Mask style | Flat pleated, 3 layers. | Flat pleated,3 layers. | Same | | |
| Design feature | Ear loop | Ear Loop and Tie-On | Similar | | |
| Material | Outer<br>facing<br>layer | Spun-bond non-woven fabric | Spun-bond polypropylene | Similar | |
| | Middle<br>layer | Melt blown non-woven fabric | Melt blown polypropylene<br>filter | Similar | |
| | Inner<br>facing<br>layer | Spun-bond non-woven fabric | Spun-bond polypropylene | Similar | |
| | Nose | Outer plastic, inner aluminum | Malleable aluminum wire | Different | |
| | piece | wire | | Note 1 | |
| | Ear<br>loops | Spandex | Polyester | Different<br>Note 1 | |
| Color | | Blue | Blue | Same | |
| Dimension (Width) | | 17.5cm±0.5cm | 17.5cm±1cm | Different<br>Note 2 | |
| Dimension<br>(Length) | | 9.5cm±0.5cm | 9.5cm±1cm | Different<br>Note 2 | |
| | OTC use | Yes | Yes | Same | |
| | Sterility | Non-Sterile | Non-Sterile | Same | |
| | Use | Single Use, Disposable | Single Use, Disposable | Same | |
| | ASTM<br>Level | F2100<br>Level 2 | Level 2 | Same | |
| | Fluid resistance<br>Performance<br>ASTM F1862 | Level 2:32 out of 32 pass at<br>120mmHg,3non-consecutive<br>lots tested. | 32 out of 32 pass at 120<br>mmHg | Similar | |
| | Particle Filtration<br>Efficiency<br>ASTM F2299 | 3 non-consecutive lots tested,<br>using a sample size of 32/lot.<br>Level 2:<br>Lot A: 98.55%<br>Lot B: 98.63%<br>Lot C: 98.53% | 98.46% | Similar Note 3 | |
| | Bacterial Filtration<br>Efficiency<br>ASTM F2101 | | 3 non-consecutive lots tested,<br>using a<br>sample size of 32/lot.<br>Level 2:<br>Lot A: 99.9%<br>Lot B: 99.875%<br>Lot C: 99.9% | 98.7% | Similar Note 3 |
| | Flammability<br>Class<br>16 CFR 1610 | Class1, 3 non-consecutive lots<br>tested, using a sample size of<br>32/lot. | Class 1<br>Non Flammable | Similar | |
| | Differential<br>Pressure (Delta -P) | | 3 non-consecutive lots tested,<br>using a sample size of 32/lot.<br>Level 2:<br>Lot A: 3.4 mm H2O/cm²<br>Lot B: 3.4 mm H2O/cm²<br>Lot C: 3.4 mm H2O/cm²<br>(EN 14683:2019, Annex C) | 4.2mmH2O/cm²<br>(MIL-M-36954C) | Similar Note 3 |
{6}------------------------------------------------
{7}------------------------------------------------
| Biocompatibility | ISO10993-5 and<br>ISO10993-10;Under the<br>conditions of the studies<br>employed, the device is<br>non-cytotoxic, non-sensitizing,<br>and non-irritating. | ISO10993-5 and<br>ISO10993-10;<br>Under the conditions of the<br>studies employed, the<br>device is non-cytotoxic,<br>non-sensitizing, and<br>non-irritating. | Same |
|------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
|------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
# Note 1
The difference in the materials of the current nose piece and the ear loop were different from the predicate device. The biocompatibility evaluation test of the subject devices have been performed on the final finished device.
# Note 2
Compare with the predicate and reference device, the different of the physical feature or size does not affect the intended use of the subject device.
# Note 3
For the Performance testing, the test results are not identical to each other, but they are similar and they both meet the requirement of Level 2 medical mask according to the ASTM F 2100.
#### 8. Summary of Non-Clinical Tests Performed
Non-clinical tests were conducted to verify that the proposed device met all design specifications as to the predicate device. The test results demonstrated that the proposed device complies with the following standards and the requirements stated in the Guidance for Industry and FDA Staff: Surgical Masks - Premarket Notification [510(k)] Submission issued on March 5, 2004:
- > ASTM F2299-03, Standard Test Method for Determining the Initial Efficiency of Materials Used in Surgical face masks to Penetration by Particulates Using Latex Spheres.
- A EN 14683:2019, Annex C. Method for determination of breathability (differential pressure)
- > ASTM F1862/ASTM F1862M-17, Standard test method for resistance of Surgical face masks to penetration by synthetic blood (Horizontal projection of fixed volume at a known velocity)
{8}------------------------------------------------
- ア ASTM F2101-19, Standard Test Method For Evaluating The Bacterial Filtration Efficiency (BFE) Of Surgical face mask Materials, Using A Biological Aerosol Of Staphylococcus Aureus.
- > 16 CFR Part 1610, Standard for the flammability of clothing textiles.
- > ISO 10993-5:2009, Biological Evaluation Of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity
- > ISO 10993-10:2010, Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.
|--|
| Test item | Purpose | Acceptance | Test results |
|------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|--------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| (Performance Level 2) | | criteria | |
| | | (Level 2) | |
| Bacterial<br>filtration<br>efficiency (BFE)<br>ASTM F2101-19 | The purpose of this testing<br>was to evaluate the Bacterial<br>filtration efficiency. | BFE≥98%. | 3 non-consecutive lots tested,<br>using a sample size of 32/lot.<br>Level 2:<br>Lot A: 99.9%<br>Lot B: 99.875%<br>Lot C: 99.9% |
| Differential pressure mm,<br>(Delta-P)<br>EN 14683:2019, Annex C | The purpose of this testing<br>was to evaluate the<br>Differential pressure. | Delta P<6.0H₂O/cm² | 3 non-consecutive lots tested.<br>using a sample size of 32/lot.<br>Level 2:<br>Lot A: 3.4 mm H₂O/cm²<br>Lot B: 3.4 mm H₂O/cm²<br>Lot C: 3.4 mm H₂O/cm²<br>(EN 14683:2019, Annex C) |
| Sub-micron<br>particulate<br>filtration efficiency at 0.1<br>micron.<br>ASTM F2299 | The purpose of this testing<br>was to evaluate sub-micron<br>particulate filtration efficiency. | PFE≥98%. | 3 non-consecutive lots tested.<br>using a sample size of 32/lot.<br>Level 2:<br>Lot A: 98.55%<br>Lot B: 98.63%<br>Lot C: 98.53% |
{9}------------------------------------------------
| Resistance to<br>penetration by synthetic<br>blood ASTM F1862 | The purpose of this testing<br>was to evaluate the resistance<br>to penetration by synthetic<br>blood. | 29 out of 32 pass at<br>120mmHg | Level 2:32 out of 32 pass at<br>120mmHg, 3non-consecutive<br>tested. |
|---------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|--------------------------------------------|-------------------------------------------------------------------------|
| Flammability<br>(Flame spread)<br>16 CFR Part 1610 | The purpose of this testing<br>was to evaluate the<br>flammability. | Class 1:<br>Burn time $\ge$ 3.5<br>seconds | Class 1, 3 non-consecutive<br>tested, using a sample size of<br>32/lot. |
Results: All tests were passed.
### Biocompatibility evaluation and test
Biocompatibility evaluation conducted in accordance with the FDA's 2016 guidance and
ISO10993-1:2018 supports that the subject devices are biocompatible.
The biocompatibility test includes the following tests:
- ≫ In vitro Cytotoxicity Test per ISO 10993-5:2009 Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicity.
- > Skin Sensitization Tests per ISO 10993-10:2010 Biological evaluation of medical devices —Part 10: Tests for irritation and skin sensitization
- ハ Skin Irritation Tests per ISO 10993-10:2010 Biological evaluation of medical devices-
Part 10: Tests for irritation and skin sensitization.
#### 9. Summary of Clinical Performance Test
No clinical study is included in this submission.
### 10. Final Conclusion
The conclusions drawn from the nonclinical tests demonstrate that the subject device, Medical surgical mask is as safe, as effective, and performs as well as or better than the legally marketed predicate device, Disposable Surgical Face Masks (K153496).
# 11. Summary Prepared Date
9/2/2021
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.